K093352 · Hiossen, Inc. · DZE · Nov 24, 2009 · Dental
Device Facts
Record ID
K093352
Device Name
HG III FIXTURE SYSTEM
Applicant
Hiossen, Inc.
Product Code
DZE · Dental
Decision Date
Nov 24, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HGIII Fixture System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The HGIII Fixture System is for single and two stage surgical procedures. It is not for immediate load.
Device Story
HGIII Fixture System is a titanium endosseous dental implant surgically placed into the upper or lower jaw arches. Used by dental professionals in clinical settings to support dental restorations (single/multiple-unit, cemented, screw-retained, or overdenture) or fixed bridgework. Device supports single or two-stage surgical procedures. Not intended for immediate load. Provides structural foundation for prosthetic teeth; benefits patients by restoring oral function and aesthetics.
Clinical Evidence
Bench testing only; no clinical data provided. Safety and performance validated through biocompatibility and product validation testing.
Technological Characteristics
Titanium endosseous dental implant. Designed for single or two-stage surgical placement in jaw bone. No software or electronic components.
Indications for Use
Indicated for partially or fully edentulous patients (mandible/maxilla) requiring single or multiple-unit restorations (cemented, screw-retained, or overdenture) or fixed bridgework support. Contraindicated for immediate load.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
K080744 — HG II FIXTURE SYSTEM · Osstem Implant Co., Ltd. · Jul 17, 2008
K090237 — HG II FIXTURE SYSTEM · Hiossen, Inc. · Apr 17, 2009
K082213 — GS III FIXTURE SYSTEM · Osstem Implant Co., Ltd. · Oct 14, 2008
K091678 — HG II SHORT FIXTURE SYSTEM · Osstem Implant Co., Ltd. · Feb 18, 2010
K083633 — HS II SHORT FIXTURE SYSTEM · Osstem Implant Co., Ltd. · May 1, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
# Ko93352
Image /page/0/Picture/1 description: The image shows the date November 24, 2009, printed at the top. Below the date, the name "HOSSEN" is printed in a stylized font. The text is printed in black ink on a white background.
Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date : September 23, 2009
1. Company and Correspondent making the submission:
- Submitter's Name :
HiOSSEN Inc.
Mr. Patrick Lim
- Address :
85 Ben Fairless Dr. Fairless Hills PA 19030
- Contact :
2. Device :
Trade or (Proprietary) Name : Common or usual name : Classification Name :
HGIII Fixture System Dental Implant Endosseous Dental Implant 21CFR872.3640 Class II DZE
#### 3. Predicate Device :
The HGIII Fixture System, HiOSSEN Inc, K082213
- 4. Description :
The HGIII Fixture System is dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches.
The HGIII Fixture System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
The HGIII Fixture System is substantially equivalent in design, function and intended use to the HGIII Fixture System (K082213) of HiOSSEN Inc.
5. Indication for use :
The HGIII Fixture System is indicated for use in partially or fully edentulous mandibles and
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "HOSSEN" in a stylized font. The letters are bold and have a textured appearance, as if they were drawn with a rough or grainy pen. The "O" in the word has a curved line extending from the top, adding a unique design element to the overall typography.
Hiossen Inc.
85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004
www.hiossen.com
maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
The HGШ Fixture System is for single and two stage surgical procedures. It is not for immediate load.
6. Review :
The HGIII Fixture System has same material and indication for use and similar design and technological characteristics as the predicate device.
The HGIII Fixture System has been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.
7. Conclusion :
Based on the information provided in this premarket notification HiOSSEN concludes that the HGIII Fixture System is safe and effective and substantially equivalent to the predicate device as described herein.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
NOV & 4 2009
Mr. Patrick Lim Quality Assurance/Regulatory Affairs Manager HiOSSEN Incorporated 85 Ben Fairless Drive Fairless Hills, Pensylvania 19030
Re: K093352
Trade/Device Name: HGIII Fixture System Regulation Number: 21CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: October 23, 2009 Received: October 27, 2009
Dear Mr. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Mr. Lim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Wh far
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "HOSSEN" in a stylized font. The letters are all capitalized and appear to be slightly distressed or textured. There is a swooping line above the first "O" in the word, adding a decorative element to the design.
Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com
510(k) Number K 093352
Device Name : HGIII Fixture System
Indication for use : The HGI Fixture System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including ; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
> The HGIII Fixture System is for single and two stage surgical procedures. It is not for immediate load.
Prescription Use_ X (Per 21CFR801 Subpart D) OR
Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aber DDS for Dr. Kevin Mullin
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093352
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.