K121334 · Thommen Medical AG · NHA · Aug 14, 2012 · Dental
Device Facts
Record ID
K121334
Device Name
VARIOECO
Applicant
Thommen Medical AG
Product Code
NHA · Dental
Decision Date
Aug 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Thommen VARIOeco dental implant abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
Device Story
VARIOeco is a titanium dental implant abutment system used to support crowns, bridges, and overdentures in the maxillary or mandibular arch. The device is used in a clinical dental setting by a dentist or dental specialist. It consists of a titanium abutment, a laboratory burn-out cap (POM), and an abutment screw. The abutment is attached to a Thommen System dental implant; the burn-out cap is used during the fabrication of porcelain-fused-to-metal (PFM) restorations. The output is a stable, implant-supported prosthetic restoration. By providing a secure interface between the dental implant and the final restoration, the device enables functional tooth replacement, improving patient masticatory function and aesthetics.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and technological characteristics compared to legally marketed predicate devices.
Technological Characteristics
Materials: CPTi Grade 4 (abutment), POM (burn-out sleeve). Design: Titanium abutment with burn-out cap and screw. Dimensions: 3.5, 4.0, 4.5, 5.0, 6.0 mm platform diameters. Interface: Compatible with Thommen System dental implants (ELEMENT, CONTACT, ONETIME). Sterilization: Standard methods consistent with predicates.
Indications for Use
Indicated for patients requiring support for crowns, bridges, or overdentures in the maxillary or mandibular arch using Thommen System dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
AUG
1 4 2012
# 510(k) Summary
## Thommen Medical AG
# K121334
# VARIOeco
June 27, 2012
#### ADMINISTRATIVE INFORMATION
Manufacturer Name:
Thommen Medical AG Hauptstrasse 26d CH-4437 Waldenburg, Switzerland +41 61 965 90 20 Telephone: Fax: +41 61 965 90 21
Official Contact:
Representative/Consultant:
Holger Jakob, Head of Quality Management/Regulatory Affairs
Linda K. Schulz, BSDH, RDH Kevin A. Thomas, PhD PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone: +1 (858) 792-1235 +1 (858) 792-1236 Fax: lschulz@paxmed.com Email: kthomas@paxmed.com
#### DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name: VARIOeco Common Name: Dental implant abutment Classification Name: Endosseous dental implant abutment Classification Regulation: 21 CFR 872.3630, Class II Product Code: NHA Classification Panel: Dental Products Panel Dental Devices Branch Reviewing Branch:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
### INTENDED USE
Thommen VARIOeco dental implant abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
### DEVICE DESCRIPTION
V ARIOeco abutments are titanium dental implant abutments for use with Thommen System dental implants for fabrication of porcelain-fused-to-metal (PFM) restorations. VARIOeco includes two abutment designs, one for single tooth (crown) and one for multi-tooth (bridge) restorations. A laboratory burn-out cap and abutment screw are provided together with the abutment in the VARIOeco set. Both VARIOeco abutment designs are provided in five platform diameters (3.5, 4.0, 4.5, 5.0 and 6.0 mm) corresponding to Thommen System dental implants (ELEMENT, CONTACT and ONETIME).
## EQUIVALENCE TO MARKETED DEVICE
Thommen Medical AG has submitted information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices. VARIOeco is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
Thommen Medical AG. SPI® EASY 4.0 Abutment - K071453. Thommen Medical AG, SPI® Dental Implant Abutments - K031747, Altatec GmbH, CAMLOG® Vario SR Abutments - K103252,
The subject VARIOeco abutment and predicates straight SPI EASY Abutment and straight VarioSR Abutment are all titanium abutments with a burn-out plastic sleeve for restoration fabrication.
The subject device and the predicate devices have the same intended use and have the same technological characteristics. They encompass the same range of physical dimensions, including platform diameter, abutment height and abutment/implant interface. The subject and predicate devices are packaged in similar materials and sterilized using similar methods. Any differences in the technological characteristics do not raise new issues of safety or efficacy.
{2}------------------------------------------------
| | Subject Device | Predicate Devices | | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Thommen Medical AG<br>VARIOeco | Thommen Medical AG<br>SPI® EASY 4.0<br>Abutment<br>K071453 | Thommen Medical AG<br>SPI® Dental Implant<br>Abutments<br>K031747 | Altatec GmbH<br>CAMLOG® Vario SR<br>Abutments<br>K103252 |
| Indications<br>for Use | Thommen VARIOeco<br>dental implant<br>abutments are intended<br>to be used in<br>conjunction with<br>Thommen System<br>dental implants in the<br>maxillary and/or<br>mandibular arch to<br>provide support for<br>crowns, bridges and<br>overdentures. | The Thommen SPI®<br>EASY Ø 4.0 Abutment<br>is intended to be used in<br>conjunction with SPI®<br>System dental implants<br>in the maxillary and/or<br>mandibular arch to<br>provide support for<br>crowns, bridges or.<br>overdentures. | Thommen SPI® Dental<br>Implant Abutments are<br>intended to be used in<br>conjunction with SPI®<br>System dental implants<br>in the maxillary and/or<br>mandibular arch to<br>provide support for<br>crowns, bridges or<br>overdentures. | CAMLOG® Vario SR<br>components for crown and<br>bridge restorations:<br>Occlusal screw-retained<br>crown, bridge and bar<br>constructions on<br>CAMLOG® implants (with<br>J and K article numbers) in<br>the anterior and posterior<br>region of the maxilla and<br>mandible<br>CAMLOG® Vario SR<br>components for bar<br>restorations:<br>- Anchorage of implant-<br>supported full dentures for<br>the edentulous maxilla and<br>mandible in combination<br>with 2, 4 or more<br>CAMLOG® implants (with<br>J and K article numbers) |
| Restoration | Single or multi-unit | Single or multi-unit | Single or multi-unit | Single or multi-unit |
| Abutment<br>(Ø mm) | 3.5, 4.0, 4.5, 5.0, 6.0 | 4.0 | 3.5, 4.5, 5.0, 6.0 | 3.8, 4.3, 5.0, 6.0 |
| Material | CPTi Gr 4 - abutment<br>POM - Burn-out sleeve | CPTi Gr 4 - abutment<br>POM - Burn-out sleeve | CPTi Gr 4 - abutment<br>POM - Burn-out sleeve | CPTi Gr 4 - abutment<br>Ti-6Al-4V - abutment<br>POM - Burn-out sleeve |
Overall, VARIOeco has the following similarities to the predicate devices:
- has the same intended use, .
- . uses the same operating principle,
- incorporates the same basic design, .
- incorporates the same or very similar materials, and .
- has similar packaging and is sterilized using the same materials and processes. .
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with outstretched wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 1 4 2012
THOMMEN Medical, AG C/O Ms. Linda K. Schulz PaxMed International, Limited Liability Company 11234 El Camino Real, Suite 200 San Diego, California 92130
Re: K121334
Trade/Device Name: VARIOeco Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 27, 2012 Received: June 28, 2012
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Ms. Schulz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
h For
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Summary
# Indications for Use
510(k) Number:
×121334
Device Name: VARIOeco
Indications for Use:
Thommen VARIOeco dental implant abutments are intended to be used in conjunction with Thommen System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges and overdentures.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susan Runner
(Division Sign-Off) Division of Anesthesiology; General Hospital Infection Control, Dental Devices
510(k) Number: k121334
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.