iSy Implant System implants are intended for immediate or delayed placement in the bone of the maxillary or mandibular arch. iSy Implant System Abutments are intended for use as support for crowns, bridges or overdentures. When a one-stage surgical approach is applied, the implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate.
Device Story
iSy Implant System consists of self-tapping, tapered, root-form endosseous dental implants and associated abutments; used by dentists/oral surgeons in clinical settings for tooth replacement. Implants placed in maxillary or mandibular bone; abutments support crowns, bridges, or overdentures. One-stage surgical approach allows immediate loading upon achieving primary stability. System includes gingiva formers, temporary abutments, Locator, and Esthomic abutments (various angulations). Device provides structural foundation for prosthetic restoration, restoring masticatory function and aesthetics.
Clinical Evidence
No clinical data submitted. Substantial equivalence demonstrated via non-clinical bench testing, including sterilization validation (ISO 11137-1/2, ISO 17665-1/2), biocompatibility (ISO 10993-1), and mechanical fatigue testing (ISO 14801).
Technological Characteristics
Materials: Implants (CP Ti Gr 4), Abutments/Screws (Ti-6Al-4V ELI). Design: Self-tapping, tapered, root-form. Sterilization: Radiation (ISO 11137) and Moist heat (ISO 17665). Mechanical testing: ISO 14801 static/dynamic compression-bending. No software or electronic components.
Indications for Use
Indicated for patients requiring dental implants for immediate or delayed placement in maxillary or mandibular bone, and abutments for support of crowns, bridges, or overdentures. Suitable for one-stage surgical approach with immediate loading if primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
K133510 — NEODENT IMPLANT SYSTEM · Jj Industria E Comercio DE Materiais Dentarios SA · Feb 24, 2014
K210499 — Alpha Dent Implants Dental Implants System · Alpha Dent Implants GmbH · Nov 1, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 2, 2015
Altatec GmbH c/o Ms. Linda Schulz PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K151599
Trade/Device Name: iSy® Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: II Product Code: DZE, NHA Dated: September 1, 2015 Received: September 2, 2015
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151599
Device Name
iSy® Implant System
#### Indications for Use (Describe)
iSy® Implant System implants are intended for immediate or delayed placement in the bone of the maxillary or mandibular arch. iSy" Implant System Abutments are intended for use as support for crowns, bridges or overdentures. When a one-stage surgical approach is applied, the implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate.
| Type of Use (Select one or both, as applicable) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td><td><div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Altatec GmbH iSy® Implant System
September 1, 2015
# ADMINISTRATIVE INFORMATION
| Manufacturer Name | Altatec GmbH<br>Maybachstrasse 5<br>D-71299 Wimsheim, Germany<br>Telephone: +49 7044 9445 0<br>Fax: +49 7044 9445 723 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Colleen Boswell<br>Director of RA/QA Ortho Organizers, Inc |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 (858) 792-1235<br>Fax: +1 (858) 792-1236<br>Email: lschulz@paxmed.com<br>flarson@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | iSy® Implant System |
|----------------------------|----------------------------------------------------------------------|
| Common Name | Endosseous dental implant<br>Endosseous dental implant abutment |
| Classification Name | Implant, endosseous, root form<br>Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3640 Class II |
| Product Code | DZE, NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
{4}------------------------------------------------
#### INTENDED USE
iSy Implant System implants are intended for immediate or delayed placement in the bone of the maxillary or mandibular arch. iSy Implant System Abutments are intended for use as support for crowns, bridges or overdentures. When a one-stage surgical approach is applied, the implant may be immediately loaded when good primary stability is achieved and the functional load is appropriate.
#### DEVICE DESCRIPTION
iSy Implant System implants are self-tapping, root form, tapered endosseous dental implants made of titanium. The implant has a smooth machined surface in the transgingival portion, and a Promote surface on the endosseous portion. The subject device implants are provided in two lengths (7.3 and 16 mm) and three diameters (3.8, 4.4 and 5.0 mm). The subject device includes four abutment types, gingiva former, temporary abutment, Locator® and Esthomic Abutment. Abutments are compatible with all three diameters (3.8, 4.4 and 5.0 mm). The Esthomic abutment is available with three design angulations (0°, 15°, 20°). iSy® Implant System Abutments are intended for use as support for crowns, bridges or overdentures.
#### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to : ISO 11137-1 Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose, ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices; ISO 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1; ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and static and dynamic compression-bending testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants.
Clinical data were not submitted in this premarket notification.
### EQUIVALENCE TO MARKETED DEVICE
Primary Predicate: Altatec GmbH, iSy® Implant System, K133991; Reference Predicates: Altatec GmbH, CAMLOG Implant System Modified Implants and Abutments, K083496; Altatec GmbH, CAMLOG Logfit Prosthetic System, K071213; Astra Tech AB. OsseoSpeed™ Plus. K120414: and Zest Anchors, Inc, Modification To Locator Implant Anchor, K072878; Zest Anchors, LLC, Locator RTx, K150925.
{5}------------------------------------------------
omparison of the technological characteristics of the subject device and predicate devices is shown in the Table of Substanti
quivalence below.
| | Subject Device | | Predicate Device | | |
|---------------------------------|------------------------------|----------------------------|----------------------------------------------------------|----------------------------------------|-----------------------------------------------|
| | Altatec GmbH | Altatec GmbH | Altatec GmbH | Astra Tech AB | Zest Anchors, Inc |
| | iSy® Implant System | iSy® Implant System | CAMLOG Implant System Modified<br>Implants and Abutments | OsseoSpeed™ Plus | Modification to:<br>Locator Implant<br>Anchor |
| | K133991 | K083496 | K120414 | K072878 | |
| Design | | | | | |
| Implant Length, mm | 7.3, 16 | 9, 11, 13 | 9.0, 11.0, 13.0, 16 | 6, 8, 9, 11, 13, 15, 17 | NA |
| Implant Diameter, mm | 3.8, 4.4, 5.0 | 3.8, 4.4, 5.0 | 3.3, 3.8, 4.3, 5.0, 6.0 | 3.0, 3.6, 4.2,4.8, 5.4 | NA |
| Abutment Diameter, mm | 3.8, 3.9, 4.8, 5.4, 5.9, 6.6 | 6.5 | 3.3, 3.8, 4.3, 5.0, 6.0 | 3.0, 3.6, 4.2, 4.8, 5.4 | 3.25 to 6.5 |
| Abutment Angle | Straight, 15°, 20° | Straight, up to 30° | Straight, 15°, 20° | Straight, 15°, 20°, 30° | Straight |
| Material | | | | | |
| Implant | CP Ti Gr 4 | CP Ti Gr 4 | CP Ti Gr 4 | CP Ti Gr 4 | NA |
| Abutments and<br>Abutment Screw | Ti-6Al-4V ELI; | Ti-6Al-4V ELI;<br>Zirconia | Ti-6Al-4V ELI;<br>Zirconia | Ti-6Al-4V ELI;<br>Zirconia, Gold alloy | Ti-6Al-4V ELI |
of Substantial Equivaler
o proce of this submission is System compones with be addition of wores and four
thent dosign. The addition the same elsig as the previously clased is implants and have begi
{6}------------------------------------------------
# CONCLUSION
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Overall, the subject device has the following similarities to the predicate devices:
- . has the same intended use,
- uses the same operating principle, ●
- . incorporates the same basic design,
- incorporates the same materials, and .
- has similar packaging and is sterilized using the same materials and processes. .
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.