The INNO SLA Mini Plus® Implant System is divided into two types: - Cemented Type The Cement type is indicated for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors to serve as temporary support prosthetic devices during the healing stage of permanent endosseous dental implant, such as artificial teeth, in order to restore masticating function in partially edentulous patients. - Ball Type The Ball type is designed for use in dental implant surgery. The Ball type is intended for use in partially or fully edentulous mandibles and maxillae, in support of overdentures. The use of the Ball type implants is not to exceed one hundred and eighty (180) days. The Cemented Type and Ball Type implant bodies are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
One-piece endosseous dental implant system comprising implant and abutment sections; fabricated from commercially pure titanium with S.L.A. surface treatment. Used in dental implant surgery by clinicians to restore masticating function. Cemented type serves as temporary support during healing of permanent implants; Ball type supports overdentures in partially or fully edentulous patients. Implant inserted into alveolar bone; Ball type connects to housing embedded in denture ridge via nylon inserts for removable solutions. Clinicians evaluate stability and occlusal loading to determine suitability for immediate loading. Benefits include restoration of chewing function and prosthetic support. MRI safety confirmed via literature and non-clinical assessment of titanium material properties.
Clinical Evidence
Bench testing only. Includes biocompatibility testing, chemical analysis, sterilization validation (ISO 17665-1/2), and shelf-life testing. MRI safety evaluated via literature review and assessment of titanium material properties per ASTM F2052, F2213, F2182, and F2119. Endotoxin testing performed via LAL gel-clot technique.
Technological Characteristics
One-piece root-form endosseous dental implant. Material: Commercially pure titanium (Grade 4). Surface: S.L.A. treatment. Diameters: 2.5/3.0 mm. Lengths: 10.0/12.0/14.0 mm. Sterilization: Gamma (implants) and Steam (caps). Connectivity: Mechanical connection to denture housing. No software or electronics.
Indications for Use
Indicated for partially or fully edentulous patients requiring temporary prosthetic support (Cemented Type) or overdenture support (Ball Type) in the maxilla or mandible. Cemented type specifically for missing maxillary lateral incisors or mandibular central/lateral incisors. Ball type limited to 180 days use. Both types indicated for immediate loading given sufficient primary stability.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
Cowellmedi Co., Ltd. % Priscilla Chung Official Correspondent LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine. California 92612
December 27, 2024
Re: K241127
Trade/Device Name: INNO SLA Mini Plus® Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: February 14, 2024 Received: December 3, 2024
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241127
# Device Name
INNO SLA Mini Plus® Implant System
## Indications for Use (Describe)
The INNO SLA Mini Plus® Implant System is divided into two types:
## - Cemented Type
The Cement type is indicated for use in the treatment of missing maxillary lateral incisors or the mandbular central and lateral incisors to serve as temporary support prosthetic devices during stage of permanent endosseous dental implant, such as artificial teeth, in order to restore masticating function in partially edentulous patients.
### - Ball Type
The Ball type is designed for use in dental implant surgery. The Ball type is intended for use in partially or fully edentulous mandibles and maxillae, in support of overdentures. The use of the Ball type implants is not to exeed one hundred and eighty (180) days.
The Cemented Type and Ball Type implant bodies are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (K241127)
This summary of 510(k)-substantial equivalence information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Dec 26, 2024
#### 1. Applicant / Submitter:
Cowellmedi Co., Ltd. 48, Hakgam-daero 221beon-gil, Sasang-gu, Busan, 46986. Republic of Korea Tel. + 82-51-314-2028
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine CA 92612 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### Subject Device: 3.
Proprietary Name: INNO SLA Mini Plus® Implant System Common Name: Endosseous dental implant Device Classification: II Regulation Number: 21 CFR 872.3640 Device Name: Implant, Endosseous, Root-Form Classification Product Code: DZE, NHA
#### 4. Predicate Devices:
- . Primary Predicate: S-Mini Implant System (K112540) by Neobiotech Co., Ltd.
- Reference Device: AnyOne Onestage Implant System (K210161) by MegaGen Implant Co., Ltd. .
- . Reference Device: The LOCATOR R-Tx® Attachment (K200827) System by Zest Anchors, LLC
- Reference Device: INNO SLA Submerged Implant System (K201323) by Cowellmedi Co., Ltd. .
- Reference Device: INNO SLA Submerged Implant System (K132242) by Cowellmedi Co., Ltd. .
- · Reference Device: MS System (K083067) by Osstem Implant Co., Ltd.
{5}------------------------------------------------
#### 5. Device Description:
The INNO SLA Mini Plus® Implant System has two types, cement type and ball type. The INNO SLA Mini Plus® Implant System is a one-piece endosseous dental implant which is a combination of implant and abutment sections. The implant is made of commercially pure titanium and has S.L.A. surface treatment.
#### 6. Indications for Use:
The INNO SLA Mini Plus® Implant System is divided into two types:
# - Cemented Type
The Cement type is indicated for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors to serve as temporary support prosthetic devices during the healing stage of permanent endosseous dental implant, such as artificial teeth, in order to restore masticating function in partially edentulous patients.
# - Ball Type
The Ball type is designed for use in dental implant surgery. The Ball type is intended for use in partially or fully edentulous mandibles and maxillae, in support of overdentures. The use of the Ball type implants is not to exceed one hundred and eighty (180) days.
The Cemented Type and Ball Type implant bodies are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
#### 7. Substantial Equivalence
## Substantial Equivalence Discussion
The Cement Type of INNO SLA Mini Plus® Implant System has the same device characteristics as the predicate device, S-Mini Implant System (K112540) by Neobiotech Co., Ltd. and MS System (K083067) by Osstem Implant Co., Ltd. in terms of intended use, design and use concept. The size range of the predicate device and the reference devices encompasses the subject device.
Same for the Ball Type, the INNO SLA Mini Plus® Implant System has the same device characteristics as the predicate device, S-Mini Implant System (K112540) by Neobiotech Co., Ltd in terms of intended use, design and use concept. The size range of the predicate device encompasses the subject device.
For the implants, the difference is the surface treatment; however, the biocompatibility and chemical analysis were performed and the test results support substantial equivalence.
{6}------------------------------------------------
The subject cap and ball attachments have the same intended use, design, raw material, and the technical characteristics as the predicate device. We performed the biocompatibility test and determine the subject components does not raise a question in substantial equivalence.
Lastly, we presented two references devices which are made by our company to be leveraged for biocompatibility and shelf life tests.
Based on the information and test results provided in submission, we conclude that the subject device is substantially equivalent to the predicate device.
# Comparison Chart
| | Subject Device | Predicate Device | Reference Device |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K232104 | K112540 | K083067 |
| Device Name | Mini Plus Fixture -<br>Cement Type<br>INNO SLA Mini Plus®<br>Implant System | S-Mini Implant System | MS System |
| Manufacturer | Cowellmedi Co., Ltd. | Neobiotech Co., Ltd. | Osstem Implant Co., Ltd. |
| Intended Use | The INNO SLA Mini Plus®<br>Implant System is divided<br>into two types:<br>- Cemented Type<br>The Cement type is<br>indicated for use in the<br>treatment of missing<br>maxillary lateral incisors or<br>the mandibular central and<br>lateral incisors to serve as<br>temporary support<br>prosthetic devices during<br>the healing stage of<br>permanent endosseous<br>dental implant, such as<br>artificial teeth, in order to<br>restore masticating function<br>in partially edentulous<br>patients.<br>The Cemented Type implant<br>bodies are indicated for<br>immediate loading when<br>good primary stability is<br>achieved and with | The Cemented type is<br>indicated for use in the<br>treatment of missing<br>maxillary lateral incisors or<br>the mandibular central and<br>lateral incisors to serve as<br>temporary support prosthetic<br>devices during the healing<br>phase of permanent<br>endosseous dental implant,<br>such as artificial teeth, in<br>order to restore chewing<br>function in partially<br>edentulous patients. | The MS System (Narrow<br>Ridge) is intended to use<br>in the treatment of missing<br>mandibular central and<br>lateral incisors to support<br>prosthetic device, such as<br>artificial teeth, in order to<br>restore chewing function<br>in partially edentulous<br>patients. MS System<br>(Narrow Ridge) is<br>intended for single use<br>only. It is not for<br>immediate loading. |
# 7.1. Mini Plus Fixture – Cement Type
{7}------------------------------------------------
| | appropriate occlusal<br>loading. | | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | Image: Dental implant | Image: Dental implant | Image: Dental implant |
| Diameters(Ø) | 2.5/3.0 | 2.0/2.5/3.0/3.5 | 3.0 |
| Lengths(mm) | 10.0/12.0/14.0 | 7.0/8.5/10.0/11.5/13.0/15.0 | 10.0/13.0/15.0 |
| Gingival<br>heights/cuffs(mm) | 2.0/4.0 | 2.0 | 2.5/4.0 |
| Product Code | DZE | DZE | DZE |
| Regulation | 872.3640 | 872.3640 | 872.3640 |
| Surface treatment | SLA | RBM | RBM |
| Sterility | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization |
| Angulation | 0° | 0° | 0° |
| Material | CP Ti Gr.4 | CP Ti Gr.4 | Ti 6Al 4V ELI |
| Principle of<br>operation | This product is a root-<br>type fixture which is<br>inserted in the alveolar<br>bone. It replaces the<br>functions of the missing<br>teeth as a dental implant<br>fixture. | This product is a root-type<br>fixture which is inserted in<br>the alveolar bone. It replaces<br>the functions of the missing<br>teeth as a dental implant<br>fixture. | This product is a root-type<br>fixture which is inserted in<br>the alveolar bone. It replaces<br>the functions of the missing<br>teeth as a dental implant<br>fixture. |
{8}------------------------------------------------
| | Subject Device | Predicate Device |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K232104 | K112540 |
| Device Name | Mini Plus Fixture - Ball Type<br>INNO SLA Mini Plus® Implant<br>System | S-Mini Implant System |
| Manufacturer | Cowellmedi Co., Ltd. | Neobiotech Co., Ltd |
| Intended Use | The INNO SLA Mini Plus® Implant<br>System is divided into two types:<br>- Ball Type<br>The Ball type is designed for use in<br>dental implant surgery. The Ball type<br>is intended for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of overdentures. The use of<br>the Ball type implants is not to exceed<br>one hundred and eighty (180) days.<br><br>The Ball Type implant bodies are<br>indicated for immediate loading when<br>good primary stability is achieved and<br>with appropriate occlusal loading. | The Ball type is designed for use in<br>dental implant surgery. Ball type is<br>intended for use in partially or fully<br>edentulous mandibles and maxillae,<br>in support of overdentures. Ball type<br>implants are for temporary use only. |
| Design | Image: Mini Plus Fixture - Ball Type | Image: S-Mini Implant System |
| Diameters(Ø) | 2.5/3.0 | 2.0/2.5/3.0/3.5 |
| Lengths(mm) | 10.0/12.0/14.0 | 7.0/8.5/10.0/11.5/13.0/15.0 |
| Gingival<br>heights/cuffs | 2.0/4.0 | 3.0/4.0 |
| Product Code | DZE | DZE |
| Regulation | 872.3640 | 872.3640 |
| Surface treatment | SLA | RBM |
| Sterility | Gamma Sterilization | Gamma Sterilization |
| Angulation | 0° | 0° |
| Material | CP Ti Gr.4 | CP Ti Gr.4 |
| Principle of<br>operation | This product is a root-type fixture<br>which is inserted in the alveolar bone.<br>It replaces the functions of the missing<br>teeth as a dental implant fixture. | This product is a root-type fixture<br>which is inserted in the alveolar bone.<br>It replaces the functions of the<br>missing teeth as a dental implant<br>fixture. |
| | Subject Device | Predicate Device |
| 510(k) Number | K232104 | K210161 |
| Device Name | Protection Cap<br>INNO SLA Mini Plus® Implant<br>System | Solid Post Cap<br>For AnyOne Onestage Implant<br>System |
| Manufacturer | Cowellmedi Co., Ltd. | MegaGen Implant Co., Ltd. |
| Intended Use | The INNO SLA Mini Plus® Implant<br>System is divided into two types:<br>- Cemented Type<br>The Cement type is indicated for use<br>in the treatment of missing maxillary<br>lateral incisors or the mandibular<br>central and lateral incisors to serve<br>as temporary support prosthetic<br>devices during the healing stage of<br>permanent endosseous dental<br>implant, such as artificial teeth, in<br>order to restore masticating function<br>in partially edentulous patients.<br>- Ball Type<br>The Ball type is designed for use in<br>dental implant surgery. The Ball<br>type is intended for use in partially<br>or fully edentulous mandibles and<br>maxillae, in support of overdentures.<br>The use of the Ball type implants is<br>not to exceed one hundred and<br>eighty (180) days.<br>The Cemented Type and Ball Type<br>implant bodies are indicated for<br>immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading. | The AnyOne Onestage Implant<br>System is intended to be surgically<br>placed in the maxillary or<br>mandibular arches for the purpose of<br>providing prosthetic support for<br>dental restorations (Crown, bridges,<br>and overdentures) in partially or fully<br>edentulous individuals. It is used to<br>restore a patient's chewing function<br>in the following situations and with<br>the clinical protocols:<br>-Delayed loading.<br>-Immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading.<br>Larger implants are dedicated for the<br>molar region. |
| Design | Image: Cylinder shape | Image: U shape |
| Diameters(Ø) | 4.05 mm | 5.3 mm |
| Lengths(mm) | 6.8 mm | 6.5, 7.5, 8.0, 8.5, 9.0, 9.5, 10.0, 10.5,<br>11.5 mm |
| Regulation | 872.3640 | 872.3640 |
| Sterility | Steam Sterilization | Steam Sterilization |
| Single Use | Yes | Yes |
| Material | POM | POM |
| Principle of<br>operation | The Protection Cap is used for<br>protecting a MiniPlus cemented type<br>fixture, and minimizing irritation to<br>tongue, gingiva and oral mucosa. | The Solid Post Cap is used for<br>protecting a Solid Post Abutment<br>after taking impression, and<br>minimizing irritation to tongue and<br>oral mucosa. |
# 7.2. Mini Plus Fixture - Ball Type
{9}------------------------------------------------
# 7.3. Protection Cap
{10}------------------------------------------------
# 7.4. Ball Attachment (Ball outer Cap, Ball Inner Cap)
| | Subject Device | Predicate Device |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K232104 | K200827 |
| Device Name | Ball outer cap, Ball Inner cap<br>INNO SLA Mini Plus® Implant<br>System | The LOCATOR R-Tx®<br>Attachment System…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.