The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant. The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant. The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant.
Device Story
Patient-specific endosseous dental implant abutments; fabricated via CAD/CAM technology. Inputs: clinician prescription instructions; implant interface geometry. Outputs: custom-designed abutments (straight or up to 30° angulation) for screw-retained, cement-retained, or friction-fit prosthetic attachment. Used in dental clinics by clinicians to restore chewing function. Benefits: provides customized substructure for final dental restorations; ensures mechanical compatibility with specific MIS implant platforms.
Clinical Evidence
Bench testing only. Includes static and dynamic compression-bending fatigue testing per ISO 14801 and geometric compatibility analyses for MIS implant platforms. Sterilization validated per ISO 17665-1 and ISO 20857.
Technological Characteristics
Materials: Titanium alloy, Zirconia. Design: Patient-specific CAD/CAM fabricated. Connection: Internal hex. Dimensions: Platform diameters 3.30mm to 6.0mm; heights 3.3mm to 13mm. Angulation: Straight to 30°. Sterilization: Moist heat and dry heat.
Indications for Use
Indicated for partially or completely edentulous patients requiring single or multiple tooth prosthetic support in the mandible or maxilla via endosseous implants. Compatible with MIS M4 and SEVEN implant systems. MIS short implants (6mm) restricted to straight abutments.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
ATLANTIS® Abutment for HIOSSEN ET Implant (K160626)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration".
November 30, 2017
Dentsply Sirona Karl Nittinger Senior Manager, Corporate Regulatory Affairs 221 West Philadelphia Street Suite 60W York, Pennsylvania 17401
Re: K172225
Trade/Device Name: ATLANTIS® Abutment for MIS Implant Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: August 31, 2017 Received: September 1, 2017
### Dear Karl Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
# Andrew I. Steen -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172225
Device Name ATLANTIS® Abutment for MIS Implant
### Indications for Use (Describe)
The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant.
The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant.
The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment. The abutment. The abutment. The abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant.
ATLANTIS® Abutment for MIS implant is compatible with MIS implant System. MIS short implants (6mm) are to be used only with straight abutments.
ATLANTIS® products are compatible with the implants shown in the table below.
| Implant manufacturer | MIS-IMPLANT TECHNOLOGIES INC |
|----------------------|------------------------------|
|----------------------|------------------------------|
| Trade Name | Abutment Platform Diameter | Implant Diameter |
|------------------------------------------|----------------------------|------------------|
| MIS Implant M4 & SEVEN Narrow Platform | Ø3.30 mm | Ø3.30 mm |
| MIS Implant M4 & SEVEN Standard Platform | Ø3.75 and 4.2 mm | Ø3.75 and 4.2 mm |
| MIS Implant M4 & SEVEN Wide Platform | Ø5.0 and 6.0 mm | Ø5.0 and 6.0 mm |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized, abstract shape on the left, resembling a curved leaf or wave in gray. To the right of the shape are the words "Dentsply" on the top line and "Sirona" on the bottom line, both in a simple, sans-serif font, also in gray. The overall design is clean and modern.
# SECTION 5. 510(k) SUMMARY K172225
# ATLANTIS® Abutment for MIS Implant
- 1. Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
| Contact Person: | Karl Nittinger |
|-------------------|-------------------|
| Telephone Number: | 717-849-4424 |
| Fax Number: | 717-849-4343 |
| Date Prepared: | November 30, 2017 |
#### 2. Device Name:
- Proprietary Name: .
- Classification Name: ●
- Endosseous dental implant abutment CFR Number: 21 CFR 872.3630
- Device Class: .
- II Product Code: NHA
#### 3. Predicate Device:
●
| Predicate Device Name | 510(k) | Company Name |
|-----------------------------------------------------------|---------|------------------------------|
| ATLANTIS® Abutment for<br>HIOSSEN ET<br>Implant | K160626 | DENTSPLY IMPLANTS |
| MIS Dental Implant System<br>(Reference Predicate Device) | K040807 | MIS-IMPLANT TECHNOLOGIES INC |
| MIS Short Implants<br>(Reference Predicate Device) | K103089 | MIS-IMPLANT TECHNOLOGIES INC |
ATLANTIS® Abutment for MIS Implant
#### 4. Description of Device:
The proposed ATLANTIS® Abutment for MIS Implant is an endosseous dental implant abutment. The subject device is provided for implant diameter (Ø3.3, 3.75, 4.2, 5.0 and 6.0 mm) and three designs: ATLANTIS® Abutment for MIS Implant, ATLANTIS® Crown Abutment for MIS Implant and ATLANTIS® Conus Abutment for MIS Implant, see table 5-1. All are patient-specific abutments fabricated using CAD/CAM technology at Dentsply Implant manufacturing sites. Each abutment is designed according to prescription instructions from the clinician to support a screw-retained, cement-retained or friction fit prosthesis.
{4}------------------------------------------------
| Table 5-1: Compatibility table (The ATLANTIS® Abutment Titanium, Zirconia and Gold-<br>shaded are compatible with MIS implant interface (Ø3.3, 3.75, 4.2, 5.0 and 6.0 mm) | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------|
| Implant<br>Manufacturer | Interface | ATLANTIS®<br>Abutment<br>for MIS<br>Implant | ATLANTIS®<br>Crown<br>Abutment<br>for MIS<br>Implant | ATLANTIS®<br>Conus<br>Abutment<br>for MIS<br>Implant<br>(Custom) | ATLANTIS®<br>Conus<br>Abutment for<br>MIS implant<br>(Overdenture) |
| MIS-IMPLANT<br>TECHNOLOGIES<br>INC | MIS M4 &<br>SEVEN Implant<br>System (MIS<br>implant) (Ø3.3,<br>3.75, 4.2, 5.0 and<br>6.0 mm) | Titanium,<br>Zirconia,<br>Gold-shaded<br>Titanium<br>(Gold-Hue) | Titanium,<br>Zirconia | Titanium,<br>Gold-shaded<br>Titanium<br>(Gold-Hue) | Titanium |
The coronal portion of the ATLANTIS® Abutment can be fabricated as a conventional abutment for prosthesis attachment (ATLANTIS® Abutment or ATLANTIS® Conus Abutment) or fabricated as a single tooth final restoration onto which porcelain is added (ATLANTIS® Crown Abutment). The ATLANTIS® abutment interface is compatible with the MIS implants from the MIS Implant System (K040807) and MIS implants from MIS Short Implants (K103089).
The MIS M4 & SEVEN implant interface has an internal hex connection and provided for implant platform diameter Narrow (3.30 mm), Standard (3.75 and 4.20 mm) and Wide (5.0 and 6.0 mm). The abutment height ranges from 3.3 to 13 mm, the maximum abutment height is 15 mm above implant interface and the minimum abutment height is 4 mm above the trans-mucosal collar. The abutment is provided straight and up to 30° of angulation.
{5}------------------------------------------------
- న. Indications for Use:
The ATLANTIS® Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS® Abutment to the endosseous implant.
The ATLANTIS® Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS® Crown Abutment to the endosseous implant.
The ATLANTIS® Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS® Conus Abutment to the endosseous implant.
ATLANTIS® Abutment for MIS implant is compatible with MIS implant from MIS Implant System. MIS short implants (6mm) are to be used only with straight abutments.
ATLANTIS® products are compatible with the implants shown in the table below.
| Trade Name | Abutment Platform Diameter | Implant Diameter |
|---------------------------------------------|----------------------------|------------------|
| MIS Implant M4 & SEVEN<br>Narrow Platform | Ø3.30 mm | Ø3.30 mm |
| MIS Implant M4 & SEVEN<br>Standard Platform | Ø3.75 and 4.2 mm | Ø3.75 and 4.2 mm |
| MIS Implant M4 & SEVEN<br>Wide Platform | Ø5.0 and 6.0 mm | Ø5.0 and 6.0 mm |
Implant manufacturer – MIS-IMPLANT TECHNOLOGIES INC
{6}------------------------------------------------
#### 6. Substantial Equivalence:
### Technological Characteristics
ATLANTIS® Abutment for MIS implant is a patient specific restorative device designed under the control of Dentsply Implants and manufactured by Dentsply Implants using CAD/CAM technology.
Table 5-2 and 5-3 below summarizes the differences and similarities of the subject and predicate devices.
| Table 5-2: Indications for use for the proposed and the predicate devices | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ATLANTIS® Abutment for<br>MIS Implant<br>(Proposed Device) | ATLANTIS® Abutment for<br>HIOSSEN ET Implant (K160626)<br>(Primary Predicate) | MIS Dental Implant System<br>(K040807)<br>(Reference Predicate) | MIS Short Implants<br>(K103089)<br>(Reference Predicate) | Summary of differences in the<br>indications for use |
| The ATLANTIS® Abutment is intended<br>for use with an endosseous implant to<br>support a prosthetic device in a<br>partially or completely edentulous<br>patient. It is intended for use to<br>support single and multiple tooth<br>prosthesis, in mandible or maxilla.<br>The prosthesis can be cemented or<br>screw retained to the abutment. The<br>abutment screw is intended to secure<br>the ATLANTIS® Abutment to the<br>endosseous implant.<br><br>The ATLANTIS® Crown Abutment<br>is intended for use with an endosseous<br>implant to function as a substructure<br>that also serves as the final<br>restoration, in a partially or<br>completely edentulous patient. The<br>abutment screw is intended to secure<br>the ATLANTIS® Crown Abutment to<br>the endosseous implant.<br><br>The ATLANTIS® Conus Abutment<br>is intended for use with an endosseous<br>implant to support a prosthetic device<br>in partially or completely edentulous<br>patients. It is intended for use to | The ATLANTIS® Abutment is intended<br>for<br>use with an endosseous implant to<br>support a prosthetic device in a partially<br>or completely edentulous patient. It is<br>intended for use to support single and<br>multiple tooth prosthesis, in mandible or<br>maxilla. The prosthesis can be cemented<br>or screw retained to the abutment. The<br>abutment screw is intended to secure the<br>ATLANTIS® Abutment to the<br>endosseous implant.<br><br>The ATLANTIS® Crown Abutment is<br>intended for use with an endosseous<br>implant to function as a substructure<br>that also serves as the final restoration,<br>in a partially or completely edentulous<br>patient. The abutment screw is intended<br>to secure the ATLANTIS® Crown<br>Abutment to the endosseous implant.<br><br>The ATLANTIS® Conus Abutment is<br>intended for use with an endosseous<br>implant to support a prosthetic device in<br>partially or completely edentulous<br>patients. It is intended for use to support<br>a removable multiple tooth prosthesis,<br>in the mandible or maxilla. The | The MIS Dental Implant<br>System is indicated for use in<br>surgical and restorative<br>applications for placement in the<br>bone of the upper or lower jaw<br>to provide support for prosthetic<br>devices, such as artificial teeth,<br>in order to restore the patient's<br>chewing function. | MIS Dental Implants are intended<br>to be surgically placed in the bone<br>of the upper or lower jaw arches to<br>provide support for prosthetic<br>devices, such as artificial teeth, in<br>order to restore a patient's chewing<br>function.<br><br>When a one stage surgical<br>procedure is applied, the implant<br>may be immediately loaded when<br>good primary stability is achieved<br>and the occlusal load is<br>appropriate. MIS short implants are<br>to be used only with straight<br>abutments. | The indications for use of the<br>proposed device are similar to<br>the indications for use of the<br>primary predicate device. The<br>difference is that the indications<br>for use of the primary predicate<br>device are for Hiossen ET<br>implant system.<br><br>The indications for use of the<br>reference predicate device cover<br>the entire dental system. The<br>indications for use are similar.<br>The difference between the<br>proposed device and the<br>reference predicate device is that<br>the prosthesis, in addition to<br>screw-retained restoration or<br>cement retained restoration, can<br>be attachment-retained (friction<br>fit) to the proposed device. |
{7}------------------------------------------------
| Table 5-2: Indications for use for the proposed and the predicate devices | | | | | | MIS Dental<br>Implant System<br>(K040807)<br>(Reference<br>Predicate) | MIS Short Implants<br>(K103089)<br>(Reference Predicate) | Summary of<br>differences in<br>the<br>indications<br>for use | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------|---------------------------------------------------------------|--|
| ATLANTIS® Abutment for<br>MIS implant<br>(Proposed Device) | ATLANTIS® Abutment for HIOSSEN ET<br>Implant (K160626)<br>(Primary Predicate) | | prosthesis, in the mandible or maxilla.<br>The prosthesis is attachment-retained by<br>friction fit to the abutment. The abutment<br>screw is intended to secure the<br>ALTLANTIS® Conus Abutment to the<br>endosseous implant. | | friction fit to the abutment. The abutment screw is<br>intended to secure the ATLANTIS® Conus<br>Abutment to the endosseous implant. | | | | |
| ATLANTIS® Abutment for MIS Implant<br>is compatible with MIS implant from<br>MIS Implant System. MIS short implants<br>(6mm) are to be used only with straight<br>abutments.<br>ATLANTIS® products are compatible<br>with the implants shown in the table<br>below.<br>Implant manufacturer:<br>MIS IMPLANT TECHNOLOGIES, INC | ATLANTIS® products are compatible with the<br>implants shown in the table below.<br>Implant Manufacturer:<br>HIOSSEN INC | | | | | | | | |
| Trade Name | Abutment<br>Platform<br>Diameter | Implant<br>Diameter | Trade Name | Abutment<br>Platform<br>Diameter | Implant<br>Diameter | | | | |
| MIS Implant M4<br>& SEVEN<br>Narrow Platform | Ø3.30 mm | Ø3.30 mm | HIOSSEN ET III SA<br>Fixture Mini | Ø3.5mm | Ø3.5mm | | | | |
| MIS Implant<br>M4 & SEVEN<br>Standard<br>Platform | Ø3.75 and<br>4.2 mm | Ø3.75 and<br>4.2 mm | HIOSSEN ET III SA<br>Fixture Regular | Ø4.0, 4.5,<br>5.0, 6.0,7.0 mm | Ø4.0, 4.5, 5.0,<br>6.0, 7.0 mm | | | | |
| MIS Implant<br>M4 & SEVEN<br>Wide Platform | Ø5.0 and<br>6.0 mm | Ø5.0 and<br>6.0 mm | | | | | | | |
{8}------------------------------------------------
| Table 5-3: Similarities and differences between the proposed and the predicate devices | | | | | |
|----------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Element | Proposed Device<br>ATLANTIS® Abutment<br>for MIS Implant | Primary Predicate Device<br>ATLANTIS® Abutment for<br>HIOSSEN ET Implant | Reference Predicate Device<br>MIS Dental Implant<br>System | Reference Predicate Device<br>MIS Short Implants | Summary of differences |
| 510(k) | To be assigned | K160626 | K040807 | K103089 | - |
| Prosthesis<br>Attachment | Screw-retained<br>Cement-retained<br>Friction Fit | Screw-retained<br>Cement-retained<br>Friction Fit | Screw-retained<br>Cement-retained | Screw-retained<br>Cement-retained | No difference between the proposed<br>and the primary predicate device.<br>The reference device does not<br>indicate friction fit. |
| Restoration<br>Abutment<br>Platform<br>Diameter | Single or Multi-unit<br>3.3, 3.75, 4.2, 5.0, 6.0 | Single or Multi-unit<br>3.5, 4.0, 4.5, 5.0, 6.0, 7.0 | Single or Multi-unit<br>3.3, 3.75, 4.2, 5.0, 6.0 | Single or Multi-unit<br>4.2, 5.0, 6.0 | No difference<br>The proposed device is designed to<br>fit the MIS interface.<br>Therefore, no difference between the<br>proposed device and the reference<br>predicate device. |
| Table 5-3: Similarities and differences between the proposed and the predicate devices | | | | | | | |
|----------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------------------|--|--|
| Element | Proposed Device<br>ATLANTIS® Abutment<br>for MIS Implant | Primary Predicate Device<br>ATLANTIS® Abutment for<br>HIOSSEN ET Implant | Reference Predicate Device<br>MIS Dental Implant<br>System | Reference Predicate Device<br>MIS Short Implants | Summary of differences | | |
| Abutment<br>angle | Straight, up to 30° | Straight, up to 30° | Straight, up to 30° | Straight | No difference | | |
| Connection | Internal hex connection | Internal hex connection | Internal hex connection | Internal hex connection | There is no difference between the<br>proposed device and the reference<br>predicate device. | | |
| Material<br>Implant | NA | NA | Titanium Alloy<br>Ti 6Al 4V ELI | Titanium Alloy | The proposed device is an abutment | | |
| Material:<br>Abutment | Titanium alloy, Zirconia | Titanium alloy, Zirconia | Titanium alloy | Titanium Alloy | No difference | | |
| Material:<br>Screw | Titanium alloy | Titanium alloy | Titanium alloy | Titanium alloy | No difference | | |
{9}------------------------------------------------
#### 7. Non-Clinical Performance Data
Non-clinical test data and analyses are included to support substantial equivalence:
- . Static and dynamic compression-bending testing conducted according to ISO 14801: Dental-implants Dynamic Fatigue Test for Endosseous Dental Implants.
- Geometric analyses conducted on implant bodies, abutments, and screws to support the dimensional compatibility of the ATLANTIS® Abutment for MIS Implant with the MIS Implant Technologies, Inc., Narrow (Ø3.30 mm), Standard (Ø3.75 and 4.2 mm), and Wide (Ø5.0 and 6.0 mm), M4 and SEVEN implant platforms.
- Sterilization parameters which have been validated according to ISO 17665-1: Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices, and ISO 20857: Sterilization of health care products -- Dry heat --Requirements for the development, validation and routine control of a sterilization process for medical devices are included and are as referenced in the predicate device (K160626).
### Biocompatibility
The material composition and manufacturing processing methods of the subject ATLANTIS® Abutment for MIS Implant are identical to the predicate device, ATLANTIS® Abutment for Hiossen ET Implant (K160626). Therefore, no additional biocompatibility data is included to support substantial equivalence.
#### 8. Conclusion Regarding Substantial Equivalence
The ATLANTIS® Abutment for MIS Implant is an endosseous dental implant abutment which is intended to support a prosthetic device in a partially or completely edentulous patient. The ATLANTIS® Abutment for MIS Implant has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the predicates, ATLANTIS® Abutment for HIOSSEN ET Implant (K160626), MIS Dental Implant System (K040807) and MIS Short Implants (K103089). Test data to verify the performance of the ATLANTIS® Abutment for MIS Implant has been provided with mechanical testing. The results of the test studies and dimensional compatibility analyses, combined with the design, and intended use comparison with the predicate devices, support substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.