NBL · Light-Induced Fluorescence Caries Detector

Dental · 21 CFR 872.1745 · Class 2

Overview

Product CodeNBL
Device NameLight-Induced Fluorescence Caries Detector
Regulation21 CFR 872.1745
Device ClassClass 2
Review PanelDental

Identification

A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.

Classification Rationale

Class II, subject to the following special controls:

Special Controls

*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter; (2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and (3) The labeling must include detailed use instructions with precautions that urge users to: (i) Read and understand all directions before using the device, (ii) Store probe tips under proper conditions, (iii) Properly sterilize the emitter-detector handpick before each use, and (iv) Properly maintain and handle the instrument in the specified manner and condition.

1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109; 2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and 3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.

Recent Cleared Devices (20 of 24)

Showing 20 most recent of 24 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K252464Surgical Microscope (OMS2360, OMS2380)Zumax Medical Co., Ltd.Apr 28, 2026SESE
K242103TRIOS Ready Tip3Shape TRIOS A/SApr 4, 2025SESE
K221275Qraycam PROAIOBIO Co., Ltd.Sep 4, 2024SESE
K231722SmarToothSmartooth Co., Ltd.May 1, 2024SESE
K223470C50SoproDec 14, 2023SESE
K221249L1P-1F (TRIOS 5)3Shape TRIOS A/SSep 16, 2022SESE
K183471IC-WHCD100 (Inspire)Good Doctors Co., Ltd.Jul 2, 2020SESE
K171007Fluorescence ModeCarl Zeiss Suzhou Co., Ltd.May 3, 2018SESE
K170822Electronic Caries DetectorOrtek Therapeutics, Inc.Sep 11, 2017SESE
K150672VistaCam iX ProofDurr Dental AGJan 27, 2016SESE
K121685SOPROCARESoproMar 12, 2013SESE
K112139THE CANARY SYSTEMQuantum Dental Technologies, Inc.Oct 26, 2012SESE
K111423CS 1600 INTRAORAL CAMERACarestream Health, Inc.Mar 2, 2012SESE
K111321REMOTE VIEW SOFTWARECariescan , Ltd.May 17, 2011SESE
K093705FUORECAMTherapeutic Technologies, Inc.Apr 23, 2010SESE
K092583SOPROLIFESoproJan 13, 2010SESE
K090598CARIESCAN PROCariescan , Ltd.Dec 15, 2009SESE
K090169SPECTRA FLUORESCENCE CARIES DETECTION DEVICEAir Techniques, Inc.Aug 18, 2009SESE
K063321D-CARIE/DETECTAR, MODELS N121, N123, N133, N134, N135 & N138Neks Technologies, Inc.Mar 30, 2007SESE
K051909DIAGNODENTKavo AmericaOct 21, 2005SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...