← Product Code [DZE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE) · K122855

# TL 04.1MM RN,S, SLACTIVE TIZR 6, 8, 10, 12, 14, 16MM DENTAL IMPLANTS (K122855)

_Straumann USA · DZE · Dec 12, 2012 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K122855

## Device Facts

- **Applicant:** Straumann USA
- **Product Code:** [DZE](/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE.md)
- **Decision Date:** Dec 12, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only larqe diameter implants. In cases of fully edentulous patients. 4 or more implants must be used in immediately loaded cases.

## Device Story

Endosseous dental implants; threaded design; fabricated from Roxolid (Titanium-Zirconium alloy). Used by dental surgeons for oral rehabilitation; implanted into jawbone to support prosthetic restorations (crowns, bridges, dentures). Provides functional/esthetic tooth replacement; enables immediate or early loading if primary stability is achieved. Material change from standard titanium to TiZr alloy; maintains established surgical protocols and surface treatment (SLActive).

## Clinical Evidence

Bench testing only. Fatigue testing conducted in accordance with ISO 14801:2007(E) to validate mechanical performance of the Roxolid material compared to predicate titanium implants. No clinical data provided.

## Technological Characteristics

Threaded endosseous dental implant; material: Titanium-Zirconium (TiZr) alloy (Roxolid); surface treatment: SLActive; diameters: 4.1 mm and 4.8 mm; lengths: 6.0 mm to 16 mm; sterilization: standard established protocols.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- ITI All-in-One Implant ([K983742](/device/K983742.md))
- ITI Dental Implant System Modification ([K003552](/device/K003552.md))
- ITI Dental Implant System ([K033922](/device/K033922.md))
- ITI Dental Implants ([K003271](/device/K003271.md))
- ITI Tapered Dental Implant ([K012757](/device/K012757.md))
- Dental Implant System Modification, SLActive ([K033984](/device/K033984.md))
- SLActive New Claims ([K053088](/device/K053088.md))
- Modified TE 3.3 RN Implant ([K081419](/device/K081419.md))
- Modified Straumann Dental Implants ([K083550](/device/K083550.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122855

# NOK

straumann

DEC 1 2 2012

#### 1. Applicant's Name and Address

Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext 2513 Fax Number: 978-747-0023 Contact Person: Elaine Alan Regulatory Project Manager

- 2. Date of Submission: Sept. 17, 2012
3. Name of the Device Trade Name:

> Common Name: Classification Name: Regulation Number:

Straumann Tissue Level Ø4.1 mm and Ø4.8 mm Roxolid Dental Implants TL Ø4.1 mm and Ø4.8 mm RXD Dental Implants Implant, Endosseous, Root-form \$872.3640

- 4. Legally Marketed Device to which Equivalence is Claimed (Predicate Device)
K983742, ITI All-in-One Implant K003552, ITI Dental Implant System Modification K033922, ITI Dental Implant System K003271, ITI Dental Implants K012757, ITI Tapered Dental Implant K033984, Dental Implant System Modification, SLActive K053088, SLActive New Claims K081419, Modified TE 3.3 RN Implant K083550, Modified Straumann Dental Implants

- 5. Description of the Device
The proposed Straumann Tissue Level Ø4.1 mm and Ø4.8 mm Roxolid (TiZr) Dental Implants utilize the Straumann Titanium Zirconium material, Roxolid, to the currently cleared Straumann Tissue Level Ø4.1 mm and Ø4.8 mm Titanium dental implants.

The proposed devices are Tissue Level Roxolid dental implants with apical diameters of 4.1 mm and 4.8 mm, and coronal diameters of 4.8 mm and 6.5 mm. The implants will be available in varying lengths of 6.0.to 16mm.

> National Headquarters: 60 Minuteman Road • Andover, MA 01810 Phone 800/448 8168 • 978/747 2500 • Fax 978/747 2490 • ww.straumannusa.com

{1}------------------------------------------------

#### હ. Intended Use of the Device

Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only larqe diameter implants. In cases of fully edentulous patients. 4 or more implants must be used in immediately loaded cases.

#### 7. Technological Characteristics

The body of the proposed implants has a threaded implant body design made of Straumann's Titanium Zirconium (TiZr) Alloy material with Straumann's SLActive surface treatment. The proposed changes are a material change only. There are no changes to the surface treatment, indications for use, fundamental operating principles, or sterilization processes or procedures as a result of the proposed change. No new surgical instruments are being introduced as placement of the proposed implants will follow the established surgical protocols of the currently cleared Straumann Dental Implant Systems. The technological characteristics of the proposed devices are substantially equivalent to the currently marketed devices.

#### 8. Performance Testing

Verification and validation testing were performed to ensure that the devices subject to this 510(k) Premarket Notification function as intended and that design input matches design output. Testing included:

- 1. Performance Testing
### 510(k) Summary

Straumann USA, LLC

Straumann Tissue Level 4.1 and 4.8mm Roxolid Dental Implants

{2}------------------------------------------------

- a. Fatigue Testing in accordance to ISO 14801.2007(E), Dentistry-Implants-Dynamic fatigue test for endosseous dental implants.
#### Conclusion 9.

The results from the testing conducted demonstrated that the Straumann Tissue Level Ø4.1 mm and Ø4.8 mm Roxolid Dental Implants function as intended and met the pre-determined acceptance criteria.

The Straumann Dental Implant System is a validated system. The results of the performance bench testing and risk analysis indicate that the Straumann Tissue Level Ø4.1 mm and Ø4.8 Roxolid Dental Implants are substantially equivalent to the named predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## December 12, 2012

Ms. Elaine Alan Regulatory Project Manager Straumann USA 60 Minuteman Road ANDOVER MA 01810

Re: K122855

Trade/Device Name: Straumann Tissue Level Ø4.1 mm and Ø4.8 mm Roxolid Dental Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: November 12, 2012 Received: November 13, 2012

## Dear Ms. Alan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Kwame Q. Ulmer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K 122855

## Indications for Use

## 510(k) Number (if.known): K122855

Device Name: Straumann Tissue Level Ø4.1 mm and Ø4.8 mm Roxolid Dental Implants

Indications for Use:

Straumann® dental implants are suitable for the treatment of oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients (unless specific indications and limitations are present, as stated below). Straumann® dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). When placing implants in the posterior region, we recommend using only large diameter implants. In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of DCRH, Office of Device Evaluation (ODE)

Susan Runner **DDS, MA** 2012.12.12
10:43:20 -05'00'

(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital DIVIsion of Anesthoolone of Joseph

510(k) Number:

RE: 510(k) Straumann TL 04.1 and 04.8 Roxolid Implants

22855

Straumann USA Sept. 17, 2012 Page 1 of 1

Page 15 of 112

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K122855](https://fda.innolitics.com/submissions/DE/subpart-d%E2%80%94prosthetic-devices/DZE/K122855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
