Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System
K143528 · Zimmer Dental, Inc. · NHA · May 1, 2015 · Dental
Device Facts
Record ID
K143528
Device Name
Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System
Applicant
Zimmer Dental, Inc.
Product Code
NHA · Dental
Decision Date
May 1, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Engaging, is intended for use in the creation of a customized abutment for a cement-retained crown or bridge, or a customized screw-retained restoration. The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Non-Engaging, is intended for use in the creation of customized multiple-unit restorations on implants (e.g., bars and bridges), when anti-rotation of the abutment is not necessary.
Device Story
Device consists of gold alloy abutments (engaging or non-engaging) designed for 3.1mmD dental implants; used by dental laboratories to create customized frameworks for cement-retained or screw-retained dental restorations. Engaging version features internal hex conical connection for single abutments or multiple-unit frameworks; non-engaging version features passive cylindrical interface for multiple-unit restorations. Abutments are secured to implants via separate titanium alloy screws. Device is sold non-sterile; requires end-user sterilization. Healthcare providers (dentists/prosthodontists) use the final cast restoration to support prosthetic teeth, restoring function and aesthetics for patients with missing teeth.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including static and fatigue testing per ISO 14801:2007, and MRI compatibility testing per ASTM F 2182.
Indicated for patients requiring customized abutments for cement-retained crowns/bridges, screw-retained restorations, or multiple-unit restorations (bars/bridges) on 3.1mmD dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2015
Zimmer Dental Inc. Ms. Christina Boydston Regulatory Affairs Manager 1900 Aston Ave. Carlsbad, California 92008
Re: K143528
Trade/Device Name: Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: April 2, 2015 Received: April 3, 2015
Dear Ms. Boydston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the Zimmer Dental logo on the left and the words "Indications for Use" on the right. The Zimmer Dental logo is a blue Z inside of a circle, with the words "zimmer dental" below it. The words "Indications for Use" are in black and are in a larger font than the words "zimmer dental."
510(k) Number (if known): K143528
# Device Name: Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System
Indications For Use:
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Engaging, is intended for use in the creation of a customized abutment for a cement-retained crown or bridge, or a customized screw-retained restoration.
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Non-Engaging, is intended for use in the creation of customized multiple-unit restorations on implants (e.g., bars and bridges), when anti-rotation of the abutment is not necessary.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Zimmer Dental 1900 Aston Avenue arlsbad, CA 92008 60.929.4300 (ph) 60.431.7811 (fax)
# Traditional 510(k) PRE-MARKET NOTIFICATION 510(k)
# 510(k) SUMMARY (21CFR807.92(a))
# K143528
### 1. Submitter's Information:
| Name: | Zimmer Dental Inc. | | |
|----------------|--------------------|--|--|
| Address: | 1900 Aston Ave. | | |
| | Carlsbad, CA 92008 | | |
| Phone: | 760-929-4300 | | |
| Contact: | Christina Boydston | | |
| Date Prepared: | April 29, 2015 | | |
- 2. Device Name: Trade Name:
Regulation Number: Classification Code: Classification: Device Classification Name: Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System 872.3630 NHA II Abutment, Implant, Dental, Endosseous
### 3. Predicate Device(s):
| Predicate Device No. 1 | |
|-----------------------------|---------------------------------------|
| Trade Name: | "Cast-To" Gold Abutments, Engaging |
| 510(k) Number: | K011028 |
| Regulation Number: | 872.3630 |
| Classification Code: | NHA |
| Device Classification Name: | Abutment, Implant, Dental, Endosseous |
Predicate Device No. 2 Trade Name: "Cast-To" Gold Abutments, Non-Engaging 510(k) Number: K011028 872.3630 Regulation Number: Classification Code: NHA Device Classification Name: Abutment, Implant, Dental, Endosseous
{4}------------------------------------------------
### 4. Device Description:
The purpose of an abutment with Cast-to capabilities is to satisfy customer needs of attaching cast framework prostheses to an abutment. The nonengaging Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System can be used as base to cast frameworks for multiple unit crown and bridge and bar over denture restorations by a dental laboratory using a precious metal alloy. The engaging Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System can be used as base to cast a single abutment for a cement-retained crown or bridge and frameworks for multiple unit crown and bridge and bar over denture restorations by a dental laboratory using a precious metal alloy. The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System is designed for use with Zimmer 3.1mmD Dental Implant System to support cast framework prostheses. The engaging abutment/implant interface is an internal conical connection with a hex. The non-engaging abutment/implant interface is a connection that sits on top of the implant but does also has cylindrical features that sit passively within the implant.
The abutment is composed of Gold Alloy 6019 and secured to the implant with a separate Titanium alloy screw for retention.
### న్. Indications for Use:
### Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Engaging
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Engaging, is intended for use in the creation of a customized abutment for a cement-retained crown or bridge, or a customized screw-retained restoration.
### Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Non-Engaging
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Non-Engaging, is intended for use in the creation of customized multiple-unit restorations on implants (e.g., bars and bridges), when anti-rotation of the abutment is not necessary.
{5}------------------------------------------------
### 6. Device Comparison:
Differences in technological characteristics between the subject and predicate device did not raise different questions of safety and effectiveness, and that mechanical testing was used to support a conclusion of substantial equivalence (please see non-clinical testing section for details). The subject abutment is substantially equivalent to the predicate relative to material, general design features as well as design and manufacturing process. It is fabricated from Gold alloy. The differences are the implant/abutment interface and platform sizes. The subject abutment and base device will be affixed to the implant by a retaining screw, the same manner as the predicate.
| Feature | Subject Device 1<br>Zimmer Cast-to Gold<br>Abutment for the<br>3.1mmD Dental Implant<br>System, Engaging | Predicate 1<br>"Cast-To" Gold<br>Abutments, Engaging<br>K011028 | Subject Device 2<br>Zimmer Cast-to Gold<br>Abutment for the<br>3.1mmD Dental Implant<br>System, Non-Engaging | Predicate 2<br>"Cast-To" Gold Abutments,<br>Non-Engaging<br>K011028 |
|-----------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Material | Gold Alloy 6019 | Gold Alloy 6019 | Gold Alloy 6019 | Gold Alloy 6019 |
| Implant<br>Interface | Internal Hex<br>Conical Connection | Internal Hex | Non-engaging conical<br>connection | Non-hex connection that sits<br>on the top of the implant |
| Platforms | 2.9mm | 3.5mm, 4.5mm,<br>5.7mm | 2.9mm | 3.5mm, 4.5mm |
| Maximum<br>Abutment<br>Angulation | 30° | 30° | 30° | 30° |
| Method of<br>Attachment | Retaining screw | Retaining Screw | Retaining screw | Retaining Screw |
| Retaining<br>Screw | Cat. No. CUAS<br>Cat. No. CASLC | Cat. No. MHLAS<br>Cat. No. MTWSD | Cat. No. CUAS | Cat. No. MHLAS |
### 7. Technological Characteristics
### 8. Non-Clinical Testing:
Non-clinical test data was used to support the decision of substantial equivalence. Non-clinical testing consisted of performance of fatigue and compression testing in accordance with the FDA guidance Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments and ISO 14801:2007 Dentistry - Implants -Dynamic fatigue test for endosseous dental implants. The testing indicates that the subject device is substantially equivalent to the predicate device in terms of static and fatigue testing.
{6}------------------------------------------------
In addition, the Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System will be sold non-sterile and will be sterilized by the end user. The sterilization procedures listed in the Instruction for Use were validated to provide a minimum sterility assurance level of 10°.
The subject abutment will be manufactured using biocompatible material. The subject abutment will be manufactured using the same material as the predicate which were cleared in K011028; therefore, no additional biocompatibility testing was performed on the subject device. The Biocompatibility of the raw material and manufacturing process testing of Gold Alloy 6019 was previously conducted, documented and cleared in the predicate submission (K011028). Biocompatibility testing was completed in compliance to ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
Additionally, Zimmer Dental implant systems were evaluated for interactions with magnetic fields during Magnetic Resonance Imaging (MRI) in accordance with the FDA Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment and ASTM F 2182: Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants during Magnetic Resonance Imaging. The results of these tests support the MR conditional labeling of this device.
#### 9. Clinical Testing
No clinical testing was performed. Non-clinical testing was used to support the decision of substantial equivalence.
#### 10. Conclusion
The subject device, Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System are gold alloy, abutments with an engaging or non-engaging interface and are designed to provide support for a cast framework.
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System is equivalent to predicate devices in general abutment design, manufacturing process, material, indications, mechanical performances, and risk. The subject devices are Magnetic Resonance Imaging (MRI) compatible conditionally with respect to patient safety.
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System will be manufactured from the same biocompatible material (Gold Alloy 6019) as predicate devices, will be manufactured and inspected according to approved engineering drawings, and Zimmer Dental procedures, and will be packaged in a similar manner as the predicate devices. The subject devices will be sold non
{7}------------------------------------------------
sterile and will be labeled as "non-sterile", and is intended to be sterilized by the end user. The sterilization procedures were validated to provide a minimum sterility assurance level of 10-6.
Based on the above analysis and the 510(k) submission for Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, the subject device is substantially equivalent to the predicate
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.