← Product Code [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH) · K111152

# ROTOGRAPH EVO 3D (K111152)

_Villa Sistemi Medicali S.P.A. · MUH · Jul 22, 2011 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH/K111152

## Device Facts

- **Applicant:** Villa Sistemi Medicali S.P.A.
- **Product Code:** [MUH](/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH.md)
- **Decision Date:** Jul 22, 2011
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Rotograph EVO 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dento- maxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D Images are obtained using the standard narrow beam technique. The device is to be operated and used by dentists, radiologists and other legally qualified health care professionals.

## Device Story

Rotograph EVO 3D is an extraoral digital X-ray system; utilizes cone-shaped X-ray beam and flat panel detector for 3D Cone Beam Volumetric Tomography (CBVT) and standard narrow beam technique for 2D imaging. Operated by dentists, radiologists, and qualified healthcare professionals in clinical settings. System performs image acquisition, processing, manipulation, archiving, and 3D reconstruction via computer. Output displayed on 3D viewing stations; assists clinicians in dental diagnostics, implant planning, and TMJ assessment. Benefits include enhanced visualization of dento-maxillofacial structures for improved clinical decision-making.

## Clinical Evidence

Bench testing only. No clinical data provided. Device validation and risk assessment conducted to identify design hazards; compliance with IEC 60601-1 and derivatives confirmed.

## Technological Characteristics

Extraoral digital X-ray system; utilizes flat panel detector; CBVT (Cone Beam Volumetric Tomography) and narrow beam 2D imaging; computer-based image processing and 3D reconstruction. Complies with IEC 60601-1.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- Villa Sistemi Medicali Rotograph EVO D ([K090749](/device/K090749.md))
- Planmeca ProMax 3D ([K060328](/device/K060328.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for "VALITALIA SISTEMI MEDICALI". The logo consists of the word "VALITALIA" in large, bold, sans-serif font, with the words "SISTEMI MEDICALI" stacked below it in a smaller, thinner font. Above the word "VALITALIA" is a horizontal bar with a wavy pattern.

### 510(k) Summary or 510(k) Statement

# JUL 222 2011

### KITTES

### 510 (k) Summary

 This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.

| Submitter:           | Villa Sistemi Medicali S.p.A.<br>via delle Azalee 3<br>20090 Buccinasco (MI)<br>ITALY<br>Registration # 8021091                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Paolo Casagrande Santin<br>QA Manager<br>Phone: +39 02 48859 239<br>Fax: +39 02 48859 303<br>Email: p.casagrande@villasm.com                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Designated Agent:    | Roy Torzullo<br>Del Global Technologies<br>100 Pine Aire Drive<br>Bay Shore, NY 11706<br>Registration # 9027610                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Date Prepared:       | April 18, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade Name:          | Rotograph EVO 3D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Common Name:         | Dental panoramic/tomography and cephalometric x-ray system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Name: | 872.1800 System, x-ray, extraoral source, digital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Class:               | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code:        | MUH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Devices:   | The Rotograph EVO 3D is compared with the following predicate devices:<br>• Villa Sistemi Medicali Rotograph EVO D (K090749),<br>• Planmeca ProMax 3D (K060328).                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Description: | Rotograph EVO 3D is a panoramic x-ray system utilizing digital imaging. It can be equipped<br>with a cephalostat. The device can be equipped with accessories to fulfil different diagnostic<br>needs. The images are acquired by a flat panel detector and are displayed on a monitor, and<br>image processing, manipulation, archiving, communication and 3D reconstruction (starting<br>from cross-sectional images taken using CBVT (Cone Beam Volumetric Tomography)<br>technique)) are performed via a computer.                                                                 |
| Indication for Use:  | Rotograph EVO 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral<br>structures, specifically for panoramic examinations and implantology and for TMJ studies and<br>cephalometry, and it has the capability, using the CBVT technique, to generate dento-<br>maxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat<br>panel detector, and the examined volume image is reconstructed to be viewed in 3D<br>viewing stations. |

ﻟﻢ

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo with the word "VANILLA" in bold, stylized letters at the top. Below "VANILLA" are the words "SISTEMI" and "MEDICALI" stacked on top of each other in a simple, sans-serif font. The logo appears to be for a medical systems company called Vanilla Sistemi Medicali.

2D Images are obtained using the standard narrow beam technique. The device is to be operated and used by dentists, radiologists and other legally qualified health care . Drofessionals ... .......................................................................................................................................................... Rationale for Substantial Rotograph EVO 3D has the same indication for use as the predicate devices. It shares the Equivalence: same technological characteristics as the predicate devices. Minor technological differences do not raise any new questions regarding safety or effectiveness of the devices. Safety and Effectiveness The device labeling contains operating instructions for safe and effective use of Rotograph Information: EVO 3D. The software development for this device follows documented processes for software design, verification and validation testing. Final device validation and risk assessment has been conducted, to identify potential design hazards that could cause an error or injury based on the use of this device. Appropriate steps have been taken to control all identified risks. The device has been tested for compliance to IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety. and its derivatives Rotograph EVO 3D performs the same functions in the same environment as the predicate Conclusion: devices. It shares the same technology as the predicate devices. It is based on well known technology, It is as safe and effective as the predicate devices. We believe it does not introduce any new potential safety risks and is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Paolo Casagrande Santin Ouality Assurance Manager Villa Sistemi Medicali S.p.A. via delle Azalee 3 Buccinasco, MI, 20090 ITALY

JUL 2 2 2011

Re: K111152

Trade/Device Name: Rotograph EVO 3D Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: April 18, 2011 Received: April 25, 2011

Dear Mr. Casagrande Santin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K111152

Device Name: Rotograph EVO 3D

Indications for Use:

Rotograph EVO 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specially for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dento maxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D images are obtained using the standard narrow beam technique. The device is to be operated and used by dentists, radiologists and other legally qualified healthcare professionals.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary S. Patil

510K K11152

Page 1 of 1 -

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH/K111152](https://fda.innolitics.com/submissions/DE/subpart-b%E2%80%94diagnostic-devices/MUH/K111152)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
