K965205 · American Custom Medical, Inc. · NPK · Apr 30, 1997 · Dental
Device Facts
Record ID
K965205
Device Name
TEFGEN - LS
Applicant
American Custom Medical, Inc.
Product Code
NPK · Dental
Decision Date
Apr 30, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TefGen-LS membrane is a 100% PTFE non-absorbable guided tissue membrane. TefGen-LS is an implant material which is intended to be used as a temporary space making barrier over bone or other tissue. The material is easily trimmed to a variety of shapes as required by specific cases.
Device Story
TefGen-LS is a non-absorbable, 100% polytetrafluoroethylene (PTFE) barrier membrane. It functions as a space-making implant to cover bone or tissue defects, allowing for guided tissue regeneration. The device is supplied as a sheet that clinicians trim to the required shape for specific surgical cases. It is designed to be stiff enough to maintain space while remaining supple enough to conform over defect margins. The device is used by clinicians in surgical settings. It benefits patients by providing a biocompatible, inert barrier that prevents soft tissue ingrowth into bone defects, thereby promoting osteogenesis.
Clinical Evidence
No clinical data provided. Evidence is based on literature citations demonstrating the function of PTFE membranes in treating osseous defects, including clinical reports, rat model studies, and comparative healing studies of bone defects.
Technological Characteristics
Material: 100% Polytetrafluoroethylene (PTFE). Form factor: Sheet/membrane, trimmable. Properties: Non-resorbable, biocompatible, inert. Sterilization: Not specified.
Indications for Use
Indicated for use as a temporary space-making barrier over bone or other tissue to facilitate guided tissue regeneration.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
TefGen-FD membrane
GORE-TEX augmentation membrane
Submission Summary (Full Text)
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
OCT 10 2007
Food and Drug Administration
9200 Corporate Boulevard
Rockville, Maryland 20850
Mr. Bruce G. Ruefer
President
American Custom Medical, Company
2430 N. 7th Avenue, Suite 4
Bozeman, Montana 59715
Re: K965205
Trade Name: TefGen-LSTM
Regulation Number: 872.3930
Regulation Name: Bone Grafting Material
Regulatory Class: 2
Product Code: NPK
Dated: March 12, 1997
Received: March 14, 1997
Dear Mr. Ruefer:
This letter corrects our substantially equivalent letter of April 30, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
Page 2 –
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Suzette Y. Michiu O.M.D.
Chiu Lin, Ph.D.
Director
Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Protecting and Promoting Public Health
{2}
K965205
APR 30 1997
# TefGen-LS™ 510(k) SUMMARY OF SAFETY AND EFFICACY
| Prepared by: | Bruce G. Ruefer, President |
| --- | --- |
| Date: | December 20, 1996 |
| Classification Name: | Augmentation Membrane, 76LYC |
| Common/Usual Names: | Barrier Membrane; Guided Tissue Membrane; or GTR Membrane. |
| Proprietary Name: | TefGen-LS™ Guided Tissue Membrane |
| Establishment Reg. No. | 9006936 |
| Classification: | Class II |
| Performance Standards: | Not Applicable. |
| Substantial Equivalence: | The non-absorbable augmentation membrane is manufactured by American Custom Medical and is equivalent in function to the augmentation membranes manufactured and marketed by both American Custom Medical and W.L. Gore which are currently on the market. |
| Product Description & Intended Use: | TefGen-LS membrane is a 100% PTFE non-absorbable guided tissue membrane. TefGen-LS is an implant material which is intended to be used as a temporary space making barrier over bone or other tissue. The material is easily trimmed to a variety of shapes as required by specific cases. |
| Comparative Technological Characteristics: | TefGen-LS is a biocompatible 100% PTFE membrane, is non-resorbable, is stiff enough to create a space, and is supple enough to be formed over the margins of a defect; as are the predicate devices. |
| Safety & Efficacy: | TefGen-LS is composed of 100% polytetrafluoroethylene or PTFE. PTFE is the most inert polymer known at this time. PTFE's use as an implant material in the cardiovascular area is well substantiated with over 3,000,000 PTFE vascular grafts implanted to date. Many other configurations of PTFE devices are marketed including soft tissue and cardiovascular PTFE patches, PTFE suture, PTFE barrier membranes, and PTFE ear implant devices.
PTFE has been found to pass biocompatibility assays including U.S.P. Class VI, carcinogenicity studies, hemocompatibility studies, and others. PTFE has been proven many times over to be non-reactive to body fluids and tissues making it a material of choice for biomaterial applications. |
PAGE K1
{3}
TefGen-LS™ 510(k) SUMMARY OF SAFETY AND EFFICACY (cont)
Literature:
ACM PTFE has been found to function as an acceptable augmentation membrane as noted in the following articles:
"The Use of High-Density Polytetrafluoroethylene Membrane to Treat Osseous Defects: Clinical Reports", B. Bartee, D.D.S., Implant Dentistry, 4, 1995, pgs. 21-26.
"Evaluation of a Full Density Polytetrafluoroethylene (PTFE) Film to Promote Osteogenesis in the Rat Model", J. Carr, et.al., Oral Implantology, 21, 1995, pgs. 88-95.
"Influence of Three Membrane Types on Healing of Skull Lesions", B. Crump, et.al, presented at the 1995 IADR Annual Meeting, Singapore, June, 1995.
"The Influence of Three Membrane Types on Healing of Bone Defects", B. Crump, et.al., Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, Oct. 1996, pgs. 365-374.
"High-density PTFE Membranes: Uses with Root-form Implants", J. Krauser, Dental Implantology Update, 7, pgs. 65-69, 1996.
Clinical Research Associates Newsletter, 20:4, February 1996.
Conclusion:
The TefGen-LS membrane configuration is substantially equivalent to the currently marketed TefGen-FD membrane and currently marketed GORE-TEX augmentation membrane.
PAGE K2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.