K151295 · Biodenta Swiss AG · NHA · Dec 1, 2015 · Dental
Device Facts
Record ID
K151295
Device Name
Biodenta Customized Abutment - Titanium
Applicant
Biodenta Swiss AG
Product Code
NHA · Dental
Decision Date
Dec 1, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Biodenta Customized Abutment - Titanium is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Device Story
One-piece custom titanium abutment; used with dental implants to support single/multiple tooth prostheses. Dental technician uses CAD software to design upper portion; Biodenta milling centers manufacture device. Final cement-retained restoration constructed in lab per dentist specifications. Abutment secured to implant via abutment screw. Used in dental clinics; operated by dental professionals. Provides structural support for prosthetic teeth; enables customized fit for various implant systems.
Clinical Evidence
Bench testing only. Fatigue testing conducted per FDA Class II Special Controls Guidance Document for Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. Worst-case scenario testing confirmed sufficient mechanical fatigue strength. Biocompatibility assessment determined testing unnecessary due to established history of Ti-6Al-4V ELI in dental applications.
Technological Characteristics
Material: Ti-6Al-4V ELI (ISO 3852-3, ASTM F136). One-piece custom abutment with abutment screw. CAD/CAM manufacturing process. Non-sterile, single-use. Compatible with various implant systems (Biodenta, Nobel Biocare, Biomet 3i, Dentsply, Straumann, Zimmer, Biohorizons, Osstem). Sterilization per ISO 17665-1:2006.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple tooth prostheses supported by dental implants in the maxilla or mandible.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2015
Biodenta Swiss AG Mr. David Eiler Regulatory Manager Tramstrasse 16 Berneck, St. Gallen 9442 SWITZERLAND
Re: K151295
Trade/Device Name: Biodenta Customized Abutment - Titanium Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: November 5, 2015 Received: November 23, 2015
Dear Mr. Eiler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Biodenta. The logo consists of the word "biodenta" in gray lowercase letters, with a registered trademark symbol to the right of the word. Above the word is a curved shape in two shades of green and gray.
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K151295
Device Name
Biodenta Customized Abutment - Titanium
#### Indications for Use (Describe)
The Biodenta Customized Abutment - Titanium is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. The Biodenta Customized Abutment - Titanium is compatible with the following dental implant systems:
| Implant Brand, Type | Implant Platform Name: Implant Diameter |
|---------------------------------------------------|--------------------------------------------------------------------|
| Biodenta, Bone Level and Tapered | B1: 3.5 mm; B2: 4.1, 4.8, 6.0 mm |
| Nobel Biocare, Nobel Replace straight and tapered | NP: 3.5 mm; RP: 4.3 mm; WP: 5.0 mm; 6.0: 6.0 mm |
| Nobel Biocare, NobelActive | NP: 3.5 mm; RP: 4.3, 5.0 mm |
| Biomet 3i, Certain Internal | 3.4: 3.25, 4.0 mm; 4.1: 4.0, 5.0 mm; 5.0: 5.0, 6.0 mm; 6.0: 6.0 mm |
| Dentsply, Astra Tech OsseoSpeed | 3.5/4.0: 3.5, 4.0 mm; 4.5/5.0: 4.5, 5.0 mm |
| Straumann, Bone Level | NC: 3.3 mm; RC: 4.1, 4.8 mm |
| Zimmer, Screw Vent and Screw Vent Tapered | 3.5: 3.3, 3.7, 4.1 mm; 4.5: 4.7 mm; 5.7: 6.0 mm |
| Biohorizons, Internal straight and tapered | 3.5: 3.5, 3.8 mm; 4.5: 4.0, 4.6 mm; 5.7: 5.0, 6.0, 5.8 mm; |
| Osstem (Hiossen), GS and TS | Mini: 3.5 mm; Regular: 4.0, 4.5, 5.0 mm |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Biodenta. The logo features the word "biodenta" in lowercase letters, with a stylized graphic above the text. The graphic is a curved shape in two shades of green and gray, positioned above the "bio" portion of the word. A small "R" in a circle is located to the right of the word.
## K151295
# 5 510(k) Summary
| Owner's name: | Biodenta Swiss AG |
|-----------------------------|-----------------------------------------------|
| Address: | Tramstrasse 16<br>9442 Berneck<br>Switzerland |
| Phone: | +41 71 747 11 11 |
| Fax number: | + 41 71 747 11 12 |
| Contact person: | Mr. David Eiler, Regulatory Manager |
| Date summary prepared: | November 27, 2015 |
| Trade / proprietary name: | Biodenta Customized Abutment - Titanium |
| Common name: | Endosseous dental implant abutment |
| Device classification name: | Endosseous Dental Implant Abutment |
| Product code: | NHA |
| Regulation number : | 21 CFR 872.3630 |
Legally marketed device to which equivalence is claimed (predicate device):
| 1. Company: | Biodenta Swiss Ag |
|----------------|--------------------------------------------------------------------|
| Device name: | Biodenta Customized Abutment |
| 510(k) number: | K110778 (primary predicate) |
| 2. Company: | Prismark Dentalcraft, Inc |
| Device name: | Inclusive Titanium Abutments For Astra Tech Osseospeed<br>Implants |
| 510(k) number: | K100993 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Biodenta. The logo features the word "biodenta" in gray lowercase letters. Above the word is a curved design in light gray and light green. To the right of the word is a circled R symbol.
| 3. Company: | Prismark Dentalcraft, Inc |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Device name: | Inclusive® Titanium Abutments compatible with:<br>Straumann Bone Level, Nobel Biocare Branemark, and<br>Nobel Biocare NobelActive Implants |
| 510(k) number: | K142118 |
| 4. Company: | Prismark Dentalcraft, Inc |
| Device name: | Inclusive® Titanium Abutments, compatible with Zimmer<br>Screw- Vent, Biomet 3i Certain, and Nobel Biocare<br>NobelReplace Implants |
| 510(k) number: | K142115 |
| 5. Company: | Biodenta Swiss Ag |
| Device name: | Biodenta Dental Implant System - Multi-Use Abutment |
| 510(k) number: | K123491 |
## Indications for Use:
The Biodenta Customized Abutment - Titanium is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
The Biodenta Customized Abutment - Titanium is compatible with the following dental implant systems:
110778
| Implant Brand, Type | Implant Platform Name: Implant Diameter |
|---------------------------------------------------|-----------------------------------------------------------------------|
| Biodenta, Bone Level and Tapered | B1: 3.5 mm; B2: 4.1, 4.8, 6.0 mm |
| Nobel Biocare, Nobel Replace straight and tapered | NP: 3.5 mm; RP: 4.3 mm;<br>WP: 5.0 mm; 6.0: 6.0 mm |
| Nobel Biocare, NobelActive | NP: 3.5 mm; RP: 4.3, 5.0 mm |
| Biomet 3i, Certain Internal | 3.4: 3.25, 4.0 mm; 4.1: 4.0, 5.0 mm;<br>5.0: 5.0, 6.0 mm; 6.0: 6.0 mm |
| Dentsply, Astra Tech OsseoSpeed | 3.5/4.0: 3.5, 4.0 mm; 4.5/5.0: 4.5, 5.0 mm |
| Straumann, Bone Level | NC: 3.3 mm; RC: 4.1, 4.8 mm |
| Zimmer, Screw Vent and Screw Vent Tapered | 3.5: 3.3, 3.7, 4.1 mm; 4.5: 4.7 mm; 5.7: 6.0 mm |
| Biohorizons, Internal straight and tapered | 3.5: 3.5, 3.8 mm; 4.5: 4.0, 4.6 mm;<br>5.7: 5.0, 6.0, 5.8 mm; |
| Osstem (Hiossen), GS and TS | Mini: 3.5 mm; Regular: 4.0, 4.5, 5.0 mm |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Biodenta. The logo features the word "biodenta" in gray lowercase letters. Above the word is a curved design in light gray and light green. To the right of the word is a circled R symbol.
## Device Description:
Biodenta Customized Abutment - Titanium is a one-piece custom abutment made of titanium. It utilizes an Abutment Screw for abutment retention.
The upper portion is designed by a dental technician using CAD software and manufactured at Biodenta milling centers. The final cement retained restoration is constructed in the lab according to the dentist's specifications.
The Abutment and the Abutment Screw are made of biocompatible Ti-6AI-4V ELI conforming to ISO 3852-3 and ASTM F136.
## Non-clinical Testing Data:
Fatigue testing was conducted according to FDA Guidance: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments - Guidance for Industry and FDA Staff. The worst-case scenario for the Biodenta Customized Abutment - Titanium and implant was tested. The results show that the Biodenta Customized Abutment - Titanium has sufficient mechanical strength for the intended clinical application. A biocompatibility assessment determined that biocompatibility testing was not necessary. Titanium has a long history of use in dental applications and Biodenta also has an established history of manufacturing titanium based dental devices.
The engineering of the Biodenta Customized Abutment - Titanium was carried out by measuring the original implant manufacturer's implant analogs, abutments, and abutment screws. Compatible abutments and screws were developed based on the measurement results. An enqineering and compatibility analysis was carried out to ensure that the Biodenta Customized Abutment - Titanium fits well with the original manufacturer's implants (Nobel Biocare, Nobel Replace; Nobel Biocare, NobelActive; Biomet 3i, Certain Internal; Dentsply, Astra Tech OsseoSpeed; Straumann, Bone Level; Zimmer, Screw Vent and Screw Vent Tapered; Biohorizons, Internal straight and tapered; Osstem (Hiossen), GS/TS). Additionally fatigue testing according to FDA Guidance for Industry and Staff -Class II Special Controls Guidance Document: Root- form Endosseous Dental Implants and Endosseous Dental Implant Abutments was carried out to confirm the mechanical fatigue strength of the Biodenta Customized Abutment - Titanium on the original manufacturer's implant fixture.
The Sterilization Validation was carried out according to ISO 17665-1:2006. The recommended sterilization method is provided in the Information for Use.
## Equivalence to marketed device:
While the indications for use statements between the subject and predicate device are not identical, the intended use is still the same. They are all abutments intended to be compatible with a specific subset of implants to support dental restorations. Biodenta Swiss AG demonstrated that, for the purposes of FDA's regulation of medical devices, the Biodenta Customized Abutment - Titanium is substantially equivalent to the predicate devices in intended use, material composition, fundamental scientific technology, principles of operation, and basic design.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for "biodenta". The word "biodenta" is written in gray, with a registered trademark symbol to the right of the "a". Above the word is a curved design in gray and light green.
#### Summary Substantial Equivalence Comparison to predicate devices:
| | Subject Device | Primary Predicate | Reference Predicates | | | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Biodenta Swiss AG | Biodenta Swiss AG | Prismark<br>Dentalcraft, Inc | Prismark<br>Dentalcraft, Inc | Prismark<br>Dentalcraft, Inc | Biodenta Swiss<br>AG | Pou Yu Biotechnology<br>Co., Ltd. |
| Device<br>Name | Biodenta<br>Customised<br>Abutment -<br>Titanium | Biodenta Customized<br>Abutment | Inclusive Titanium<br>Abutments For Astra<br>Tech Osseospeed<br>Implants | Inclusive® Titanium<br>Abutments compatible<br>with: Straumann Bone<br>Level, Nobel Biocare<br>Branemark , and Nobel<br>Biocare NobelActive<br>Implants | Inclusive® Titanium<br>Abutments, compatible<br>with Zimmer Screw-<br>Vent, Biomet 3i Certain,<br>and Nobel Biocare<br>NobelReplace Implants | Biodenta<br>Dental Implant<br>System - Multi-<br>Use Abutment | TDS Abutment for<br>Friadent Xive |
| 510(k)<br>Number | New device | K110778 | K100993 | K142118 | K142115 | K123491 | K103339 |
| Intended<br>use | The Biodenta<br>Customized<br>Abutment is<br>intended for use<br>with dental<br>implants as a<br>support for single<br>or multiple tooth<br>prostheses in the<br>maxilla or mandible<br>of a partially or fully<br>edentulous patient. | The Biodenta Customized<br>Abutment is intended for<br>use with dental implants as<br>a support for single or multi<br>ple tooth prostheses in the<br>maxilla or mandible of a par<br>tially or fully edentulous pat<br>ient. The Biodenta Custom<br>ized Abutment is compatible<br>with the following implant<br>systems: - Internal hex sys<br>tems with flat-to-flat dimen<br>sions of 1.78mm or greater:<br>Firadent: FRIALIT Implant,<br>XiVA Implant: 3i: Certain Inte<br>rnal Connect Type; Astra:<br>Osseospeed Implant, Osseo<br>speed TX Implant; BioHoriz<br>ons: Internal Implant Syst<br>em, Tapered Internal Impl<br>ant System, Single-Stage<br>Implant System; Life core:<br>Lifecore RENOVA Internal<br>Hex Implant System; Zimm<br>er: Tapered Screw-Vent Imp<br>lant System, Screw-Vent Im<br>plant System, AdVent Impla<br>nt System. - Nobel Biocare<br>Replace: NobelReplace Stra<br>ight, NobelReplace Tapered<br>Replace Select Tapered, Re<br>place Select Straight; for the<br>NP,RP,WPand6.0 implants.<br>- External hex systems with<br>flat-to-flat dimensions of 2.4<br>mm or greater:Nobel Bioca | The InclusiveO<br>Titanium Abutments<br>for Astra<br>OsseoSpeed’M<br>Implants are<br>premanufactured<br>prosthetic<br>components directly<br>connected to<br>endlosseous dental<br>implants and are<br>intended<br>for use as an aid in<br>prosthetic<br>rehabilitation. They<br>are compatible with<br>the Astra Tech<br>OsseoSpeed" 3.0,<br>3.5, 4.0, 4.5, 5.0<br>implants. | Inclusive® Titanium<br>Abutments are<br>premanufactured<br>prosthetic<br>components directly<br>connected to<br>endosseous dental<br>implants and are<br>intended for use as<br>an aid in prosthetic<br>rehabilitation.<br>Inclusive® Titanium<br>Abutments are<br>compatible with:<br>- Straumann: Bone<br>Level NC and RC<br>implant sizes<br>- Nobel Biocare:<br>Branemark RP size<br>implant<br>- Nobel Biocare<br>NobelActive NP and<br>RP internal<br>connection implants | Inclusive® Titanium<br>Abutments are<br>premanufactured<br>prosthetic components<br>directly connected to<br>endosseous dental<br>implants and are<br>intended for use as an<br>aid in prosthetic<br>rehabilitation.<br>Inclusive® Titanium<br>Abutments are<br>compatible with:<br>- Biomet: 3i Certain<br>internal hex implants in<br>3.4, 4.1, 5.0, 6.0 mm<br>sizes<br>- Nobel Biocare:<br>NobelReplace straight<br>and tapered internal<br>connection<br>implants in NP, RP, WP,<br>6.0 mm sizes<br>- Zimmer: Screw-Vent<br>and Tapered Screw-<br>Vent internal hex<br>implants in 3.5,<br>4.5, 5.7 mm sizes | Biodenta<br>Dental Implant<br>System Multi<br>Use Abutments<br>are intended for<br>terminal or<br>intermediate<br>abutment<br>support for<br>fixed or<br>removable<br>crown,<br>bridgework and<br>to retain<br>overdentures. | TDS Abutment for Friadent<br>Xive is intended for use<br>with dental implants as a<br>support for single or<br>multiple tooth prostheses<br>in the maxilla or mandible<br>of a partially or fully<br>edentulous patient.<br>This device is compatible<br>with the following implant<br>systems which have an<br>internal hex with flat-to-flat<br>dimensions of 1.78mm or<br>greater: Firadent: FRIALIT<br>Implant, XiVA Implant; 3i:<br>Internal Connect Type;<br>Astra: Osseoispeed<br>Implant, Osseospeed TX<br>Implant; BioHorizons:<br>Internal Implant System,<br>Tapered Internal Implant<br>System, Single-Stage<br>Implant System, Laser-<br>lok®&3.0 implant system;<br>Lifecore: , Lifecore:<br>RENOVATM Internal Hex<br>Implant System; Zimmer:<br>Tapered Screw-Vent<br>Implant System, Screw-<br>Vent Implant System,<br>AdVent Implant System;<br>Osstem: GS System;<br>Nobel Biocare: Active<br>Implant. |
| | Subject Device | Primary Predicate | Reference Predicates | | | | |
| Company | Biodenta Swiss AG | Biodenta Swiss AG | Prismark<br>Dentalcraft, Inc | Prismark<br>Dentalcraft, Inc | Biodenta Swiss<br>AG | Pou Yu Biotechnology<br>Co., Ltd. | |
| Device<br>Name | Biodenta<br>Customised<br>Abutment -<br>Titanium | Biodenta Customized<br>Abutment | Inclusive Titanium<br>Abutments For Astra<br>Tech Osseospeed<br>Implants | Inclusive® Titanium<br>Abutments, compatible<br>with: Straumann Bone<br>Level, Nobel Biocare<br>Branemark , and Nobel<br>Biocare NobelActive<br>Implants | Biodenta<br>Dental Implant<br>System - Multi-<br>Use Abutment | TDS Abutment for<br>Friadent Xive | |
| 510(k)<br>Number | New device | K110778 | K100993 | K142118 | K123491 | K103339 | |
| Compatibl<br>e implant<br>types | - Dentsply, AstraTech<br>OsseoSpeed<br>- Biomet 3i, Certain<br>Internal<br>- Zimmer, Screw Vent<br>- Nobel Biocare,<br>NobelReplace<br>- Nobel Biocare,<br>NobelActive<br>- Straumann, Bone<br>Level<br>- BioHorizons, Internal<br>- Osstem, GS/TS<br>- Biodenta, Bone Level<br>and Tapered | - Dentsply, AstraTech<br>OsseoSpeed<br>- Biomet 3i, Certain<br>Internal<br>- Zimmer, Screw Vent<br>- Nobel Biocare,<br>NobelReplace<br>- Nobel Biocare,<br>NobelActive<br>- Straumann, Bone<br>Level | - Dentsply, AstraTech<br>OsseoSpeed | - Nobel Biocare,<br>NobelActive<br>- Straumann, Bone<br>Level<br><br>- Biomet: 3i Certain<br>internal<br>- Zimmer: Screw-Vent<br>- Nobel Biocare:<br>NobelReplace | - Biodenta, Bone<br>Level and Tapered | - Dentsply, AstraTech<br>OsseoSpeed<br>- Biomet 3i, Certain<br>Internal<br>- Zimmer, Screw Vent<br>- Nobel Biocare,<br>NobelReplace<br>- Nobel Biocare,<br>NobelActive<br>- Straumann, Bone<br>Level<br>- BioHorizons, Internal<br>- Osstem, GS/TS | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Biodenta. The logo consists of the word "biodenta" in gray, with a registered trademark symbol to the right of the word. Above the word is a curved design in gray and green.
#### Custom Design
| maximal<br>Abutment<br>Angulation | 30° | 30° | 30° | 30° | 30° | 30° | 30° |
|-----------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|-------------------|------------------------------------------------------------|
| Abutment<br>Diameter | 4.5 - 15 mm | 5 - 10 mm | max. 9.4 mm | max. 9.4 mm | max. 9.4 mm | 4.5 - 5.0 mm | 5 - 10 mm |
| Abutment<br>Height | 5.0 - 12.3 mm | 3.0 - 7.5 mm | max 12.45 mm | max 12.45 mm | max 12.45 mm | 2.0 - 5.5 mm | 3.0 - 7.5 mm |
| Abutment<br>fixation | Abutment<br>Screw | Abutment<br>Screw | Abutment<br>Screw | Abutment<br>Screw | Abutment<br>Screw | Abutment<br>Screw | Abutment<br>Screw |
| CAD/CAM<br>Processing | Milled in Biodenta<br>milling center<br>under QSR control | Milled in Biodenta<br>milling center under<br>QSR control | Milled in manufact.<br>milling center under<br>QSR control | Milled in manufact.<br>milling center under<br>QSR control | Milled in manufact.<br>milling center under<br>QSR control | NA | Milled in manufact.<br>milling center under<br>QSR control |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for "biodenta". The logo features the word "biodenta" in gray, with a stylized graphic above it. The graphic consists of two curved shapes, one in light gray and the other in light green, positioned to suggest movement or growth.
| | Subject Device | Primary Predicate | Reference Predicates | | | | | |
|---------------------|--------------------------------------------------|---------------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|-----------------------------------|--|
| Company | Biodenta Swiss AG | Biodenta Swiss AG | Prismark<br>Dentalcraft, Inc | Prismark<br>Dentalcraft, Inc | Prismark<br>Dentalcraft, Inc | Biodenta Swiss<br>AG | Pou Yu Biotechnology<br>Co., Ltd. | |
| Device<br>Name | Biodenta<br>Customised<br>Abutment -<br>Titanium | Biodenta Customized<br>Abutment | Inclusive Titanium<br>Abutments For Astra<br>Tech Osseospeed<br>Implants | Inclusive® Titanium<br>Abutments compatible<br>with: Straumann Bone<br>Level, Nobel Biocare<br>Branemark , and Nobel<br>Biocare NobelActive<br>Implants | Inclusive® Titanium<br>Abutments, compatible<br>with Zimmer Screw-<br>Vent, Biomet 3i Certain,<br>and Nobel Biocare<br>NobelReplace Implants | Biodenta<br>Dental Implant<br>System - Multi-<br>Use Abutment | TDS Abutment for<br>Friadent Xive | |
| 510(k)<br>Number | New device | K110778 | K100993 | K142118 | K142115 | K123491 | K103339 | |
| Material | | | | | | | | |
| Abutment +<br>Screw | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | |
| Sterile / Reuse | | | | | | | | |
| Sterile | Delivered Non<br>Sterile | Delivered Non<br>Sterile | Delivered Non<br>Sterile | Delivered Non<br>Sterile | Delivered Non<br>Sterile | Delivered Non<br>Sterile | Delivered Non<br>Sterile | |
| Reusable | no | no | no | no | no | no | no | |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Biodenta. The logo consists of the word "biodenta" in gray, with a registered trademark symbol to the right of the word. Above the word is a curved shape that is gray on top and light green on the bottom.
(Blank Page)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.