INCLUSIVE TITANIUM ABUTMENTS, COMPATIBLE WITH DENTSPLY IMPLANTS ANKYLOS C/X
K141923 · Prismatik Dentalcraft, Inc. · NHA · Jan 6, 2015 · Dental
Device Facts
Record ID
K141923
Device Name
INCLUSIVE TITANIUM ABUTMENTS, COMPATIBLE WITH DENTSPLY IMPLANTS ANKYLOS C/X
Applicant
Prismatik Dentalcraft, Inc.
Product Code
NHA · Dental
Decision Date
Jan 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Inclusive Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. Inclusive Titanium Abutments are compatible with Dentsply Implants Ankylos® C/X in 3.5, 4.5, 5.5, 7.0 mm sizes.
Device Story
Inclusive Titanium Abutments are premanufactured prosthetic components designed to connect to endosseous dental implants; they provide support for prosthetic restorations. The device is placed over the implant shoulder and secured with a screw. Used in dental clinics by dental professionals during prosthetic rehabilitation. The abutment acts as an interface between the implant and the final prosthesis (crowns, bridges, or overdentures). By providing a stable connection, the device enables the restoration of dental function and aesthetics for the patient.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including reliability calculation, fatigue strength testing, and static load failure testing of finished assembled implant/abutment systems.
Technological Characteristics
Material: Titanium alloy (Ti-6AL-4V ELI) per ASTM F-136. Design: Machined endosseous dental implant abutment. Connection: Unigrip. Implant seat: Taper. Screw seat: Taper. Sizes: 3.5, 4.5, 5.5, 7.0 mm. Sterilization: Validated per ANSI/AAMI ST79 and ISO 17665-1.
Indications for Use
Indicated for use as an aid in prosthetic rehabilitation for patients requiring support for prosthetic restorations connected to endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Atlantis Abutment for Dentsply Ankylos Implant (K101004)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure in profile, with three overlapping heads suggesting a sense of community or interconnectedness. The figure is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6,2015
Prismatik Dentalcraft, Incorporated Mr. Brandon Shepard Regulatory Affairs & Quality Assurance Specialist 2212 Dupont Drive, Suite P Irvine, CA 92612
Re: K141923 Trade/Device Name: Inclusive® Titanium Abutments, compatible with Dentsply Implants Ankylos® C/X Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 5, 2014 Received: December 8, 2014
Dear Mr. Shepard,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
Page 2 - Mr. Shepard
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin Keith Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo features a colorful triangle on the left side, with the word "PRISMATIK" in bold, black letters to the right of the triangle. Below "PRISMATIK" is the text "DENTALCRAFT, INC." in a smaller, black font.
## 004 Indications for Use Statement
510(k) Number (if known): K141923
Device Name: Inclusive® Titanium Abutments, compatible with Dentsply Implants Ankylos® C/X
Indications for Use: Inclusive Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
> Inclusive Titanium Abutments are compatible with Dentsply Implants Ankylos® C/X in 3.5, 4.5, 5.5, 7.0 mm sizes.
> > Type of Use
Prescription Use: Yes > No | (Part 21 CFR 801 Subpart D)
Over-the-Counter Use: Yes | | No (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo features a triangular prism on the left side, with each face of the prism displaying a different color. To the right of the prism is the company name, "PRISMATIK" in bold, followed by "DENTALCRAFT, INC." in a smaller font.
## 005 510(k) Summary
[As Required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements 21 CFR 807.92.
### SUBMITTER INFORMATION A.
B.
| Company Name: | PRISMATIK DENTALCRAFT, INC. |
|------------------------|-------------------------------------------------------------------------------------------|
| Company Address: | 2212 Dupont Dr., Suite P<br>Irvine, CA 92612 |
| Company Phone / Fax: | (949) 225-1269 / (978) 313-0850 |
| Contact Person: | Primary Contact:<br>● Brandon Shepard, <ins>RA/QA Specialist</ins><br>(949) 225-1243 |
| | Secondary Contact:<br>● Marilyn Pourazar, <ins>Sr. Director RA/QA</ins><br>(949) 225-1269 |
| Date Summary Prepared: | December 5, 2014 |
DEVICE IDENTIFICATION
| Trade/Proprietary Name: | Inclusive <sup>®</sup> Titanium Abutments,<br>compatible with Dentsply Implants Ankylos <sup>®</sup> C/X |
|-------------------------|----------------------------------------------------------------------------------------------------------|
| Common Name: | Endosseous Dental Implant Abutment |
| Regulation Number: | 872.3630 |
| Product Code: | NHA |
| Device Class: | 2 |
| Review Panel: | Dental |
{4}------------------------------------------------
#### C. IDENTIFICATION OF PREDICATE DEVICE
Trade/Proprietary Name: Atlantis Abutment for Dentsply Ankylos Implant (K101004) Inclusive Titanium Abutment Blanks (K083192)
#### D. DEVICE DESCRIPTION
Inclusive Titanium Abutments are endosseous implant abutments which are placed into the dental implant to provide support for a prosthetic restoration. The abutment is placed over the implant shoulder and is mounted into the implant with a screw. Abutments and screws are made of titanium alloy (Ti-6AL-4V ELI) and meet ASTM F-136 Standard. Inclusive Titanium Abutments are compatible with Dentsply Implants Ankylos C/X in 3.5 mm, 4.5 mm, 5.5 mm, and 7.0 mm Ø sizes. Physical and technological characteristics of the proposed device are summarized in the comparison table below.
#### E. INDICATIONS FOR USE
Inclusive Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
Inclusive Titanium Abutments are compatible with Dentsply Implants Ankylos C/X in 3.5, 4.5, 5.5, 7.0 mm sizes.
#### F. SUBSTANTIAL EQUIVALENCE
Inclusive Titanium Abutments, compatible with Dentsply Implants Ankylos C/X, are substantially equivalent to the Atlantis Abutment for Dentsply Ankylos Implant (K101004) and the Inclusive Titanium Abutment Blanks (K083192). They are substantially equivalent in intended use, materials, as well as design, technological characteristics and performance.
#### NON-CLINICAL TESTING G.
Non-clinical test data was used to evaluate the proposed device's safety and effectiveness, and determine substantial equivalence with predicate devices. Clinical testing was not necessary to establish substantial equivalency of the device.
Non-clinical testing was performed in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" and it consisted of testing finished assembled implant/abutment systems of the worst case scenario, through Reliability Calculation and Testing, as well as Fatigue Strength Testing and Static Load Failure Testing.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo features a colorful triangular shape on the left side, resembling a prism. To the right of the prism is the company name, "PRISMATIK" in bold, followed by "DENTALCRAFT, INC." in a smaller font size.
In addition, sterilization validation information and recommended sterilization method based on ANSI/AAMI ST79 and ISO 17665-1 is provided in the Information for Use.
The testing performed demonstrated implant to abutment compatibility and has established substantial equivalency of the proposed device with predicate devices.
#### CONCLUSION H.
The results of the nonclinical testing performed, demonstrate that Inclusive Titanium Abutments, compatible with Dentsply Implants Ankylos C/X are as safe, as effective, and perform as well as the predicate devices. Therefore the proposed device is substantially equivalent with the predicate devices cleared for the same intended use.
(See Comparison Table Below)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo features a colorful triangle on the left side, with the word "PRISMATIK" in bold, black letters to the right of the triangle. Below "PRISMATIK" is the text "DENTALCRAFT, INC." in a smaller, non-bold font.
| PREDICATES | | PROPOSED | | |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Atlantis Abutment for<br>Dentsply Ankylos<br>Implant | Inclusive Titanium<br>Abutment Blanks | Inclusive Titanium<br>Abutments,<br>compatible with<br>Dentsply Implants<br>Ankylos C/X | Similarities /<br>Differences<br>of Devices | |
| Manufacturer | Astra Tech | Prismatik | Prismatik | - |
| 510(k) No. | K101004 | K083192 | K141923 | - |
| Indications<br>for Use | The Atlantis Abutment<br>is intended for use with<br>an endosseous implant<br>to support a prosthetic<br>device in a partially or<br>completely edentulous<br>patient. It is intended for<br>use to support single and<br>multiple tooth<br>prosthesis, in the<br>mandible or maxilla.<br>The prosthesis can be<br>cement or screw retained<br>to the abutment. The<br>abutment screw is<br>intended to secure the<br>abutment to the<br>endosseous implant. | The Inclusive<br>Titanium Abutment<br>Blank is intended to be<br>used in conjunction<br>with endosseous<br>implants in the<br>maxillary and/or<br>mandibular arch to<br>provide support for<br>crowns, bridges or<br>overdenture<br>prostheses. The<br>prosthesis can be<br>cement retained to the<br>abutment. The<br>abutment screw is<br>intended to secure the<br>abutment to the<br>endosseous implant. | Inclusive Titanium<br>Abutments are<br>premanufactured<br>prosthetic<br>components directly<br>connected to<br>endosseous dental<br>implant and are<br>intended for use as<br>an aid in prosthetic<br>rehabilitation. | Same<br>Intended<br>Use |
| Dimensions of<br>Abutment | 11.5° Taper,<br>2.84 Gage Point | Cylinder Ø 9.43mm | Cylinder Ø 9.43mm<br>11.5° Taper,<br>2.84 Gage Point | Same |
| Abutment<br>Screw | Integrated fastening<br>screw (mobile, not<br>removable) | Detached and<br>removable | Detached and<br>removable | Same |
| Platform<br>Compatibility | Ankylos C/X<br>3.5mm, 4.5mm,<br>5.5mm, 7.0mm | Various (Straumann<br>Bone Level,<br>NobelActive,<br>Branemark) | Ankylos C/X<br>3.5mm, 4.5mm,<br>5.5mm, 7.0mm | Same |
| Connection | Unigrip | Hexagon | Unigrip | Same |
| Material | Ti-6AL-4V ELI<br>(ASTM F-136) | Ti-6AL-4V ELI<br>(ASTM F-136) | Ti-6AL-4V ELI<br>(ASTM F-136) | Same |
| Design /<br>Construction | Machined | Machined | Machined | Same |
| Abutment<br>Angle | 0°-30° | 0°-20° | 0°-30° | Same |
| Implant Seat | Taper | Taper | Taper | Same |
| Screw Seat | Taper | Flat | Taper | Same |
# Comparison of Devices
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.