Anorganic Bone Mineral with Collagen in Delivery Applicator
K171008 · Collagen Matrix, Inc. · NPM · Oct 30, 2017 · Dental
Device Facts
Record ID
K171008
Device Name
Anorganic Bone Mineral with Collagen in Delivery Applicator
Applicant
Collagen Matrix, Inc.
Product Code
NPM · Dental
Decision Date
Oct 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Anorganic Bone Mineral with Collagen in Delivery Applicator is intended for use in dental surgery. The products may be used in surgical procedures such as: - Augmentation or reconstructive treatment of alveolar ridge - Filling of infrabony periodontal defects - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of maxillary sinus floor - Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
Device Story
Device consists of porous bovine-derived anorganic bone mineral and bovine type I collagen; pre-loaded into plastic delivery applicator. Used in dental surgery by clinicians to deliver bone graft material to defect sites. Applicator facilitates placement in difficult-to-reach areas; plunger force 4.1-4.5 N. Device is single-use, non-pyrogenic, and gamma-irradiated. Provides scaffold for bone regeneration; aids in ridge preservation and defect filling. Clinical benefit derived from ease of delivery and material biocompatibility.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (cytotoxicity, irritation, sensitization), sterilization validation (ISO 11137-1), and functionality testing of the delivery applicator. Animal performance data from predicate devices (canine intraoral, rabbit femoral condyle, rat subcutaneous) supports the bone mineral and collagen components.
Technological Characteristics
Porous bovine bone mineral and bovine Achilles tendon type I collagen. Delivery applicator: plastic. Sterilization: gamma irradiation (SAL 10^-6). Biocompatibility per ISO 10993-12/18 and USP <661>. Plunger force: 4.5 N ± 1.3 N (preformed) or 4.1 N ± 1.1 N (moldable).
Indications for Use
Indicated for dental surgery patients requiring alveolar ridge augmentation/reconstruction, filling of infrabony periodontal defects, defects post-root resection/apicoectomy/cystectomy, extraction sockets, maxillary sinus floor elevation, or peri-implant defects in conjunction with GTR/GBR procedures.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the next line.
October 30, 2017
Collagen Matrix, Inc. Danielle Lindner Senior Regulatory Affairs Associate II 15 Thornton Road Oakland, New Jersey 07436
Re: K171008
Trade/Device Name: Anorganic Bone Mineral with Collagen in Delivery Applicator Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: September 26, 2017 Received: September 28, 2017
Dear Danielle Lindner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mary S. Runner -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Anorganic Bone Mineral with Collagen in Delivery Applicator
Indications for Use (Describe)
Anorganic Bone Mineral with Collagen in Delivery Applicator is intended for use in dental surgery. The products may be used in surgical procedures such as:
- · Augmentation or reconstructive treatment of alveolar ridge
- · Filling of infrabony periodontal defects
- · Filling of defects after root resection, apicoectomy, and cystectomy
- · Filling of extraction sockets to enhance preservation of the alveolar ridge
- · Elevation of maxillary sinus floor
- · Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- · Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|--|--|--|--|
| | | | | |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
K171008
# Applicant Information
| Applicant Name: | Collagen Matrix, Inc. |
|---------------------|------------------------------------------------------------------------------------------------------------|
| Owner Operator No.: | 9043463 |
| Address: | 15 Thornton Road<br>Oakland, New Jersey 07436 |
| Telephone: | (201) 405-1477 |
| Fax: | (201) 405-1355 |
| Contact Person: | Peggy Hansen<br>Senior VP, Quality and Regulatory Affairs<br>(201) 405 -1477<br>phansen@collagenmatrix.com |
| Date Prepared: | September 25, 2017 |
## Name of Device
| Trade Name: | Anorganic Bone Mineral with Collagen in Delivery Applicator |
|----------------------|---------------------------------------------------------------------------------------|
| Common Name: | Bone Grafting Material |
| Classification Name: | Bone Grafting Material, Animal Source<br>21 CFR 872.3930<br>NPM<br>Class II<br>Dental |
# Legally Marketed Devices to Which Substantial Equivalence is Claimed
| Predicate Device(s) | Anorganic Bone Mineral |
|---------------------|---------------------------------------------|
| | Anorganic Bone Mineral with Collagen |
| | Anorganic Bone Mineral with Collagen Blocks |
K043034
{4}------------------------------------------------
| Reference Predicates: |
|-------------------------------------------------------|
| Porcine Anorganic Bone Mineral in Delivery Applicator |
| K162158 |
| 21 CFR 872.3930 |
| Bone Grafting Material |
| Class II |
| NPM |
| Geistlich Bio-Oss Pen® |
| K120601 |
| 21 CFR 872.3930 |
| Bone Grafting Material |
| Class II |
| NPM |
### Description of the Device
Anorganic Bone Mineral with Collagen in Delivery Applicator are porous bone mineral matrices consisting of calcium phosphate derived from bovine bone with the addition of bovine type I collagen, pre-loaded into a delivery applicator for ease of placement in the defect site. The anorganic bone mineral component is produced by removal of the organic components from bovine bone. The type I collagen component is derived from bovine Achilles tendon. Anorganic Bone Mineral with Collagen in Delivery Applicator is sterilized by gamma irradiation. The products are non-pyrogenic and for single use only.
#### Indications for Use
Anorganic Bone Mineral with Collagen in Delivery Applicator is intended for use in dental surgery. The products may be used in surgical procedures such as:
- . Augmentation or reconstructive treatment of alveolar ridge
- Filling of infrabony periodontal defects
- Filling of defects after root resection, apicoectomy, and cystectomy ●
- . Filling of extraction sockets to enhance preservation of the alveolar ridge
- . Elevation of maxillary sinus floor
- Filling of periodontal defects in conjunction with products intended for Guided Tissue . Regeneration (GTR) and Guided Bone Regeneration (GBR)
- . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
{5}------------------------------------------------
## Summary/Comparison of Technical Characteristics
The delivery applicator has been designed to deliver the anorganic bone mineral products to the intended treatment area that may be more difficult to reach. The significant modifications that were made are (i) additional sizes offered to the user in a pre-loaded applicator and (ii) new packaging configuration. The key device characteristics, environment for use, performance specifications, principals of operation, mechanism of action and technological characteristics remain unchanged.
The addition of the delivery applicator does not affect the product performance of the bone grafting material, and therefore does not affect substantial equivalence when comparing the subject device to its predicate device. The table below summarizes the comparison of technical characteristics.
| Feature | Anorganic Bone Mineral with Collagen<br>in Delivery Applicator (Subject Device) | Anorganic Bone Mineral with Collagen<br>(K043034) |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | - Augmentation or reconstructive treatment<br>of alveolar ridge<br>- Filling of infrabony periodontal defects<br>- Filling of defects after root resection,<br>apicoectomy, and cystectomy<br>- Filling of extraction sockets<br>- Elevation of maxillary sinus floor<br>- Filling of periodontal and peri-implant<br>defects in conjunction with products<br>intended for GTR and GBR | - Augmentation or reconstructive treatment<br>of alveolar ridge<br>- Filling of infrabony periodontal defects<br>- Filling of defects after root resection,<br>apicoectomy, and cystectomy<br>- Filling of extraction sockets<br>- Elevation of maxillary sinus floor<br>- Filling of periodontal and peri-implant<br>defects in conjunction with products<br>intended for GTR and GBR |
| Material | Anorganic Bone Mineral with Purified Type<br>I Collagen | Anorganic Bone Mineral with Purified Type<br>I Collagen |
| Source | Bovine bone (anorganic bone mineral) with<br>bovine tendon (collagen) | Bovine bone (anorganic bone mineral) with<br>bovine tendon (collagen) |
| Form: | Preformed Plug & Moldable Plug | Block |
| Sizes | 0.25 cc, 0.5 cc, 1.0 cc | 0.25 g, 0.5 g, 1.0 g, 2.0 g |
| Biocompatibility | Biocompatible for the cleared indications<br>for use | Biocompatible for the cleared indications<br>for use |
| Packaging | Plastic delivery applicator in a single blister<br>tray with Tyvek lid | Glass vial with plastic screw cap in a single<br>blister with Tyvek lid |
| Plunger Force | 4.5 N ± 1.3 N (preformed plug)<br>4.1 N ± 1.1 N (moldable plug) | Not applicable |
| Sterility | Gamma irradiation, SAL 10⁻⁶ | Gamma irradiation, SAL 10⁻⁶ |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic |
| Single Use | Single Use Only | Single Use Only |
{6}------------------------------------------------
## Non-Clinical Performance Testing
To demonstrate substantial equivalence, the following non-clinical performance testing on the subject device was performed:
- Biocompatibility of Sterile Finished Device: cytotoxicity, irritation, and sensitization .
- . Biocompatibility of Delivery Applicator: cytotoxicity, physicochemical attributes of a polymeric material, containers - plastics, physicochemical tests USP <661>. physicochemical (material/chemical) characterization testing of non-aged and 36 month real time aged delivery applicators with extraction conditions and methods in accordance with ISO 10993-12 and ISO 10993-18.
- Biocompatibility of Anorganic Bone Mineral and Purified Type I Collagen (previously . submitted in the company's own predicate device premarket notification and applicable to the subject device): cytotoxicity, sensitization, acute systemic toxicity, genotoxicity, pyrogenicity, implantation, subacute/subchronic toxicity
- Sterilization validation was performed in accordance with ISO 11137-1 Sterilization of health care products - Radiation
- Bench testing of delivery applicator functionality and customer assessment. ●
- Animal performance testing of the Anorganic Bone Mineral derived from porcine bone . tissue (K140714) in an intraoral defect in a canine model and the Anorganic Bone Mineral with and without Collagen (K043034) in a rabbit femoral condyle defect model.
- Animal performance testing of the Purified Type I Collagen derived from bovine Achilles ● tendon in a rat subcutaneous implantation study.
### Conclusion
The results of the in vitro product characterization studies, functionality testing, animal performance testing, and biocompatibility studies show that the Anorganic Bone Mineral with Collagen in Delivery Applicator is substantially equivalent to its predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.