← Product Code [DZE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE) · K200906

# MINICONE Implant (K200906)

_Medentika GmbH · DZE · Oct 16, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K200906

## Device Facts

- **Applicant:** Medentika GmbH
- **Product Code:** [DZE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE.md)
- **Decision Date:** Oct 16, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

MINICONE Implants are intended for the stabilization of removable dentures.

## Device Story

MINICONE is a one-piece endosseous dental implant system designed for stabilizing removable dentures. The device consists of a 2.6 mm diameter tapered implant available in 10 mm and 12 mm lengths. It features a roughened surface (sandblasted/acid-etched) for osseointegration and a machined neck. The coronal portion incorporates an Optiloc® attachment element, coated with Titanium Nitride (TiN) for wear resistance, which connects to the denture matrix. The device is surgically placed by a dental professional in a clinical setting. By providing a stable anchor point in the jawbone, the implant allows for the secure attachment of overdentures or partial dentures, thereby improving patient chewing function and denture retention. The system is provided sterile for single use.

## Clinical Evidence

Bench testing only. Testing included insertion torque, wear testing of the Optiloc® attachment element across various angulations (0°, 10°, 20°), and dynamic fatigue strength testing per ISO 14801. Sterilization was validated per ISO 11137-1 and ISO 11137-2. Accelerated aging tests confirmed a five-year shelf life.

## Technological Characteristics

Material: Titanium Grade 5 ELI (Ti-6Al-4V) per ASTM F136. Surface: Sandblasted/acid-etched. Retention element: TiN coated. Dimensions: 2.6 mm diameter; 10/12 mm lengths. Design: One-piece implant with coronal micro-thread and 3 mm machined collar. Sterilization: Beta irradiation (min 25 kGy, SAL 10^-6).

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- Straumann Mini implants ([K191895](/device/K191895.md))

## Reference Devices

- MDI MII One-piece Implant 2.9 mm ([K081653](/device/K081653.md))
- TiUnite Implants (Nobel Biocare) ([K050705](/device/K050705.md))
- Medentika Abutment System ([K180564](/device/K180564.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 16, 2020

MEDENTiKA GmbH % Jennifer Jackson Director, Regulatory Affairs Straumann USA. LLC 60 Minuteman Road Andover, Massachusetts 01801

Re: K200906

Trade/Device Name: MINICONE Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: September 15, 2020 Received: September 16, 2020

Dear Jennifer Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K200906

Device Name:

MINICONE Implant

Indications for Use (Describe)

MINICONE are dental implants that are intended for the stabilization of removable dentures.

MINICONE Implants 02.6 mm are suitable for oral endosteal implantation in the upper and lower jaw of fully or partially edentulous patients. The implants can be placed with immediate function when good primary stability is achieved. Furthermore, they are to be used in combination with the corresponding prosthetic Optilo® matrix system and individual new or existing Optiloc® compatible overdentures or partial dentures.

For mandibular restorations, at least 4 MINICONE Implants should be placed.

For maxillary restorations, at least 6 MINICONE Implants should be placed.

Type of Use (Select one or both, as applicable) 図Prescription Use (Part 21 CFR 801 Subpart D)

□Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### MINICONE

#### 510(k) Summary 5

#### 5.1 Submitter's Contact Information

| Submitter:                          | Straumann USA, LLC (on behalf of MEDENTIKA GmbH)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052                             |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                     | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023<br>E-mail: jennifer.jackson@straumann.com |
| Prepared By &<br>Alternate Contact: | Dr. Gerhard Polzer<br>Head of Regulatory Affairs<br>MEDENTIKA GmbH<br>Phone number: +49 7229 6991216                                                                            |
| Date of Submission:                 | October 14, 2020                                                                                                                                                                |

#### 5.2 Name of the Device

| Trade Names:           | MINICONE                  |
|------------------------|---------------------------|
| Common Name:           | Endosseous Dental Implant |
| Classification Name:   | Endosseous Dental Implant |
| Regulation Number:     | 21 CFR 872.3640           |
| Device Classification: | II                        |
| Product Code(s):       | DZE                       |
| Review Branch:         | Dental                    |

#### Predicate Device(s) 5.3

Primary Predicate:

- . K191895 — Straumann Mini implants

{4}------------------------------------------------

### MINICONE

510(k) Summary

Reference Devices:

- K081653 MDI MII One-piece Implant 2.9 mm .
- K050705 TiUnite Implants (Nobel Biocare) .
- K180564 Medentika Abutment System .

#### 5.4 Device Description

MINICONE consists of 2 MINICONE tapered implants with an external diameter of 2.6 mm and lengths of 10 and 12 mm, as well as related accessories.

The implants are manufactured utilizing Titanium Grade 5 ELI material (Ti-4Al-6V) and are finished with a roughened surface (sandblasted/acid etched). The implant neck is machined, and the attachment element of the implants is acting as a retention feature for dentures. This retention feature is coated using a Titanium Nitride (TiN) to obtain a more wear resistant surface and has the Optiloc® geometry which is connected to the denture. The roughened surface, implant neck, and retention feature are shown in Figure 1.

Image /page/4/Figure/10 description: The image shows a dental implant with labels pointing to different parts of the implant. The top of the implant is labeled as the "Retention Feature". The middle of the implant is labeled as the "Implant Neck", and the bottom of the implant is labeled as the "Roughened Surface".

Figure 1 - MINICONE Implant

MINICONE Implants Ø2.6mm are suitable for oral endosteal implantation in the upper and lower jaw of fully or partially edentulous patients.

The implants can be placed with immediate function when good primary stability is achieved. The MINICONE implants are intended for the stabilization of removable dentures. The removable dentures are connected to the MINICONE implants through the incorporated Optiloc® attachment element.

#### 5.5 Intended Use

MINICONE Implants are intended for the stabilization of removable dentures.

{5}------------------------------------------------

### MINICONE

510(k) Summary

#### Indications for Use 5.6

MINICONE are dental implants that are intended for the stabilization of removable dentures. MINICONE Implants ø2.6 mm are suitable for oral endosteal implantation in the upper and lower iaw of fully or partially edentulous patients. The implants can be placed with immediate function when good primary stability is achieved. Furthermore, they are to be used in combination with the corresponding prosthetic Optiloc® matrix system and individual new or existing Optiloc® compatible overdentures or partial dentures.

For mandibular restorations, at least 4 MINICONE Implants should be placed.

For maxillary restorations, at least 6 MINICONE Implants should be placed.

### 5.7 Technological Characteristics

The technological characteristics of the subject devices are compared to the primary predicate (K191895) and reference devices in Table 1. There are no significant changes related to the implant surface treatment, fundamental operating principles, sterilization processes or procedures between the subject devices and the predicate devices cleared under K191895. The subject implant diameter is 0.2 mm larger than the primary predicate device and the thread design of the subject implant incorporates a coronal micro-thread. Both devices are made of Titanium alloys, whereas MINICONE implants are manufactured from Titanium-6Aluminum-4Vanadium (Titanium Grade 5 ELI according to ASTM F136) and the reference devices from Titanium-Zirconium (Roxolid®). Furthermore, there is no difference regarding the Optiloc® attachment element incorporated on the top of the implants.

The reference predicate K180564 is included for reference to MRI compatibility. The reference device K081653 is included for reference to the implant material Titanium grade 5 ELI (Ti-6Al-4V). The reference device K050705 is included for reference to the machined collar of the implant which can be placed up to 1.5 mm subcrestally. The machined collar of the subject device can also be placed up to 1.5 mm subcrestally. The remaining 1.5 mm of the machined collar is intended for gingival support.

The Indications for Use of the subject device are similar to the Indications for Use of the primary predicate and reference devices, K191895 and K081653. The Indications for Use for K050705 includes single tooth restorations which is not applicable for the subject device. The Indications for Use are substantially equivalent.

{6}------------------------------------------------

# MINICONE

## 510(k) Summary

| FEATURE                                   | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | PRIMARY PREDICATE<br>DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | REFERENCE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                 | REFERENCE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number                                  | K200906                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K191895                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | K081653                                                                                                                                                                                                                                                                                                                                                                                                                          | K050705                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for Use                       | MINICONE Implants are intended<br>for the stabilization of removable<br>dentures.<br>MINICONE Implants are suitable<br>for oral endosteal implantation in<br>the upper and lower jaw of fully or<br>partially edentulous patients.<br>The implants can be placed with<br>immediate function when good<br>primary stability is achieved.<br>Furthermore, they are to be used<br>in combination with the<br>corresponding prosthetic Optiloc®<br>matrix system and individual new<br>or existing Optiloc® compatible<br>overdentures or partial dentures.<br>For mandibular restorations, at<br>least 4 MINICONE Implants<br>should be placed. For maxillary<br>restorations, at least 6 MINICONE<br>Implants should be placed. | STRAUMANN MINI IMPLANTS are<br>intended for the stabilization of<br>removable dentures.<br>STRAUMANN MINI IMPLANTS are<br>suitable for oral endosteal<br>implantation in the upper and lower<br>jaw of fully or partially edentulous<br>patients.<br>The implants can be placed with<br>immediate function when good<br>primary stability is achieved.<br>Furthermore, they are to be used in<br>combination with the corresponding<br>prosthetic Optiloc® matrix system<br>and individual new or existing<br>Optiloc® compatible overdentures or<br>partial dentures.<br>For mandibular restorations, at least<br>4 STRAUMANN MINI IMPLANTS<br>should be placed. For maxillary<br>restorations, at least 6<br>STRAUMANN MINI IMPLANTS<br>should be placed. | The MII Implant is intended to<br>support single unit or multi-unit<br>restorations in both long-term and<br>temporary applications throughout<br>the maxillary and mandibular<br>arches.<br>TheMII implant is indicated for<br>immediate loading when good<br>primary stability is achieved.<br>Additionally, this device will permit<br>stability and long-term fixation of<br>upper and lower dentures in<br>edentulous cases | Nobel Biocare TiUnite® Implants are<br>root-form endosseous implants<br>intended to be surgically placed in<br>the bone of the upper or lower jaw<br>arches to provide support for<br>prosthetic devices, such as an<br>artificial tooth, in order to restore<br>patient esthetics and chewing<br>function. Nobel Biocare TiUnite®<br>Implants are indicated for single or<br>multiple unit restorations in splinted<br>or non-splinted applications. Nobel<br>Biocare TiUnite® Implants may be<br>placed immediately and put into<br>immediate function providing that<br>the initial stability requirements<br>detailed in the surgical manuals are<br>satisfied.<br>TiUnite® implants are indicated for<br>use in soft bone whenever<br>immediate or early loading is<br>applied. The TiUnite® implants are<br>preferred in these soft bone<br>indications because bone formation<br>is more rapid and greater than on<br>machined surface implants resulting<br>in better maintenance of initial<br>implant stability, faster and stronger<br>osseointegration, and higher<br>success rates. |
| Implant Diameter                          | 2.6 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 2.4 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 1.8 mm                                                                                                                                                                                                                                                                                                                                                                                                                           | 3.5, 4.3, 5.0 and 6.0 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Design                                    | One-piece implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | One-piece implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | One-piece implant                                                                                                                                                                                                                                                                                                                                                                                                                | Two-piece implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Implant Length                            | 10 and 12 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 10, 12, and 14 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 10, 13, 15, and 18 mm                                                                                                                                                                                                                                                                                                                                                                                                            | 8, 10, 11.5, 13, and 16 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Material                                  | Ti-6Al-4V (Titanium grade 5 ELI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Ti- 15Zr (Roxolid)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ti-6Al-4V                                                                                                                                                                                                                                                                                                                                                                                                                        | CP Titanium Grade 4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Retention Element<br>Coating              | TiN coated at the retention<br>element                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | TiN coated at the retention element                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | TiN coated at the retention element                                                                                                                                                                                                                                                                                                                                                                                              | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| FEATURE                                   | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | PRIMARY PREDICATE<br>DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | REFERENCE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                 | REFERENCE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| K Number                                  | K200906                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K191895                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | K081653                                                                                                                                                                                                                                                                                                                                                                                                                          | K050705                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Endosseous Surface<br>Treatment           | Blasting and acid etching                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | SLA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Blasting and acid etching                                                                                                                                                                                                                                                                                                                                                                                                        | TiUnite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Machined Collar                           | 3 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                                                                                                                                                              | 0.75 and 1.5 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Abutment-to-<br>restoration<br>connection | Optiloc® geometry<br>(Optimized Ball attachment<br>element)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Optiloc geometry<br>(Optimized ball attachment element)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | O-Ball and tapered abutment design                                                                                                                                                                                                                                                                                                                                                                                               | Internal Tri-channel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Type of<br>recommended<br>restoration     | Stabilization of removable<br>dentures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Stabilization of removable dentures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Support of single-unit or multi-unit<br>restorations for fixed and removable<br>applications                                                                                                                                                                                                                                                                                                                                     | Support of single-unit or multi-unit<br>restorations for fixed and removable<br>applications                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sterilization Method                      | Beta irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Gamma irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Gamma irradiation                                                                                                                                                                                                                                                                                                                                                                                                                | Irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{7}------------------------------------------------

# MINICONE

## 510(k) Summary

Table 1 – Comparison of subject device versus predicate/reference devices

{8}------------------------------------------------

### MINICONE

510(k) Summary

#### Performance Testing 5.8

Insertion torque testing and wear testing were conducted according to the FDA quidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and demonstrated the MINICONE implants are equivalent to the predicate and reference devices.

Wear testing is performed to evaluate the performance of the Optiloc® attachment element regarding its retention properties. The retention force loss of all Optilo® matrices (white, yellow, green, blue, red, black ring) on MINICONE implants with 0°, 10° and 20° angulation passed the acceptance criteria and demonstrated substantial equivalence.

Fatigue testing of the MINICONE implants was conducted according to ISO 14801. The dynamic fatigue strength of the MINICONE implants demonstrated substantial equivalence. No new issues of biocompatibility are raised for the subject devices. The titanium grade 5 ELI material and the TiN coating were previously cleared per K081653.

The sterilization process for the MINICONE implants was validated in accordance with ISO 11137-1:2006, Sterilization of health care products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, 2006-04-05. The subject devices are cleaned, disinfected and sterilized via beta irradiation at a dose of 25 kGy (2.5 Mrad) minimum after packaging. A sterility assurance level (SAL) of 10-6 had been validated according to the standard cited above. The validation method used was the over kill bioburden method in accordance with ISO 11137-2:2013, Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose.

The subject devices are single use and are provided sterile. The shelf life is five years. Shelf life determination has been conducted in accelerated aging tests.

#### Conclusion

The documentation submitted in this premarket notification demonstrates the MINICONE implants are substantially equivalent to the primary predicate and reference devices.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K200906](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K200906)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
