← Product Code [DZE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE) · K963401

# TPS APICAL IMPLANT (K963401)

_Steri-Oss, Inc. · DZE · Nov 5, 1996 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K963401

## Device Facts

- **Applicant:** Steri-Oss, Inc.
- **Product Code:** [DZE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE.md)
- **Decision Date:** Nov 5, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The implant is indicated for use in the mandible and maxilla for use as an artificial root structure for abutments for bridgework and denture retention.

## Device Story

Endosseous dental implant; serves as artificial root for natural dentition replacement; supports prosthetic crowns/abutments for masticatory function. Implant features threaded design; 3.8 mm width; lengths 8-18 mm; titanium coating on half of surface. Used in clinical dental settings by practitioners to restore patient dentition.

## Clinical Evidence

No clinical data; performance data not applicable.

## Technological Characteristics

Material: Titanium; threaded geometry; 3.8 mm width; lengths 8-18 mm; flat superior aspect; partial titanium coating; sterile.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- Steri-Oss Apical Implants

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

STERI-OSS
A BAUSCH &amp; LOMB Company
Original 510(k)-
TPS Apical Implant
NOV - 5 1996
19
K963401

# Section 6

## 510(k) Summary

### Manufacturer Information:

Submitter's Name: Steri-Oss

Address: 22895 Eastpark Drive
Yorba Linda, CA 92887
U.S.A.

Contact's Name: Paul Gasser
Regulatory Affairs Specialist

Phone: 714-282-6515

Date Prepared: August 1996

### Device Names:

Common Name: Endosseous dental implant

Trade Name: TPS Apical Implant

Classification Name: Endosseous implant

### Predicate Device:

Substantial equivalence is claimed to Steri-Oss’ Apical Implants.

### Device Description:

How device functions: The Steri-Oss implant functions by serving as a replacement for the root portion of natural dentition, upon which a prosthetic may be constructed.

Page 1 of 2
Steri-Oss Inc. - August 1996
Steri-Oss Inc. 22895 Eastpark Drive Yorba Linda, CA 92687 • (714) 282-6515 Fax: (714) 998-9236 • (800) 993-8100

{1}

Original 510(k)
TPS Apical Implant
20

## Device Description (cont.):

Scientific concepts: Natural dentition is composed of a root (subgingival) and a crown (supragingival). Designs in existence are intended to mimic this structure to aid the patient in restoring natural masticatory function. The implant is designed to serve as the root of the artificial tooth and the abutment/prosthetic is designed to serve as the crown.

Characteristics: The implant is threaded with a 3.8 mm width and an 8, 10, 12, 14, 16 and 18 mm length. The superior aspect is flat. Half of the implant is coated with titanium.

## Intended Use:

The implant is indicated for use in the mandible and maxilla for use as an artificial root structure for abutments for bridgework and denture retention.

## Comparison to Predicate:

The following table provides a comparison of the technological characteristics of the Steri-Oss implant to the predicate.

|  Item | Predicate | TPS Apical Implant  |
| --- | --- | --- |
|  Material | Titanium and HA | Titanium  |
|  Width (mm) | 3.8 | Same  |
|  Length (mm) | 8 - 18 | Same  |
|  Geometry | Threaded | Same  |
|  Sterility | Sterile | Same  |

## Performance Data:

Not applicable.

Page 2 of 2
Steri-Oss Inc. - August 1996

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K963401](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K963401)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
