Dental Implant Abutments are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses. Abutments can be used in single tooth replacements and multiple tooth restorations. The Dental Implant Abutments OKTAGON® are intended to be compatible to OKTAGON® implants (Dental Implant OKTAGON®) with diameters 3.3mm, 4.1mm and 4.8mm in the variation Regular Platform, Wide Platform and Tapered Design with the lengths 8mm, 10mm, 12mm and 14mm.
Device Story
OKTAGON® Dental Implant Abutment system; provides support for prosthetic reconstructions (crowns, bridges, prostheses) in partially or fully edentulous patients. System comprises abutments and corresponding screws; connects to OKTAGON® endosseous implants via internal octagon and metric thread. Used by dental clinicians in clinical settings. Abutments serve as interface between dental implant and prosthetic restoration; facilitates screw-retained or cemented restorations. Benefits include restoration of dental function and aesthetics for patients requiring tooth replacement.
Clinical Evidence
Bench testing only. Fatigue testing conducted in accordance with ISO 14801 and FDA guidance for root-form endosseous dental implants and abutments. No clinical data provided.
Technological Characteristics
Materials: Titanium Grade 4, Titanium Alloy, POM. Connection: Internal octagon, metric thread. Standards: ISO 14801 (fatigue), ISO 7405, ISO 10993-1 (biocompatibility), ISO 5832-2/3, ASTM F67, ASTM F136, ISO 14971, AAMI ST79, ISO 17665-1. Dimensions: Compatible with 3.3mm, 4.1mm, 4.8mm diameter implants; lengths 8mm-14mm.
Indications for Use
Indicated for partially or fully edentulous patients requiring prosthetic reconstructions, including individual crowns, bridges, partial or total prostheses, for single or multiple tooth restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Abutment, Dental, Endosseous implants (RN synOcta 1.5 mm Abutment) (K073628)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 07 2014
## 510(k) Traditional Dental Endosseous Implant Abutment OKTAGON®
## Section #5 510((k) Summary
### 1. Applicant's Name and Address
Hager & Meisinger GmbH Hansemannstraße 10 41468 Neuss, Germany Phone: (0049) 2131 2012-0 Fax: (0049) 2131 2012- 222 Contact Person: Wiebke Stolten
Management product approval and product validation (Regulatory Affairs)
#### 2. Date prepared
Date prepared:
06/11/2014
#### 3. Name of the device
Trade Name: Common Name: Classification Name: Product Code: Regulation No: Class: Panel:
Dental Implant Abutment OKTAGON® Endosseous dental implant abutments Endosseous dental implant abutment NHA 872.3630 . ll Dental
#### 4. Predicate Devices:
| 510(k No.) | Manufacturer | Trade Name |
|------------|--------------|----------------------------------------------------------------------|
| K091701 | Straumann | Dental Abutment (Modified Dental<br>Implant Abutment) |
| K073628 | Straumann | Abutment, Dental, Endosseous<br>implants (RN synOcta 1.5 mm Abutment |
#### 5. Device Description:
The OKTAGON® Dental Implant System is an integrated system of endosseous dental implants which are designed to support prosthetic devices for partially of fully edentulous patients. The devices covered in this submission are abutments in different version including the corresponding screws.
Endosseous Dental Implant Abutment OKTAGON®, Hager & Meisinger GmbH 510(k) Traditional ·
{1}------------------------------------------------
## 510(k) Traditional Dental Endosseous Implant Abutment OKTAGON®
## Section #5 510(k) Summary
The abutments are made of Titanium Grade 4, Titanium Alloy or POM; the ' connection to the implants is achieved by an internal octagon and a metric thread.
#### 6. Indications for use:
Dental Implant Abutments are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.
Abutments can be used in single tooth replacements and multiple tooth restorations.
The Dental Implant Abutments OKTAGON® are intended to be compatible to OKTAGON® implants (Dental Implant OKTAGON®) with diameters 3.3mm, 4.1mm and 4.8mm in the variation Regular Platform, Wide Platform and Tapered Design with the lengths 8mm, 10mm, 12mm and 14mm."
#### 7. Performance tests and used standards / Clinical data
For Dental Implant Abutments OKTAGON® a performance test (fatigue test) has been conducted.
The performed test fulfills the requirements listed in ISO 14801 and the Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Abutments, Chapter 8. The following standards have been considered for Dental Implant Abutments OKTAGON® during development, production and market surveillance: ISO 14801, ISO 7405, ISO 10993-1, ISO 5832-2, ASTM F67, ISO 14971, ISO 5832-3, ASTM F136. AAMI ST79. ISO 17665-1.
| Manufacturer | Hager & Meisinger<br>GmbH | Straumann AG | Straumann AG |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K132214 | K091701 | K073628 |
| Indications for Use | Abutments are intended to be<br>placed into dental implants to<br>provide support for prosthetic<br>reconstructions. Prosthetic<br>applications can include<br>individual crowns, bridges,<br>partial or total prostheses. | Abutments are placed into<br>dental implants to provide<br>support for prosthetic<br>restorations such as crowns<br>and bridges.<br>Abutments can be used in<br>single tooth replacements and | Abutments are placed into<br>dental implants to provide<br>support for prosthetic<br>restorations such as crowns<br>and bridges.<br>Abutments can be used in<br>single tooth replacements and |
#### 8. Basis for substantial equivalence
Endosseous Dental Implant Abutment OKTAGON®, Hager & Meisinger GmbH 510(k) Traditional
{2}------------------------------------------------
# 510(k) Traditional Dental Endosseous Implant Abutment OKTAGON®
## Section #5 510(k) Summary
| Manufacturer | Hager & Meisinger GmbH | Straumann AG | Straumann AG |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K132214 | K091701 | K073628 |
| | Abutments can be used in single tooth replacements and multiple tooth restorations. | multiple tooth restorations. | multiple tooth restorations. |
| Description | The OKTAGON® Dental Implant System is an integrated system of endosseous dental implants which are designed to support prosthetic devices for partially of fully edentulous patients. The devices covered in this submission are abutments in different version including the corresponding screws. | The Straumann Dental Implant System is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, abutments and surgical and prosthetic parts and instruments. The devices covered in this submission are abutments. | The Straumann Dental Implant System is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, abutments and surgical and prosthetic parts , and instruments. The device covered in this submission is an abutment.<br>The coronal portion of the abutment is seated in the implant with a screw. The abutment is used for screw-retained and cemented restorations. |
| Material | Titanium Grade 4, Titanium Alloy, synthetic material | Not detailed in submission, acc. to catalogue same material used | Not detailed in submission, acc. to catalogue same material used |
| Connection | octagonal anti-rotation device | octagonal anti-rotation device | octagonal anti-rotation device |
The intended use for Endosseous Dental Implant Abutment OKTAGON® is identical to the named predicated devices. The abutments have the same indications for use, material composition and the connection to implants is equivalent. In addition the principal design including measurements of abutments is identical to the previously cleared predicated devices.
Based on the information provided in the summary, we conclude Dental Implant Abutments OKTAGON® are substantially equivalent to the legally marketed predicate devices described.
(
Endosseous Dental Implant Abutment OKTAGON®, Hager & Meisinger GmbH 510(k) Traditional
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three legs, representing health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### July 7, 2014
Hager & Meisinger GMbh Ms. Wiebke Stolten Management, Regulatory Affairs Hansemannstrasse 10 Neuss, D-41468 GERMANY
Re: K132214
Trade/Device Name: Dental Implant Abutment OKTAGON® Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implants abutment Regulatory Class: II Product Code: NHA Dated: June 2, 2014 Received: June 3, 2014
Dear Ms. Stolten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Stolten
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MaryDner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### 510(k) Traditional Dental Endosseous Implant Abutment OKTAGON®
#### Section #4 Indications for Use Statement
510(k) Number (if known): K132214
Device Name: Dental Implant Abutment OKTAGON®
Indications for use:
Dental Implant Abutments are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.
Abutments can be used in single tooth replacements and multiple tooth restorations.
The Dental Implant Abutments OKTAGON® are intended to be compatible to OKTAGON® implants (Dental Implant OKTAGON®) with diameters 3.3mm, 4.1mm and 4.8mm in the variation Regular Platform, Wide Platform and Tapered Design with the lengths 8mm, 10mm, 12mm and 14mm.
Over-The-Counter Use AND/OR Prescription Use X (Part 21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/11 description: The image shows the words "Sheena A. Greens" and "2014.07.04 -04'00'". The words "A. Greens" and "07.04" are in a decorative font. The words are stacked on top of each other.
Page 4.1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.