K160244 · Thommen Medical AG · NHA · Jun 2, 2016 · Dental
Device Facts
Record ID
K160244
Device Name
VARIOunite
Applicant
Thommen Medical AG
Product Code
NHA · Dental
Decision Date
Jun 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Thommen Medical VARIOflex abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures. Thommen Medical VARIOtemp abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
Device Story
VARIOunite is a straight titanium dental implant abutment; features retention grooves; used for permanent or temporary screw-retained or cement-retained restorations. VARIOflex configuration includes a burn-out sleeve for permanent restoration fabrication; VARIOtemp configuration includes a fabrication screw for temporary restoration. Device is prepable to appropriate occlusal height using retention rings for guidance. Available in indexed (crowns) or non-indexed (bridges) connections; five platform diameters (3.5, 4.0, 4.5, 5.0, 6.0 mm). Used by dental professionals in clinical settings to support prosthetic restorations on Thommen System implants. Benefits include standardized interface for restorative support and flexibility in fabrication.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by non-clinical bench testing, including sterilization validation (ISO 17665-1, ISO/TR 17665-2) and biocompatibility evaluation (ISO 10993-1).
Technological Characteristics
Material: CPTi (Commercially Pure Titanium). Design: Straight, prepable abutment with indexed or non-indexed connection. Dimensions: 3.5, 4.0, 4.5, 5.0, 6.0 mm diameters. Sterilization: Moist heat (ISO 17665-1).
Indications for Use
Indicated for patients requiring dental implant-supported restorations (crowns, bridges, overdentures) in the maxillary or mandibular arch using Thommen System dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, often associated with medicine and healthcare. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2016
Thommen Medical AG % Ms. Linda K. Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K160244
Trade/Device Name: VARIOunite Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 3, 2016 Received: May 4, 2016
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image contains a signature, which includes a stylized design and the name "Tina Kiang". Below the name, there is a small "-S". The signature is in black ink and appears to be handwritten. The background is white.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160244
Device Name
VARIOunite
Indications for Use (Describe)
#### VARIOflex
Thommen Medical VARIOflex Abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
#### VARIOtemp
Thommen Medical VARIOtemp Abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary Thommen Medical AG VARIOunite
May 3, 2016
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | Thommen Medical AG<br>Neckarsulmstrasse 28<br>2540 Grenchen, Solothurn<br>Switzerland | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|
| | Telephone: | +41 61 965 90 20 | |
| | Fax: | +41 61 965 90 21 | |
| Official Contact | Daniel Snetivy, PhD<br>CTO | | |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | | |
| | Telephone: | +1 (858) 792-1235 | |
| | Fax: | +1 (858) 792-1236 | |
| | Email: | lschulz@paxmed.com<br>flarson@paxmed.com | |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | VARIOunite |
|---------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulation | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
{4}------------------------------------------------
## PREDICATE DEVICE
The primary predicate device is K121334, VARIOeco, Thommen Medical AG. The reference predicate device is K120414. OsseoSpeed™ Plus, Astra Tech AB.
### INDICATIONS FOR USE
#### VARIOflex
Thommen Medical VARIOflex abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
#### VARIOtemp
Thommen Medical VARIOtemp abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
#### DEVICE DESCRIPTION
VARIOunite is a straight titanium dental implant abutment with retention groves that can be used for either permanent or temporary screw-retained or cement-retained restorations. Labeled as VARIOflex, it is provided with a burn-out sleeve for fabrication of a permanent restoration. Labeled as VARIOtemp, it is provided with a fabrication screw for fabrication of a temporary restoration. VARIOunite can be prepared to the appropriate occlusal height prior to fabrication of the restoration using the retention rings for guidance.
V ARIOunite is available with an indexed connection for crowns or a non-indexed connection for bridges. It is available in five platform diameters (3.5, 4.0, 4.5, 5.0 and 6.0 mm) corresponding to existing platform diameters of the Thommen System dental implants.
### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included sterilization validation according to ISO 17665-1 Sterilization of health care products — Moist heat — Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO/TR 17665-2 Sterilization of health care products — Moist heat — Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1, and biocompatibility evaluation according to ISO 10993-1 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
Clinical data were not submitted in this premarket notification.
{5}------------------------------------------------
## EQUIVALENCE TO MARKETED DEVICE
A comparison of the technological characteristics of the subject device and the primary predicate device K121334 is provided in the following table.
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Thommen Medical AG<br>VARIOunite<br>K160244 | Thommen Medical AG<br>VARIOeco<br>K121334 | Astra Tech AB<br>OsseoSpeed Plus<br>K120414 |
| Indications<br>for Use | VARIOflex<br>Thommen Medical VARIOflex<br>abutments are intended to be used<br>in conjunction with Thommen<br>System dental implants in the<br>maxillary and/or mandibular arch<br>to provide support for crowns,<br>bridges and overdentures.<br>VARIOtemp Thommen Medical<br>VARIOtemp abutments are<br>intended to be used in conjunction<br>with Thommen System dental<br>implants in the maxillary and/or<br>mandibular arch to provide support<br>for crowns, bridges and<br>overdentures. | Thommen VARIOeco dental<br>implant abutments are intended to<br>be used in conjunction with<br>Thommen System dental implants<br>in the maxillary and/or mandibular<br>arch to provide support for crowns,<br>bridges and overdentures. | Astra Tech Implant System Plus<br>abutments are intended to be used<br>in conjunction with Astra Tech<br>Implant System Plus in fully or<br>partially edentulous maxillary<br>and/or mandibular arches to<br>provide support for crowns,<br>bridges or overdentures. |
| Design | | | |
| Abutment Design | Straight abutment,<br>prepable | Straight abutment,<br>non-prepable | Straight abutment,<br>prepable |
| Abutment<br>Diameter (mm) | 3.5, 4.0, 4.5, 5.0, 6.0 | 3.5, 4.0, 4.5, 5.0, 6.0 | 3.0, 3.6, 4.2, 4.8, 5.4 |
| Restoration | Single or multi-unit | Single or multi-unit | Single or multi-unit |
| Implant<br>Connection | Internal | Internal | Internal |
| Material | | | |
| Abutment | CPTi | CPTi | Titanium Alloy |
The subject device VARIOunite abutment is substantially equivalent to the primary predicate, VARIOeco K121334, in basic design and function. Both are straight titanium abutments indicated for screw-retained or cement-retained prostheses, have a precision burn-out sleeve for fabrication of the permanent prosthesis, have the same implant platform diameters, attach to the same corresponding implants in the Thommen Implant System, and are for single or multi-unit restorations. When VARIOunite is provided with a fabrication screw and labeled as VARIOtemp, it is used for a direct temporary restoration where as the primary predicate
{6}------------------------------------------------
VARIOeco uses a separate temporary abutment. VARIOtemp is substantially equivalent to the reference predicate K120414 in abutment design and function. The subject VARIOunite abutment prepable design is substantially equivalent to the prepable design of the reference predicate K120414.
# CONCLUSION
Overall, the subject device has the following similarities to the predicate devices:
- has the same intended use, ●
- uses the same operating principle,
- incorporates the same basic design, ●
- incorporates the same or very similar materials, and
- has similar packaging and is to be sterilized using the same processes.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.