Abutment Design is intended as an aid to the restoration of chewing function in partially or fully edentulous mandibles and maxillae. Abutment Design is intended for use by a dental practitioner or dental laboratory staff for designing the patient specific component of a two-piece, one-piece, or hybrid dental implant abutment. The resulting abutment design is intended to be used by the manufacturer of an endosseous dental implant abutment to create the final device.
Device Story
Abutment Design is a software-only CAD tool for dental practitioners and laboratory staff. It accepts 3D scan data (DCM, STL) of teeth, impressions, or stone models. Users design patient-specific abutment components using integrated CAD tools, constrained by manufacturer-provided parameters (e.g., post height, angulation, gingival margin) for 510(k)-cleared abutment systems. The software ensures only cleared systems are used. Output is a digital file (proprietary or STL) sent to a manufacturer to produce the final physical abutment. The device operates on standard Windows PCs. By facilitating precise, patient-specific design, it aids in restoring chewing function for edentulous patients.
Clinical Evidence
No clinical data. Bench testing only, including software, hardware, and integration verification and validation performed per FDA software guidance. Cybersecurity analysis conducted per FDA guidance.
Technological Characteristics
Software-only device programmed in C# and Delphi. Operates on Windows 7/8/8.1/10 (64-bit) PCs. Requires 8GB RAM, 1GB DirectX 11 video memory, and 500GB HDD. Supports DCM and STL file formats. Connectivity via Internet/TCP-IP for server-based design parameters. No patient contact.
Indications for Use
Indicated for use by dental practitioners or laboratory staff to design patient-specific components for two-piece, one-piece, or hybrid endosseous dental implant abutments in partially or fully edentulous patients.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
October 7, 2020
3Shape A/S Jenny Axel Regulatory Affairs Specialist Holmens Kanal 7 Copenhagen 1060 DENMARK
Re: K200100
Trade/Device Name: Abutment Design Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: PNP Dated: July 13, 2020 Received: July 15, 2020
# Dear Jenny Axel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200100
Device Name Abutment Design
#### Indications for Use (Describe)
Abutment Design is intended as an aid to the restoration in partially or fully edentulous mandibles and maxillae. Abutment Design is intended for use by a dental practitioner or dental laboratory staff for designing the patient specific component of a two-piece, or hybrid dental implant abutment. The resulting abutment design is intended to be used by the manufacturer of an endosseous dental implant abutment to create the final device.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 35
# 510(K) SUMMARY K200100
#### Submitter Information
| Company Name: | 3Shape A/S |
|--------------------------------|---------------------------------------------|
| Company Address: | Holmens Kanal 7<br>DK-1060 Copenhagen K |
| Company Phone:<br>Company Fax: | +45 7027 2620<br>+45 7027 2621 |
| Contact Person: | Jenny Axel<br>Regulatory Affairs Specialist |
| Date Summary Prepared: | October 7, 2020 |
## Device Identification
| 510(k) number: | K200100 |
|-------------------------|------------------------------------|
| Trade/proprietary Name: | Abutment Design |
| Regulation Number: | 872.3630 |
| Classification: | Class 2 |
| Product Code: | PNP |
| Regulation Name: | Endosseous dental implant abutment |
#### Primary Predicate Device
The primary predicate device is 3Shape Abutment Designer™ (K151455) manufactured by 3Shape A/S.
The Abutment Design software for abutments (K200100), based on the information and supporting documentation provided, has the same intended use, scientific concept, and technical characteristics as the primary predicate device (K151455).
Both software devices are used by dental professionals for the design of endosseous dental implant abutments, and the devices ensure only FDA cleared abutment systems can be used.
Therefore, the Abutment Design software (K200100) and the predicate (K151455) are found to be similar in their intended use, supported anatomic areas and the available relevant features and functionalities.
{4}------------------------------------------------
## Indications for Use
Abutment Design is intended as an aid to the restoration of chewing function in partially or fully edentulous mandibles and maxillae. Abutment Design is intended for use by a dental practitioner or dental laboratory staff for designing the patient specific component of a two-piece, one-piece, or hybrid dental implant abutment. The resulting abutment design is intended to be used by the manufacturer of an endosseous dental implant abutment to create the final device.
#### Device Description
The Abutment Design software receives scan data containing topographical characteristics of real teeth, dental impressions, or stone models. The software provides the user with the ability to create matching endosseous dental implant abutments using computer aided design.
The Abutment Design is restricted to be used with 510(k) cleared abutment systems, and the design output from Abutment Design (the patient specific part of the abutment) must be manufactured according to the instruction of the selected abutment system.
The Abutment Design software includes design parameters from the 510(k) cleared abutment systems such as implant type, maximum and minimum dimensions (e.g., abutment post height, gingival height, angulation, gingival margin diameter, etc.). The design parameters, provided by abutment system manufacturers, for an abutment system are available via a 3Shape server when documentation of the 510(k) clearance of said implant system is presented to 3Shape.
Abutment Design has no patient contact being a software only device.
# Scientific Concept
The underlying scientific concept of the Abutment Design software is to apply digital imaging tools for computer aided design, CAD, of abutments.
The system supports the following types of digital data: DCM and STL
#### Summary of the technological characteristics
Abutment Design™ is a software only device programmed in C# and Delphi and has the following PC/laptop hardware requirements equivalent to the reference device:
| Item | Minimum Requirements<br>Abutment Design (K200100) | Minimum Requirements<br>3Shape Abutment Designer™<br>(K151455) |
|-------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------|
| OS: | Windows 7, 8, 8.1 or 10 (64-bit) | Windows 7 32-bit Professional* |
| RAM: | 8 GB | 4 GB |
| Monitor<br>Resolution: | 1920x1080 pixels | 1440x900 pixels |
| Video Card<br>Memory: | 1GB DirectX 11 | 512MB DirectX 10<br>(1GB DirectX 10) |
| Available<br>HDD Space: | 500 GB (1TB if used as a standalone<br>system or a server with the order<br>folder) | 250 GB |
{5}------------------------------------------------
| CPU: | Intel Core i7 or equivalent | Intel Core i5 or equivalent |
|------------|--------------------------------------------|--------------------------------------------|
| Network: | Network Internet connection | Internet connection |
| USB ports: | USB 2.0 port for 3Shape desktop<br>scanner | USB 2.0 port for 3Shape desktop<br>scanner |
| Mouse: | With the wheel button support | Mouse with wheel button support |
| 3D Mouse: | (Optional) 3DConnexion<br>SpaceMouse™ Pro | (Recommended)3DConnexion<br>SpaceMouse™ |
The Abutment Design software has the same intended uses and technical characteristics as the Abutment Designer™ software (K151455) also manufactured by 3Shape A/S:
| Feature name | Abutment<br>Design<br>K200100 | Abutment<br>Designer ™<br>K151455 | Identical to<br>Predicate |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 1. Graphical UI | Yes | Yes | Yes |
| 2. Windows OS platform | Yes | Yes | Yes |
| 3. Uses standard PC<br>hardware | Yes | Yes | Yes |
| 4. Digitally imports<br>topography of teeth by 3D<br>Scan | Yes | Yes | Yes |
| 5. Uses 3D CAD design<br>tools | Yes | Yes | Yes |
| 6. Patient specific<br>abutment design | Yes | Yes | Yes |
| 7. Implant Bar design | No | Yes | No, outside<br>indications for use |
| 8. Export to remote milling<br>machine by internet | Yes | Yes | No, the subject<br>device workflow<br>requires<br>instructed user<br>actions |
| 9. Network Protocol | Internet/TCP-IP | Internet/TCP-IP | Yes |
| 10. Intended users | Dental<br>practitioners<br>and dental<br>technicians | Dental<br>practitioners<br>and dental<br>technicians | Yes |
| 11. Output type | Digital<br>encrypted or<br>non-encrypted | Digital<br>encrypted or<br>non-encrypted | Yes |
| | | | |
| | proprietary or<br>.STL file only of<br>the patient-<br>specific<br>abutment<br>component, not<br>including the<br>abutment-to-<br>implant<br>connection<br>interface.<br><br>If encryption is<br>active or not<br>will solely be<br>determined by<br>the 510(k)<br>clearance of the<br>abutment<br>system. | proprietary<br>or .STL file only<br>of the patient-<br>specific<br>abutment<br>component, not<br>including the<br>abutment-to-<br>implant<br>connection<br>interface.<br><br>If encryption is<br>active or not will<br>solely be<br>determined by<br>the 510(k)<br>clearance of the<br>abutment<br>system. | |
| 12. Device submission<br>includes pre-manufactured<br>prosthetics* | No | No | Yes |
{6}------------------------------------------------
* Endosseous dental implant abutments as per 21CFR872.3630
# Nonclinical Testing
Software, hardware, and integration verification and validation testing was performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued on May 11, 2005.
The validation suite includes validation of implemented mitigations related to device hazards identified in the risk management procedures.
The cybersecurity analysis was performed in accordance with the FDA Guidance Document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", issued on October 02, 2014.
All test results have been reviewed and approved, showing the Abutment Design™ to be substantially equivalent to the primary predicate device.
## Clinical Testing
Clinical testing is not a requirement and has not been performed.
#### Conclusion
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for 3Shape. The text "3shape" is written in a sans-serif font, with the "3" being slightly separated from the rest of the word. To the right of the word is a red triangle pointing to the right, with a smaller triangle cut out of the center.
Based on a comparison of intended use, indications, scientific concept, features and technical data, and test results, the Abutment Design software is found to be as safe and effective as the primary predicate device. Therefore, Abutment Design is found to be substantially equivalent with the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.