Resolve 2.0 is recommended for use on dental restorations, external to the mouth, prior to insertion into the mouth. After applying Resolve 2.0 to cleanse the restoration, the surface of the restoration should be washed thoroughly with water, and then air dried before placement in the mouth. For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.
Device Story
Resolve 2.0 is a non-alcohol based chlorhexidine gluconate gel used as a cleaning agent for dental restorations. The device is supplied in syringes with floc-tipped applicators for precise delivery. It is applied to restorations external to the mouth, followed by water rinsing and air drying before placement. Used by licensed dentists in a clinical setting. The gel's water solubility and abrasive action facilitate the removal of contaminants from the restoration surface, potentially improving the quality of the final dental restoration placement.
Clinical Evidence
Bench testing only. Evidence includes comparative analysis of physical properties (pH, water solubility, cleansing activity), shelf-life testing (accelerated aging at 37°C), and biocompatibility testing per ISO 10993-5. Both Resolve 2.0 and the predicate exhibited 'Moderate' cytotoxicity reactivity zones up to 1.0 cm.
Technological Characteristics
Non-alcohol based gel containing chlorhexidine gluconate and silicate (sand). pH 6. Water soluble. Thin liquid consistency. Delivered via syringe and floc-tipped applicator. Biocompatibility evaluated per ISO 10993-5. Shelf life established via accelerated aging at 37°C.
Indications for Use
Indicated for cleansing dental restorations external to the mouth prior to insertion. For use by licensed dentists only.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
Predicate Devices
Consepsis, Ultradent Products (k925375)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a central emblem. The emblem consists of a stylized representation of three human profiles facing right, stacked on top of each other, symbolizing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2015
Denali Corporation Jan G. Stannard President 134 Old Washington Street Hanover, Massachusetts 02339
Re: K152322
Trade/Device Name: Resolve 2.0 Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity varnish Regulatory Class: Class II Product Code: LBH Dated: November 4, 2015 Received: November 9, 2015
Dear Ms. Jan G. Stannard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
510 (k) Number 152322
Device Name
Resolve 2.0
Indications for Use:
Resolve 2.0 is recommended for use on dental restorations, external to the mouth, prior to insertion into the mouth. After applying Resolve 2.0 to cleanse the restoration, the surface of the restoration should be washed thoroughly with water, and then air dried before placement in the mouth.
For Use only by a Licensed Dentist. Rx Use Only. Non-sterile.
Please do not write below this line. Continue on another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
or
区
Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use
{4}------------------------------------------------
## denali corporation 134 Old Washington Street
Hanover, MA 023339 USA
# T 781.826.9190
F 781.826.4465
denalicorporation.com
### 510 (k) Summary
December 10, 2015
| ADDRESS | DENALI CORPORATION<br>134 Old Washington Street<br>Hanover, MA 02339-1629 | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| OWNER/CONTACT PERSON | Dr. Jan G. Stannard<br>TEL:<br>FAX: | 781-826-9190<br>781-826-4465<br>j.stannard@denalicorporation.com |
| TRADE NAME | Resolve 2.0 | |
| COMMON NAME | External Cleansing Solution, Cavity Varnish | |
| CLASSIFICATION NAME | Cavity Varnish, Class II (21 CFR 872.3260, Product Code PME) | |
| REGISTRATION | 3006367836 | |
| PRIMARY PREDICATE DEVICE | Consepsis, Ultradent Products (k925375) | |
| EQUIVALENCE | The predicate product has been found substantially equivalent<br>under the 510(k) premarket notification process as Class II Dental<br>Devices under CFR PME 872.3260. | |
| DEVICE DESCRIPTION | Resolve 2.0 is a thin, non-alcohol based gel containing chlorhexidine gluconate.<br>This thin gel because of its water solubility, abrasive action and low viscosity when<br>applied to dental restorations, acts as a cleaning agent. This gel is provided in easy<br>to use syringes and floc-tipped applicator tips to precisely deliver the gel<br>to the intended areas.<br>For Use only by a Licensed Dentist. Rx Use Only. Non-sterile. | |
| INTENDED USE | Resolve 2.0 is recommended for use on dental restorations, external to the<br>mouth, prior to insertion into the mouth. After applying Resolve 2.0 to cleanse the<br>restoration, the surface of the restoration should be washed thoroughly with<br>water, and then air dried before placement in the mouth.<br>For Use only by a Licensed Dentist. Rx Use Only. Non-sterile. | |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Denali Corporation. The logo features a blue and black geometric shape with a blue sphere and a curved line around it. Next to the logo is the text "denali corporation" in gray. Below the company name is the address "134 Old Washington Street, Hanover, MA 02339 USA".
T 781.826.9190 F 781.826.4465 denalicorporation.com
#### TECHNOLOGICAL CHARACTERISTICS SUMMARY
Resolve 2.0 has the same technological characteristics as Consepsis including: design, composition, biocompatibility, performance, ageing, intended use, physical characteristics, including pH, shelf life, water solubility, and cleansing activity, as summarized in the Table below.
| PROPERTY | Resolve 2.0 | Consepsis* |
|-----------------------------------------------------|----------------------------------------|----------------------------------------|
| Cleansing Activity | Yes | Yes |
| Contains Chlorhexidine gluconate | Yes | Yes |
| Water Soluble | Yes | Yes |
| pH | 6 | 6 |
| Contains silicate (sand) | Yes | Yes |
| Consistency | Thin liquid | Thin liquid |
| Pigmented | Yes | Yes |
| Ageing/Shelf Life Test - Accelerated Ageing at 37°C | PASS, 2.0+ year shelf life established | PASS, 2.0+ year shelf life established |
BIOCOMPATIBILITY
Resolve 2.0 was found to be biocompatible, and is consistent in formulation to Consepsis, in its limited intended uses, biocompatibility, and properties, Resolve 2.0 was evaluated per ISO 10993-5 "Biological Evaluation of Medical Devices". Resolve 2.0 and Consepsis according to ISO 10993-5 both had cytotoxicity values of "Moderate", exhibiting a reactivity zone extending from the specimen up to 1.0 cm. According to these results Resolve 2.0 was found to be equivalent in biocompatibilty compared to the predicate product Consepsis.
SUBSTANTIAL Resolve 2.0 is substantially equivalent in design, composition, biocompatibility EQUIVALENCE according to ISO 10993, performance, ageing, limited intended uses, and safety DETERMINATION and effectiveness to Consepsis. This assessment is based upon a comparison AND SUMMARY of the composition and physical properties of water solubility, pH, shelf-life and cleansing activity to Consepsis, as well as Biocompatibility evaluation. The results of these tests are consistent with values of the predicate product.
CONCLUSIONS Resolve 2.0 has been found to be substantially equivalent in design, composition, ageing, biocompatibility, performance, with limited intended uses to Consepsis.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.