The TSX Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to minimally support the implant (minimum 1mm circumferential and 2mm apical) and provide good primary stability. The 3.1mmD TSX Implants should be splinted to additional implants when used in the pre-molar region and should not be used in the molar region.
Device Story
TSX Implants are endosseous root-form dental implants; manufactured from titanium alloy (Ti 6Al-4V ELI). Device features tapered screw-type design; hybrid surface (dual acid-etched + MTX textured surface). Used by dental clinicians in clinical settings for single or multiple tooth replacement; immediate or delayed loading. Implants placed into prepared osteotomy; primary stability achieved via macro design features. Output is a stable foundation for prosthetic attachment. Benefits include restoration of chewing function and tooth replacement. Device is supplied sterile via gamma radiation.
Clinical Evidence
No clinical data were included in this submission. Evidence is based on bench testing, including ISO 14801 fatigue testing, biocompatibility assessment per ISO 10993-1, sterility validation per ISO 11137, and SEM/EDS surface analysis.
Indicated for patients requiring replacement of one or more missing teeth in the maxilla or mandible. Suitable for immediate loading or conventional healing periods, and immediate placement post-extraction or tooth loss, provided sufficient alveolar bone volume exists. Contraindicated for 3.1mmD implants in the molar region; 3.1mmD implants in the pre-molar region must be splinted.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Biomet 3i LLC Mariela Cabarcas Sr. Regulatory Affairs Specialist 4555 Riverside Drive Palm Beach Gardens, Florida 33410
Re: K220978
Trade/Device Name: TSXTM Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: August 11, 2022 Received: August 12, 2022
Dear Mariela Cabarcas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known)
K220978
Device Name
TSXTM Implants
Indications for Use (Describe)
The TSX Implants are designed for use in the maxilla or mandible for immediate loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to minimally support the implant (minimum 1mm circumferential and 2mm apical) and provide good primary stability.
The 3.1mmD TSX Implants should be splinted to additional implants when used in the pre-molar region and should not be used in the molar region.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete. and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (6/20) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 |
|----------------------|-------------|----------------------------------------|
|----------------------|-------------|----------------------------------------|
{3}------------------------------------------------
## TSX™ Implants 510(k) Summary 09/12/2022
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92
#### I. Submitter Information:
| | Name: | Biomet 3i LLC |
|------|-----------------------------|-----------------------------------------------------------|
| | Address: | 4555 Riverside Drive<br>Palm Beach Gardens, Florida 33410 |
| | Phone: | (561) 776-6923 |
| | Fax: | (561) 514-6316 |
| | Contact Person: | Mariela Cabarcas |
| | Job Title: | Sr. Regulatory Affairs Specialist |
| | Email: | mariela.cabarcassanclemente@ZimVie.com |
| II. | Proprietary Trade Name: | TSXTM Implants |
| III. | Device Classification Name: | Implant, Endosseous, Root-Form (21 CFR 872.3640) |
| IV. | Regulatory Class: | Class II |
| V. | Product Code: | DZE |
| VI. | Reviewing Branch: | Dental Devices Branch |
| VII. | Predicate Devices: | |
| | Primary predicate device: | |
| | Trade Name: | Tapered Screw-Vent Implant, 4.1mmD |
| | 510(k): | K072589 |
- 21 CFR 872.3640 Regulation Number: ●
- Product Code: DZE
- . SE Date: 10/04/2007
### Reference Device:
| Trade Name: | Zimmer 3.1mmD Dental Implant System |
|--------------------|-------------------------------------|
| 510(K): | K142082 |
| Regulation Number: | 21 CFR 872.3640 |
- Prooduct Code: DZE, NHA
{4}------------------------------------------------
- . SE: 10/28/2014
### Reference Device:
- . Trade Name: ScrewVent Implant; Tapered ScrewVent Implant
- . 510(K): K013227
- Regulation Number: 21 CFR 872.3640 .
- Product Code: DZE
- 11/19/2001 ● SE:
### Reference Device:
| • Trade Name: | 3i T3 Dental Implants |
|----------------------|-----------------------|
| • 510(K): | K122300 |
| • Regulation Number: | 21 CFR 872.3640 |
| • Product Code: | DZE |
| • SE: | 01/30/2013 |
### Reference Device:
| • Trade Name: | 3i T3 Short Implants |
|----------------------|----------------------|
| • 510(K): | K150571 |
| • Regulation Number: | 21 CFR 872.3640 |
| • Product Code: | DZE |
| • SE: | 11/20/2015 |
#### VIII. Product Description:
TSX Dental Implants are manufactured from biocompatible titanium alloy. TSX Dental Implants surface consists of a Contemporary Hybrid Surface Treatment (Dual Acid Etch + MTX™ Textured Surface). The implants are available in different platform diameters and feature an internal hex connection for mating with associated Zimmer Dental and Biomet 3i internal connection restorative components.
For specific product descriptions, please refer to individual product labels.
| Implant<br>Connection | Implant Diameter<br>(mm) | Implant Platform<br>(mm) | Implant Length (mm) |
|---------------------------|--------------------------|--------------------------|---------------------|
| Conical +<br>Internal Hex | 3.1 | 2.9 | 8, 10, 11.5, 13, 16 |
| Internal Hex | 3.7 | 3.5 | 8, 10, 11.5, 13, 16 |
| | 4.1 | 3.5 | 8, 10, 11.5, 13, 16 |
| | 4.7 | 3.5 | 8, 10, 11.5, 13, 16 |
| | 5.4 | 4.5 | 8, 10, 11.5, 13, 16 |
| | 6.0 | 4.5 | 8, 10, 11.5, 13, 16 |
### TSXTM Implants:
{5}------------------------------------------------
NOTE: The Zimmer Dental and Biomet 3i Zirconia-based abutments have not been tested for compatibility with the TSX Implant and therefore should not be used.
#### IX. Indications for Use:
The TSX Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to minimally support the implant (minimum 1mm circumferential and 2mm apical) and provide good primary stability.
The 3.1mmD TSX Implants should be splinted to additional implants when used in the premolar region and should not be used in the molar region.
#### X. Summary of the Technological Characteristics:
The TSX Implants are similar to that of the predicate devices and reference devices listed below in terms of intended use, indications for use, operating principle, fundamental scientific technology, implant connection, and material. The technological characteristics such as implant diameters, length, connection and platform geometry, as well as mating abutments are similar to that of the predicate devices and reference devices, except for Zimmer Dental Zirconia-based abutments which have not been tested for compatibility with the subject devices and therefore should not be used. The subject device features a tapered design with a similar screw-type design as the predicate and reference devices. Like the Predicate Device (K072589) and Reference Devices (K013227, K142082), TSX Implants are manufactured from titanium alloy (Ti 6Al-4V ELI) per ASTM F136. The primary change from the predicate devices is the external tapered thread geometry, the hybrid surface (Dual acid-etched + MTX Surface), the new 5.4mm(D) size in 8.0, 10.0, 11.5. 13.0. and 16.0mm lengths, and the one style packaging Inner Vial for all diameters and lengths, with a Titanium sleeve based on implant size, as well as an anodized alignment pin.
The subject devices have indications for use that are similar to the predicate devices. The differences include additional information around immediate placement in extractions and natural loss of teeth. This additional language is in the indications for use for the reference devices in K142082. The intended use of the subject devices is not changed by the addition of this additional language.
A substantial equivalence comparison between the Subject Device and the Predicate and Reference Devices is provided in Table 5a below.
{6}------------------------------------------------
| Table 5a: General Device Comparison | | | | | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | Primary Predicate Device<br>Device (K072589) | Reference Device #1<br>(K142082) | Reference Device #2<br>(K122300) | Reference Device #3<br>(K013227) | Subject Device |
| | Tapered Screw-Vent<br>Implant<br>Zimmer Dental | Zimmer Eztetic, 3.1mmD<br>Dental Implant System<br>Zimmer Dental | 3i T3 Dental Implants<br>Biomet 3i | ScrewVent® and Tapered<br>ScrewVent®<br>Sulzer Dental, Incorporated<br>(Became Zimmer Dental) | TSX Implants<br>Biomet 3i |
| Indications for<br>Use | The Tapered Screw-Vent<br>Implants are designed for<br>use in the maxilla or<br>mandible for immediate<br>loading or for loading after<br>a conventional healing<br>period. Implants may be<br>used to replace one or<br>more missing teeth.<br>Immediate loading is<br>indicated when there is<br>good primary stability and<br>an appropriate occlusal<br>load. | Zimmer 3.1mmD Dental<br>Implants are designed for<br>use in the anterior maxilla<br>or mandible for immediate<br>loading or for loading after<br>a conventional healing<br>period. Implants may be<br>used to replace one or<br>more missing teeth.<br>Immediate loading is<br>indicated when there is<br>good primary<br>stability and an appropriate<br>occlusal load.<br>Zimmer 3.1mmD Dental<br>Implants may be placed<br>immediately following an<br>extraction or loss of natural<br>teeth provided there is<br>sufficient volume of<br>alveolar bone to minimally<br>support the implant<br>(minimum 1mm<br>circumferential and 2mm | 3i T3 Dental Implants are<br>intended for surgical<br>placement in the upper or<br>lower jaw to provide a<br>means for prosthetic<br>attachment in single tooth<br>restorations and in partially<br>or fully edentulous spans<br>with multiple single teeth<br>utilizing delayed or<br>immediate loading, or as a<br>terminal or intermediary<br>abutment for fixed or<br>removable bridgework,<br>and to retain overdentures.<br>3i T3 Dental Implants are<br>intended for immediate<br>function on single tooth<br>and/or multiple tooth<br>applications when good<br>primary stability is<br>achieved, with appropriate<br>occlusal loading, in order<br>to restore chewing<br>function. | The ScrewVent and<br>Tapered Screw Vent<br>Implants are intended for<br>surgical implantation in<br>edentulous or partially<br>edentulous mandibles or<br>maxillae for attachment of<br>complete denture<br>prostheses, or as a terminal<br>or intermediary abutment<br>for fixed or removable<br>bridgework, or as a<br>freestanding single tooth<br>replacement.<br>In patients with an<br>edentulous mandible,<br>ScrewVent and Tapered<br>ScrewVent Implants may<br>be loaded immediately<br>when at least four implants<br>are placed between the<br>metal foramina and rigidly<br>splinted with a bar.<br>ScrewVent® and Tapered<br>ScrewVent® implants are | The TSX Implants are<br>designed for use in the<br>maxilla or mandible for<br>immediate loading or for<br>loading after a<br>conventional healing<br>period. Implants may be<br>used to replace one or more<br>missing teeth. Immediate<br>loading is indicated when<br>there is good primary<br>stability and an appropriate<br>occlusal load. Implants<br>may be placed immediately<br>following an extraction or<br>loss of natural teeth<br>provided there is sufficient<br>volume of alveolar bone to<br>minimally support the<br>implant (minimum 1mm<br>circumferential and 2mm<br>apical) and provide good<br>primary stability.<br>The 3.1mmD TSX<br>Implants should be splinted |
| Table 5a: General Device Comparison | | | | | |
| Comparison | Primary Predicate Device<br>Device (K072589) | Reference Device #1<br>(K142082) | Reference Device #2<br>(K122300) | Reference Device #3<br>(K013227) | Subject Device |
| | Tapered Screw-Vent<br>Implant<br>Zimmer Dental | Zimmer Eztetic, 3.1mmD<br>Dental Implant System<br>Zimmer Dental | 3i T3 Dental Implants<br>Biomet 3i | ScrewVent® and Tapered<br>ScrewVent®<br>Sulzer Dental, Incorporated<br>(Became Zimmer Dental) | TSX Implants<br>Biomet 3i |
| | | apical).<br>The Zimmer 3.1mmD<br>Dental Implants should be<br>splinted to additional<br>implants when used in the<br>pre-molar region and<br>should not be used in the<br>molar region. | | placed into a prepared<br>osteotomy and once<br>stability is achieved, the<br>implant is restored with a<br>compatible restorative<br>device. | to additional implants<br>when used in the pre-molar<br>region and should not be<br>used in the molar region. |
| Operating<br>Principle | The Tapered Screw-Vent<br>Implant Dental Implants<br>are placed into a prepared<br>osteotomy and once<br>stability is achieved, the<br>implant is restored with a<br>compatible restorative<br>device. | The Tapered Screw-Vent<br>Implant Dental Implants are<br>placed into a prepared<br>osteotomy and once stability<br>is achieved, the implant is<br>restored with a compatible<br>restorative device. | The 3i T3 Dental Implants<br>achieve their intended<br>purpose based upon their<br>macro design features,<br>which maximize primary<br>stability at time of<br>placement. | The Tapered Screw-Vent<br>Implant Dental Implants<br>are placed into a prepared<br>osteotomy and once<br>stability is achieved, the<br>implant is restored with a<br>compatible restorative<br>device. | The TSX Dental Implants<br>achieve their intended<br>purpose based upon their<br>macro design features,<br>which maximize primary<br>stability at time of<br>placement. |
| Fundamental<br>Scientific<br>Technology | Endosseous Dental<br>Implants;<br>Screw-type designs | Endosseous Dental<br>Implants;<br>Screw-type designs | Endosseous Dental<br>Implants;<br>Screw-type designs | Endosseous Dental<br>Implants;<br>Screw-type designs | Endosseous Dental<br>Implants;<br>Screw-type designs |
| Material | Titanium Alloy (Ti-6Al-4V<br>ELI) Per ASTM 136 | Titanium Alloy (Ti-6Al-4V<br>ELI) Per ASTM 136 | Commercially Pure<br>Titanium (CP4)<br>Per ASTM F67 | Titanium Alloy (Ti-6Al-4V<br>ELI) Per ASTM 136 | Titanium Alloy (Ti-6Al-4V<br>ELI) Per ASTM 136 |
| Table 5a: General Device Comparison | | | | | |
| Comparison | Primary Predicate Device<br>Device (K072589) | Reference Device #1<br>(K14082) | Reference Device #2<br>(K122300) | Reference Device #3<br>(K013227) | Subject Device |
| | Tapered Screw-Vent<br>Implant<br>Zimmer Dental | Zimmer Eztetic, 3.1mmD<br>Dental Implant System<br>Zimmer Dental | 3i T3 Dental Implants<br>Biomet 3i | ScrewVent® and Tapered<br>ScrewVent®<br>Sulzer Dental, Incorporated<br>(Became Zimmer Dental) | TSX Implants<br>Biomet 3i |
| Surface<br>Topography | Single stage surface:<br>MTX<br>Grit Blast using<br>Hydroxyapatite blast<br>media | Single stage surface:<br>MTX<br>Grit Blast using<br>Hydroxyapatite blast media | Contemporary hybrid<br>surface:<br>Grit Blast using<br>Hydroxyapatite blast<br>media followed by<br>Osseotite Dual Acid<br>Etching | Single stage surface:<br>MTX<br>Grit Blast using<br>Hydroxyapatite blast media | Contemporary hybrid<br>surface:<br>hybrid surface: Osseotite<br>Dual Acid Etching surface<br>from the coronal surface to<br>approximately 1.5mm<br>down the implant collar,<br>followed by MTX<br>Grit Blast using<br>Hydroxyapatite blast media |
| Implant Body<br>Diameter<br>Range | 4.1mm | 3.1mm | 3.25mm, 4.0mm, 5.0mm<br>and 6.0mm | 3.7mm, 4.7mm, 6.0mm | 3.1mm, 3.7mm, 4.1mm,<br>4.7mm, 5.4mm (new),<br>6.0mm |
| Seating<br>Platform<br>Diameter | 3.5mm | 2.9mm | 3.25mm, 4.0mm, 5mm and<br>6mm | 3.5mm, 4.5mm, 5.7mm | 2.9mm, 3.5mm, 4.5mm |
| Implant<br>Length | 8.0mm, 10.0mm, 11.5mm,<br>13.0mm, 16.0mm | 8.0mm, 10.0mm, 11.5mm,<br>13.0mm, 16.0mm | 8.5mm, 10mm, 11.5mm,<br>13mm and 15.0mm | 8mm, 10.0mm, 13.0mm,<br>16.0mm | 8.0mm, 10.0mm, 11.5mm,<br>13.0mm, 16.0mm |
| Platform<br>Geometry | Platform switched only | Platform switched only | Platform Switched and<br>Non-Platform Switched<br>Implants | Platform switched only | Platform switched only |
| Implant<br>External<br>Geometry<br>Design | Tapered only | Tapered only | Tapered / and<br>Parallel Walled | Tapered only | Tapered only |
| Table 5a: General Device Comparison | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.