ATRISORB® GTR Barrier is indicated for the surgical treatment of periodontal defects to aid in the regeneration and integration of tissue components in guided tissue regeneration procedures. ATRISORB® GTR Barrier is not intended for use in defects outside the indications statement.
Device Story
ATRISORB® GTR Barrier is a bioabsorbable, flowable polymeric formulation (poly(DL-lactide) in N-methyl-2-pyrrolidone) used in guided tissue regeneration (GTR) procedures. The device acts as a physical barrier to isolate surgical sites from gingival connective tissue and epithelium, allowing periodontal ligament and alveolar bone cells to regenerate. It is applied either by pre-forming a barrier using a specialized case kit (extraoral) or by direct application over bone grafting material (in situ). The barrier solidifies upon contact with tissue fluids or saline. It remains intact during the critical wound healing phase and subsequently bioabsorbs via hydrolysis over several months, eliminating the need for a second surgical removal procedure. Used by periodontal surgeons in clinical settings, the device provides a film covering over defects, facilitating tissue regeneration and integration.
Clinical Evidence
Pivotal multicenter clinical trial (n=162) compared ATRISORB® GTR Barrier to GORE-TEX® Periodontal Material in Class II furcation defects over 52 weeks. Primary endpoints included vertical/horizontal attachment level gain, pocket depth, and gingival index. Results showed ATRISORB® was statistically equivalent to the predicate, with mean vertical attachment gain 0.38 mm greater (p<0.001) and horizontal attachment gain 0.04 mm greater (p=0.017) than the predicate. Safety profile was benign with no sensitivity or immune reactions reported.
Technological Characteristics
Synthetic, flowable, bioabsorbable polymer formulation consisting of poly(DL-lactide) (PLA) dissolved in N-methyl-2-pyrrolidone (NMP). Provided in single-patient dose packs. Barrier formation utilizes a case with porous pads and spacers or direct application. Terminally sterilized by irradiation. Non-pyrogenic and biocompatible.
Indications for Use
Indicated for surgical treatment of Class II furcations, 2 and/or 3 walled intra-osseous defects, and dehiscence defects in patients aged 25-75 in good general health to aid in periodontal tissue regeneration.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
GORE-TEX® Periodontal Material
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date "SEP 8 1998". The month is abbreviated to three letters. The day is a single digit, and the year is four digits.
2579 MIDPOINT DRIVE FORT COLLINS, CO 80525-4417 U.S.A.
Image /page/0/Picture/2 description: The image shows the logo for Atrix Laboratories, Inc. The logo features a stylized sunburst design above the company name. The sunburst is composed of several thick, black lines radiating upwards and outwards from a central point. Below the sunburst, the word "ATRIX" is printed in a bold, sans-serif font, and beneath that, the words "LABORATORIES, INC" are printed in a smaller, sans-serif font.
K982865
PHONE: (970) 482-5868
EMAIL: atrixlab@frii.com http://www.atrixlabs.com
# 510(k) Summary of Safety and Effectiveness
## 1. General Information
| Manufacturer: | Atrix Laboratories, Inc.<br>2579 Midpoint Drive<br>Fort Collins, CO 80525-4417<br>Telephone: 970-482-5868<br>Fax: 970-482-9735 |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Elaine M. Gazdeck, R.A.C.<br>Vice President, Regulatory Affairs/Quality |
| Device Generic Name: | Bioabsorbable GTR Barrier Kit (used to form a<br>Bioabsorbable GTR Barrier) |
| Device Trade Name: | ATRISORB® Bioabsorbable Guided Tissue<br>Regeneration (GTR) Barrier Kit (used to form an<br>ATRISORB® Bioabsorbable GTR Barrier) |
| 510(k) Clearance Number<br>for Unmodified Device: | K955838 |
| 510(k) Clearance Number<br>for Modified Device: | K |
| Date Prepared: | July 1, 1998 |
| Predicate Device: | GORE-TEX® Periodontal Material |
## 1. Description
#### 1.1. Formulation
The ATRISORB® polymer formulation is a sterile, synthetic flowable polymeric formulation which is bioabsorbable and has been developed for GTR procedures by Atrix Laboratories, Inc. It consists of a polymer of lactic acid, poly(DL-lactide) (PLA), dissolved in N-methyl-2-pyrrolidone (NMP).
{1}------------------------------------------------
#### 2.2. Device
The finished device is provided in two different packaging configurations. The first configuration is the ATRISORB® GTR Barner Kit (case kit) used to form either with the barrier-forming case (extraoral) or by direct application over bone grafting material. The second configuration is the ATRISORB® GTR Barrier multipack (multipack) used only for the direct application technique.
The ATRISORB® GTR Barrier Kit configuration contains a foil-pouched, single-patient use dose pack containing 0.5 g of ATRISORB® polymer formulation, a pouched tray that contains a barrier forming case and three units of 0.9% sodium chloride solution (pH range 3.5 - 6.0), and an instructions for use booklet.
The barrier forming case has two sections of porous pads held in the cavities of the case and two spacers that are loose in the case. A label with a 1 mm grid pattern is affixed to the underside of the case. This grid is visible through the case and allows an area on the inside of the case to be used for trimming the formed barrier
The ATRISORB® GTR Barrier multipack configuration consists of three foilpouched, single patient use dose packs containing 0.5 g of the ATRISORB® polymer formulation, three 18 gauge cannulae, and an instructions for use booklet.
## 3. Mechanics
The ATRISORB® GTR barrier functions as a guided tissue regeneration barrier by isolating the regenerative surgical site from the adjacent gingival connective tissue and epithelium. This facilitates population of the surgical site with cells from the periodontal ligament and adjacent alveolar bone that lead to regeneration.
The ATRISORB® polymer formulation is a flowable solution. A preformed GTR barner, with consistent thickness, is prepared by using a barrier forming kit. The flowable formulation partially precipitates when it is "sandwiched" between two porous pads which are wetted with sodium chloride solution contained in a barrier forming case. The barrier thickness is determined by the thickness of the spacers which hold the two porous pads apart. The barrier, which is prepared in approximately three to six minutes, can be trimmed to the required size and morphology of the defect. The formed ATRISORB® GTR barner is applied to the defect
Another method of forming the barrier can be accomplished in situ where the tissue fluid aids in the precipitation of the barrier at the defect site. For defects where bone replacement graft material has been applied, the flowable formulation can be applied directly over the grafting material. Extraoral formation (formed in the case) is required when grafting materials are not used, as the unprecipitated formulation would flow into the defect.
{2}------------------------------------------------
With either method of application, the formed barrier is a film which covers the defect. Once in place, it continues to solidify and the resulting barrer precludes the migration of epithelial or connective tissue cells through it, and simultaneously isolates the periodontal compartment, thereby facilitating regeneration. The chemistry of the ATRISORB® polymer formulation is such that, after setting, the polymer component subsequently bioabsorbs through the process of hydrolysis. Bioabsorption occurs over the course of several months. The barrier is present and intact during the critical period when periodontal regenerative cells are proliferating during the wound healing cascade from adjacent periodontal sources.
## 4. Intended Use
The clinical indications for the ATRISORB® GTR Barrier are for GTR procedures in the surgical treatment of Class II furcations, or 2 and/or 3 walled intra-osseous defects and dehisence defects to aid in the regeneration and integration of periodontal tissue components.
The indications for the ATRISORB® GTR barrier are equivalent to the indications for the predicate device, expanded polytetrafluoroethylene (e-PTFE) GORF-TEX® Periodontal Material (W.L. Gore & Assoc., Flagstaff, AZ).
The indications that differ are for bioabsorbability and elimination of a second surgical procedure to remove a non-bioabsorbable material. Results of preclinical and clinical trials support the claim of bioabsorbability and good tissue compatibility.
The predicate device, GORE-TEX® Periodontal Material, is indicated as an implantable material to aid in the healing of periodontal defects. The GORE-TEX® Periodontal Material is non-bioabsorbable and requires a second surgical procedure to remove it.
# 5. Summary of In Vitro Studies
#### 5.1. Cytotoxicity
The study evaluated the toxic effects of the ATRISORB® GTR barrier upon mature monolayer tissue cultures of L929 cells. The ATRISORB® GTR barrier is mildly cytotoxic; however, these results are not consistent with results from more meaningful, in vivo implantation and clinical studies on the barrier.
#### 5.2. Sensitization
A guinea pig maximization test, based on the method of Magnusson and Kligman (1970), was used to evaluate the potential of the ATRISORB® polymer formulation to elicit delayed dermal contact sensitization. The ATRISORB® polymer formulation is not a skin sensitizer
{3}------------------------------------------------
#### Genotoxicity 5.3.
Genotoxicity
The study evaluated the mutagenicity of an extract of the ATRISORB® The study evaluated the mutagencity of an extract of the mulation is not mutagenic.
polymer formulation. The ATRISORB® polymer formulation is not mutagenic.
## Pyrogenicity - Limulus Amebocyte Lysate Test (LAL) 5.4.
Pyrogenicity - Limules Almaded the concentration of bacterial endotoxins in an extract
The LAL test estimated the concentration of bacterial non-The LAL test estimated the concentration of becamer is nonpyrogenic according to this test method.
#### Hemolysis 5.5.
The ATRISORB® GTR barrier causes hemolysis in an in vitro test system, The ATRISORB® GTR Dannel Causes hemolysis in patients treated with the barrier.
but it is highly unlikely to cause hemolysis in patients begolvice potential using but it is highly unlikely to cause nemorysis nemolytic potential using
The study evaluated the ATRISORR® GTR barner's hemolytic potential using The study evaluated the ATKION City of the Buffering capacity. The barrier
human blood in an in vitro system with the black and the NMP human blood in an in vitro system title blood path and the NMP does not have significant contact with the blood personal.
concentrations in this assay may have been far greater than that which could be attained in the human body.
Based upon this analysis, the ATRISORB® GTR berner's in vitro hemolytic Based upon this analysis, the ATTATCOND of the barrier in accordance with the barrier's intended use.
#### Sterilization 5.6.
The ATRISORB® GTR Barrier Kit and ATRISORB® GTR Barrier are The ATRISORD® GTK Banier Tat and Arradiation. The cannulae in the terminally sterilized by Costing are provided sterile by the manufacturer and are not further processed.
#### Bioburden 5.7.
Bioburden determination is performed prior to sterilization following USP procedures.
{4}------------------------------------------------
# 6. Summary of In Vivo Implantation Studies
#### Subchronic Implantation 6.1.
Subchronic implantation
The study evaluated the biocompatibility of the rat . During the course o The study evaluated the blocompanism of the rest. During the course of the when surgically implanted subcularied around any of the test implantation sites study, no toxicity was observed around any of the toot major and the locked to the really was observed
which could be directly related to the barrier. No overt toxicity was which could bing the course of the study.
The subcutaneous implantation of the ATRISORB® GTR barner in rats for a
rt a many and the country to this taxicity The subcatures and not result in toxicity.
### Chronic Implantation and Biodegradation 6.2.
Chronic Imprantation and Libers of the ATRISORB® GTRISORB® GTR
The study evaluated (1) the degradation kinetics of the controls The study evaluated (1) the degradation hiriclib the barner The controls for
barrier, and (2) response of tissue in contact with the barner of the test and barrier, and (2) response of tissue in concess winding of the test and
the study were USP plastic strips. Histological examination of the test and the study were USP plastic surpos. This signs of biologically significant
control excised implant sites showed no significant difference between control excised implant sites showed no significant difference between the
changes or abnormal tissue. There was no significant dissue effects. changes or abnormal tissue. There was no significant and control sites .
ATRISORB® GTR barrier sites and control sites with respect to tissue effects.
The study demonstrated that the ATRISORB® GTR barrier does not ellicit The study demonstrated the the ATRIONDO Shipping on rabbits for up
toxicity (dermal or systemic) when subcutaneously implanted in rabbits for up toxicity (dermal or systemic) when subculariously implication of the ATRISORB® GTR barrier should
to twelve months and biodegradation of the ATRISORB® GTR barrier should be complete by 13-14 months.
# 7. Summary of Animal Efficacy Studies
#### Histology Study 7.1.
This study compared the safety, clinical results, and histologic effects of This study compared the Sately, cirinour returns in six adult beagle dogs
ATRISORB® GTR barriers to sham operated controls in six adult beagle dogs ATRISORD® GTR Barnors is unam of them on tall furcation defects.
A periodontal surgeon selected and prepared 16 study sites as follows: a full A periodontal surgeon selector and propares was reflected, plaque and thickness ilap created after giftgirds increasing devices, and roots planed by hand. calculus were removed with ultrasonic were applied to eight test sites, all
Preformed ATRISORB® GTR barriers were applied to eight test sites, all Preformed ATRISORD® OTTS Barner . Post-surgery periodontal maintenance
study sites were closed and sutured. Post-surgery periodontal maintenance sites were closed and Sularou. T est only inse to all study sites for consisted on topical applications or shed using a modified Bass technique wo months, then the teen were bradned as a graintegrity, clinical conditions, with brushes soaked in Pendox® crainment of the study. All of the animals were sacrificed after six months of observation.
The ATRISORB® GTR barrier demonstrated clinically and statistically The ATRIONDO OTHe depth, horizontal furcation depth, and improved superior results in position as a more significantly increased the
{5}------------------------------------------------
amount of periodontal regeneration (i.e., new bone, cementum, and amount of periodontal regeneration (f.e., hew bone, before, before, and toxicity
periodontal ligament formation) in the furcation defect. No signs of toxicity were observed.
### Biodegradation and Safety Study 7.2.
Biodegradation and Safety Study
The object of this study was to evaluate the potential for clinical application, The object of this study was to evaluate the potential hor and the polymer
clinical management, safety and biocompatibility of a biodegradable polymer clinical management, salety and blockhpanding or cated to facilitate groend thissue
barrier preformed in a barrier forming device to facilitate in the barrier preformed in a barner forming device to fallance a maxilla in the dog.
regeneration at periodontal defects in the study and 17 naturally occurring defects regeneration at periodontal detects in the study and 17 naturally occurring defects
Seven beagle dogs were used in the study and 17 naturally were treated. Twelve periodontal furcation defects were treated surgically were treated. Twelve pendontal furcation if the barriers formed with three
with ATRISORB® GTR barriers to determine if the bartiers formed with three with ATRISORB® GTR barners to determine if the barners for the world.
prototype kits compared favorably to 5 sham operated controls in terms of
the supers of the company to safety and biocompatibility.
The ATRISORB® GTR barrier was well tolerated in all periodontal test sites
in the ATRISOR was and continues inflammation compared to control The ATRISORB® GTR barner was well tolerated in all por
and produced no increases in localized inflammation compared to control sites.
Use of the ATRISORB® GTR barrier in treating periodontal defects in the dog
ing and the collection birthlaniael evidence of adverse healing and Use of the ATRISORB® GTTA banier in treating for a resulted in no clinical or histological evidence of dover of the end of one vear.
Histological examination confirmed that barriers were completely absorbed at
the examination and confinent and spicule found in one out of twelve Histological examination confilmed that bamore word in one out of twelve
12 months with the exception of one small spicule found in one out of twelve test sites.
#### Pilot Study 7.3.
Pliot Study
This study was a pilot study designed to examine four compositions of This study was a pilot study designed to oxepare the ATRISORB® GTR ATRISORB® polymer formulation used to propare tissue regeneration in the periodontal region.
Thirty-nine defects were treated in six beagle dogs. All dogs were terminated I hirty-hine defects were treated in six bearier placement. The study
between 9 and 12 months following initial bearier placement, The study between 9 and 12 months lollowing think barner pecome reconstituted on root
showed that new periodontal supporting tissues became reconstituted on root showed that hew penodonial burgically-induced and naturally-occurring and furcation sunasse in ssirg. The ATRISORB® GTR barrier.
#### Feasibility Study 7.4.
The objective of this study was to evaluate the safety and potential of a The objective of this study was to overse to facilitate periodontal synthetically formulated nowable polymer same to the canine and
{6}------------------------------------------------
molar regions of dogs over a period of 3 months. Two dogs were used with a moral regions of "dogeier sites and 8 control sites) being treated.
The results of this study demonstrated the potential safety and the ability of l he results of this study demonstration the polymers to facilitate periodontal tissue regeneration.
## 8. Summary of Pivotal Trial
A pivotal clinical trial under an FDA Investigational Device Exemption was conducted A pivotal cillinear that chuck and efficacy of the ATRISORB® GTR Barrier Kit in a multicenter clinical trial.
The primary object of the study was to demonstrate that the ATRISORB® GTR The primally object of the study was to GRE-TEX® periodontal material for guided tissue regeneration (GTR) treatment of Class II furcation defects in humans. The tissue regeneration (GTTY) a cannertical and horizontal attachment level, with primally clinical end points were vig pocket depth, bleeding on probing, gingival margin location, and gingival index.
The investigation was a multicenter study with a paralled design. Individuals with a The investigation was a mailies national with either ATRISORB® GTR barrier or Glass in furcation delect were troulou with on randomized treatment assignment. OUtcomes of healing were evaluated at regular intervals up to one year after the Outcomes of nealing were comparisons of clinical results between the two products were made in order to demonstrate substantial equivalency between the were made in order to riement GORE-TEX® periodontal material in guided tissue regeneration treating Class II furcation defects.
One hundred sixty-two subjects were randomized and entered into this 52-week one handroo oxly the sabjects are a learner and 80 subjects received GORE-TEX® periodontal material. Subjects were selected if they met the following entry criteria:
- A Class II furcation with vertical pocket depth []5 mm and attachment loss at the site and the horizontal depth of the furcation was 03 mm
- Between the ages of 25 and 75 .
- . in good general health
Radiographs were taken within the two months prior to Screening or at the Screening or Baseline visit, and provided evidence of crestal alveolar bone loss associated with periodontitis.
The study demonstrated that ATRISORB® GTR barner is substantially equivalent to the predicate device, GORE-TEX® periodontal material for the treatment of periodontal defects using guided tissue regeneration procedures.
{7}------------------------------------------------
The safety results over the 52-week duration of the study support the benign The salety results over the ose of the ATRISORB® GTR barrier.
## Summary of the Statistical Analysis of the Data from the Pivotal Clinical 8.1. Trial
The following tabulation presents the data for ATRISORB® GTR barrier vs. GORE-TEX® periodontal material at Baseline and Week 52. Vertical attachment level gain significantly satisfied the condition of clinical attachment level gall regility wo treatment groups at Week 52. The mean equivalence between the the treated with ATRISORB® GTR barner was 0.38 mm greater than the mean for subjects treated with GORE-TEX® periodontal material. The null hypothesis that GORE-TEX® periodontal material was clinically superior by at least 0.5 mm was rejected with a p-value of less than 0.001. The one-sided confidence interval for the difference of less than 0.00 . - TRUSORB® GTR barrier and the GORE-TEX® periodontal between the ATTRICO any difference greater than 0.08 mm in favor of GORE-TEX® periodontal material can be ruled out with 95% confidence.
Horizontal attachment level gain significantly satisfied the condition of clinical equivalence between the two treatment groups at Week 52. The mean increase at Week 52 for subjects treated with ATRISORB® GTR barrier was slightly greater (0.04 mm) than the mean for subjects treated with GORE-TEX® periodontal material. The null hypothesis that GORE-TEX® periodontal material was clinically superior by at least 0.5 mm was rejected with a p-value of 0.017. The one-sided confidence interval for the difference in treatment effects indicates that any difference greater than 0.38 mm can be ruled out with 95% confidence.
Pocket depth improvement significantly satisfied the condition of clinical equivalence for the two treatment groups at Week 52. The mean decrease at Week 52 was 0.18 mm greater for ATRISORB® GTR barrier subjects than the mean decrease for subjects treated with GORE-TEX® periodontal material.
{8}------------------------------------------------
| Measurement | Time Point | ATRISORB® GTR Barrier | | | GORE-TEX® Periodontal Material | | | GORE-TEX® Periodontal Material Minus ATRISORB® GTR Barrier Difference | P-value* | One-sided* Confidence Interval |
|-----------------------------|-----------------------|-----------------------|-------|------|--------------------------------|-------|------|-----------------------------------------------------------------------|----------|--------------------------------|
| | | N | Mean | S.E. | N | Mean | S.E. | | | |
| Vertical Attachment Level | Baseline**<br>Week 52 | 76 | 5.82 | 0.21 | 73 | 5.74 | 0.20 | -0.08 | 0.009 | 0.30 |
| | | 66 | 2.00 | 0.19 | 64 | 1.63 | 0.20 | -0.38 | <0.001 | 0.08 |
| Horizontal Attachment Level | Baseline**<br>Week 52 | 76 | 4.57 | 0.14 | 73 | 4.77 | 0.15 | 0.20 | 0.010 | 0.34 |
| | | 66 | 2.11 | 0.20 | 64 | 2.06 | 0.16 | -0.04 | 0.017 | 0.38 |
| Pocket Depth | Baseline**<br>Week 52 | 76 | 5.55 | 0.11 | 73 | 5.55 | 0.10 | -0.00 | 0.024 | 0.43 |
| | | 66 | 2.26 | 0.16 | 64 | 2.08 | 0.15 | -0.18 | 0.001 | 0.19 |
| Gingival Margin | Baseline**<br>Week 52 | 76 | -0.26 | 0.19 | 73 | -0.19 | 0.19 | 0.07 | <0.001 | 0.15 |
| | | 66 | -0.26 | 0.15 | 64 | -0.45 | 0.18 | -0.20 | 0.002 | 0.20 |
| Bleeding on Probing | Baseline**<br>Week 52 | 76 | 1.34 | 0.09 | 72 | 1.17 | 0.09 | -0.18 | <0.001 | 0.12 |
| | | 66 | 0.85 | 0.14 | 64 | 0.81 | 0.11 | -0.04 | 0.001 | 0.25 |
| Plaque Index | Baseline**<br>Week 52 | 76 | 0.68 | 0.08 | 73 | 0.84 | 0.09 | 0.15 | 0.015 | 0.42 |
| | | 66 | 0.15 | 0.12 | 64 | 0.30 | 0.12 | 0.15 | 0.018 | 0.42 |
| Gingival Index | Baseline**<br>Week 52 | 76 | 1.32 | 0.08 | 72 | 1.19 | 0.08 | -0.12 | <0.001 | 0.10 |
| | | 66 | 0.48 | 0.13 | 63 | 0.38 | 0.11 | -0.10 | <0.001 | 0.17 |
Gingival margin recession significantly satisfied the condition of clinical equivalence for the two treatment groups at all time points. The mean increase at Week 52 for subjects treated with ATRISORB® GTR barner was 0.2 mm less than the mean increase for subjects treated with GORE-TEX® periodontal material.
Bleeding on probing score significantly satisfied the condition of clinical equivalence for the two treatment groups at all time points. The mean change at Week 52 for subjects treated with ATRISORB® GTR barrier was slightly greater (0.04 mm - 0.06 mm) than the mean change for subjects treated with GORE-TEX® periodontal material.
Gingival index score significantly satisfied the condition of clinical equivalence for the two treatment groups at all time points. The mean change at Week 52 for subjects treated with ATRISORB® GTR barrier was slightly greater (0.07 - 0.10) than the mean change for subjects treated with GORE-TEX® periodontal material.
Plaque index score significantly satisfied the condition of clinical equivalence for the two treatment groups at all time points. The mean change at Week 52 for subjects treated with ATRISORB® GTR barrier was 0.15 less than the mean change for subjects treated with GORE-TEX® periodontal material.
{9}------------------------------------------------
#### Postoperative Complications 8.2.
Postoperative symptoms or complications that were reported were those commonly associated with oral or periodontal surgical procedures, such as intra oral swelling associated with offer of penodonial ourgroup without treatment or in some cases and pain which usually rodorroules. These events occurred in both test and control treatments.
There were no sensitivity.reactions or immune responses associated with the use of the ATRISORB® GTR barrier.
## 9. Direct Application
Case studies were conducted with the direct application of the barrier over bone Gase Studies were conteen subjects with a total of twenty-two periodontal defects of grarting matchan. Eightedn and it was determined that the material was suitable varying comploxity word troutes and determined by clinical measurements and reentry surgeries. These cases were followed and no adverse sequela were observed. encry Surgerior "The direct application of the ATRISORB® GTR Barrier has These using and is a suitable alternative to forming the barrier in the precipitation case.
#### 10. Conclusions
The claims for the ATRISORB® GTR Barrier Kit and the ATRISORB® GTR Barrier are equivalent to the claims for the predicate device, GORE-TEX® Periodontal Material. Additional claims for the ATRISORB® GTR Barrier are for bioabsorbability and elimination of a second surgical procedure to remove the non-bioabsorbable GORE-TEX® Periodontal Material. The features of the ATRISORB® GTR Barrier compared to the GORE-TEX® Periodontal Material are described below.
Human clinical trials have demonstrated that the ATRISORB® GTR Barrier is equivalent to the non-bioabsorbable GORE-TEX® Periodontal Material as an implantable material intended to aid in the healing of periodontal defects.
The ATRISORB® GTR Barrier is bioabsorbed, eliminating the need for a second surgical procedure such as that required to remove the non-bioabsorbable membrane, GORE-TEX® Periodontal Material.
Preclinical and clinical trial studies have demonstrated the ATRISORB® GTR Barrier to be safe and effective in the treatment of periodontal defects when formed either using the barrier forming case or the direct application technique.
The ATRISORB® GTR Barrier is non-pyrogenic and is biocompatible.
In vitro, animal. and clinical studies have demonstrated that the ATRISORB® GTR Barrier is safe and effective for its stated indications and is substantially equivalent to the predicate device, GORE-TEX® Periodontal Material.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird in flight, formed by three curved lines that also resemble human profiles facing to the right.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville, Maryland 20850
OCT 1 0 2007
Elyse Wolff, MT (ASCP) Regulatory Project Leader Atrix Laboratories, Incorporated 2579 Midpoint Drive Fort Collins, Colorado 80525-4417
Re: K982865
Trade Name: ATRISORB® Bioabsorbable Guided Tissue Regeneration (GTR) Barrier Regulation Number: 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: 2 Product Code: NPK Dated: July 1, 1998 Received: August 13, 1998
Dear Ms. Wolff:
This letter corrects our substantially equivalent letter of September 8, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
## Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Spitte y. Michluams.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Image /page/11/Picture/7 description: The image is a black and white circular logo. The logo has the text "FDA Centennial" in the center. The logo also has the text "1906-2006" at the top and is surrounded by stars and text.
Protecting and Promoting Public Health
{12}------------------------------------------------
Page
1 of 1
# K982865
-K955838 (Unmodified Device) 510(k) Number (if knc wn):
ATRISORB® GTR Barrier Device Name:
Indications For Use:
ATRISORB® GTR Barrier is indicated for the surgical treatment of periodontal defects to aid in the regeneration and integration of tissue components in guided tissue regeneration procedures. ATRISORB® GTR Barrier is not intended for use in defects outside the indications statement.
(PLEASE DO NOT VRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runne
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices Kh 510(k) Number
Prescription Use V (Per 21 CFR 801.109) --
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.