← Product Code [EBF](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF) · K102703

# BUILD-IT TOTAL CORE (K102703)

_Pentron Clinical · EBF · Jan 5, 2011 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF/K102703

## Device Facts

- **Applicant:** Pentron Clinical
- **Product Code:** [EBF](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF.md)
- **Decision Date:** Jan 5, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Build-It Total Core is indicated for use as a core build up material on vital and non-vital teeth, sealing root canal openings, cavity floor liner, post cementation, restorative material and dentin replacement material.

## Device Story

Build-It Total Core is a dual-cure, self-adhesive composite dental material with continuous fluoride release. It utilizes 4-MET technology to adhere to dentin and cut enamel without requiring a separate bonding agent. Used by clinicians in dental settings for core build-ups, post cementation, and restorative procedures. The material is applied directly to the tooth structure to provide structural support and restoration. It benefits patients by simplifying the restorative process through its self-adhesive properties and dual-cure mechanism, which allows for efficient placement and curing.

## Clinical Evidence

No clinical data. Evidence consists of biocompatibility studies and bench testing comparing performance characteristics (water absorption, solubility, flexural and compression strength, linear expansion, fluoride release, and bond strength) against predicate devices.

## Technological Characteristics

Dual-cure, self-adhesive composite resin material. Features 4-MET technology for bonding to dentin and enamel. Includes continuous fluoride release. Formulated as a fiber-reinforced resin composite.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- Build-It F.R. ([K000211](/device/K000211.md))
- Cement-It All Purpose ([K060698](/device/K060698.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kl02703.

Image /page/0/Picture/1 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a curved line above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are written in a smaller, sans-serif font. The logo is in black and white.

JAN - 5 2011

Section III - 510(k) Summary of Safety and Effectiveness

# Submitter:

Sybron Dental Specialties, Inc. 1717 W. Collins Avenue ! Orange, California 92867 (714) 516-7602 - Phone (714) 516-7472 - Facsimile Wendy Garman - Contact Person

## Date Summary Prepared:

December 2010

### Device Name:

- · Trade Name Build-It Total Core
- Common Name Dental Composite Restorative Material
- · Classification Name Tooth Shade Resin Material, per 21 CFR § 872.3690

# Devices for Which Substantial Equivalence is Claimed:

- Build-It F.R., K000211, Pentron Clinical
- · Cement-It All Purpose, K060698, Pentron Clinical

# Device Description:

Build-It Total Core is a dual-cure, self-adhesive composite with a continuous fluoride release. Build-It Total Core utilizes 4-MET technology to adhere to dentin and cut enamel without a separate bonding agent. Ideal flow characteristics, esthetic and contrast shades, and outstanding physical properties make Build-It Total Core an indispensable tool in every clinician's armamentarium.

# Intended Use of the Device: (

Build-It Total Core is indicated for use as a core build up material on vital and non-vital teeth, sealing root canal openings, cavity floor liner, post cementation, restorative material and dentin replacement material. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{1}------------------------------------------------

# Substantial Equivalence:

Build-It Total Core is substantially equivalent to two other legally marketed devices in the United States. Build-It Totol Core functions in a manner similar to and is intended for the same use as Build-It F.R, manufactured by Pentron Clinical. Its self-adhering properties are equivalent to Cement-It All Purpose, currently marketed by Pentron Clinical as Breeze. Build-It Total Core is similar to Build-It F.R in that it is a fiber reinforced, dual cure post and core build-up resin composite material which is used for tooth restorations. Build-It Total Core differs from Build-It F.R in that it utilizes 4-MET technology to adhere to dentin and cut enamel without a separate bonding agent. Build-it Total Core is similar to Cement-It All Purpose, a self-adhering dental cement currently marketed by Pentron Clinical as Breeze, in that it has a similar formulation and it has self-adhering properties.

# Non-Clinical Test Data

A biocompatibility study was completed, which demonstrates that the material is safe for its intended use. This 510(k) submission also includes data from bench testing used to ēvaluate performance characteristics of Build-It Total Core compared to the predicate devices, Build-It F.R. and Cement-It All Purpose currently marketed by Pentron Clinical as Breeze. The characteristics evaluated include water absorption and solubility, flexural and compression strength, linear expansion in water, fluoride release, and bond strength.

#### Clinical Testing

Clinical testing has not been conducted on this product.

#### Conclusion:

Based upon the biocompatibility studies and similar technological/performance characteristics as compared to the predicate devices, the clinical performance of Build-It Total Core is deemed to be substantially equivalent to the predicate devices, Build-It F.R. and Cement-It All Purpose currently marketed by Pentron Clinical as Breeze.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Wendy Garman Director, Regulatory Affairs Sybron Dental Specialties, Incorporated 1717 W. Collins Avenue Orange, California 92867

Re: K102703

Trade/Device Name: Build-It Total Core Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: December 6, 2010 Received: December 7, 2010

JAN - 5 2011

Dear Ms. Garman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Garman

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K102703

JAN - 5 2011

Device Name: Build-It Total Core

Indications For Use:

Build-It Total Core is indicated for use as a core build up material on vital and non-vital teeth, sealing root canal openings, cavity floor liner, post cementation, restorative material and dentin replacement material.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

Over-The-Counter Use ___________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation |         |
|--------------------------------------------------|---------|
| (Division Sign-Off)                              |         |
| Division of Anesthesiology, General Hospital     |         |
| Infection Control, Dental Devices Page 1 of 1    |         |
| 1 510(k) Number:                                 | K182703 |

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF/K102703](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/EBF/K102703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
