AZTECH 70

K984524 · Villa Sistemi Medicali S.P.A. · EHD · Jan 14, 1999 · Dental

Device Facts

Record IDK984524
Device NameAZTECH 70
ApplicantVilla Sistemi Medicali S.P.A.
Product CodeEHD · Dental
Decision DateJan 14, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesPediatric

Indications for Use

extra oral source X-ray system for dental radiographic examination and diagnosis of diseases of the teeth.

Device Story

Aztech 70 is a wall-mounted dental X-ray generator; operates at 70 kVp and 8 mA; filters X-rays via 2 mm Al equivalent. Device features a tubehead on a spring-balanced scissors arm. Operator selects exposure times via VCA timer in either automatic (based on tooth, patient type, and size) or manual mode (0.02 to 3.2 seconds). Microprocessor controls exposure, compensates for line voltage fluctuations via non-linear mathematical algorithm, and displays error codes on LED screen. Safety features include dead man switch, feedback sensors, and independent hardware backup timer. X-ray emission signaled by acoustic and optical devices. Used in dental clinics by clinicians to produce images for diagnostic purposes; aids in identifying dental diseases.

Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and technical comparison of design specifications, safety features, and operational parameters against the predicate device.

Technological Characteristics

Dental X-ray generator; 70 kVp, 8 mA; 2 mm Al filtration. Single-phase, self-rectifying high voltage generator. Microprocessor-controlled exposure timing and line voltage compensation. Safety features: dead man switch, feedback sensors, hardware backup timer. Wall-mounted with spring-balanced scissors arm. LED display for time and error codes. Software algorithm performs non-linear compensation for line voltage fluctuations.

Indications for Use

Indicated for dental radiographic examination and diagnosis of diseases of the teeth in patients requiring extra oral X-ray imaging.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a logo for Villa Sistemi Medicali. The logo consists of a stylized letter "V" in white against a black square. Below the square, the words "VILLA," "SISTEMI," and "MEDICALI" are stacked vertically in a simple, sans-serif font. The overall design is clean and modern. # Annex 1 - 510 (k) SUMMARY # 510(k) summary for Aztech 70 # Identification | Applicant | Villa Sistemi Medicali S.p.A.<br>Via delle Azalee 3,<br>20090 BUCCINASCO - Milan- Italy | |----------------------------------|----------------------------------------------------------------------------------------------------------------------| | Contact Person | dr. Francesco Attuati | | Telephone (applicant) | + 39 2 488591 | | Official Correspondent | Chicago X-Ray Systems, Inc.<br>Wheeling, IL 60090 | | Contact Person | Al SOSA - President | | Telephone<br>(contact person) | 847 - 459 3889 | | Initial distributor in the<br>US | The Aztech Group, Inc.<br>1401 Walnut Street - suite 565<br>Boulder - Colorado 80302<br>Registration Number: 1722809 | | Contact Person | Robert Padgett - President | | Manufacturing site | Villa Sistemi Medicali S.p.A.<br>Via delle Azalee 3,<br>20090 BUCCINASCO - Milan - Italy | Trade name: Aztech 70 Common name: Aztech 70 with VCA timer Classification name: according to 21 CFR 872-1800, Aztech 70 device is in Class II. Substantial equivalent device: the proposed equipment is defined as Substantially Equivalent (SE) to the Aztech 65 with CCD timer (K950667). This assumption is based on the comparison table contained on the following page. {1}------------------------------------------------ | | Aztech 65 with CCD<br>Timer (K950667) | Aztech 70 with VCA<br>Timer | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | Intended use | extra oral source X-ray<br>system for dental<br>radiographic examination<br>and diagnosis of diseases<br>of the teeth | extra oral source X-ray<br>system for dental<br>radiographic examination<br>and diagnosis of diseases of<br>the teeth | | High Voltage value | 65 kVp | 70 kVp | | Tube current | 8 mA | 8 mA | | Tube insert | CEI OCX 70-G | CEI OCX 70-G | | H.V. type: | Single phase, self<br>rectifying | Single phase, self rectifying | | X-Ray exposure<br>time control | Microprocessor<br>Controlled | Microprocessor Controlled | | Compensation of<br>Line Voltage<br>Fluctuations | Yes, automatically by<br>software algorithm | Yes, automatically by<br>software algorithm | | Safety features | Dead man command | Improved controls, with<br>feedback sensors<br>Dead man command<br>Safety backup timer (HW) | | Signaling devices | Acoustic and visual signal | Acoustic and visual signal<br>Optional remote signaling | The main differences between the proposed equipment and the substantially equivalent one are listed in the following points: - · the application of line voltage to the tubehead is microprocessor controlled - · the ON/OFF switching of the main is controlled through-out the complete cycle - · the user is requested to "enable" the emission by pushing the specific button; this feature, that can be selected by the service engineer, is particularly useful for the remote activation. - the timer is ready to accept an optional remote signaling of X-Ray emission - a safety back up timer, capable of inhibitin X-ray emission, independent from the microprocessor control, is provided in order to increase safety. # Aztech 70 description {2}------------------------------------------------ Aztech 70 is a Dental X-Ray generator; its primary use is for intra oral image receptor radiology. with a peak voltage of 70 kVp and a tube current of 8 mA. The soft X-Ray are filtered by 2 mm eq. of Al. The high voltage generator is enclosed in a plastic cover, with an apparently cylindrical aspect. The beam limiting device is formed by a circular cone with a maximum diameter of 6 cm. The tubehead is mounted on a oval shaped scissors arm, spring balanced. The scissors arm has a length of about 60 cm for each arm. The arm is mounted on an extension arm, that on the standard version has a length of 90 cm: optional configurations are available using extension arm of different lengths (75 and 60 cm). The unit is marketed as a wall mount permanent equipment. The electronic control of the VCA timer offers operator the possibility to select between automatic and manual selection of exposure times. In the automatic mode, selected exposure times according to the tooth. patient type (adult or child) and patient size to be filmed can be selected. With the manual mode of the VCA timer, the operator has the possibility to choose exposure times in steps ranging from 0.02 up to 3.2 seconds in discrete step increments. The exposure time is displayed on the timer main box by a three digit LEDs display, both for automatic and manual selection. This display is also used to show error codes, to inform the operator about the various possible faults. The microprocessor has also the job to compensate the line voltage fluctuations by changing the exposure time. This feature is accomplished by a mathematical algorithm. with a non linear function. X-Rav exposures are signaled by both acoustic and optical devices. ## Indication for use. The indication for use of the Aztech 70 is: extra oral source X-ray system for dental radiographic examination and diagnosis of diseases of the teeth.. {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 4 1999 Al Sosa Official Correspondent The Aztech Groups, Inc. 1401 Walnut Street, Suite 565 Boulder, Colorado 80302 K984524 Aztech 70 Dated: December 9, 1998 Received: December 21, 1998 Regulatory class: II 21 CFR 872.1800/Procode: 90 EHD Dear Mr. Sosa: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Re: If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 #### 510 (K) NUMBER (IF KNOWN): K984524 DEVICE NAME: AZTECH 70 ### INDICATIONS FOR USE: The Aztech 70 is intended for extra oral source X-ray system for dental radiographic examination and diagnosis of deseases of the teeth. ### (PLEASE DO NOT WRITE BELOW – CONTINUE ON ANOTHER PAGE IF NEEDED.) | | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |------------------------------------------|--------------------------------------------------------|--------------------------------------------------| | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter-Use<br>(Optional Format 1-2-96) | (Division Sign-Off) Division of Reproductive, Abdominal, ENT Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...