K200623 · Sk Bioland Co., Ltd. · NPL · Oct 2, 2020 · Dental
Device Facts
Record ID
K200623
Device Name
OssGuide
Applicant
Sk Bioland Co., Ltd.
Product Code
NPL · Dental
Decision Date
Oct 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OssGuide is recommended for: - Simultaneous use of GBR-membrane (OssGuide) and implants; - Augmentation around implants placed in immediate extraction sockets; - Augmentation around implants placed in delayed extraction sockets; - Localized ridge augmentation for later implantation; - Alveolar ridge reconstruction for prosthetic treatment; - Aid regeneration of bone defects after root resection, cystectomy, removal of retained teeth; - Guided bone regeneration in dehiscence defects
Device Story
OssGuide is a sterile, single-use, bioresorbable collagen membrane derived from porcine pericardium. Used in dental surgery by clinicians to facilitate guided bone regeneration (GBR). The membrane acts as an occlusive physical barrier, placed between the mucoperiosteal flap and the bone defect to prevent epithelial and connective tissue ingrowth. It provides a scaffold for bone regeneration and is resorbed by host tissue within 24 weeks. Clinicians trim the membrane to fit the defect, placing it dry or hydrated, and secure it with sutures or tacks. The device benefits patients by promoting bone healing in dental defects and around implants.
Clinical Evidence
No human clinical data. Evidence consists of in vitro characterization (physicochemical, tensile strength, suture pull-out) and an in vivo animal study in beagle dogs (n=20). The animal study compared OssGuide to Bio-Gide/Bio-Oss, demonstrating sufficient GBR effect for alveolar bone regeneration with complete biodegradation by 24 weeks.
Technological Characteristics
Bioresorbable membrane made of purified porcine pericardium collagen. Porous, fibrous, bilayer structure. Dimensions: 15x20mm to 30x40mm. Sterilized via gamma irradiation. Complies with ASTM F2212 (physicochemical), ISO 10993 (biocompatibility), ISO 11607-2 (packaging), and ISO 22442 (animal tissue control).
Indications for Use
Indicated for patients requiring guided bone regeneration (GBR) in dental procedures, including implant site augmentation (immediate/delayed extraction sockets), localized ridge augmentation, alveolar ridge reconstruction, and bone defect regeneration following root resection, cystectomy, or tooth removal.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2020
Sk bioland Co., Ltd. % Sanglok Lee Manager WISE COMPANY Inc. RM #303. The Sky Balley-303. 142. Gasan Digital 1-Ro Geumcheon-gu, Seoul 08507 Republic of Korea
Re: K200623
Trade/Device Name: OssGuide Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: September 1, 2020 Received: September 3, 2020
Dear Sanglok Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200623
Device Name OssGuide
Indications for Use (Describe)
OssGuide is recommended for:
- · Simultaneous use of GBR-membrane (OssGuide) and implants;
- · Augmentation around implants placed in immediate extraction sockets;
- · Augmentation around implants placed in delayed extraction sockets;
- · Localized ridge augmentation for later implantation;
- · Alveolar ridge reconstruction for prosthetic treatment;
- · Aid regeneration of bone defects after root resection, cystectomy, removal of retained teeth;
- · Guided bone regeneration in dehiscence defects
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### SK bioland Co., Li
162, Gwahaksaneop 3-ro, Ochang-eup, Cheongwon-gu, Cheongju-si, Chungcheongbuk-do, 28125 Republic of Korea
Image /page/3/Picture/2 description: The image shows the logo for SK bioland. The logo consists of the letters "SK" in red, with a stylized orange and red graphic above the letters. To the right of the letters "SK" is the word "bioland" in orange. The logo is simple and modern, and the colors are bright and eye-catching.
### 510(k) Summary - K200623
#### 01. Date of Submission: October 2, 2020
#### 02. Applicant
Company name: SK bioland Co., Ltd. Address: 162, Gwahaksaneop 3-ro, Ochang-eup, Cheongwon-gu, Cheongju-si, Chungcheongbuk-do, Republic of Korea TEL: +82 43 240 8633 FAX: +82 43 240 8699 Email: yjchae@sk.com
#### 03. Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
#### 04. Proposed Device Identification
Trade Name: OssGuide Common Name: Absorbable collagen membrane Classification Name: Barrier, Animal Source, Intraoral Product Code: NPL Panel: Dental Regulation Number: 21 CFR 872.3930 Device Class: Class II
#### 05. Indication for use
OssGuide is recommended for:
- Simultaneous use of GBR-membrane (OssGuide) and implants;
- . Augmentation around implants placed in immediate extraction sockets;
- . Augmentation around implants placed in delayed extraction sockets;
- . Localized ridge augmentation for later implantation;
- Alveolar ridge reconstruction for prosthetic treatment;
- Aid regeneration of bone defects after root resection, cystectomy, removal of retained teeth;
- Guided bone regeneration in dehiscence defects ●
#### 06. Predicate devices
510(k) Number: K050446 Device Name: BIO-GIDE® Manufacturer: Ed. Geistlich Soehne Ag fuer Chemische Industrie
#### 07. Device Description
OssGuide (Absorbable collagen membrane) is made of purified collagen without further cross-linking or chemical treatment. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. The OssGuide is a collagen membrane obtained by a standardized controlled
{4}------------------------------------------------
SK biola
162, Gwahaksaneop 3-ro, Ochang-eup, Cheongwon-gu, Cheongju-si, Chungcheongbuk-do, 28125 Republic of Korea
Image /page/4/Picture/2 description: The image contains the logo for SK bioland. The logo consists of the letters "SK" in red, followed by the word "bioland" in orange. Above the letters, there is an abstract butterfly-like shape with red and orange wings. The logo is simple and modern, with a focus on the company's name and a visual representation of nature.
manufacturing process. The OssGuide has a porous and fibrous microstructure that prevents the ingrowth of fibrous tissue into the bone defect.
OssGuide is packed by a packaging material composed of a transparent side and is easy to use by open system of easy peel type. It is easy to check the OssGuide, because one side of packaging is transparent closed by double packing system in order to protect OssGuide from microbial formation and moisture penetration. Then this packaged OssGuide is sterilized by gamma irradiation. The packaging contains the OssGuide and one Tyvek material template.
The surgical procedure consists of placing of an occlusive physical barrier between the connective tissue and the bone defect to prevent the migration of the epithelial and connective tissue cells into the defect. The periodontal defect or bone defect is exposed by a mucoperosteal flap and basic surgical procedures (e.g. curettage) are performed. The clinician should perform thorough debridement and efficient planing of the defect. Space-making material such as autologous bone or bone substitute may be used to fill the defect. OssGuide can be placed either dry or hydrated. OssGuide can be trimmed to the size and shape of the defect in a dry or wet state using scissors. A piece of sterile Tyvek may be used as a template for membrane sizing and trimming. OssGuide should overlap the walls of the defect by at least 2 mm to allow complete bone contact and to prevent gingival connective tissue invasion below the material. The membrane may be secured in place with sutures, tacks, or by means of sufficient contact with the mucoperiosteal flap to avoid displacement and micro motion during healing. To facilitate proper healing, ensure that there is tension-free primary closure of the soft tissue flap covering the membrane.
OssGuide a bioresorbable scaffold that is eventually resorbed by host tissue. OssGuide is substantially resorbed within 24 weeks after implantation.
The OssGuide is available to the United Sates market in 5 sizes as shown below. They are supplied sterile and intended for single use only.
| Model No. | Size | Packaging<br>(ea/box) |
|-----------|-----------|-----------------------|
| TG-1 | 15 x 20mm | 1 |
| TG-2 | 20 x 30mm | 1 |
| TG-3 | 30 x 40mm | 1 |
| TG-4 | 15 x 30mm | 1 |
| TG-5 | 25 x 25mm | 1 |
#### 08. Non-Clinical Test Conclusion
The determination of substantial equivalence was based on an assessment of non-clinical performance data obtained from in vitro characterization studies, an in vivo animal study, biocompatibility testing.
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate device in biochemical, mechanical properties. The test results demonstrated that the proposed device complies with the following standards:
- · Physicochemical Characterization per ASTM F2212
- Tensile strength testing
- · Suture pull-out strength testing
{5}------------------------------------------------
# SK bioland
162, Gwahaksaneop 3-ro, Ochang-eup, Cheongwon-gu, Cheongju-si, Chungcheongbuk-do, 28125 Republic of Korea
Image /page/5/Picture/2 description: The image contains the logo for SK bioland. The logo consists of the letters "SK" in red, followed by the word "bioland" in orange. Above the text is a stylized butterfly-like shape, with the left side in red and the right side in orange.
Validation of packaging, shelf life, transport and sterilization are performed as following:
- · Validation of Primary and Secondary barrier packaging per ISO 11607-2
- Validation of shelf life per ASTM F1980 using age accelerated and real-time aged samples
- · Validation of transport packaging per ASTM D4169
- · Validation of sterilization parameters per ISO 11137
Biocompatibility of the subject device was evaluated in accordance with ISO 10993-1 as followings:
- · Cytotoxicity per ISO 10993-5
- · Sensitization per ISO 10993-10
- · Irritation per ISO 10093-10
- · Subchronic Toxicity per ISO 10993-11
- · Genotoxicity per ISO 10993-3
- · Implantation per ISO 10993-6
- · Material mediated pyrogenicity per USP
Results indicate that the device's biocompatibility profile is acceptable.
The pyrogenicity of OssGuide was tested in vivo using the Rabbit Pyrogen Test as per USP <151>, and the test article was judged as nonpyrogenicity sample testing plan, the endotoxin (LAL) testing will be conducted on every batch of OssGuide products.
The manufacturing control of animal tissue components is performed as following:
- · Controls on sourcing, collection, and handling per ISO 22442-2
- · Viral Inactivation per ISO 22442-3
Animal study was conducted in the alveolar bone defect of beagle dogs to evaluate the in vivo performance and degradation of the OssGuide, supporting substantial equivalence. The study was conducted on 20 animals followed for 4, 8, 12 and 24 weeks. When comparing OssGuide with the predicate device, Bio-Gide with Bio-Oss bone graft material as a positive control, it showed sufficient GBR effect for the alveolar bone regeneration on the alveolar bone defect of beagle dogs. OssGuide maintained their contour until 8 weeks, lost most of its integrity after 12 weeks, and disappeared by biodegradation at 24 weeks.
#### 09. Substantially Equivalent Conclusion
#### Table 1: Substantial Equivalence Comparison
| Product<br>Name | SUBJECT Device<br>OssGuide | PREDICATE Device<br>BIO-GIDE® (K050446) | Equivalence Discussion |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Product<br>code | NPL | NPL | Identical |
| Regulatory<br>class | Class II | Class II | Identical |
| Regulation<br>Number | 21 CFR 872.3930 | 21 CFR 872.3930 | Identical |
| Intended use | OssGuide is recommended<br>for; | Bio-Gide® is recommended<br>for; | Equivalent<br>The indications for the su<br>bject device are a subset |
| | - Simultaneous use of GBR-<br>membrane (OssGuide)<br>and implants;<br>- Augmentation around implants placed in immediate extraction sockets;<br>- Augmentation around implants placed in delayed extraction sockets;<br>- Localized ridge augmentation for later implantation;<br>-Alveolar ridge reconstruction for prosthetic treatment;<br>- Aid regeneration of bone defects after root resection, cystectomy, removal of retained teeth;<br>- Guided bone regeneration in dehiscence defects | - Simultaneous use of GBR-<br>membrane (BIO-GIDE) and implants;<br>- Augmentation around implants placed in immediate extraction sockets;<br>- Augmentation around implants placed in delayed extraction sockets;<br>- Localized ridge augmentation for later implantation;<br>-Alveolar ridge reconstruction for prosthetic treatment;<br>- Aid regeneration of bone defects after root resection, cystectomy, removal of retained teeth;<br>- Guided bone regeneration in dehiscence defects; and<br>- Guided tissue regeneration procedures in periodontal defects. | of the indications for the predicate devices. |
| Mode of Action | OssGuide functions as a barrier when applied between bone graft material and soft tissue. The membrane serves as a bioresorbable scaffold that is eventually remodeled, resorbed, and replaced by host tissue. | Bio-Gide® functions as a barrier when applied between bone graft material and soft tissue. The membrane serves as a bioresorbable scaffold that is eventually remodeled, resorbed, and replaced by host tissue. | Identical |
| Operating<br>Principles | Cell-Occlusive<br>Implantable<br>Resorbable<br>Biocompatible | Cell-Occlusive<br>Implantable<br>Resorbable<br>Biocompatible | Identical |
| Material | Purified collagen | Purified collagen | Identical<br>Equivalent |
| Collagen<br>source | Porcine pericardium | Porcine peritoneum | Using the tissue extracted from porcine is the same, but the parts of tissue used are different. |
| Form | Membrane | Membrane | Identical |
| Structure | Bilayer structure | Bilayer structure | Identical |
| Color | White to off-white | White to off-white | Identical<br>Equivalent |
| Product<br>Size | 15x20mm, 20x30mm,<br>30x40mm,<br>15x30mm, 25x25mm | 13x25mm, 25x25mm,<br>40x50mm | Both devices are provided in clinically relevant sizes for intra-oral surgical procedures |
| Resorption<br>time | Substantially resorbed by<br>24 Weeks | Substantially resorbed by<br>24 Weeks | Identical |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Identical |
| Singe Use/<br>Reuse | Single use only | Single use only | Identical |
| Packaging | Double pouch pack | Double blister pack | Identical |
| Shelf -life | 36 months | 36 months | Identical |
{6}------------------------------------------------
# SK bioland Co., Ltd.
162, Gwahaksaneop 3-ro, Ochang-eup, Cheongwon-gu, Cheongju-si, Chungcheongbuk-do, 28125 Republic of Korea
Image /page/6/Picture/2 description: The image shows the logo for SK bioland. The logo consists of the letters "SK" in red, followed by the word "bioland" in orange. Above the letters "SK" is a stylized butterfly shape, with the left wing in red and the right wing in orange. The logo is simple and modern, and the colors are bright and eye-catching.
{7}------------------------------------------------
## SK bioland Co., Ltd.
162, Gwahaksaneop 3-ro, Ochang-eup, Cheongwon-gu, Cheongju-si, Chungcheongbuk-do, 28125 Republic of Korea
Image /page/7/Picture/2 description: The image shows the logo for SK bioland. The logo consists of the letters "SK" in red, followed by the word "bioland" in orange. Above the letters "SK" is a stylized butterfly-like shape with red and orange colors. The logo is simple and modern, and the colors are bright and eye-catching.
The proposed device, OssGuide, is determined to be Substantially Equivalent (SE) to the predicate devices, BIO-GIDE® (K050446) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.