The High Retention Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed hybrid restorations that can be attached with a snap-in system.
Device Story
System provides rigid connection for fixed/detachable hybrid dentures to endosseous dental implants; accommodates implant divergence up to 20° per implant and 40° between implants. Components include titanium alloy abutments (TiN coated), PEEK inserts, denture attachment housings, and processing/seating tools. Used by dental professionals in clinical settings to attach restorations via snap-in mechanism. Benefits patient by providing secure, retentive support for dental prostheses, restoring masticatory function.
Clinical Evidence
Bench testing only. Mechanical denture retention testing demonstrated retention force exceeding that of predicate K151789. Compatibility verified via OEM drawings and engineering analysis.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) with TiN coating. Components: Abutment housings, PEEK inserts, processing/seating tools. Interface: Conical, external hex, internal hex, internal multi-lobe. Sterilization: Identical to predicates. Connectivity: None.
Indications for Use
Indicated for patients requiring fixed, partial, or full arch restorations supported by endosseous dental implants in the mandible or maxilla to restore masticatory function.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zest Anchors, LLC % Melissa Burbage RA Sr Specialist PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K213391
Trade/Device Name: High Retention Attachment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 15, 2021 Received: October 15, 2021
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K213391
Device Name
High Retention Attachment System
#### Indications for Use (Describe)
The High Retention Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory finction. It is used in fixed hybrid restorations that can be attached with a snap-in system.
The High Retention Attachment System is compatible with the following implants:
| Implant Manufacturer | Implant System | Implant Diameter |
|----------------------|----------------------------------------------------------|--------------------------------------------------|
| Ace Surgical | Infinity OCTAGON | 3.3, 4.1, 4.8 |
| | Infinity TRI-CAM | 3.5, 4.3, 5.0 |
| | Infinity Internal Hex | 3.7, 4.1, 4.7, 5.1 |
| BioHorizons | Tapered Internal | 3.0 |
| | Tapered Plus | 3.8 |
| | Laser-Lok | 3.0 |
| Biomet 3i | 3iT3, NanoTite, OSSEOTITE | 3.25, 3.75, 4.0 |
| Blue Sky Bio | Quattro | 3.3, 4.1, 4.8 |
| | One Stage | 3.3, 4.1, 4.8 |
| Camlog | SCREW-LINE ROOT-LINE 2 | 3.3, 3.8, 4.3, 5.0 |
| | SCREW-LINE | 3.3, 3.8, 4.3, 5.0 |
| Dentsply | Astra Tech OsseoSpeed EV | 3.6, 4.2, 4.8 |
| | Xive Frialit-2 | 3.4, 3.8, 4.5, 5.5 |
| | Ankylos C/X | 3.5 |
| | Astra Tech OsseoSpeed TX | 3.5, 4.0, 4.5, 5.0 |
| Hiossen, Inc. | SS | 3.5, 4.0, 4.5 |
| | ET, TS | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 |
| | US | 3.5, 3.3, 4.5, 4.0, 5.0, 6.0, 7.0 |
| Implant Direct | Swish Plus | 4.1, 4.8 |
| | SwishTapered | 4.1, 4.8 |
| | InterActive | 3.2, 3.7, 4.3, 5.0 |
| | ReActive | 3.7, 4.2, 4.7, 5.7 |
| | RePlant | 3.5, 4.3, 5.0 |
| | Legacy 1, 2, 3, 4 | 3.7, 4.2 |
| | Legacy 2, 3, 4 | 3.2, 4.7, 5.2 |
| | Legacy 1, 3 | 5.7 |
| | Legacy 2, 4 | 5.7, 7.0 |
| | Legacy 1 | 4.7 |
| IDS | MegaGen Any Ridge | 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0 |
| Keystone | PrimaConnex | 3.5, 4.1, 5.0 |
| | Genesis | 3.5 |
| | TILOBEMAXX | 7.0, 8.0, 9.0 |
| MIS | C1 | 3.3, 3.75, 4.2 |
| | V3 | 3.3, 3.9, 4.3, 5.0 |
| | SEVEN | 3.3 |
| | M4 | 3.3 |
| Nobel | Replace, Replace Select, NobelSpeedy | 3.5, 4.3, 5.0 |
| | NobelActive, NobelParallel CC, NobelReplace CC | 3.5, 4.3, 5.5 |
| | Branemark, Nobel Speedy, Groovy | 3.3, 3.75, 4.0, 5.0 |
| OCO Biomedical | Engage | 3.25, 4.0, 5.0 |
| Southern Implants | Tri-Nex | 3.5, 4.3, 5.0 |
| | Tri-MAX7 | 7.0 |
| Straumann | Roxolid SLActive, Roxolid SLA | 3.3, 4.1, 4.8 |
| Zimmer Dental | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | 3.7, 4.1, 4.7, 6.0 |
| | Spline Reliance Cylinder | 3.25, 4.0, 5.0 |
| | Spline Twist | 3.75 |
{3}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of I
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
### 510(k) Summary
# K213391
# High Retention Attachment System
# Zest Anchors, LLC
January 10, 2022
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Zest Anchors, LLC<br>2875 Loker Ave. East<br>Carlsbad, CA 92010 |
|-------------------|-----------------------------------------------------------------|
| Telephone: | +1 (442) 244-4835 |
| Official Contact | Mark Stavro, Regulatory Affairs Director |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Representative/Consultant | Melissa Burbage<br>Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: mburbage@paxmed.com<br>kthomas@paxmed.com; flarson@paxmed.com |
### DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | High Retention Attachment System |
|----------------------|----------------------------------------------------------------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices<br>(OHT1) Dental Devices (DHT1B) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K151789, LOCATOR F-Tx Attachment System, Zest Anchors, LLC
Reference Devices K072878, LOCATOR Implant Anchor Abutment, Zest Anchors, LLC K200827, LOCATOR R-Tx Attachment System, Zest Anchors, LLC
{5}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
The High Retention Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed hybrid restorations that can be attached with a snap-in system.
The High Retention Attachment System is compatible with the following implants:
| Implant Manufacturer | Implant System | Implant Diameter |
|----------------------|----------------------------------------------------------|--------------------------------------------------|
| Ace Surgical | Infinity OCTAGON | 3.3, 4.1, 4.8 |
| | Infinity TRI-CAM | 3.5, 4.3, 5.0 |
| | Infinity Internal Hex | 3.7, 4.1, 4.7, 5.1 |
| BioHorizons | Tapered Internal | 3.0 |
| | Tapered Plus | 3.8 |
| | Laser-Lok | 3.0 |
| Biomet 3i | 3iT3, NanoTite, OSSEOTITE | 3.25, 3.75, 4.0 |
| Blue Sky Bio | Quattro | 3.3, 4.1, 4.8 |
| | One Stage | 3.3, 4.1, 4.8 |
| Camlog | SCREW-LINE ROOT-LINE 2 | 3.3, 3.8, 4.3, 5.0 |
| | SCREW-LINE | 3.3, 3.8, 4.3, 5.0 |
| Dentsply | Astra Tech OsseoSpeed EV | 3.6, 4.2, 4.8 |
| | Xive, Frialit-2 | 3.4, 3.8, 4.5, 5.5 |
| | Ankylos C/X | 3.5 |
| | Astra Tech OsseoSpeed TX | 3.5, 4.0, 4.5, 5.0 |
| Hiossen, Inc. | SS | 3.5, 4.0, 4.5 |
| | ET, TS | 3.5, 4.0, 4.5, 5.0, 6.0, 7.0 |
| | US | 3.5, 3.3, 4.5, 4.0, 5.0, 6.0, 7.0 |
| Implant Direct | Swish Plus | 4.1, 4.8 |
| | SwishTapered | 4.1, 4.8 |
| | InterActive | 3.2, 3.7, 4.3, 5.0 |
| | ReActive | 3.7, 4.2, 4.7, 5.7 |
| | RePlant | 3.5, 4.3, 5.0 |
| | Legacy 1, 2, 3, 4 | 3.7,4.2 |
| | Legacy 2, 3, 4 | 3.2, 4.7, 5.2 |
| | Legacy 1, 3 | 5.7 |
| | Legacy 2, 4 | 5.7, 7.0 |
| | Legacy 1 | 4.7 |
| IDS | MegaGen Any Ridge | 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0 |
| Keystone | PrimaConnex | 3.5, 4.1, 5.0 |
| | Genesis | 3.5 |
| | TILOBEMAXX | 7.0, 8.0, 9.0 |
| MIS | C1 | 3.3, 3.75, 4.2 |
| | V3 | 3.3, 3.9, 4.3, 5.0 |
| | SEVEN | 3.3 |
| | M4 | 3.3 |
| Nobel | Replace, Replace Select, NobelSpeedy | 3.5, 4.3, 5.0 |
| | NobelActive, NobelParallel CC, NobelReplace CC | 3.5, 4.3, 5.5 |
| | Branemark, Nobel Speedy, Groovy | 3.3, 3.75, 4.0, 5.0 |
| OCO Biomedical | Engage | 3.25, 4.0, 5.0 |
| Southern Implants | Tri-Nex | 3.5, 4.3, 5.0 |
| | Tri-MAX7 | 7.0 |
| Straumann | Roxolid SLActive, Roxolid SLA | 3.3, 4.1, 4.8 |
| Zimmer Dental | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | 3.7, 4.1, 4.7, 6.0 |
| | Spline Reliance Cylinder | 3.25, 4.0, 5.0 |
| | Spline Twist | 3.75 |
{6}------------------------------------------------
510(k) Summary Page 3 of 4
## SUBJECT DEVICE DESCRIPTION
The High Retention Attachment System is a system that provides rigid connection of fixed partial and full arch restorations (fixed/detachable hybrid dentures) to endosseous dental implants. It is designed to accommodate a path of insertion on implants to accommodate a divergence of up to 20° per implant and no more than 40° of divergence between implants. The components are similar to the LOCATOR Implant Anchor Abutment, cleared in K072878; however, the retention and removal of the inserts is similar to that of the LOCATOR F-Tx Attachment System cleared in K151789.
The High Retention Attachment System consists of abutment housings, inserts, laboratory processing tools and seating and removal tools. Abutments, attachment housings and inserts are Class II subject devices (Product Code NHA) and laboratory processing tools and seating and removal tools are Class I accessories (Product Code NDP). The abutments are provided in various cuff heights with the implant/abutment connection specific to the OEM implant. The subject device abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications (UNS R56401) and are TiN (titanium nitride) coated.
### PERFORMANCE DATA
Non-clinical testing data referenced/relied upon to demonstrate substantial equivalence included: sterilization validation and biocompatibility testing. Both the material and Titanium Nitride (TiN) coating have the same specifications, processes, and are manufactured in the same facilities as the predicate F-Tx Attachment System (K151789) and predicate LOCATOR device (K072878).
Non-clinical testing data submitted to demonstrate substantial equivalence included: mechanical denture retention testing. The mechanical testing demonstrated that the retention force of the High Retention Attachment System was greater than retention force of the predicate device K151789.
The subject abutment-implant interface of the High Retention Attachment System is verified to be compatible through either OEM signed abutment drawings, OEM implant drawings, or established compatibility demonstrated with the use of LOCATOR abutments through mutual contractual agreements with the OEM, Once the specifications and tolerances have been identified, the R-Tx abutment-implant interface is then verified through engineering analysis and documented per Zest internal procedures for Line Extensions. No performance testing is required, as demonstrated for the predicate LOCATOR R-Tx Attachment System (K200827).
{7}------------------------------------------------
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device | Comparison |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| | Zest Anchors, Inc.<br>High Retention<br>Attachment System<br>K213391 | Zest Anchors, LLC<br>LOCATOR® F-Tx<br>Attachment System<br>K151789 | Zest Anchors, Inc.<br>LOCATOR® Implant<br>Anchor Abutment<br>K072878 | Zest Anchors, Inc.<br>LOCATOR R-Tx®<br>Attachment System<br>K200827 | |
| Indications<br>for Use | The High Retention<br>Attachment System is<br>designed to support fixed,<br>partial or full arch<br>restorations on endosseous<br>dental implants in the<br>mandible or maxilla for<br>the purpose of restoring<br>masticatory function. It is<br>used in fixed hybrid<br>restorations that can be<br>attached with a snap-in<br>system. | The LOCATOR® F-Tx<br>Attachment System is<br>designed to support fixed,<br>partial or full arch<br>restorations on endosseous<br>dental implants in the<br>mandible or maxilla for<br>the purpose of restoring<br>masticatory function. It is<br>used in fixed hybrid<br>restorations that can be<br>attached with a snap-in<br>system. | The Locator Implant<br>Anchor Abutment for<br>Endosseous Dental<br>Implants is appropriate for<br>use with overdentures or<br>partial dentures retained in<br>whole or in part by<br>endosseous implants in<br>the mandible or maxilla. | The LOCATOR R-Tx®<br>Attachment System is<br>designed for use with<br>overdentures or partial<br>dentures, retained in<br>whole or in part, by<br>endosseous implants in<br>the mandible or maxilla. | Same as<br>Predicate and<br>Reference |
| Design | | | | | |
| Abutment<br>Platform<br>Diameter | 2.3 mm to 7.0 mm,<br>Multiple Systems | 3.0 mm to 7.0 mm;<br>Multiple Systems | 2.3 mm to 7.0 mm,<br>Multiple Systems | 3.0 mm to 7.0 mm;<br>Multiple Systems | Same as<br>Reference |
| Abutment<br>Angle | Straight | Straight | Straight | Straight | Same |
| Abutment/<br>Implant<br>Interface | Conical, External Hex,<br>Internal Hex, Internal<br>Multi Lobe | Conical, External Hex,<br>Internal Hex, Internal<br>Multi Lobe | Conical, External Hex,<br>Internal Hex, Internal<br>Multi Lobe | Conical, External Hex,<br>Internal Hex, Internal<br>Multi Lobe | Same |
| Divergence<br>Allowance | 20°/40°<br>(except when not allowed<br>by the implant<br>manufacturer) | 20° | 20° | 30°/60° | Same |
| Prosthesis<br>Attachment<br>Type | PEEK Insert retained in<br>Denture Attachment<br>Housing | PEEK Retention Ball<br>attached to Denture<br>Attachment Housing | Nylon Insert retained in<br>Denture Attachment<br>Housing | Nylon Male<br>Retention Cap | Same as<br>Predicate |
| Materials | | | | | |
| Abutment | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Same |
| Abutment<br>Coating | TiN | TiN, TiCN | TiN | TiN, TiCN | Same |
| Prosthetic<br>Retention | PEEK | PEEK | Nylon | Nylon | Same as<br>Predicate |
# EQUIVALENCE TO MARKETED DEVICES
### CONCLUSION
The subject devices, the primary predicate device and reference devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.