Temporary anchorage screws are intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of the teeth. It is used temporarily and is removed after orthodontic treatment has been completed.
Device Story
Self-tapping orthodontic anchorage screw; provides fixed anchorage point for orthodontic/pre-prosthetic appliances; facilitates tooth movement. Used in dental practices by qualified dentists. Device implanted temporarily (avg 6 months); removed after treatment. Input: manual surgical placement into bone. Output: mechanical anchorage point. Benefits: stable support for orthodontic appliances. Supplied non-sterile; requires steam sterilization before use.
Clinical Evidence
Bench testing only. No clinical or animal studies performed. Biocompatibility testing (ISO 10993-5, -10, -11) and mechanical performance testing (ASTM F543) demonstrate equivalence to predicate.
Technological Characteristics
Materials: Stainless Steel (ISO 5832-1/ASTM F138) and Titanium Alloy (ISO 5832-3/ASTM F136). Self-tapping screw design. Dimensions: 1.4-2mm diameter; 6-14mm length. Non-active, non-sterile, steam-sterilizable. Mechanical performance per ASTM F543.
Indications for Use
Indicated for patients requiring orthodontic tooth movement; provides fixed anchorage point for orthodontic appliances. Used temporarily and removed post-treatment.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Bomei Co., Ltd. % Chih-Hao Kao General Manager Voler Biotech Consulting Co., Ltd. No. 3-1, Ln. 58, Hejiang St., Zhongshan Dist., Taipei City 10480 TAIWAN
Re: K202278
Trade/Device Name: OBS Anchorage Screw, Biokey Anchorage Screw Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: OAT Dated: December 2, 2020 Received: December 11, 2020
Dear Chih-Hao Kao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202278
Device Name
OBS Anchorage Screw, Biokey Anchorage Screw
Indications for Use (Describe)
Temporary anchorage screws are intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of the teeth. It is used temporarily and is removed after orthodontic treatment has been completed.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
# OBS Anchorage Screw/Biokey Anchorage Screw
#### 1. Submitter
#### (a) Contact person
Name: Howard Kao Title: General Manager E-mail: leon.howard01@gmail.com Office address: 4F, No. 1492-2, Chunrih Rd., Taoyuan Dist., Taoyuan City, 33051, TAIWAN Tel: (O) +886-933-872-972 Date prepared: January 6, 2021
#### Manufacturer information (b)
| Submitter name: | BOMEI CO., LTD. |
|-------------------------|----------------------------------------------------------------------------|
| Submitter address: | 4F, No. 1492-2, Chunrih Rd., Taoyuan Dist.,<br>Taoyuan City, 33051, TAIWAN |
| Submitter phone number: | +886-3-355-4989 |
| Submitter fax number: | +886-3-346-0897 |
#### 2. Device Name and Classification
| Product Name: | OBS Anchorage Screw, Biokey Anchorage Screw |
|-----------------------|---------------------------------------------|
| Classification Name: | Implant, Endosseous, Orthodontic |
| Common or Usual Name: | Endosseous dental implant |
| Classification Panel: | Dental |
| Regulation Number: | 21 CFR 872.3640 |
| Device Class: | Class 2 |
| Product Code: | OAT |
{4}------------------------------------------------
#### 3. Predicate Device(s)
K152297 OBS Anchorage Screw
#### 4. Device Description
The OBS/Biokey Anchorage Screw, consisting of stainless steel (SUS-316LVM) and titanium alloy (Ti-6Al-4V ELI), is a self-tapping screw with various sizes for applications in the orthodontic field. It is intended to serve as a fixed anchorage point for the attachment of orthodontic and pre-prosthetic appliances, in order to facilitate the orthodontic movement of teeth. The OBS/Biokey Anchorage Screw and associated accessories are supplied non-sterile and should be sterilized before use. The devices are used temporarily with the intention to be removed after orthodontic treatment. The average temporary implantation period for the anchorage screw is six months. Screws are intended for single-use only.
The purpose of this submission is to implement six additional specification models: BM, AT, AY, CH, FH, and HH. The BD model was previously cleared under K152297.
The intended use, materials, manufacturing process, and sterilization method of the OBS/Biokey Anchorage Screw for BD, BM, AT, AY,CH, FH, and HH are the same. The differences are the type and dimensions (diameter and length). Different kinds of OBS/Biokey Anchorage Screw are manufactured to meet market needs. The BD model is available in diameters of 1.5. 2mm and lengths of 6. 8. 10. 12. 14mm. The BM and AT models are available in diameters of 1.6, 2mm and lengths of 6, 8, 10, 12, 14mm. The AY model is available in diameter of 1.4mm and lengths of 6, 8, 10mm. The CH, FH, and HH models are available in diameters of 1.4, 1.6, 2mm and lengths of 6, 8, 10, 12, 14mm.
#### 5. Indications for Use
Temporary anchorage screws are intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of the teeth. It is used temporarily and is removed after orthodontic treatment has been completed.
{5}------------------------------------------------
### 6. Comparison of technological characteristics with the predicate device
We compared the modified device with the existing device (K152297) in the Comparison Table below.
| Device<br>Name | OBS Anchorage Screw<br>(Existing) | OBS Anchorage Screw/Biokey Anchorage<br>Screw<br>(Modified) | Comparison |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Applicant | BOMEI | BOMEI | Identical |
| 510(k) Number | K152297 | K202278 | - |
| Regulation No. | 21 CFR 872.3640 | 21 CFR 872.3640 | Identical |
| Product Code | OAT | OAT | Identical |
| Trade Name | OBS Anchorage Screw | OBS Anchorage Screw<br>Biokey Anchorage Screw | Different<br>Add new trade name |
| Material | Stainless Steel (ISO 5832-1:2007 /<br>ASTM F138-13)<br>Titanium Alloy (ISO 5832-3:1996 / ASTM<br>F136-13) | Stainless Steel (ISO 5832-1:2007 /<br>ASTM F138-13)<br>Titanium Alloy (ISO 5832-3:1996 / ASTM<br>F136-13) | Identical |
| Device<br>Name | OBS Anchorage Screw<br>(Existing) | OBS Anchorage Screw/Biokey Anchorage<br>Screw<br>(Modified) | Comparison |
| Intended<br>Use | Temporary anchorage screws are intended to<br>provide a fixed anchorage point for<br>attachment of orthodontic appliances to<br>facilitate the orthodontic movement of the<br>teeth.<br>It is used temporarily and is removed after<br>orthodontic treatment has been completed. | Temporary anchorage screws are intended to<br>provide a fixed anchorage point for<br>attachment of orthodontic appliances to<br>facilitate the orthodontic movement of the<br>teeth.<br>It is used temporarily and is removed after<br>orthodontic treatment has been completed. | Identical |
| Model | BD | BD, BM, AT, AY, CH, FH, HH | Different<br>Add new six specifications |
| Sterility | Supplied non-sterile; steam sterilize before<br>use | Supplied non-sterile; steam sterilize before<br>use | Identical |
| Type | Square collar, Mushroom<br>head, None/Round/Slot hole<br>(BD) | Square collar, Mushroom<br>head, None/Round/Slot hole<br>(BD)<br>Hexagonal collar, Mushroom<br>head, None/Round/Slot hole<br>(BM)<br>Hexagonal collar, Oval head,<br>None/Round/Slot hole (AT)<br>Hexagonal collar, Slight head,<br>None/Round hole (AY) | Different |
| | | Cap head, None/Round/Slot<br>hole (CH)<br>Flat head, None/Round/Slot<br>hole (FH)<br>Hook head, None/Round/Slot<br>hole (HH) | |
| Length (mm) | 8, 10, 12, 14 | 6, 8, 10, 12, 14 | Different<br>Add length of 6mm |
| Diameter (mm) | 1.5, 2 | 1.4, 1.5, 1.6, 2 | Different<br>Add diameters of 1.4mm,<br>1.6mm |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Device<br>Name | OBS Anchorage Screw<br>(Existing) | OBS Anchorage Screw/Biokey Anchorage<br>Screw<br>(Modified) | Comparison |
|------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Non-thread<br>region<br>(mm) | 0mm | 0, 2, 4mm | Different<br>Add non-thread region<br>lengths of 2mm, 4mm<br>(except AY)<br>In the non-thread region,<br>there is no pitch on the<br>screw. |
| Energy sources | Non-active implantable devices | Non-active implantable devices | Identical |
| Where used | Dental practices | Dental practices | Identical |
| Target<br>population | Professional use only - qualified dentists<br>Strictly reserved to specialized and trained<br>users. | Professional use only - qualified dentists<br>Strictly reserved to specialized and trained<br>users. | Identical |
{9}------------------------------------------------
| Device<br>Name | OBS Anchorage Screw<br>(Existing) | OBS Anchorage Screw/Biokey Anchorage<br>Screw<br>(Modified) | Comparison |
|----------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------|
| Biocomp.<br>Testing | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11 | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11 | Identical |
| Sterilization<br>Validation<br>Testing | ISO 11737-1, ISO 11737-2, ISO 17665-<br>1, ISO 17665-2, AAMI/ANSI ST79<br>for steam sterilization | ISO 11737-1, ISO 11737-2, ISO 17665-<br>1, ISO 17665-2, AAMI/ANSI ST79<br>for steam sterilization | Identical |
| Mechanical<br>Performance<br>Testing | ASTM F543 | ASTM F543 | Identical |
{10}------------------------------------------------
| Device<br>Name | OBS Anchorage Screw<br>(Existing) | OBS Anchorage Screw/Biokey Anchorage<br>Screw<br>(Modified) | Comparison |
|----------------|-----------------------------------|-------------------------------------------------------------|------------|
|----------------|-----------------------------------|-------------------------------------------------------------|------------|
{11}------------------------------------------------
#### 7. Performance data
- Non-clinical performance testing
The following performance data were provided in support of the substantial equivalence determination.
- Biocompatibility testing in accordance to ISO 10993-5, ISO 10993-10, and ISO 10993-11 and referenced from K152297:
- Cytotoxicity: ISO 10993-5 Biological evaluation of medical . devices - Part 5: Tests for In Vitro cytotoxicity
- Sensitization and Irritation: ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- . Systemic toxicity: ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- Sterilization validating testing in accordance to ISO 11737-1, ISO 11737-2, ISO 17665-1, ISO 17665-2, and AAMI/ANSI ST79 for steam sterilization and referenced from K152297:
- · ISO 11737-1 Sterilization of medical devices Microbiological Methods Part 1: Determination of the population of microorganisms on product, 2ed
- ISO 11737-2 Sterilization of medical devices Microbiological Methods Part 2: Tests of sterility performed in the definition, validation, and maintenance of a sterilization process
- ISO 17665-1 Sterilization of health care products -- Moist Heat --Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices
- ISO 17665-2 Sterilization Of Health Care Products Moist Heat -Part 2: Guidance On The Application Of ISO 17665-1
- AAMI/ANSI ST79 Comprehensive Guide to Steam Sterilization . and Sterility Assurance in Health Care Facilities
- ������������������������������������������������������������������������������������������������������������������������������������������������������������������������������ Mechanical Performance Testing has been performed with the predicate devices in accordance to ASTM F543 used comparatively to demonstrate VOL004_8
{12}------------------------------------------------
substantial equivalence.
- · ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws
- Animal Studies .
- ・ None
- Clinical Studies .
- None .
The results of this testing indicate that the OBS/Biokey Anchorage Screw met acceptance criteria and is substantially equivalent to the predicate device.
#### 8. Conclusions
In comparison to the legally marketed device, the OBS/Biokey Anchorage Screw has the same intended use, materials, manufacturing process, and sterilization method. The differences between the subject device and the predicate device are the type and dimensions (diameter and length).
However, testing data such as mechanical performance testing provided in the submission show that these differences do not raise issues in performance. Based on the information provided in this premarket notification, BOMEI concludes that the OBS/Biokey Anchorage Screw is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.