← Product Code [NTK](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/NTK) · K172007

# CamX Triton HD Proxi Head (K172007)

_Duerr Dental AG · NTK · Nov 22, 2017 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-b%E2%80%94diagnostic-devices/NTK/K172007

## Device Facts

- **Applicant:** Duerr Dental AG
- **Product Code:** [NTK](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/NTK.md)
- **Decision Date:** Nov 22, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1745
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The CamX Triton HD Proxi Head is a diagnostic aid for the detection of interproximal caries lesions above the gingiva and for monitoring the progress of such lesions

## Device Story

CamX Triton HD Proxi Head is a handheld dental diagnostic device used by dentists to detect interproximal caries. The device consists of a toothbrush-sized handpiece with a Proxi head, connected via USB to a PC running imaging software (e.g., DBSWIN). Operation involves placing a single-patient disposable sheath and an autoclave-able spacer over the distal end, then positioning the handpiece over the teeth. The device transilluminates tooth enamel using infrared light (850 nm). Sound enamel transmits light, while caries lesions interrupt transmission, creating bright areas on the captured image (inverted compared to some predicates). A CMOS sensor converts the optical signal into an electrical signal, which is transmitted to the PC. The software displays the image in monochrome, allowing the clinician to visualize suspected decay. This aids in clinical decision-making regarding treatment planning and monitoring of lesion progression, potentially benefiting patients through early detection of caries.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including LED illumination/output verification and image quality comparisons against the DIAGNOcam 2170 predicate. Biocompatibility testing was performed per ISO 10993-5 and ISO 10993-10. Sterilization validation was conducted per ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO 14937.

## Technological Characteristics

Handheld dental device; transillumination principle using two 850 nm infrared LEDs; CMOS image sensor; USB 5V power; autoclave-able spacer; disposable camera cover. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 80601-2-60. Connectivity via USB to PC-based software.

## Regulatory Identification

A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.

## Special Controls

*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.

1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;

2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and

3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.

## Predicate Devices

- DIAGNOcam 2170 ([K123402](/device/K123402.md))
- DIFOTI USB2.0 System ([K043068](/device/K043068.md))
- Transillumination Cable Ti2200 ([K071429](/device/K071429.md))

## Reference Devices

- Spectra Fluorescence Caries Detection Aid System ([K090169](/device/K090169.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 22, 2017

Duerr Dental AG % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114

Re: K172007

Trade/Device Name: CamX Triton HD Proxi Head Regulation Number: 21 CFR 872.1745 Regulation Name: Laser fluorescence caries detection device Regulatory Class: Class II Product Code: NTK Dated: October 5, 2017 Received: October 10, 2017

Dear Daniel Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172007

Device Name CamX Triton HD Proxi Head

Indications for Use (Describe)

The CamX Triton HD Proxi Head is a diagnostic aid for the detection of interproximal caries lesions above the gingiva and for monitoring the progress of such lesions

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary, DUERR DENTAL AG, CamX Triton HD Proxi Head K172007

This 510(k) is being submitted in accordance with the requirements of 21 CFR §807.92.

#### Date Summary Prepared: 1.

September 21, 2017

#### 2. Submitter's Identification:

DUERR DENTAL AG Höpfigheimer Strasse 17 74321 Bietigheim-Bissingen Deutschland/Germany Phone: + 49 (0) 7142 70 5-0 Fax: + 49 (0) 7142 705-500 E-Mail info@duerr.de Internet: www.duerrdental.com

## Contact:

Oliver Lange Head of Quality Management DUERR Dental AG Höpfigheimer Strasse 17 D-74321 Bietigheim-Bissingen Deutschland / Germany Phone: + 49 (0) 7142 / 705-0 Fax: + 49 (0) 7142 / 705-500 Email: lange.o@duerr.de

## U.S. Contact:

Suzanne Lucas Air Techniques, Inc. 1295 Walt Whitman Road Melville, NY 11747 Tel: 516-214-5514 Email: slucas@airtechniques.com

#### 3. Device:

Trade /Proprietary Name: Device: Regulation Description: Medical Specialty: Product Code: NTK Regulation Number: Device Class: 2

CamX Triton HD Proxi Head Caries Detector, Laser Light, Transmission Laser fluorescence caries detection device. Dental 872.1745

{4}------------------------------------------------

#### 4. Predicate Devices:

| DIAGNOcam 2170 (K123402); Kaltenbach & Voigt GmbH,                      |                                             |
|-------------------------------------------------------------------------|---------------------------------------------|
| DIFOTI USB2.0 System (K043068), ELECTRO-OPTICAL SCIENCES, INC.          |                                             |
| Transillumination Cable Ti2200 K071429), Sybron Dental Specialties, Inc |                                             |
| For all of these predicates:                                            |                                             |
| Device:                                                                 | Caries Detector, Laser Light, Transmission  |
| Regulation Description:                                                 | Laser fluorescence caries detection device. |
| Medical Specialty:                                                      | Dental                                      |
| Product Code:                                                           | NTK                                         |
| Regulation Number:                                                      | 872.1745                                    |
| Device Class:                                                           | 2                                           |

#### 5. Device Description:

The CamX Triton HD "Proxi" aids in the detection and diagnosis of proximal caries. It consists of a toothbrush-sized handpiece and a "Proxi" head. A USB cable connects the handpiece to a personal computer with PACS software such as DBSWIN to enable communication between a PC computer and the handpiece.

After a camera cover (single patient disposable sheath, K132953) is placed over the distal end, and an autoclave-able spacer is installed, the Handpiece is positioned over the teeth to be examined. The camera functions by transilluminating sound tooth enamel with infrared light. Areas that spread and reflect the light (e.g. caries lesions) show up clearly delimited bright areas. A digital camera converts the object situation into an electrical signal, sent it over USB to a computer, converted into an image (by imaging software) and presented on a monitor in monochrome colors to illustrate suspected areas of decay.

#### 6. Indications for use:

The CamX Triton HD Proxi Head is a diagnostic aid for the detection of interproximal caries lesions above the gingiva and for monitoring the progress of such lesions.

#### 7. Summary of the technological characteristics of the device compared to the predicate devices:

Duerr Dental's CamX Triton HD Proxi Head is substantially equivalent in terms of indications for use and the technology to the predicate devices which are currently in commercial distribution.

Table 1 below, summarizes the technological characteristics of CamX Triton HD Proxi Head vs. the predicate devices.

{5}------------------------------------------------

## Summary of the Technological Characteristics

| Descriptive<br>Information | CamX Triton HD Proxi Head (new<br>product)                                                                                                                                                                                                        | DIAGNOcam 2170<br>(K123402)                                                                                                                                                                                                                                                                                                                                                 | DIFOTI USB 2.0 System<br>(K043068)                                                                                                                                                                                                   | Transillumination<br>Cable TI2200<br>(K071429)                                                                                                                                                                                                                              |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use         | The CamX Triton HD Proxi Head is a<br>diagnostic aid for the detection of<br>interproximal caries lesions above<br>the gingiva and for monitoring the<br>progress of such lesions                                                                 | The DIAGNOcam 2170 is a<br>diagnostic aid for the detection  of<br>open or incipient caries  lesions<br>above the gingiva and for<br>monitoring the progress of such<br>lesions.<br>Indications:<br>- Detection of smooth surface<br>caries<br>- Detection of occlusal caries<br>- Detection of proximal caries<br>- Detection of secondary caries<br>- Detection of cracks | The DIFOTI USB 2.0 System (DIFOTI<br>System for Dental Examinations,<br>Model B) is indicated for detection<br>of frank or incipient caries lesions<br>above the gum line, and for<br>monitoring the progression of such<br>lesions. | The TI2200<br>Transillumination Cable is a<br>diagnostic aid used to<br>locate decay, calculus,<br>fracture lines, endodontic<br>orifices, cracks and fissures<br>underneath the tooth<br>surface utilizing a fiber<br>optic cable and handle<br>attached to a light source |
| Design                     | Handheld device                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                                   |
| Functional<br>Principle    | Transillumination<br>It makes use of the tooth  structure<br>which has the ability  of light<br>transmission. If the light<br>transmission is interrupted due  to<br>caries lesions a bright area  appears<br>(inverse to e.g.<br>DIAGNOcam 2170) | Transillumination<br>It makes use of the tooth structure<br>which has the ability  of light<br>transmission. If the light<br>transmission is interrupted due to<br>caries lesions a dark shadow<br>appears.                                                                                                                                                                 | Transillumination<br>It makes use of the tooth  structure<br>which has the ability  of light<br>transmission. If the light<br>transmission is interrupted due to<br>caries lesions a dark shadow<br>appears.                         | Transillumination<br>It makes use of the tooth<br>structure which has the<br>ability of light transmission.<br>If the light transmission is<br>interrupted due to caries<br>lesions a dark  shadow<br>appears.                                                              |
| Device<br>Components       | Handheld device with USB<br>cable and software                                                                                                                                                                                                    | Identical                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                            | Handheld device with<br>cable and unit.                                                                                                                                                                                                                                     |
| Descriptive<br>Information | CamX Triton HD Proxi Head (new<br>product)                                                                                                                                                                                                        | DIAGNOcam 2170<br>(K123402)                                                                                                                                                                                                                                                                                                                                                 | DIFOTI USB 2.0 System<br>(K043068)                                                                                                                                                                                                   | Transillumination<br>Cable TI2200<br>(K071429)                                                                                                                                                                                                                              |
| Light Source               | Two LED are used to generate the<br>exact wavelength being detectable<br>by the CMOS sensor.                                                                                                                                                      | An internal laser diode is used to<br>generate the exact wavelength<br>being detectable by the CCD<br>sensor.                                                                                                                                                                                                                                                               | An internal laser diode is used to<br>generate the exact wavelength<br>being detectable by the CCD<br>sensor.                                                                                                                        | External light source                                                                                                                                                                                                                                                       |
| Installation               | The computer based<br>installation enables the<br>customer to update the<br>software                                                                                                                                                              | The computer based installation<br>enables the customer to update<br>the firm- and software                                                                                                                                                                                                                                                                                 | The computer based installation<br>enables the customer to update<br>the firm- and software                                                                                                                                          | Independent (no<br>software)                                                                                                                                                                                                                                                |
| Power Source               | USB-5V                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                         |
| Compatibility              | USB connection                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                            | Specific connection                                                                                                                                                                                                                                                         |
| Compliance to<br>Standards | IEC 60601-1                                                                                                                                                                                                                                       | IEC60601-1, UL60601-1                                                                                                                                                                                                                                                                                                                                                       | IEC60601-1, UL60601-1                                                                                                                                                                                                                | Not specified                                                                                                                                                                                                                                                               |
| Autoclaveable              | Yes (distance spacer of the<br>product)                                                                                                                                                                                                           | Yes (tip of the product)                                                                                                                                                                                                                                                                                                                                                    | No (tip of the product is<br>disposable)                                                                                                                                                                                             | No                                                                                                                                                                                                                                                                          |
| Portable                   | No                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                                   |
| Software                   | The software is a computer<br>based software, which is<br>controlling:<br>- Show / display the pictures<br>- Store / save the pictures<br>- Live stream<br>- Steering of camera function                                                          | The software consists of a<br>product firmware and a<br>computer based software,<br>which is controlling:<br>- Show / display the pictures<br>- Store / save the pictures<br>- Life stream<br>- Steering of camera function                                                                                                                                                 | The software consists of a<br>product firmware and a<br>computer based software,<br>which is controlling:<br>- Show / display the pictures<br>- Store / save the pictures<br>- Life stream<br>- Steering of camera function          | No software                                                                                                                                                                                                                                                                 |
| Intended Users             | Dentist                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                                   |
| Wavelength                 | 850 nm                                                                                                                                                                                                                                            | 788 nm                                                                                                                                                                                                                                                                                                                                                                      | 670 nm                                                                                                                                                                                                                               | Not specified                                                                                                                                                                                                                                                               |
| Output power               | 2 sources (together ~4,5mW)<br>max. 4,1 mW / cm² at a distance<br>of 7 mm (spacer)                                                                                                                                                                | 2 sources (each ~2mW)<br>max. 21 mW / cm²                                                                                                                                                                                                                                                                                                                                   | 2 sources (left side ~1mW,<br>right side ~3mW) max. 41<br>mW/cm²                                                                                                                                                                     | Not specified                                                                                                                                                                                                                                                               |

{6}------------------------------------------------

{7}------------------------------------------------

#### 8. Discussion of Differences:

- i. Functional Principle: The main function is based on transillumination, same as the predicate devices. The only difference is the image information of the CamX Triton HD Proxi is displayed inverted in comparison to the predicate devices.
- ii. Light source : CamX Triton HD Proxi uses 2 infrared LEDs whereas other devices use a laser light source. The difference has geometrical and constructional/ design reasons. The predicate devices use light guides to bring the laser light to the tooth surface, in contrast to the CamX Triton HD Proxi has a direct illumination by LED in the distal end.
- iii. Wavelength: The wavelength which is used (850 nm) is a little bit higher than the wavelength used by the predicate devices. The wavelength of the CamX Triton HD Proxi is optimized for the ability of light transmission of the sound enamel and for the sensitivity of the used CMOS image sensor.
- iv. Output power: The output power for CamX Triton HD Proxi is 4,5 mW and for DIAGNOcam and DIFOTI the output power is about 4 mW (the two sources in addition, each).

#### 9. Non-Clinical Data and Performance Testing

The purpose of the evaluation was to demonstrate substantial equivalence based on similar performance between the new device, the CamX Triton HD Proxi and latest predicate device, DIAGNOcam 2170. Key performance attributes tested and compared include:

- a. LED Illumination and output
- b. Image Quality

Validation and verification test results showed that new device and the predicate device are equivalent, and that illumination and image quality of potential caries detection products are similar for both Duerr Dental AG's CamX Triton HD Proxi and KaVo's DIAGNOcam camera. Duerr Dental AG's CamX Triton HD Proxi complies with IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (edition 3) - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, IEC 60601-1-2:2014 (edition 4) - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests, and IEC 80601-2-60:2012 (first edition) Medical electrical equipment - Part 2-60: Particular requirements for basic safety and essential performance of dental equipment. Sterilization validation testing was successfully performed according to ANSI/AAMI/ISO 17665-1, Annex D and the validation approach outlined in ANSI/AAMI/ISO 14937, Annex D (Approach 3).

{8}------------------------------------------------

## 10. Biocompatibility Assessment

There are no biocompatibility issues known to be associated with any of the materials used to manufacture the CamX Triton HD Proxi. Patient contact Distance Spacer component was tested and complies with ISO 10993- 10:2002 Standard and Amendment 1, "Biological Evaluation of Medical Devices, Part 10-Tests for Irritation and Delayed-Type Hypersensitivity", and ISO 10993-5, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity. By design, many of the components of the CamX Triton HD Proxi are isolated from patient contact by a camera cover and through a distance spacer. Components that have no patient contact are:

- Hand piece housing ●
- Lens window-
- Handpiece control buttons -
- Umbilical -

Components that contact the patient are:

- Camera cover*
- 8 mm Spacer*

*These materials are currently in use and market distribution with the similar device "Spectra Fluorescence Caries Detection Aid System" device, 510K# K090169.

- 11. Clinical Data: Not required for a finding of substantial equivalence.
## 12. Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate device in terms of technology, performance and indications for use, Duerr Dental AG concludes that the CamX Triton HD Proxi is substantially equivalent to the predicate device as described herein.

The differences between the new device and the predicate device shown in the comparison table above do not raise any new questions about safety and effectiveness and so we consider it substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-b%E2%80%94diagnostic-devices/NTK/K172007](https://fda.innolitics.com/submissions/DE/subpart-b%E2%80%94diagnostic-devices/NTK/K172007)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
