ZIMMER ONE PIECE IMPLANT, 3.0MM ANGLED, MODELS ZOP30A10, ZOP30A11 AND ZOP30A13
K071235 · Zimmer Dental, Inc. · DZE · May 30, 2007 · Dental
Device Facts
Record ID
K071235
Device Name
ZIMMER ONE PIECE IMPLANT, 3.0MM ANGLED, MODELS ZOP30A10, ZOP30A11 AND ZOP30A13
Applicant
Zimmer Dental, Inc.
Product Code
DZE · Dental
Decision Date
May 30, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 3.0mm One-Piece Implants are indicated for the support and retention of fixed single tooth and fixed partial denture restorations in the mandibular central and lateral incisor and maxillary lateral incisor regions of partially edentulous jaws. The 3.0mm One-Piece implant must be splinted if two or more are used adjacent to each other. The 3.0mm implant may be immediately restored with a temporary prosthesis that is not in functional occlusion.
Device Story
Zimmer One-Piece Implant, 3.0mm, Angled is an endosseous dental implant combining implant and abutment sections. Constructed of titanium alloy; features tapered design with double-lead threads for osseointegration. Abutment is pre-prepared, contoured, and angled at 17 degrees to facilitate esthetic restoration. Used by dental professionals in clinical settings for tooth replacement. Implant is surgically placed into the jawbone; provides structural support for fixed restorations. Benefits include simplified restoration process due to pre-prepared margin and angled abutment geometry.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Titanium alloy endosseous dental implant. One-piece design (implant and abutment integrated). Features tapered body, double-lead threads, and 17-degree angled abutment. Dimensions: 3.0mm diameter. Non-software device.
Indications for Use
Indicated for support and retention of fixed single tooth and fixed partial denture restorations in mandibular central/lateral incisor and maxillary lateral incisor regions of partially edentulous patients. Requires splinting if multiple implants are adjacent. May be immediately restored with non-functional temporary prosthesis.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
K083474 — ZIMMER ANGLED CONTOUR ZIRCONIA ABUTMENT · Zimmer Dental, Inc. · Dec 23, 2008
K080162 — UNO - ONE PIECE SCREW-TYPE DENTAL IMPLANT · Mis Implants Technologies , Ltd. · Apr 22, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for Zimmer Dental. The logo consists of a large, stylized letter Z inside a circle, followed by the word "zimmer" in a smaller font. Below "zimmer" is the word "dental" in an even smaller, italicized font.
| Zimmer Dental | |
|-------------------|--|
| 1900 Aston Avenue | |
Carisbad, CA 92008 760 929 4300 (ph)
760.431.7811 (fax)
510k No .: K07123
12.7-1
### MAY \$ 0 2007 Special 510(k): Device Modification PRE-MARKET NOTIFICATION 510(k)
# 510(k) SUMMARY (21CFR807.92(a))
- Submitter's Information: پسپ
| Name: | Zimmer Dental Inc. |
|----------------|--------------------|
| Address: | 1900 Aston Ave. |
| | Carlsbad, CA 92008 |
| Phone: | 760-929-4300 |
| Contact: | William Fisher |
| Date Prepared: | May 2, 2007 |
- Zimmer® One-Piece Implant, 3.0mm, Angled Device Name: ನು
Device Classification Name: Endosseous Dental Implant
Predicate Device(s): Zimmer® One-Piece Implant, 3.0mm, Straight 3. (cat, no. ZOP37S13) Zimmer® Screw Vent® Mini™ 3.3mm (cat. no. SVMB13) + Zimmer® Contour Hex-Lock™ Abutment, 3.5mm (cat. no. ZOA342A)
#### Device Description: 4.
The Zimmer One-Piece Implant, 3.0mm, Angled is a one-piece endosseous r no Limplant which is a combination of implant and abutment sections. The implant is composed of titanium alloy. The abutment portion is pre-prepared and contoured for esthetic restoration. The abutment portion is angled by 17 degrees. The abutment portion of the implant features a pre-prepared margin to facilitate the restoration process. The implant section is designed for ease of implantation and with greater surface area for osseointegration. The implant implantation and with facilitate osseointegration. In addition, the implant section is tapered with double-lead threads.
- 5. Intended Use:
The 3.0mm One-Piece Implants are indicated for the support and retention of fixed single tooth and fixed partial denture restorations in the mandibular central nxed single tooth and maxillary lateral incisor regions of partially edentulous and lateral moloor and handlery plant must be splinted if two or more are used jaws. The U.Omin One TTbe 3.0mm implant may be immediately restored with a temporary prosthesis that is not in functional occlusion.
{1}------------------------------------------------
| 1446<br><br><br>ABARANTA PARTY PERSONAL PROPERTY | .<br>.<br>. | 4-4-4-4-6-4-8-4-4 |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| .<br>1 | " - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - | 1 |
Zimmer® One-Piece, 3.7mm, Angled; 510(k) Summary cont'd:
の 2017年 10:00 【 10:00 【 10:00 【 10:00
#### Device Comparison: ଚ.
The new device is equivalent in design with Predicate. The new device is The new device to oquiration to the Predicate cleared in K052997. It differs a ulmentsional modifion to abutment portion of the new device is angled from the Predicato in the materials, general structure, and function in the by 17 degrees. The matonale, goins the same as the Predicate Device. enoosseous implant of the same bone tap, drills, and driver as Predicate. The new implant allinessence is used as listed in K052997 previously cleared. It has a matching angled implant try-in
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 3 0 2007
Mr. William Fisher Regulatory Affairs Associate Zimmer Dental, Incorporated 1900 Aston Avenue Carlsbad, California 92008
Re: K071235
Trade/Device Name: Zimmer® One-Piecc Implant, 3.0mm, Angled Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: May 02, 2007 Received: May 03, 2007
Dear Mr. Fisher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{3}------------------------------------------------
## Page 2 - Mr. Fisher
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as sct forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sudite y. Michie Oms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
K071235 510(k) Number (if known):
Device Name: Zimmer® One-Piece Implant, 3.0mm, Angled
Indications For Use:
The 3.0mm One-Piece Implants are indicated for the support and retention of fixed single tooth and fixed partial denture restorations in the mandibular central and lateral incisor and maxillary lateral incisor regions of partially edentulous jaws. The 3.0mm One-Piece implant must be splinted if two or more are used adjacent to each other. The 3.0mm implant may be immediately restored with a temporary prosthesis that is not in functional occlusion.
Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Keir Mulvey for MSR
n of Anesthesiology, General Hospital, on Control, Dental Devices
Number: K071235
Page 12.6-1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.