Transfer & Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
Device Story
Transfer & Angled Abutment is a titanium alloy dental component used to support prosthetic restorations like crowns, bridges, or over-dentures. The device features a mini connection structure with an 11° Morse taper; some models include a 2.08mm hex connection for anti-rotation. Angled versions provide a 17° angle for prosthetic alignment. The device is used by dental professionals in a clinical setting to connect the implant to the final prosthesis. It is secured via an abutment screw. The device benefits patients by providing a stable interface for tooth replacement, facilitating functional and aesthetic restoration.
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is supported by bench testing, specifically fatigue testing conducted in accordance with the FDA guidance document for Class II Special Controls for Root-form Endosseous Dental Implants and Endosseous Dental Abutments.
Technological Characteristics
Material: Titanium Alloy (Ti-6Al-4V) with TiN coating. Connection: 11° Morse taper internal connection, mini connection. Features: 4.0mm diameter, 17° angle (angled version), hex/non-hex configurations. Energy source: None (mechanical). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring prosthetic restoration support (crowns, bridges, or over-dentures) via dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 5, 2016
OSSTEM Implant Co., Ltd % David Kim RA/OA Manager HIOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K153015
Trade/Device Name: Transfer & Angled Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 3, 2016 Received: November 3, 2016
Dear David Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo and address of OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" slightly offset and larger than the rest of the name. Below the company name is the address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. To the left of the company name is a smaller, stylized version of the word "OSSTEM" in orange.
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
### Indications for Use Statement
510(k) Number K 153015
3
Device Name : Transfer & Angled Abutment
Indication for use :
Transfer & Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
Prescription Use (Per 21CFR801 Subpart D) OR
Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# STEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
### 510(k) Summary
Date : November 23, 2016
| - Submitter's Name : | OSSTEM Implant Co., Ltd. |
|-------------------------------|----------------------------------------------|
| - Address : | 66-16, Bansong-ro 513beon-gil, Haeundae-gu, |
| | Busan, Republic of Korea |
| - Contact : | Mr. Hee Kwon Son |
| - Phone: | +82 51 850 2575 |
| - Correspondent's Name: | HIOSSEN Inc. |
| - Address: | 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact: | DAVID KIM |
| - Phone: | 267 759 7031 |
| Device : | |
| Trade or (Proprietary) Name : | Transfer & Angled Abutment |
| Common or usual name : | Dental Abutment |
| Regulation Number: | 21 CFR 872.3630 |
| Regulation Name: | Endosseous Dental Implant Abutment |
| Regulatory Class: | Class II |
| Product Code: | NHA |
### 3. Predicate Device:
2.
Primary predicate
ET Prosthetic System, OSSTEM CO., LTD., K130662 ET SS Implant System, OSSTEM CO., LTD., K120847
- 4. Description :
The Transfer & Angled Abutment is device made of titanium alloy intended for use as an aid in prosthetic restoration. It consists of Abutments and Abutment Screws. Surfaces are partially TiN coated.
The Transfer & Angled Abutment has mini connection structure that is consist two types of connection structure (Non-HEX and HEX (including A type hex)
Non-HEX connection of Transfer & Angled Abutment has a 11° Morse taper structure Hex Mini connection of Transfer & Angled Abutment has a 11° Morse taper structure and 2.08mm Hex connection structure.
Angled Abutment two types of Hex connection (A type hex, B type hex) to set direction (Double hex function)
Hex connection structure has function for anti-rotation
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, gray font. The logo is simple and modern.
# OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Item | Content | |
|----------------------|-----------------------|-----------------------------------------------------------|
| Transfer<br>Abutment | Description | Use for marking general cement-type prosthesis |
| | Material | Titanium Alloy (Ti-6Al-4V) / TiN-coating (middle coating) |
| | Diameter | Ø 4.0mm |
| | Gingival<br>heights | 1.0mm, 2.0mm, 3.0mm, 4.0mm, 5.0mm |
| | type of<br>connection | 11° Morse taper internal connection |
| | connection | Mini Connection |
| | Post Height | 5.5mm, 7.0mm |
| Angled Abutment | Description | Use for marking general cement-type prosthesis |
| | Material | Titanium Alloy (Ti-6Al-4V) / TiN-coating |
| | Diameter | Ø 4.0mm |
| | Gingival<br>heights | 2.0mm, 4.0mm |
| | type of<br>connection | 11° Morse taper internal connection |
| | connection | Mini Connection |
| | Angle | 17° |
### - Substantial Equivalence Matrix
| Part Name | Transfer Abutment | Primary predicate:<br>Transfer Abutment |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| | Proposed | Predicate |
| 510K | - | K130662 |
| Material | Titanium Alloy | Titanium Alloy |
| Design | Image: Transfer Abutment Proposed | Image: Transfer Abutment Predicate |
| Dimension<br>(Diameter) | 4.0mm | 4.6, 5.0, 6.0, 7.0mm |
| Coating | TiN coating | TiN coating |
| SE | The subject device and the predicate device (K130662) are the same except for<br>dimensional changes. There are no changes in function and intended use between<br>the subject device and the predicate devices. No biocompatibility testing,<br>sterilization validation or packaging validation needed to be performed because<br>the material, coating, manufacturing process and packaging are identical between<br>the predicate and subject device (K130662). | |
| Reason for<br>change | Addition Art# of diameter 4.0mm | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. There is a small orange superscript "6" in the upper right corner of the logo.
# OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Part Name | Angled Abutment | Primary predicate:<br>Angled Abutment |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| | Proposed | Predicate |
| 510K | - | K120847 |
| Material | Titanium Alloy | Titanium Alloy |
| Design | Image: Angled Abutment Proposed | Image: Angled Abutment Predicate |
| Dimension<br>(Diameter) | 4.0mm | 4.3, 4.5, 5.0, 5.5, 6.0mm |
| Angled | 17° | 17° |
| Coating | TiN coating | TiN coating |
| SE | The subject device and the predicate device (K120847) are the same except for<br>dimensional changes. There are no changes in function and intended use between<br>the subject device and the predicate devices. No biocompatibility testing,<br>sterilization validation or packaging validation needed to be performed because<br>the material, coating, manufacturing process and packaging are identical between<br>the predicate and subject device (K120847). | |
| Reason for<br>change | Addition Art# of diameter 4.0mm | |
### 5. Indication for use:
Transfer & Angled Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
- 6. Summary of nonclinical testing
Fatigue testing was considered according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment" with the worst case scenario.
### 7. Summary of clinical testing
No clinical studies were submitted to support substantial equivalence.
- 8. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification OSSTEM CO., LTD. concludes that Transfer & Angled Abutment is substantially equivalent to the predicate devices as described herein
{6}------------------------------------------------
K153015
Image /page/6/Picture/1 description: The image shows the logo for Osstem Implant. The logo is orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. There is a small orange circle with the number 6 in the upper right corner of the logo.
# OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.