INCLUSIVE TITANIUM ABUTMENTS COMPATIBLE WITH: ZIMMER, BIOMET 3I, AND NOBEL BIOCARE IMPLANT
K142115 · Prismatik Dentalcraft, Inc. · NHA · Nov 13, 2014 · Dental
Device Facts
Record ID
K142115
Device Name
INCLUSIVE TITANIUM ABUTMENTS COMPATIBLE WITH: ZIMMER, BIOMET 3I, AND NOBEL BIOCARE IMPLANT
Applicant
Prismatik Dentalcraft, Inc.
Product Code
NHA · Dental
Decision Date
Nov 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Inclusive® Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
Device Story
Inclusive Titanium Abutments are premanufactured prosthetic components used in dental clinics to support prosthetic restorations (crowns, bridges, overdentures). The device is placed over the shoulder of an endosseous dental implant and secured using an abutment screw. Operated by dental professionals, the abutment serves as the interface between the implant and the final prosthesis. By providing a stable connection to the implant, the device facilitates the restoration of dental function and aesthetics for patients. The abutments are machined from titanium alloy and are compatible with specific internal hex or trilobe implant platforms.
Clinical Evidence
Bench testing only. Evidence includes reliability calculations, fatigue testing, and static load failure testing of finished assembled implant/abutment systems in accordance with FDA guidance for root-form endosseous dental implants and abutments. Sterilization validation provided per ANSI/AAMI ST79 and ISO 17665-1.
Technological Characteristics
Material: Titanium grade 23 (Ti-6AL-4V ELI) per ASTM F-136. Design: Machined, premanufactured prosthetic components. Connection: Internal hex or internal trilobe. Dimensions: Compatible with specific Zimmer, Biomet 3i, and Nobel Biocare platforms. Sterilization: Validated method per ANSI/AAMI ST79 and ISO 17665-1.
Indications for Use
Indicated for use as an aid in prosthetic rehabilitation for patients requiring support for crowns, bridges, or overdenture prostheses in the maxillary or mandibular arch, connected to compatible Zimmer Screw-Vent, Biomet 3i Certain, or Nobel Biocare NobelReplace endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Inclusive Abutment for Zimmer, 3i and Nobel Biocare Implants (K073217)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13,2014
Prismatik Dentalcraft, INC Brandon Shepard Regulatory Affairs & Ouality Assurance Specialist 2212 Dupont Dr. Suite P Irvine, California 92612
Re: K142115
Trade/Device Name: Inclusive® Titanium Abutments, compatible with Zimmer Screw-Vent, Biomet 3i Certain, and Nobel Biocare NobelReplace Implants Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: II Product Code: NHA Dated: October 10, 2014 Received: October 14, 2014
Dear Mr. Shepard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, mA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
### 510(k) Number: K142115
Device Name: Inclusive® Titanium Abutments, compatible with Zimmer Screw-Vent, Biomet 3i Certain, and Nobel Biocare NobelReplace Implants
### Indications for Use:
Inclusive® Titanium Abutments are premanufactured prosthetic components directly
connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
Inclusive® Titanium Abutments are compatible with:
- Biomet: 3i Certain internal hex implants in 3.4, 4.1, 5.0, 6.0 mm sizes -Nobel Biocare: NobelReplace straight and tapered internal connection implants in NP, RP, WP, 6.0 mm sizes
- Zimmer: Screw-Vent and Tapered Screw-Vent internal hex implants in 3.5. 4.5. 5.7 mm sizes
Prescription Use 1 (Part 21 CFR 801 Subpart D)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a gradient of colors, including pink, blue, and yellow. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." in a sans-serif font.
| 005 | 51 |
|-----|----|
|-----|----|
# 10(k) Summary
[As Required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements 21 CFR 807.92.
#### SUBMITTER INFORMATION A.
B.
| Company Name: | PRISMATIK DENTALCRAFT, INC. |
|------------------------|-------------------------------------------------------------------------------------------------------------------|
| Company Address: | 2212 Dupont Dr., Suite P<br>Irvine, CA 92612 |
| Company Phone / Fax: | (949) 225-1269 / (978) 313-0850 |
| Contact Person: | Primary Contact:<br>• Brandon Shepard, (949) 225-1243<br>Secondary Contact:<br>• Marilyn Pourazar, (949) 225-1269 |
| Date Summary Prepared: | August 1, 2014 |
DEVICE IDENTIFICATION
| Trade/Proprietary Name: | Inclusive <sup>®</sup> Titanium Abutments<br>compatible with: Zimmer Screw-Vent, Biomet 3i<br>Certain, and Nobel Biocare NobelReplace Implants |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Endosseous Dental Implant Abutment |
| Regulation Number: | 872.3630 |
| Product Code: | NHA |
| Device Class: | 2 |
| Review Panel: | Dental |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a rainbow gradient. To the right of the triangle is the company name, "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line. The text is in a blue sans-serif font.
#### C. IDENTIFICATION OF PREDICATE DEVICE
Trade/Proprietary Name:
- · Inclusive 110 Abutment for Zimmer, 3i and Nobel Biocare Implants (K073217)
- · NobelProcera Ti Abutment (K091756)
- · BIOMET 3i Dental Abutments and Restorative Components (K072642)
- · Zimmer Ti Prepable Abutment (K092403).
#### D. DEVICE DESCRIPTION
Inclusive Titanium Abutments are endosseous implant abutments which are placed into the dental implant to provide support for a prosthetic restoration. The abutment is placed over the implant shoulder and is fastened into the implant with a screw. Abutments are made of titanium grade 23 (Ti-6AL-4V ELI) and meet ASTM F-136 Standard.
#### E. INDICATIONS FOR USE
Inclusive® Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
Inclusive® Titanium Abutments are compatible with:
- Biomet: 3i Certain internal hex implants in 3.4, 4.1, 5.0, 6.0 mm sizes -
- -Nobel Biocare: NobelReplace straight and tapered internal connection implants in NP, RP, WP, 6.0 mm sizes
- -Zimmer: Screw-Vent and Tapered Screw-Vent internal hex implants in 3.5, 4.5, 5.7 mm sizes
#### NON-CLINICAL TESTING F.
Non-clinical test data was used to evaluate the device's safety and effectiveness, and determine substantial equivalence with predicate devices.
Non-clinical testing was performed in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" and it consisted of Reliability Calculation, Fatigue Testing and Static Load Failure Testing of finished assembled implant/abutment systems.
In addition, sterilization validation information and a recommended sterilization method based on ANSI/AAMI ST79 and ISO 17665-1 is provided in the Information for Use.
The testing performed demonstrated implant to abutment compatibility and established that the predicate device is as safe, as effective, and performs as well as the predicate device(s).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections of different colors, including pink, purple, blue, green, and yellow. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
#### SUBSTANTIAL EQUIVALENCE G.
Inclusive® Titanium Abutments, compatible with Zimmer Screw-Vent, Biomet 3i Certain and Nobel Biocare NobelReplace Implants are substantially equivalent to the Inclusive™ Abutment for Zimmer, 3i and Nobel Biocare Implants (K073217) and the OEM predicates, (K091756, K072642, and K092403), identified in Section C above. They are substantially equivalent in intended use, materials, design and performance.
(See Comparison Tables below).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections with different colors, including pink, purple, blue, and yellow. The company name is written in a sans-serif font, with "PRISMATIK" on the first line and "DENTALCRAFT, INC." on the second line.
| | Predicate Device (1) | Predicate Device (2) | Proposed Device | |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| | Ti Prepable Abutment<br>(K092403) | Inclusive Abutment for<br>Zimmer, 3i, and Nobel Biocare<br>Implants (K073217) | Inclusive Titanium<br>Abutments compatible<br>with: NobelReplace<br>Implants | Similarities and<br>Differences |
| Manufacturer | Zimmer Dental | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | N/A |
| Dimensions of<br>Abutment | Hex Dimensions:<br>2.4mm Across Flats<br>3.0mm Across Flats | Cylindrical Diameter: 9.4mm<br>Hex Dimensions:<br>2.4mm Across Flats<br>3.0mm Across Flats | Cylindrical Diameter: 9.4mm<br>Hex Dimensions:<br>2.4mm Across Flats<br>3.0mm Across Flats | Same |
| Abutment Screw<br>Manufacturer | OEM (Zimmer) | OEM (Zimmer) | Inclusive (Prismatik) | Equivalent screw<br>from different<br>manufacturer |
| Dimensions of<br>Abutment Screw | Length 8.2mm; 1-72 UNF-2A<br>Thread | Length 8.2mm; 1-72 UNF-2A<br>Thread | Length 8.2mm; 1-72 UNF-2A<br>Thread | Same |
| Indications for Use | The Ti Prepable Abutments are<br>designed for use as a terminal or<br>intermediate abutment for cement<br>retained prosthesis. It allows for<br>preparation for the crown to be<br>attached. It can be used for a<br>single or multiple-unit restoration.<br>The abutment is intended to be<br>prepared and placed with patient<br>specific margins. | The Inclusive Abutment is<br>intended to be used in<br>conjunction with endosseous<br>implants in the maxillary and/or<br>mandibular arch to provide<br>support for crowns, bridges or<br>overdenture prostheses. The<br>prosthesis can be cement<br>retained to the abutment. The<br>abutment screw is intended to<br>secure the abutment to the<br>endosseous implant. | Inclusive Titanium<br>Abutments are<br>premanufactured prosthetic<br>components directly<br>connected to endosseous<br>dental implants and are<br>intended for use as an aid in<br>prosthetic rehabilitation. | Same Intended<br>Use. Different<br>wording in the<br>Indications for<br>Use |
| Platform<br>Compatibility | Zimmer Dental Screw-Vent<br>3.5, 4.5, 5.7mm | Zimmer Dental Screw-Vent<br>3.5, 4.5, 5.7mm | Zimmer Dental Screw-Vent<br>3.5, 4.5, 5.7mm | Same |
| Connection | Internal Hex | Internal Hex | Internal Hex | Same |
| Design/Construction | Machined | Machined | Machined | Same |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity | Same |
| Abutment Angle | 0° | 0°-20° | 0°-30° | Similar; increased<br>angulation range |
| Implant Seat | Flat | Flat | Flat | Same |
| Screw Seat | Flat | Flat | Flat | Same |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | Same |
| | Predicate Device (1) | Predicate Device (2) | Proposed Device | |
| | Biomet 3i Dental Abutments &<br>Restorative Components (K072642) | Inclusive Abutment for<br>Zimmer, 3i, and Nobel Biocare<br>Implants (K073217) | Inclusive Titanium<br>Abutments compatible with:<br>Biomet 3i Certain Implants | Similarities and<br>Differences |
| Manufacturer | Biomet | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | N/A |
| Dimensions of<br>Abutment | Hex Dimensions:<br>2.2mm Across Flats<br>2.7mm Across Flats | Cylindrical Diameter: 9.4mm<br>Hex Dimensions:<br>2.2mm Across Flats<br>2.7mm Across Flats | Cylindrical Diameter: 9.4mm<br>Hex Dimensions:<br>2.2mm Across Flats<br>2.7mm Across Flats | Same |
| Abutment Screw<br>Manufacturer | OEM (Biomet 3i) | OEM (Biomet 3i) | Inclusive (Prismatik) | Equivalent screw<br>from different<br>manufacturer |
| Dimensions of<br>Abutment Screw | Length: 8.4mm; M1.6 Thread | Length: 8.4mm; M1.6 Thread | Length: 8.4mm; M1.6 Thread | Same |
| Indications for Use | Biomet 3i Dental Abutments are<br>intended for use as accessories to<br>endosseous dental implant to support<br>a prosthetic device in a partially or<br>completely edentulous patient. A<br>dental abutment is intended for use to<br>support single and multiple tooth<br>prosthesis, in the mandible or maxilla.<br>The prosthesis can be screw retained<br>or cement retained. | The Inclusive Abutment is<br>intended to be used in<br>conjunction with endosseous<br>implants in the maxillary and/or<br>mandibular arch to provide<br>support for crowns, bridges or<br>overdenture prostheses. The<br>prosthesis can be cement<br>retained to the abutment. The<br>abutment screw is intended to<br>secure the abutment to the<br>endosseous implant. | Inclusive Titanium Abutments<br>are premanufactured prosthetic<br>components directly connected<br>to endosseous dental implants<br>and are intended for use as an<br>aid in prosthetic rehabilitation. | Same Intended<br>Use. Different<br>Indications for<br>Use statement. |
| Platform<br>Compatibility | Biomet 3i Certain<br>3.4, 4.1, 5.0, 6.0mm | Biomet 3i Certain<br>3.4, 4.1, 5.0, 6.0mm | Biomet 3i Certain<br>3.4, 4.1, 5.0, 6.0mm | Same |
| Connection | Internal Hex | Internal Hex | Internal Hex | Same |
| Design/Construction | Machined | Machined | Machined | Same |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity | Same |
| Abutment Angle | 0°-15° | 0°-20° | 0°-30° | Similar; increased<br>angulation range |
| Implant Seat | Flat | Flat | Flat | Same |
| Screw Seat | Flat | Flat | Flat | Same |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | Same |
| | Predicate Device (1) | Predicate Device (2) | Proposed Device | |
| | NobelProcera Ti Abutment<br>(K091756) | Inclusive Abutment for<br>Zimmer, 3i, and Nobel<br>Biocare Implants<br>(K073217) | Inclusive Titanium<br>Abutments compatible<br>with: Nobel Biocare<br>NobelReplace Implants | Similarities and<br>Differences |
| Manufacturer | Nobel Biocare | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | N/A |
| Dimensions of<br>Abutment | Connection Length: 2.9mm | Cylindrical Diameter: 9.4mm<br>Connection Length: 2.9mm | Cylindrical Diameter: 9.4mm<br>Connection Length: 2.9mm | Same |
| Abutment Screw<br>Manufacturer | OEM (Nobel Biocare) | OEM (Nobel Biocare) | Inclusive (Prismatik) | Equivalent screw from<br>different manufacturer |
| Dimensions of<br>Abutment Screw | Length 8.3mm; M1.8 Thread | Length 8.3mm; M1.8 Thread | Length 8.3mm; M1.8 Thread | Same |
| Indications for Use | The NobelProcera Ti<br>Abutments are<br>premanufactured prosthetic<br>components directly connected<br>to endosseous dental implants<br>and are intended for use as an<br>aid in prosthetic rehabilitation. | The Inclusive Abutment is<br>intended to be used in<br>conjunction with endosseous<br>implants in the maxillary<br>and/or mandibular arch to<br>provide support for crowns,<br>bridges or overdenture<br>prostheses. The prosthesis can<br>be cement retained to the<br>abutment. The abutment<br>screw is intended to secure the<br>abutment to the endosseous<br>implant. | Inclusive Titanium<br>Abutments are<br>premanufactured prosthetic<br>components directly<br>connected to endosseous<br>dental implants and are<br>intended for use as an aid in<br>prosthetic rehabilitation. | Same Intended Use.<br>Different Indications<br>for Use statement. |
| Platform<br>Compatibility | Nobel Biocare NobelReplace<br>NP, RP, WP, 6.0 | Nobel Biocare NobelReplace<br>NP, RP, WP, 6.0 | Nobel Biocare NobelReplace<br>NP, RP, WP, 6.0 | Same |
| Connection | Internal Trilobe | Internal Trilobe | Internal Trilobe | Same |
| Design/Construction | Machined | Machined | Machined | Same |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity | Same |
| Abutment Angle | 0°-30° | 0°-20° | 0°-30° | Similar; increased<br>angulation range |
| Implant Seat | Flat | Flat | Flat | Same |
| Screw Seat | Flat | Flat | Flat | Same |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy | Same |
## Comparison of Predicate and Proposed Devices: Zimmer Screw-Vent Platform
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with different colors. The text "PRISMATIK DENTALCRAFT, INC." is on the right in a blue sans-serif font.
## Comparison of Predicate and Proposed Devices: Biomet 3i Certain Platform
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangular prism on the left and the company name on the right. The prism is made up of various colors, including pink, purple, blue, and green. The company name is written in a sans-serif font.
## Comparison of Predicate and Proposed Devices: Nobel Biocare NobelReplace Platform
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.