PaX-Reve3D Plus is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral and craniofacial anatomy on a real time basis by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the anatomic structures by acquiring 360° rotational image sequences of oral and craniofacial area for a precise treatment planning in adult and pediatric care . The device is operated and used by physicians, dentists, and x-ray technicians.
Device Story
PaX-Reve3D Plus is a dental diagnostic imaging system utilizing an extra-oral X-ray tube and CMOS digital detector to capture 360° rotational image sequences. The system performs panoramic, cephalometric, and computed tomography (CT) imaging of oral and craniofacial anatomy. Operated by physicians, dentists, or X-ray technicians in a clinical setting, the device uses computer reconstruction to transform X-ray data into cross-sectional images. These images provide diagnostic details for surgical and treatment planning. The system features a compact design and uses LDCP logic circuitry for image processing. Healthcare providers view the reconstructed images to assess anatomical structures, facilitating precise clinical decision-making for dental and craniofacial procedures.
Clinical Evidence
Bench testing only. Performance verified through electrical, mechanical, and environmental safety testing per IEC 60601-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and IEC 60601-2-44. EMC testing conducted per IEC 60601-1-2. Compliance with FDA guidance for Solid State X-ray Imaging Devices confirmed.
Technological Characteristics
Extra-oral X-ray system with CMOS digital detector. Energy source: 50-100 kV tube voltage, 2-10 mA tube current. Focal spot size: 0.5 mm. Total filtration: 2.8 mm Al. Imaging modes: panoramic, cephalometric, CT. Connectivity: DICOM 3.0 compatible. Logic: LDCP logic circuit. Standards: IEC 60601 series.
Indications for Use
Indicated for adult and pediatric patients requiring panoramic, cephalometric, or cross-sectional imaging of oral and craniofacial anatomy for treatment planning.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
# OCT 2 2 2019
# 510(k) Summary
1402124
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800.
Date
، ﺗ
July 23, 2010
## Manufacturer
Vatech Co., Ltd.
23-4, Seogu-Dong, Hwaseong-Si, Gyeonggi-Do, 445-170, Korea Republic
Tel: +82-31-679-2081
Fax: +82-31-379-9587
Contact person: Mr. Choi Hyuk-jun
#### United States Sales Representative (U.S. Designated agent)
VATECH America. 333 Meadowlands Parkway #303, Secaucus, NJ 07094, USA Tel: +832-623-2099 Fax: +713-464-8880 Contact person: Mr. Dave Kim
### Trade/Proprietary Name:
PaX-Reve3D Plus
#### Common Name:
Dental Computed Tomography X-ray System
#### Classification Name:
X-ray, Tomography, Computed, Dental (21CFR 892.1750, Product code OAS, Class2)
{1}------------------------------------------------
#### Description:
PaX-Reve3D Plus is a diagnostic imaging system which consists of multiple image acquisition modes; panorama, cephalometric, and computed tomography for implantation. Specifically designed for dental radiography of the oral and craniofacial anatomy, PaX-Reve3D Plus is equipped with extra-oral x-ray detector, panoramic radiography with an extra-oral x-ray tube, cephalometric radiography and computed tomographic radiography. The computed tomography is a system based on CMOS digital X-ray detector. CMOS CT detector is used to capture scanned images in 3D for obtaining diagnostic information for craniofacial surgery or other treatments. The device can also be operated as the panoramic and cephalometire dental x-ray system based on CMOS X-ray detector.
#### Indication for use:
PaX-Reve3D Plus is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral and craniofacial anatomy on a real time basis by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the anatomic structures by acquiring 360° rotational image sequences of oral and craniofacial area for a precise treatment planning in adult and pediatric care . The device is operated and used by physicians, dentists, and x-ray technicians.
#### Predicate Device:
| Manufacturer | : E-WOO Technology Co., Ltd. |
|---------------|---------------------------------------|
| Device | : PaX-Reve3D |
| 510(k) Number | : K090171 (Decision Date - 4/30/2009) |
#### Substantial Equivalence:
The PaX-Reve3D Plus described in this 510(k) has the same intended use and similar technical characteristics as PaX-Reve3D of E-WOO Technology Co., Ltd.
| Characteristic | Proposed<br>Vatech Co., Ltd.<br><i>PaX-Reve3D Plus</i> | Predicate<br>E-WOO Technology Co., Ltd.<br><i>PaX-Reve3D</i> |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| 510(k) number | | K090171 |
| Indications<br>for use | PaX-Reve3D Plus is a computed<br>tomography x-ray system intended<br>to produce panoramic,<br>cephalometric or cross-sectional<br>images of the oral and craniofacial<br>anatomy on a real time basis by<br>computer reconstruction of x-ray<br>image data from the same axial<br>plane taken at different angles. It<br>provides diagnostic details of the<br>anatomic structures by acquiring<br>360° rotational image sequences of<br>oral and craniofacial area for a<br>precise treatment planning in adult<br>and pediatric care . The device is<br>operated and used by physicians,<br>dentists, and x-ray technicians. | PaX-Reve3D is Computed<br>Tomography X-Ray System.<br>Real time - image acquisition.<br>Especially, advanced digital<br>imaging process allows<br>considerably efficient diagnosis,<br>all kind of information<br>management, real-time sharing of<br>image information on network.<br>Furthermore PaX-Reve3D is<br>equipped with the Flat Panel<br>Detector, CT sensor to capture a<br>3D X-ray Computed tomography<br>scanned image. |
| Performance<br>Specification | Panoramic, cephalometric and computed<br>tomography | Panoramic, cephalometric and<br>computed tomography |
| Input Voltage | 110V/230V~ | 110V/220V~ |
| Tube Voltage | 50-100 kV | 40-90kV |
| Tube Current | 2~10 mA | 2-10mA |
| Focal Spot Size | 0.5 mm | 0.5mm |
| Exposure Time | 0.5-24 s (Various) | 0.5s-24s (Various) |
| Size of Imaging<br>Volume | 15 x 15 cm<br>12 x 8 cm<br>8 x 6 cm<br>5 x 5 cm | 14 x 12 cm<br>10 x 6 cm<br>8 x 6 cm<br>5 x 5 cm |
| Slice Width | 0.1mm min. | 0.1mm min. |
| Total Filtration | 2.8mmAl | 2.8mmAl |
| | 2.5 lp/mm - CT | 2.5 lp/mm - CT |
| Pixel<br>Resolution | 5 lp/mm -- Panorama | 4.5 lp/mm -- Panorama |
| | 3.94 lp/mm - Cephalometric | 3.94 lp/mm - Cephalometric |
| | 200 $\mu$ m - CT | 200 $\mu$ m - CT |
| Pixel Size | 100 $\mu$ m - Panorama | 96 $\mu$ m - Panorama |
| | 127 $\mu$ m - Cephalometric | 127 $\mu$ m - Cephalometric |
Vatech Co., Ltd.
{2}------------------------------------------------
## 510(k) Submission -- PaX-Reve3D Plus
・
.
Vatech Co., Ltd.
·
----------------
{3}------------------------------------------------
510(k) Submission - Pax-Reve3D Plus
| Image Receptor | CT with Flat Panel Detector | CT with Flat Panel Detector |
|------------------------------|-----------------------------|-----------------------------|
| Chin Rest | Equipped Headrest | Equipped Headrest |
| Performance<br>Specification | Computed tomography | Computed tomography |
| Mechanical | Compact design | Compact design |
| Electrical | LDCP logic circuit | LDCP logic circuit |
| Software | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible |
| Anatomical<br>Sites | Maxillofacial | Maxillofacial |
The indications for use, material, form factor, performance, and safety characteristics between PaX-Reve3D Plus and the predicate device are the same. The primary difference is cosmetic, structure and component used only. Accordingly we can claim the substantially equivalence of PaX-Reve3D Plus to the predicate device.
#### Safety, EMC and Performance Data:
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 (A1+A2, 1995), IEC 60601-1-3(Ed.I, 2006), IEC 60601-2-7 (1998), IEC 60601-2-28 (Ed.1, 1993), IEC 60601-2-32 (Ed.3, 2007), and IEC 60601-2044 (Ed.2+A1, 2002) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2.
Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.
#### . Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Vatech Co., Ltd. concludes that PaX-Reve3D Plus is safe and effective and substantially equivalent to predicate device as described herein.
END
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
VATECH Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Consultant VATECH America 333 Meadowlands Paekway, #303 SECAUCUS NJ 07094
DCT 2 2 2010
Re: K102124 Trade/Device Name: Dental Computed tomography X-ray system Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: July 23, 2010 Received: July 29, 2010-
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(K) Number (if known):
102/24
Device Name: PaX-Reve3D Plus
DCT 2 2 2010
Classification: Dental Computed tomography X-ray system
Indications for Use:
PaX-Reve3D Plus is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral and craniofacial anatomy on a real time basis by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the anatomic structures by acquiring 360° rotational image sequences of oral and craniofacial area for a precise treatment planning in adult and pediatric care . The device is operated and used by physicians, dentists, and x-ray technicians.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use_ AND/OR (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation(ODE) | OIVD |
|-------------------------------------------------------|------|
|-------------------------------------------------------|------|
| (Division Sign-Off) |
|------------------------------------------------------------|
| Division of Radiological Devices |
| Office of In Vitro Diagnostic Device Evaluation and Safety |
| 510K | K102124 |
|------|---------|
|------|---------|