K203464 · Terrats Medical SL · NHA · Aug 4, 2021 · Dental
Device Facts
Record ID
K203464
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Aug 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with C-Base abutments are to be sent to a Terrats Medical validated milling center for manufacture.
Device Story
DESS Dental Smart Solutions C-Base abutments are two-piece titanium alloy (Ti-6Al-4V ELI) components designed to support custom CAD/CAM zirconia superstructures for dental restorations. The system includes engaging and non-engaging designs for single or multi-unit restorations. Clinicians prescribe custom abutment designs; digital files are sent to a Terrats Medical validated milling center for fabrication of the zirconia superstructure. The C-Base features a 4.7 mm post height, SelectGrip surface treatment for improved cement adhesion, and gold anodization. The device is used in dental clinics; the clinician attaches the C-Base to the endosseous implant using an included titanium alloy screw. The system provides functional and esthetic rehabilitation for patients with missing teeth. The device is supplied non-sterile and requires sterilization by the end-user.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including sterilization validation (ISO 17665-1/2), biocompatibility (ISO 10993-5/12), and reverse engineering analysis of OEM implant interfaces to confirm compatibility.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) for abutments/screws; Y-TZP (ISO 13356) for zirconia superstructures. Surface treatments: Gold anodization and SelectGrip. Design: Two-piece CAD/CAM base system. Connectivity: Digital workflow via validated milling centers. Sterilization: End-user steam sterilization.
Indications for Use
Indicated for patients requiring prosthetic restorations supported by endosseous dental implants in the maxillary or mandibular arch. Compatible with specific implant systems including Ankylos C/X, Astra Tech EV, Astra Tech OsseoSpeed, Biomet 3i Certain, Biomet 3i OSSEOTITE, FRIADENT XIVE, NobelActive, NobelReplace/NobelParallel Conical, NobelReplace (Internal tri-channel), Nobel Brånemark System, Osstem TS, Straumann Bone Level, Straumann Tissue Level, and Zimmer Screw Vent/Tapered Screw-Vent.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Terrats Medical SL % Floyd Larson President PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K203464
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 16, 2021 Received: July 19, 2021
Dear Floyd Larson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203464
Device Name
#### DESS Dental Smart Solutions
#### Indications for Use (Describe)
DESS Dental Smart Solutions abutments are in conjunction with endosseus dental implants in the maxillary or mandbular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with C-Base abutments are to be sent to a Terrats Medical validature. Compatible Implant Systems
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform |
|-----------------------------------------------------|---------------------------|-------------------|
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 mm |
| Astra Tech EV | 3.6 | 2.9 mm |
| | 4.2 | 3.5 mm |
| | 4.8 | 4.1 mm |
| | Astra Tech OsseoSpeed™ | 3.5/4.0 |
| 4.5/5.0 | | 4.5/5.0 mm |
| Biomet 3i Certain® | 3.25 | 3.45 mm |
| | 4.0 | 4.1 mm |
| | 5.0 | 5.0 mm |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 mm |
| | 3.75, 4.0 | 4.1 mm |
| | 5.0 | 5.0 mm |
| FRIADENT XIVE® | 3.4 | 3.4 mm |
| | 3.8 | 3.8 mm |
| | 4.5 | 4.5 mm |
| | 5.5 | 5.5 mm |
| NobelActive®,<br>NobelReplace/NobelParallel Conical | 3.5 | NP (3.5 mm) |
| | 4.3, 5.0 | RP (3.9 mm) |
| | 5.5 | WP (5.1 mm) |
| NobelReplace® (Internal tri-channel) | 3.5 | NP (3.5 mm) |
| | 4.3 | RP (4.3 mm) |
| | 5.0 | WP (5.0 mm) |
| | 6.0 | 6.0 (6.0 mm) |
| Nobel Brånemark System® | 3.3 | NP (3.5 mm) |
| | 3.75, 4.0 | RP (4.1 mm) |
| Osstem TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular (3.35 mm) |
| Straumann® Bone Level | 3.3 | NC (3.3 mm) |
| | 4.1/4.8 | RC (4.1/4.8 mm) |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN (4.8 mm) |
| | 4.8 | WN (6.5 mm) |
| Zimmer Screw Vent®/ Tapered Screw-<br>Vent® | 3.3, 3.7, 4.1 | 3.5 mm |
| | 4.7 | 4.5 mm |
| | 6.0 | 5.7 mm |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person it to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Terrats Medical SL DESS Dental Smart Solutions
### K203464
#### August 4, 2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda, 75-99<br>Barberà del Vallès<br>Barcelona, Spain | |
|---------------------------|--------------------------------------------------------------------------------------|--------------------|
| | Telephone | +34 935 646 006 |
| | Fax | +34 935 646 006 |
| Official Contact | Roger Terrats, COO | |
| Representative/Consultant | Floyd G. Larson, MS, MBA | |
| | Kevin Thomas, PhD | |
| | PaxMed International, LLC | |
| | 12264 El Camino Real, Suite 400 | |
| | San Diego, CA 92130 | |
| | Telephone: | +1-858-792-1235 |
| | Fax: | +1-858-792-1236 |
| | Email: | FLarson@paxmed.com |
| | | KThomas@paxmed.com |
### DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name Common Name Regulation Number Regulation Name Regulatory Class Product Code
Classification Panel Reviewing Division
DESS Dental Smart Solutions Dental implant abutment 21 CFR 872.3630 Endosseous dental implant abutment Class II NHA
Dental Products Panel DHT1B: Division of Dental Devices
### PREDICATE DEVICE INFORMATION
Primary Predicate Device K191986, DESS Dental Smart Solutions, Terrats Medical SL
Additional Predicate Device K170588, DESS Dental Smart Solutions, Terrats Medical SL
{4}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with C-Base abutments are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant Systems | | |
|-----------------------------------------|---------------------------|-------------------|
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform |
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 mm |
| Astra Tech EV | 3.6 | 2.9 mm |
| | 4.2 | 3.5 mm |
| | 4.8 | 4.1 mm |
| Astra Tech OsseoSpeed™ | 3.5/4.0 | 3.5/4.0 mm |
| | 4.5/5.0 | 4.5/5.0 mm |
| Biomet 3i Certain® | 3.25 | 3.45 mm |
| | 4.0 | 4.1 mm |
| | 5.0 | 5.0 mm |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 mm |
| | 3.75, 4.0 | 4.1 mm |
| | 5.0 | 5.0 mm |
| FRIADENT XIVE® | 3.4 | 3.4 mm |
| | 3.8 | 3.8 mm |
| | 4.5 | 4.5 mm |
| | 5.5 | 5.5 mm |
| NobelActive® | 3.5 | NP (3.5 mm) |
| NobelReplace/NobelParallel Conical | 4.3, 5.0 | RP (3.9 mm) |
| | 5.5 | WP (5.1 mm) |
| NobelReplace® (Internal tri-channel) | 3.5 | NP (3.5 mm) |
| | 4.3 | RP (4.3 mm) |
| | 5.0 | WP (5.0 mm) |
| | 6.0 | 6.0 (6.0 mm) |
| Nobel Brånemark System® | 3.3 | NP (3.5 mm) |
| | 3.75, 4.0 | RP (4.1 mm) |
| Osstem TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular (3.35 mm) |
| Straumann® Bone Level | 3.3 | NC (3.3 mm) |
| | 4.1/4.8 | RC (4.1/4.8 mm) |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN (4.8 mm) |
| | 4.8 | WN (6.5 mm) |
| Zimmer Screw Vent®/ Tapered Screw-Vent® | 3.3, 3.7, 4.1 | 3.5 mm |
| | 4.7 | 4.5 mm |
| | 6.0 | 5.7 mm |
{5}------------------------------------------------
### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to expand the DESS Dental Smart Solutions abutment system cleared under K170588 and K191986 to add an additional series of titanium base components for previously cleared OEM implant platform compatibilities. The new components, referred to as C-Base engaging and C-Base non-engaging abutments, are available in a range of abutment gingival heights and abutment platform diameters. This submission includes abutments compatible with 33 implant platforms from 13 implant compatibilities have been cleared in previous Terrats Medical submissions. No new implant compatibilities are added in this submission. Screws used with the subject device C-Base abutments were cleared previously, with the exception of two screws added in this submission. The two screws added in this submission are specific to the Zimmer Screw Vent® / Tapered Screw-Vent® Implants and Ankylos C/X Implants and are not for use with any other previously cleared implant bodies.
The subject device DESS Dental Smart Solutions C-Base abutments are similar to TiBase Abutments cleared in K170588 and K191986. C-Base abutments are two-piece abutments designed to support a custom CAD/CAM zirconia superstructure on which a single-unit or multiunit restoration may be placed. The ceramic superstructure produced through CAD/CAM is the second part of the two-piece abutment. The C-Base also may support a ceramic hybrid abutment (direct restoration) in which the crown is included in the design of the zirconia superstructure. They are available either in designs that engage with the anti-rotational feature of the implant or in non-engaging designs for multi-unit restorations. The C-Base post is 4.7 mm high. The gingival height of the abutment (distance from implant platform to abutment platform) ranges from 0.3 mm to 3.0 mm. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. C-Base abutments are made of titanium alloy (Ti-6Al-4V) with anodization and a SelectGrip® surface, described below. When used for a direct crown, a POM burn-out sleeve, an exempt laboratory component that is not a subject of this submission, is available for laboratory fabrication of the prosthesis.
When the C-Base abutment is used with a CAD/CAM zirconia superstructure, or for a direct restoration, design parameters are identical to those cleared in K191986 and in K170588, except that the minimum post height for the subject device is 4.7 mm. The superstructure or direct restoration design parameters are:
Minimum wall thickness – 0.4 mm Minimum post height for single-unit loading - 4.7 mm Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm Zirconia superstructures and direct restorations are not intended for angulation correction.
Manufacture of the CAD/CAM zirconia superstructure is to be performed at a Terrats Medical validated milling center, defined as a facility that is registered with FDA as a manufacturer or contract manufacturer.
Each abutment is supplied with the appropriate abutment screw for attachment to the corresponding implant. DESS Dental Smart Solutions screws are designed to attach the abutment or restoration to the implant. With the exception of two new subject device screws, all screws were cleared in previous Terrats Medical submissions. As discussed above, the two screws added in this submission are specific to the Zimmer Screw Vent® / Tapered Screw-Vent® Implants and Ankylos C/X Implants and are not for use with any other previously cleared implant bodies.
All subject device abutments and subject device screws are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Vanadium ELI (Extra
{6}------------------------------------------------
Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). Zirconia superstructures for C-Base abutments are made of Y-TZP conforming to ISO 13356 Implants for surgery -Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).
Abutments are colored gold by an anodization process in which the abutment is submerged in an electrolytic solution and exposed to an electric current to achieve the gold color. They also are treated with the SelectGrip® surface to improve adhesion of the cement that is used to attach the superstructure or restoration to the C-Base abutment. The gold anodized surface treatment is identical to that cleared in K191986 and the SelectGrip surface treatment is identical to that cleared in K191986 and K170588. All the subject device components are manufactured from the same materials, are treated with the same surface treatments and are manufactured in the same facilities using the same manufacturing processes as used for the previously cleared predicate devices in K191986 and K170588.
## PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1 and ISO 17665-2 referenced from the primary predicate device and the sponsor's additional predicate device; biocompatibility according to ISO 10993-5 and ISO 10993-12 referenced from the primary predicate device and the sponsor's additional predicate device; and reverse engineering analysis of OEM implant bodies, OEM abutment, and OEM abutment screws to confirm compatibility, referenced from the primary predicate device and the additional predicate device. The two screws added in this submission are specific to the Zimmer Screw Vent® / Tapered Screw-Vent® Implants and Ankylos C/X Implants. Because the minor differences between the designs of these two subject device screws and corresponding screws cleared in K170588 and K191986 are not related to implant compatibility, no new OEM analysis was needed. No clinical data were included in this submission.
### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the TiBase abutments of the primary predicate device and the sponsor's additional predicate device listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the additional predicate device.
Subject device abutments are substantially equivalent in intended use to the TiBase abutments of the primary predicate device cleared in K191968 and the additional predicate device cleared in K170588. All are intended for use with endosseous dental implants in the maxilla and mandible to provide functional and esthetic rehabilitation. The Indications for Use Statement (IFUS) for the subject device is identical to that of the primary predicate K191986 and the additional predicate K170588, except for the specific name of the subject device and the list of compatible OEM implants.
All subject device abutments are similar in design and are identical in materials and technological characteristics to the TiBase abutments of the primary predicate K191986. All are titanium base abutments intended to be completed by attaching a zirconia superstructure fabricated from Y -TZP conforming to ISO 13356 Implants for surgery - Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP) that is manufactured at a validated milling center - a facility that is registered with FDA as a manufacturer and is approved by Terrats Medical as a contract manufacturer.
{7}------------------------------------------------
The only differences between the subject devices and the TiBase abutments of the primary predicate device are slight dimensional differences in the abutment to which the zirconia superstructure is cemented. These differences include a slightly greater post height for the subject devices (4.7 mm vs. 4.2 mm), the addition of retention grooves to the post of the subject device and a slight taper in the coronal portion of the subject device. These differences are accommodated by corresponding differences in the design of the zirconia superstructure. These differences do not affect the worst-case design parameters of the corresponding superstructures and do not have any effect on substantial equivalence.
All implant compatibilities for subject devices are included among those for primary predicate devices in K191986 and for additional predicate devices in K170588. Because compatibility has been demonstrated in these predicate submissions, this submission does not include compatibility analysis.
The SelectGrip® surface on subject C-Base abutments is identical to the SelectGrip surface on equivalent abutments cleared in the primary predicate K191986 and in the additional predicate K170588.
The gold anodized surface on subject C-Base abutments is identical to the anodized surface on Aurum Abutments of the primary predicate K191986.
The two subject device screws are substantially equivalent in design, materials, and technological characteristics to those cleared in the primary predicate K191986 and additional predicate K170588. The new screw designs incorporate only changes related to the new geometries of the subject abutments and are not related to any mating features with the OEM devices.
Digital files for all CAD/CAM superstructures must be sent to a validated milling center for manufacture. DESS C-Base abutments are for fabrication of straight custom abutments only.
The subject device is to be sterilized by the end-user, the same as primary predicate device K191986 and additional predicate device K170588.
All the subject device components are manufactured from the same materials and in the same facilities using the same manufacturing processes as used for the Terrats Medical components previously cleared in K191986 and K170588. Therefore, no new biocompatibility testing has been performed, as the subject device is substantially equivalent to the predicate devices in K191986 and K170588 with regard to materials and processing.
Minor differences in the designs, dimensions or sizes among the subject device, the primary predicate device, and the additional predicate device do not affect substantial equivalence. These minor differences do not impact substantial equivalence because the only differences are in the portion of the abutment to which the CAD/CAM zirconia superstructure is attached.
## CONCLUSION
The subject device, the primary predicate device, and the sponsor's additional predicate device have the same intended use, have similar technological characteristics, and are made of the same materials. The subject device and the primary predicate device encompass the same range of physical dimensions, are packaged in the same materials, and are to be sterilized using the same methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{8}------------------------------------------------
| Subject Device | Indications for Use Statement | | |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-------------------|
| DESS Dental Smart<br>Solutions<br>Terrats Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction with<br>endosseous dental implants in the maxillary or mandibular arch to provide support for<br>prosthetic restorations. | | |
| | All digitally designed custom abutments for use with C-Base abutments are to be sent to<br>a Terrats Medical validated milling center for manufacture. | | |
| | Compatible Implant Systems | | |
| | Compatible Implant System | Implant Body<br>Diameter, mm | Implant Platform |
| | Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 mm |
| | | 3.6 | 2.9 mm |
| | Astra Tech EV | 4.2 | 3.5 mm |
| | | 4.8 | 4.1 mm |
| | Astra Tech OsseoSpeed™ | 3.5/4.0 | 3.5/4.0 mm |
| | | 4.5/5.0 | 4.5/5.0 mm |
| | | 3.25 | 3.45 mm |
| | Biomet 3i Certain® | 4.0 | 4.1 mm |
| | | 5.0 | 5.0 mm |
| | Biomet 3i OSSEOTITE® | 3.25 | 3.4 mm |
| | | 3.75, 4.0 | 4.1 mm |
| | | 5.0 | 5.0 mm |
| | FRIADENT XIVE® | 3.4 | 3.4 mm |
| | | 3.8 | 3.8 mm |
| | | 4.5 | 4.5 mm |
| | | 5.5 | 5.5 mm |
| | NobelActive® | 3.5 | NP (3.5 mm) |
| | NobelReplace/NobelParallel<br>Conical | 4.3, 5.0 | RP (3.9 mm) |
| | | 5.5 | WP (5.1 mm) |
| | | 3.5 | NP (3.5 mm) |
| | NobelReplace®<br>(Internal tri-channel) | 4.3 | RP (4.3 mm) |
| | | 5.0 | WP (5.0 mm) |
| | | 6.0 | 6.0 (6.0 mm) |
| | Nobel Brånemark System® | 3.3 | NP (3.5 mm) |
| | | 3.75, 4.0 | RP (4.1 mm) |
| | Osstem TS | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular (3.35 mm) |
| | Straumann® Bone Level | 3.3 | NC (3.3 mm) |
| | | 4.1/4.8 | RC (4.1/4.8 mm) |
| | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN (4.8 mm) |
| | | 4.8 | WN (6.5 mm) |
| | Zimmer Screw Vent®/<br>Tapered Screw-Vent® | 3.3, 3.7, 4.1 | 3.5 mm |
| | | 4.7 | 4.5 mm |
| | | 6.0 | 5.7 mm |
| Primary Predicate<br>Device | Indications for Use Statement | | |
| 191986,<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction with<br>endosseous dental implants in the maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed custom abutments for use with TiBase abutments or Pre-milled<br>(Blank) abutments are to be sent to a Terrats Medical validated milling center for<br>manufacture. | | |
| Compatible Implant Systems | | | |
| Compatible Implant System | Implant Body Diameter, mm | Implant<br>Platform | |
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 mm | |
| Astra Tech EV | 3.6 | 2.9 mm | |
| | 4.2 | 3.5 mm | |
| | 4.8 | 4.1 mm | |
| Astra Tech OsseoSpeed™ | 3.0 | 3.0 mm | |
| | 3.5/4.0 | 3.5/4.0 mm | |
| | 4.5/5.0 | 4.5/5.0 mm | |
| Biomet 3i Certain® | 3.25 | 3.45 mm | |
| | 4.0 | 4.1 mm | |
| | 5.0 | 5.0 mm | |
| Biomet 3i OSSEOTITE® | 3.25 | 3.4 mm | |
| | 3.75, 4.0 | 4.1 mm | |
| | 5.0 | 5.0 mm | |
| Camlog | 3.3 | 3.3 mm | |
| | 3.8 | 3.8 mm | |
| | 4.3 | 4.3 mm | |
| | 5.0 | 5.0 mm | |
| FRIADENT XIVE® | 3.4 | 3.4 mm | |
| | 3.8 | 3.8 mm | |
| | 4.5 | 4.5 mm | |
| | 5.5 | 5.5 mm | |
| MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 mm | |
| | 3.0 | 3.0 (3.0 mm) | |
| NobelActive®,<br>NobelParallel Conical | 3.5 | NP (3.5 mm) | |
| | 4.3, 5.0 | RP (3.9 mm) | |
| | 5.5 | WP (5.1 mm) | |
| NobelReplace® Trilobe | 3.5 | NP (3.5 mm) | |
| | 4.3 | RP (4.3 mm) | |
| | 5.0 | WP (5.0 mm) | |
| | 6.0 | 6.0 (6.0 mm) | |
| Nobel Brånemark System® | 3.3 | NP (3.5 mm) | |
| | 3.75, 4.0 | RP (4.1 mm) | |
| | 5.0 | WP (5.1 mm) | |
| Osstem TS | 3.5 | Mini (2.8 mm) | |
| | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular (3.35 mm) | |
| Straumann® Bone Level | 3.3 | NC (3.3 mm) | |
| | 4.1/4.8 | RC (4.1/4.8 mm) | |
| Straumann® Tissue Level | 3.3 | NNC (3.5 mm) | |
| | 3.3, 4.1, 4.8 | RN (4.8 mm) | |
| | 4.8 | WN (6.5 mm) | |
| Zimmer Screw Vent®/ Tapered<br>Screw-Vent® | 3.3, 3.7, 4.1 | 3.5 mm | |
| | 4.7 | 4.5 mm | |
| | 6.0 | 5.7 mm | |
Table of Substantial Equivalence – Indications for Use Statement
{9}------------------------------------------------
{10}------------------------------------------------
| Additional<br>Predicate<br>Device | Indications for Use Statement | | |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-------------------|
| K170588,<br>DESS<br>Dental<br>Smart<br>Solutions<br>Terrats<br>Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction with<br>endosseous dental implants in the maxillary or mandibular arch to provide support for<br>prosthetic restorations.<br>All digitally designed custom abutments for use with TiBase or Pre-milled Blank are to be<br>sent to a Terrats Medical validated milling center for manufacture. | | |
| | Compatible Implant Systems | | |
| | Implant System | Implant Diameter | Platform Diameter |
| | Compatibility | (mm) | (mm) |
| | 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | OsseoSpeed™ | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 |
| | FRIADENT XiVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 |
| | NobelActive® | 3.5, 4.3, 5.0 | NP, RP |
| | NobelReplace Conical | 3.5, 4.3, 5.0 | NP, RP |
| | Nobel Replace Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP |
| | Brånemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP |
| | Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC |
| | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN, WN |
| | Tapered Screw-Vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 |
{11}------------------------------------------------
| Table of Substantial Equivalence – Technological Characteristics |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.