SPLINE DENTAL IMPLANT SYSTEMS
Device Facts
| Record ID | K962106 |
|---|---|
| Device Name | SPLINE DENTAL IMPLANT SYSTEMS |
| Applicant | Calcitek, Inc. |
| Product Code | DZE · Dental |
| Decision Date | Aug 29, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
For use in edentulous mandibles or maxillae for attachment of complete denture prostheses, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement.
Device Story
Spline Dental Implant System; endosseous dental implant; used in edentulous mandibles or maxillae; placed by licensed dentist; provides attachment for complete dentures, fixed/removable bridgework, or single tooth replacement; modified apical portion enables self-tapping; supplied sterile; prosthetic components unchanged; interface unchanged; available in textured and untextured surfaces.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Endosseous dental implant; modified apical geometry for self-tapping; available in textured and untextured surfaces; supplied sterile; materials unchanged from predicate.
Indications for Use
Indicated for patients with edentulous mandibles or maxillae requiring complete denture prostheses, fixed or removable bridgework, or single tooth replacement.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- Spline Dental Implant System (K962106)
Related Devices
- K964768 — THREADLOC DENTAL IMPLANT SYSTEMS · Sulzer Calcitek, Inc. · Feb 25, 1997
- K981562 — SPLINE DENTAL IMPLANT SYSTEM - SPLINE XI IMPLANT · Sulzer Calcitek, Inc. · Jul 30, 1998
- K133510 — NEODENT IMPLANT SYSTEM · Jj Industria E Comercio DE Materiais Dentarios SA · Feb 24, 2014
- K050712 — 4.5.* 6.0MM DENTAL IMPLANT AND 6.0*6.0 MM DENTAL IMPLANT · Bicon, Inc. · Apr 22, 2005
- K013494 — 3.75MM AND 5.0MM SPLINE TWIST IMPLANT · Sulzer Dental, Inc. · Jan 18, 2002