K143758 · Clinical House Europe GmbH · DZE · Aug 7, 2015 · Dental
Device Facts
Record ID
K143758
Device Name
PerioType Rapid
Applicant
Clinical House Europe GmbH
Product Code
DZE · Dental
Decision Date
Aug 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PerioType Rapid Implants are threaded, root-form dental implants intended for use in the upper and lower jaw to support prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function to partially or fully edentulous patients for loading after healing or immediate loading provided that stability requirements are satisfied.
Device Story
PerioType Rapid is a threaded, root-form endosseous dental implant. It is surgically placed into the upper or lower jaw by a dental professional to serve as an anchor for prosthetic devices (artificial teeth). The device restores aesthetics and chewing function for patients with partial or full tooth loss. It supports both delayed loading (after healing) and immediate loading protocols, provided clinical stability requirements are met. The device functions as a mechanical support structure within the alveolar bone.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Threaded, root-form dental implant. Class II device. Intended for surgical implantation in the jawbone. No specific materials, software, or electronic components described.
Indications for Use
Indicated for partially or fully edentulous patients requiring dental implants in the upper or lower jaw to support prosthetic devices for restoration of aesthetics and chewing function. Suitable for loading after healing or immediate loading if stability requirements are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Related Devices
K072980 — NOBELPERFECT CONICAL CONNECTION HEXAGONAL IMPLANTS · Nobel Biocare AB · Dec 20, 2007
K083205 — NOBELACTIVE 8.5 MM & 18.0 MM · Nobel Biocare AB · Feb 13, 2009
K063718 — SFB AND CFB IMPLANTS - 5.0 PLATFORM · Nobel Biocare AB · Jan 5, 2007
K070007 — SPI ELEMENT PLATFORM O 4.0 MM · Thommen Medical AG · Jan 10, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus-like symbol with three human profiles facing right, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2015
Clinical House Europe Gmbh % Judi Weissinger Regulatory Consultant Weissinger Solutions. Inc. 9360 West Flamingo Road, Suite #110-553 Las Vegas, Nevada 89147
Re: K143758
Trade/Device Name: Periotype Rapid Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: June 29, 2015 Received: July 1, 2015
Dear Judi Weissinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Erin - Keith -S
Erin Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Clinical House Europe. The logo features a gray, stylized image of a Greek temple above the text "CLINICAL HOUSE EUROPE." The text is also gray and appears to be in all capital letters. The background of the image is white.
## PerioType Rapid
Premarket Notification / 510(k) Submission
Indications for Use Statement
Indications for Use
510(k) Number (if known): _K143758
ത്ര Device Name: PerioType Rapid
PerioType Rapid Implants are threaded, root-form dental implants intended for use in the upper and lower jaw to support prosthetic devices, such as artificial teeth, in order to restore aesthetics and chewing function to partially or fully edentulous patients for loading after healing or immediate loading provided that stability requirements are satisfied.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.