The BellaTek Encode Emergence Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.
Device Story
Two-piece healing abutment (abutment and retaining screw) used in dental implant procedures; facilitates gingival tissue healing and creates emergence profile for final restoration. Device fixated on implant via retaining screw in one-stage or two-stage surgical protocols; remains in mouth until soft tissue develops. Features machined markings on occlusal surface for identification during intraoral scanning or impression taking. Used by dentists in clinical settings. Benefits patient by shaping soft tissue for final prosthesis. Machined from Titanium Alloy (Ti-6Al-4V ELI); color anodized pink for aesthetics. Provided sterile via gamma irradiation.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: sterilization validation (ISO 11137-2, SAL 10^-6), biocompatibility (ISO 10993-1, ISO 10993-5), and MRI compatibility assessment. Shelf life of 5 years leveraged from predicate K173374.
Technological Characteristics
Two-piece healing abutment and retaining screw; material: Titanium Alloy Ti-6Al-4V ELI (ASTM F136). Features machined identification markings on occlusal surface. Color anodized pink. Sterilized via gamma irradiation. MR conditional. Available in various diameters, emergence profiles, and heights.
Indications for Use
Indicated for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration in patients requiring dental implant prosthetic rehabilitation.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Certain BellaTek Encode Healing Abutments and Certain EP Healing Abutments (K072642)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Biomet 3i LLC Krupal Patel Principal Regulatory Specialist 4555 Riverside Drive Palm Beach Gardens, Florida 33410
Re: K212730
Trade/Device Name: BellaTek Encode Emergence Healing Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 19, 2021 Received: November 22, 2021
Dear Krupal Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): K212730
Device Name: BellaTek Encode Emergence Healing Abutments
# Indications for Use:
The BellaTek Encode Emergence Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# BellaTek Encode Emergence Healing Abutments 510(k) Summary K212730 12/21/2021
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92
### I. Submitter Information:
| Name: | Biomet 3i LLC |
|----------|-----------------------------------------------------------|
| Address: | 4555 Riverside Drive<br>Palm Beach Gardens, Florida 33410 |
| Phone: | (561) 776-6923 |
| Fax: | (561) 514-6316 |
Contact Person: Krupal Patel Job Title: Principal Regulatory Specialist Email: krupal.patel(@zimmerbiomet.com
- II. Proprietary Trade Name: BellaTek Encode Emergence Healing Abutments
- III. Device Classification Name: Endosseous Dental Implant Abutment (21 CFR 872.3630)
- IV. Regulatory Class: Class II
- V. Product Code: NHA
#### Predicate Devices: VI.
Primary predicate device:
- . TSV BellaTek Encode Healing Abutments (K173374)
Reference device:
- . Eztetic BellaTek Encode Healing Abutments (K170013)
- Certain BellaTek Encode Healing Abutments and Certain EP Healing Abutments referenced in Biomet 3i Dental Abutments and Restorative Components (K072642)
#### VII. Product Description:
The BellaTek Encode Emergence Healing Abutment is a two-piece healing abutment (abutment with a retaining screw) designed to facilitate gingival tissue healing before a final restoration is placed. It consists of an abutment and a retaining screw that are assembled and packaged together and provided sterile. Both components are machined from Titanium Alloy Ti-6Al-4V ELI
{4}------------------------------------------------
(ASTM F136). They are available in pre-defined diameters, emergence profiles and heights to accommodate varying patient anatomies. The shelf life of the TSV BellaTek Encode Healing Abutment is 5 years from the date of manufacture and they are intended for single use only. The device is packaged in a blister tray with Tyvek Lid and sold sterile. The device is sterilized using the gamma irradiation method. The BellaTek Encode Emergence Healing Abutments are color anodized for aesthetic purposes.
# Indications for Use: VIII.
The BellaTek Encode Emergence Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.
# IX. Summary of the Technological Characteristics:
The BellaTek Encode Emergence Healing Abutments are designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final prosthesis. They have the added design feature of machined markings for identification when taking an abutment level impression or an intraoral scan/digital impression. Identification and orientation information is captured in the intraoral scan or model scan.
The BellaTek Encode Emergence Healing Abutments consist of three variations differing primarily on the implant connection: TSV BellaTek Encode Emergence Healing Abutments, Eztetic BellaTek Encode Emergence Healing Abutments, and Certain BellaTek Encode Emergence Healing Abutments. The retaining screw is used to secure the abutment to the implant. The subject device seats on the lead-in bevel (TSV and Eztetic) or the seating surface (Certain) of the implant connection. A substantial equivalence comparison of subject and predicate device is provided in Table 1.
The BellaTek Encode Emergence Healing Abutments and retaining screws are color anodized pink for improved aesthetics. The primary predicate device does not have color anodization, however the reference device in K072642 does have color anodization in a manner similar to the subject devices. The landmark pattern on the occlusal surface of the BellaTek Encode Emergence Healing Abutments represents a simplified "code scheme" that achieves same functionality as the predicate device. The BellaTek Encode Emergence Healing Abutments are offered in additional emergence profiles and slightly different abutment heights to accommodate varying patient anatomy as compared to the predicate device. Risk Analysis was conducted in accordance to ISO 14971 to evaluate the effect of the modifications on the subject device. The Biocompatibility evaluation, MRI assessment and Design Validation studies have shown that the device changes, including emergence profile sizes, full color anodization and machined marking designs, do not raise new risks and it concludes that the differences between the predicate and the subject device do not impact the substantial equivalence of the product.
{5}------------------------------------------------
| Feature | Subject Device<br>BellaTek Encode<br>Emergence Healing<br>Abutment | Primary Predicate<br>TSV BellaTek<br>Encode Healing<br>Abutment<br>(K173374) | Reference Device 1<br>Eztetic BellaTek<br>Encode Healing<br>Abutment<br>(K170013) | Reference Device 2<br>Certain BellaTek<br>Encode Healing<br>Abutment & Certain<br>EP Healing<br>Abutment<br>(K072642) |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for<br>Use | The BellaTek<br>Encode Emergence<br>Healing Abutments<br>are intended for use<br>as an accessory to<br>endosseous dental<br>implants during<br>endosseous and<br>gingival healing to<br>prepare gingival<br>tissue for acceptance<br>of a final abutment<br>and restoration. | The TSV BellaTek<br>Encode Healing<br>Abutments are<br>intended for use as<br>an accessory to<br>endosseous dental<br>implants during<br>endosseous and<br>gingival healing to<br>prepare gingival<br>tissue for<br>acceptance of a<br>final abutment and<br>restoration. | The Eztetic<br>BellaTek Encode<br>Healing Abutments<br>are intended for use<br>as an accessory to<br>endosseous dental<br>implants during<br>endosseous and<br>gingival healing to<br>prepare gingival<br>tissue for acceptance<br>of a final abutment<br>and restoration. | Biomet 3i Dental<br>Abutments are<br>intended for use as<br>accessories to<br>endosseous dental<br>implant to support a<br>prosthetic device in a<br>partially or<br>completely<br>edentulous patient. A<br>dental abutment is<br>intended for use to<br>support single and<br>multiple tooth<br>prosthesis, in the<br>mandible or maxilla.<br>The prosthesis can<br>be screw retained or<br>cement retained.<br>Restorative<br>Components:<br>• Temporary<br>Healing<br>Abutments are<br>intended for use<br>to shape and<br>maintain the soft<br>tissue opening<br>during healing.<br>• Castable<br>restorative<br>components are<br>intended for use<br>as accessories to<br>endosseous dental<br>implants to aid in<br>the fabrication of<br>dental prosthetics.<br>• Screw |
{6}------------------------------------------------
| Feature | Subject Device<br>BellaTek Encode<br>Emergence Healing<br>Abutment | Primary Predicate<br>TSV BellaTek<br>Encode Healing<br>Abutment<br>(K173374) | Reference Device 1<br>Eztetic BellaTek<br>Encode Healing<br>Abutment<br>(K170013) | Reference Device 2<br>Certain BellaTek<br>Encode Healing<br>Abutment & Certain<br>EP Healing<br>Abutment<br>(K072642) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating<br>Principle | The BellaTek<br>Encode Emergence<br>Healing Abutment<br>aids in prosthetic<br>rehabilitation by<br>supporting the<br>surrounding gingival<br>tissue during the<br>healing period. The<br>healing abutment is<br>fixated on the<br>implant using a<br>retaining screw<br>immediately after<br>implant placement in<br>a one-stage surgical<br>protocol. In a two-<br>stage surgical<br>protocol, the healing<br>abutment is fixated<br>on the implant<br>following the bone<br>healing period. The<br>soft tissue is sutured<br>around the healing<br>abutment and the<br>device remains in the<br>mouth until the soft<br>tissue fully develops. | The TSV BellaTek<br>Encode Healing<br>Abutment aids in<br>prosthetic<br>rehabilitation by<br>supporting the<br>surrounding<br>gingival tissue<br>during the healing<br>period. The healing<br>abutment is fixated<br>on the implant<br>using a retaining<br>screw immediately<br>after implant<br>placement in a one-<br>stage surgical<br>protocol. In a two-<br>stage surgical<br>protocol, the<br>healing abutment is<br>fixated on the<br>implant following<br>the bone healing<br>period. The soft<br>tissue is sutured<br>around the healing<br>abutment and the<br>device remains in<br>the mouth until the<br>soft tissue fully<br>develops. | The Eztetic<br>BellaTek Encode<br>Healing Abutment<br>aids in prosthetic<br>rehabilitation by<br>supporting the<br>surrounding gingival<br>tissue during the<br>healing period. The<br>healing abutment is<br>fixated on the<br>implant using a<br>retaining screw<br>immediately after<br>implant placement<br>in a one-stage<br>surgical protocol. In<br>a two-stage surgical<br>protocol, the healing<br>abutment is fixated<br>on the implant<br>following the bone<br>healing period. The<br>soft tissue is sutured<br>around the healing<br>abutment and the<br>device remains in<br>the mouth until the<br>soft tissue fully<br>develops. | components are<br>intended for use<br>as accessories to<br>endosseous dental<br>implants for<br>retention of screw<br>retained<br>abutments to the<br>dental implant.<br>The Certain<br>BellaTek Encode<br>Healing Abutment<br>aids in prosthetic<br>rehabilitation by<br>supporting the<br>surrounding gingival<br>tissue during the<br>healing period. The<br>healing abutment is<br>fixated on the<br>implant using a<br>retaining screw<br>immediately after<br>implant placement in<br>a one-stage surgical<br>protocol. In a two-<br>stage surgical<br>protocol, the healing<br>abutment is fixated<br>on the implant<br>following the bone<br>healing period. The<br>soft tissue is sutured<br>around the healing<br>abutment and the<br>device remains in the<br>mouth until the soft<br>tissue fully develops. |
{7}------------------------------------------------
| Feature | Subject Device<br>BellaTek Encode<br>Emergence Healing<br>Abutment | Primary Predicate<br>TSV BellaTek<br>Encode Healing<br>Abutment<br>(K173374) | Reference Device 1<br>Eztetic BellaTek<br>Encode Healing<br>Abutment<br>(K170013) | Reference Device 2<br>Certain BellaTek<br>Encode Healing<br>Abutment & Certain<br>EP Healing<br>Abutment<br>(K072642) |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Identification | Scheme<br>(Machined<br>Markings) | Scheme<br>(Machined<br>Markings) | Scheme<br>(Machined<br>Markings) | Scheme<br>(Machined<br>Markings) |
| Emergence<br>Profile (mm) | TSV Abutments:<br>3.8, 4.5, 5.0, 5.5, 5.7,<br>6.5, 6.8 and 7.5<br>Eztetic Abutments:<br>3.7 and 4.5<br>Certain Abutments:<br>3.8, 4.1, 5.0, 6.0, 6.8<br>and 7.5 | 3.8, 5.0, 5.6, 6.0<br>and 6.8 | 3.8 and 5.0 | 3.8, 4.1, 5.0, 5.6, 6.0,<br>6.8 and 7.5 |
| Abutment<br>Height (mm) | TSV Abutments:<br>3.0, 5.0 and 7.0<br>Eztetic Abutments:<br>3.0, 5.0 and 7.0<br>Certain Abutments:<br>3.0, 5.0 and 7.0 | 3.0, 5.0 and 7.0 | 3.0, 4.0, 6.0 and 8.0 | 3.0, 4.0, 6.0 and 8.0 |
| Restorative<br>Platform<br>Diameter (mm) | TSV Abutments:<br>3.5, 4.5 and 5.7<br>Eztetic Abutments:<br>2.9<br>Certain Abutments:<br>3.4, 4.1, 5.0 and 6.0 | 3.5, 4.5. 5.7 | 2.9 | 3.4, 4.1, 5.0 and 6.0 |
| Mating Implant | TSV Abutments:<br>Zimmer Dental TSV<br>Dental Implant<br>System<br>Eztetic Abutments:<br>Zimmer Dental<br>Eztetic Dental<br>Implant System<br>Certain Abutments:<br>Biomet 3i Certain<br>(Internal Hex) Dental<br>Implant System | Zimmer Dental<br>TSV Dental<br>Implant System | Zimmer Dental<br>Eztetic Dental<br>Implant System | Biomet 3i Certain<br>(Internal Hex)<br>Dental Implant<br>System |
| Color<br>Anodization | Pink color<br>anodization | None | None | Certain EP Healing<br>Abutment include<br>color anodized<br>surfaces |
| Abutment and<br>Screw Material | Titanium Alloy<br>(Ti-6Al-4V ELI) | Titanium Alloy<br>(Ti-6Al-4V ELI) | Titanium Alloy<br>(Ti-6Al-4V ELI) | Titanium Alloy<br>(Ti-6Al-4V ELI) |
| Packaging | Bubble Tray<br>packaging<br>configuration | Bubble Tray<br>packaging<br>configuration | Bubble Tray<br>packaging<br>configuration | Bubble Tray<br>packaging<br>configuration |
| Sterilization | Supplied Sterile | Supplied Sterile | Supplied Sterile | Supplied Sterile |
{8}------------------------------------------------
| Feature | Subject Device<br>BellaTek Encode<br>Emergence Healing<br>Abutment | Primary Predicate<br>TSV BellaTek<br>Encode Healing<br>Abutment<br>(K173374) | Reference Device 1<br>Eztetic BellaTek<br>Encode Healing<br>Abutment<br>(K170013) | Reference Device 2<br>Certain BellaTek<br>Encode Healing<br>Abutment & Certain<br>EP Healing<br>Abutment<br>(K072642) |
|------------|--------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Method | (Gamma radiation) | (Gamma radiation) | (Gamma radiation) | (Gamma radiation) |
| Shelf Life | 5 years | 5 years | 5 years | 5 years |
| Single Use | Yes | Yes | Yes | Yes |
# X. Non-Clinical Testing:
Non-clinical testing data submitted or relied upon to demonstrate substantial equivalence included radiation sterilization validation according to ISO 11137-2, demonstration a sterility assurance level (SAL) of 106, biological evaluation according to ISO 10993-1 demonstrating acceptable biocompatibility and accelerated and real time aging studies by reference to K173374 demonstrating a shelf life of five years.
The subject devices have less surface area and are made out of same materials as compared to previously tested Biomet 3i devices that were used for MR compatibility testing. Hence the MR testing data is leveraged and the subject devices are labeled as MR conditional.
Risk analysis was conducted in accordance to ISO 14971 to evaluate the effect of the modifications on the subject device. The changes in design from the predicate devices have been verified through biocompatibility evaluation, MRI compatibility assessment and design validation study. The changes to the subject device do not identify new risks. nor do the changes impact the risk profile of the device compared to the predicate devices. Revised labeling was provided to ensure proper placement of the device due to differences in emergence profile. Design control activities conclude that the differences between the predicate and the subject device do not impact the substantial equivalence of the product. In addition, confirmatory cytotoxicity testing was performed via ISO 10993-5, which also demonstrated substantial equivalence.
No clinical data were included in this submission.
### XI. Conclusion:
The subject devices have demonstrated substantial equivalence to the predicate devices in that they utilize same materials and fundamental designs and also have the same intended use and principles of operation.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.