NEODENT IMPLANT SYSTEM

K123022 · Jjgc Industria E Comercio DE Materiais Dentarios S · DZE · May 17, 2013 · Dental

Device Facts

Record IDK123022
Device NameNEODENT IMPLANT SYSTEM
ApplicantJjgc Industria E Comercio DE Materiais Dentarios S
Product CodeDZE · Dental
Decision DateMay 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Titamax WS implant is indicated for a delayed loading protocol. The Facility implant is indicated for replacement of maxillary lateral incisors, mandibular incisors or retention of overdentures.

Device Story

Threaded, self-tapping, root-form endosseous dental implants; Morse taper abutment interface; commercially pure titanium or titanium alloy; grit-blasted and acid-etched surface. Used by dental professionals in clinical settings to support prosthetic teeth (cement- or screw-retained). Implants surgically placed in jaw bone; provide foundation for artificial teeth to restore chewing function. Benefits include improved mastication and prosthetic stability. System includes various diameters and lengths for specific clinical needs.

Clinical Evidence

No clinical data submitted. Substantial equivalence demonstrated via non-clinical bench testing, including engineering analysis, dimensional analysis, and comparative surface area analysis.

Technological Characteristics

Commercially pure titanium or titanium alloy; grit-blasted and acid-etched surface. Threaded, self-tapping, root-form design with Morse taper abutment interface. Various diameters (2.9mm to 6.0mm) and lengths (5.0mm to 16.0mm).

Indications for Use

Indicated for patients requiring dental restoration in upper or lower jaw. Titamax WS indicated for delayed loading. Facility implant indicated for replacement of maxillary lateral incisors, mandibular incisors, or retention of overdentures.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Summary Neodent Implant System ## 510(k) Summary ## MAY 1 7 2013 ## JJGC Indústria e Comércio de Materiais Dentários SA Neodent Implant System ## K123022 ### May 6, 2013 ## ADMINISTRATIVE INFORMATION Manufacturer Name: JJGC Indústria e Comércio de Materiais Dentários SA Av. Juscelino Kubitschek de Oliveira, 3291 - CIC Curitiba - PR - CEP, 81270-200, Brazil Telephone: +55 41 2169 4071 +55 41 2169 4041 Fax: Official Contact Representative/Consultant Daniel Lecuona General Manager Allison C. Komiyama, Ph.D. Linda K. Schulz, BSDH, RDH PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone: +1 (858) 792-1235 +1 (858) 792-1236 Fax: akomiyama@paxmed.com Email: lschulz@paxmed.com ## DEVICE NAME AND CLASSIFICATION Trade/Proprietary Name: Common Name: Classification Regulations: Product Code Classification Panel: Reviewing Branch: Neodent Implant System Endosseous dental implant Endosseous dental implant abutment 21 CFR 872.3640 . DZE NHA Dental Products Panel Dental Devices Branch {1}------------------------------------------------ ### 510(k) Summary #### INTENDED USE The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Titamax WS implant is indicated for a delayed loading protocol. The Facility implant is indicated for replacement of maxillary lateral incisors, mandibular incisors or retention of overdentures. #### DEVICE DESCRIPTION The Neodent Implant System implants included in this submission are threaded, self-tapping, root form, endosseous dental implants with a Morse taper abutment interface. They are designed to provide support for prosthetic teeth to restore chewing function. They are made of commercially pure titanium or titanium alloy, with a surface that is grit blasted and acid etched. In addition, multiple abutments made of titanium alloy are provided for each implant for both cement-retained and screw-retained prostheses. The CM Drive implant comes in three diameters (3.5, 4.3 and 5.0 mm), each in five lengths (8.0, 10.0, 11.5, 13.0, and 16.0 mm). The Titamax WS implant design comes in three diameters (4.0, 5.0, and 6.0 mm) with corresponding platform diameters. The 4.0 mm diameter Titamax WS implant comes in one length (6.0 mm). The 5.0 and 6.0 mm diameter Titamax WS implants come in two lengths (5.0 and 6.0 mm). The Facility implant design comes in one diameter (2.9 mm), and in five lengths (8, 10, 12, 14, and 16 mm). All abutments included in this submission are straight and have a Morse taper connection. #### EQUIVALENCE TO MARKETED DEVICE The subject device is substantially equivalent to the following predicate devices: K 101945, K 070601, K 080598, K 073343, K 092594, K 081653, K 050712, and K 0 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 1 0 1 8 The Neodent Implant System expands the currently cleared system to include new designs of implants and their corresponding cement- or screw-retained abutments. The implants of the subject Neodent Implant System have a similar design, identical material and identical surface as those cleared under K 101945. The new Titamax WS implant is similar in indications, design and dimensions to the Bicon devices cleared in K092035, K050712, and K010185. The new Facility implant is similar in indications, design and dimensions to the Intra-Lock devices cleared in K070601 and K080598, the Ace Surgical Supply devices cleared in K073343 and K092594, and the IMTEC Corporation devices cleared in K081653. The abutments of the subject Neodent Implant System have a similar design and identical material as those cleared under K101945. Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: engineering analysis, dimensional analysis and comparative surface area analysis. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. Clinical data were not submitted in this premarket notification. {2}------------------------------------------------ ## 510(k) Summary ## CONCLUSIONS The subject device and the predicate devices have the same intended use, have the same technological characteristics, and are made of the same materials. The subject device and predicate devices encompass the same range of physical dimensions, including diameter and length of the implants, and diameter, height and angle of the abutments. The subject and predicate devices are packaged in similar materials and sterilized using similar methods. Any differences in the technological characteristics do not raise new issues of safety or efficacy. The device is safe and effective for its intended use and performs as well as or better than the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered on the image and is the only element present. The text is likely a header or title for a document or website. Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three overlapping wing or feather shapes. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ## May 17, 2013 JJGC Indústria e Comércio de Materiais Dentários SA C/O Allison C. Komiyama Ph.D. Senior Regulatory Specialist PaxMed International, Limited Liability Company 11234 El Camino Real. Suite 200 SAN DIEGO CA 92130 Re: K123022 Trade/Device Name: Neodent Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: April 18, 2013 Received: April 19, 2013 Dear Dr. Komiyama: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Dr. Komiyama Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Kwame O. Ulmer -S for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number: 1123022 Device Name: Neodent Implant System Indications for Use: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Titamax WS implant is indicated for a delayed loading protocol. The Facility implant is indicated for replacement of maxillary lateral incisors, mandibular incisors or retention of overdentures. Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Ision Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Division of Antost. Dental Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...