SFI-BAR (R) IMPLANT ADAPTER SPI (R) ELEMENT PF 0 4.0 / H 3.0 / H 4.0 / H 5.0 / H 6.0, SFI-BAR (R) IMPLANT ADAPTER STRA..
Applicant
Cendres & Metaux SA
Product Code
NHA · Dental
Decision Date
Sep 12, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The SFI-Bar® is intended to be used with the implant manufacturer's (TABLE A) implant to provide support for fixation of overdentures.
Device Story
SFI-Bar® is a dental abutment system providing a connection between dental implants and removable overdentures. It consists of an implant adapter (abutment) and a stress-free bar. The adapter is screwed into the dental implant; the bar provides primary splinting and stabilization of implants. Used in the oral cavity by dental professionals. The device counters dislodging forces, distributes shear forces, and provides resilience compensation for the denture. It supports immediate loading in the mandible if specific criteria are met: implant manufacturer permits immediate loading, no guided bone regeneration required, minimum 35 Ncm insertion torque, and pull-off strength < 20 N during osseointegration. The system benefits patients by providing a secure, stable fixation for removable overdentures.
Clinical Evidence
No clinical data. Performance was established through non-clinical testing, including validation of sterilization, evaluation of manual disinfection procedures, and functional testing based on literature research. The evaluation followed FDA guidance for root-form endosseous dental implants and abutments.
Technological Characteristics
Materials: Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (ASTM F 136, UNS R 56401). Design: Machined, screw-retained implant adapter and stress-free bar. Connectivity: Mechanical interface with specific dental implant platforms (Thommen, Neoss, Straumann). Manufacturing: CNC machined, validated cleaning process. Non-sterile packaging.
Indications for Use
Indicated for patients requiring fixation of removable overdentures supported by compatible dental implants (Thommen SPI Element, Neoss ProActive, or Straumann ITI systems).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. To the left of the text is a stylized letter "C" that is thicker than the text. The text and the letter "C" are in black and the background is white.
#### 2011 12
K1113910
## Attachment: 510(k) Summary
## Applicant's Name and Address
Submitter:
CENDRES+METAUX SA Rue de Boujean 122 2501 Biel/Bienne, Switzerland Phone: +41 58 360 20 00 +41 58 360 20 10 Fax:
Contact Person:
Tanja Bongni Consultant Regulatory Affairs
Date of Submission:
May 09, 2011
#### Name of the Device
Trade Name: Common Name: Classification Name: Regulation Number:
SFI-Bar® Abutment. Dental. Endosseous implants Endosseous Dental Implant Abutment 21 CFR 872.3630
#### Legally Marketed Device to which Equivalence is Claimed (Predicate Device)
Predicate Device(s):
K073628, K083876, K102382
#### Description of the Device
Device Description:
The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar® consists of an implant adapter (abutment) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.
The implant adapter (abutment) fit the Thommen SPI® Element Platform Ø 4.0 mm / the Neoss ProActive Implant Ø 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm and the Straumann dental implants / ITI Dental Implant System® Standard Ø 4.1 and Ø 4.8 mm / Standard Plus Ø 4.1 mm and Ø 4.8 mm / Tapered Effect Ø 4.1 and Ø 4.8 mm and Regular Neck (RN) Ø 4.8 mm.
The implants (min. 2) in the mandible can be fitted with the SFI-Bar® immediately after implantation, provided the following criteria are met:
- Implant manufacturers permit immediate loading in their -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------system.
- No necessity for simultaneous guided bone regeneration; -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------implants surrounded on all sides by local bone.
- Implant insertion torque min. 35 Ncm.
- All parts are sterilized or disinfected.
- Pull-of strength during osseointegration < 20 N.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. To the left of the text is a stylized letter "M". The text is in a sans-serif font and is black. The logo appears to be for a company that deals with ashes and metals.
K111390
#### Intended Use of the Device:
The SFI-Bar® is intended to be used with the implant manufacturer's (TABLE A) implant to provide support for fixation of overdentures.
TABLE A Compatible Commercial Implant Manufacturers:
| Implant Company | Implant System | Implant Platform<br>Diameter |
|--------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Thommen Medical | SPI® Element<br>Platform | 4.0 mm |
| Neoss | Neoss ProActive<br>Implant | 3.5 / 4.0 / 4.5 / 5.0 /<br>5.5 mm |
| Institut Straumann | ITI Dental Implant<br>System® | Standard 4.1 and 4.8<br>mm /<br>Standard Plus 4.1 and<br>4.8 mm /<br>Tapered Effect 4.1 and<br>4.8 mm /<br>Regular Neck (RN) 4.8<br>mm |
### Summary Technological Characteristics:
Comparison /Compatibility Substantially Equivalence: The proposed labelling change to the SFI-Bar® is substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate device.
The material of the implant adapters conform to ASTM F 136, Wrought Titanium-6Aluminium-4Vanadium ELI Alloy for Surgical Implant applications (UNS R 56401). The parts for the SFI-Bar® System are manufactured from wires.
The proposed implant adapters are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate device.
To ensure compatibility the following process was carried out: The implant adapters are developed and manufactured in close cooperation with the implant companies (see TABLE A, column "Implant Company").
There are Quality Agreements between CENDRES+METAUX and the implant companies in place. Those agreements handle among other things the Design Control, Change Control, Complaint Handling and Post Market Surveillance.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo for Cendres+ Metaux. The logo consists of a stylized letter "M" on the left, followed by the words "CENDRES+" on the top line and "METAUX" on the bottom line. The text is in a simple, sans-serif font and is all in black.
・・
·
-
K11.1390
.
:
.
:
:
TABLE B summarizes the substantial equivalence comparison to the predicate device:
| TABLE B Substantial Equivalence Comparison to Predicate Devices: | | |
|------------------------------------------------------------------|--|--|
| | | |
.
| Attribute | Candidates | Predicate Device | Predicate Device |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | SFI-BarⓇ | SFI-BarⓇ<br>K083876,<br>K102382 | RN synOctaⓇ<br>K073628 |
| Design / construction | Machined, screw-retained | Machined, screw-retained | Machined, screw-retained |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity |
| Platform compatibility | ITI Dental Implant<br>System®: Standard Ø 4.1 mm and Ø 4.8 mm /<br>Standard Plus Ø 4.1 mm and Ø 4.8 mm/<br>Tapered Effect Ø 4.1 mm and Ø 4.8 mm /<br>Regular Neck (RN) Ø 4.8 mm | ITI Dental Implant<br>System®: Standard Ø 4.1 mm and Ø 4.8 mm /<br>Standard Plus Ø 4.1 mm and Ø 4.8 mm/<br>Tapered Effect Ø 4.1 mm and Ø 4.8 mm/<br>Regular Neck (RN) Ø 4.8 mm | ITI Dental Implant<br>System®:<br>Standard Ø 4.8 mm RN<br>and Ø 6.5 mm WN /<br>Standard Plus Ø 4.8 mm<br>and Ø 6.5 mm WN/<br>Tapered Effect Ø 4.8 mm<br>and Ø 6.5 mm WN |
| | Neoss ProActive Implant:<br>Ø 3.5 / 4.0 / 4.5 / 5.0 /<br>5.5 mm | Neoss ProActive Implant:<br>Ø 3.5 / 4.0 / 4.5 / 5.0 /<br>5.5 mm | |
| | Thommen Implant System:<br>SPI® Element Platform Ø<br>4.0 mm | Thommen Implant System:<br>SPI® Element Platform Ø<br>4.0 mm | |
| Device Material | Wrought Titanium-<br>6Aluminium-4Vanadium<br>ELI Alloy for Surgical<br>Implant applications | Wrought Titanium-<br>6Aluminium-4Vanadium<br>ELI Alloy for Surgical<br>Implant applications | Titanium abutment,<br>Ceramicor sleeve |
| Manufacturer | CENDRES+METAUX SA | CENDRES+METAUX SA | Institut Straumann AG |
| Indication for Use | The SFI-Bar® is intended<br>to be used with the<br>implant manufacturer's<br>(see TABLE A) implant to<br>provide support for fixation<br>of overdentures. | The SFI-Bar® is intended<br>to be used with the<br>implant manufacturer's<br>(see TABLE A) implant to<br>provide support for fixation<br>of overdentures. | Abutments are intended to<br>be placed into dental<br>implants to provide<br>support for prosthetic<br>reconstructions such as<br>crowns, bridges and<br>overdentures. |
| Operating principle /<br>Basic Design | Impression taking:<br>optional,<br>preassembled (plug-in<br>connection).<br>Abutment implant<br>connection: screw fixation.<br>Connecting principle to<br>overdenture: retentive<br>system.<br>Bar fixation on implant:<br>screwed.<br>Function: stabilization and<br>primary splinting of<br>implants.<br>Countering forces that<br>would dislodge the | Impression taking:<br>optional,<br>preassembled (plug-in<br>connection).<br>Abutment implant<br>connection: screw fixation.<br>Connecting principle to<br>overdenture: retentive<br>system.<br>Bar fixation on implant:<br>screwed.<br>Function: stabilization and<br>primary splinting of<br>implants.<br>Countering forces that<br>would dislodge the | Impression taking:<br>required.<br>Abutment Implant<br>connection: Screw fixation.<br>Bar connection: soldered.<br>Connecting principle to<br>overdenture: retentive<br>system.<br>Bar fixation on implant:<br>screwed.<br>Countering forces that<br>would dislodge the |
| | | | |
| Attribute | Candidates | Predicate Device | Predicate Device |
| | denture, distribution of<br>shear forces, resilience<br>compensation. | denture, distribution of<br>shear forces, resilience<br>compensation. | denture, distribution of<br>shear forces, resilience<br>compensation. |
| | Cleaning procedures for<br>patient: common<br>procedure for oral hygiene. | Cleaning procedures for<br>patient: common<br>procedure for oral hygiene. | Cleaning procedures for<br>patient: common<br>procedure for oral hygiene. |
| | Patient handling:<br>common cleaning and<br>insertion of denture. | Patient handling:<br>common cleaning and<br>insertion of denture. | Patient handling:<br>common cleaning and<br>insertion of denture. |
| Shelf life | 95% after 10 years | 95% after 10 years | 95% after 10 years |
| Packaging, materials and<br>processes | Produced on process<br>orientated CNC machines.<br>The last step is a validated<br>cleaning process (same<br>processes). | Produced on process<br>orientated CNC machines.<br>The last step is a validated<br>cleaning process (same<br>processes). | Produced on process<br>oriented CNC machines.<br>The last step is a validated<br>cleaning process (same<br>process). |
| | Packaging: dental blister,<br>non-sterile. | Packaging: dental blister,<br>non-sterile. | Packaging:<br>non-sterile. |
| Abutment / implant<br>adapter used in context<br>with the immediate<br>loading. | Yes (this submission). | No. | Yes. Abutment cleared for<br>immediate loading. |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the text "CENDRES + MÉTAUX". To the left of the text is a stylized letter "C". The text is in a sans-serif font and is black. The logo appears to be for a company or organization named Cendres + Métaux.
K111390
Performance Data:
Validation of Sterilization for the metallic components, evaluation of the intended manual disinfection procedure with SFI-Bar® Retention inserts G. application and functional testing based on literature research have been conducted to evaluate the performance characteristics of the SFI-Bar®. The test methods used were the same as in the predicate device. Testing has shown that the SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar®. The acceptance criteria were met.
Summary of Testing to Demonstrate Safety and Effectiveness / Conclusion:
Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. Non-clinical testing consisted of analysis of platforms to identify worst-case test samples. Fatigue testing was not done as the basic design is the same than the predicate device. The evaluation was based on FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." Application and functional tests are based on the predicate device and literature research. The summary of technological characteristics indicate that the device is safe and effective for its intended use and performs as well or better than the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tanja Bongni Consultant Regulatory Affairs Cendres-Metaux SA Rue de Boujean 122 Biel/ Bienne Switzerland 2501
SEP 12 2011
Re: K111390
Trade/Device Name: SFI-Bar® Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: August 23, 2011 Received: August 29, 2011
Dear Ms. Bongni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Ms. Bongni
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MicdicalDevices/Safety/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cendres+Metaux. The logo consists of a stylized letter "C" on the left, followed by the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. The text is in a simple, sans-serif font and is aligned to the right of the "C".
KIIBTO
# Attachment: Indications for Use Statement
Device Name:
SFI-Bar®
#### Indications for Use:
The SFI-Bar® is intended to be used with the implant manufacturer's implant (TABLE a) to provide support for fixation of overentures.
| Implant Company | Implant System | Implant Platform Diameter |
|--------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Institut Straumann | ITI Dental Implant System® | Standard 4.1 and 4.8 mm /<br>Standard Plus 4.1 and 4.8 mm /<br>Tapered Effect 4.1 and 4.8 /<br>Regular Neck (RN) 4.8 mm |
| Thommen Medical | SPI® Element Platform | 4.0 mm |
| Neoss | Neoss ProActive Implant | 3.5 / 4.0 / 4.5 / 5.0 / 5.5 mm |
TABLE a Compatible Commercial Implant Manufacturers
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Suan Suany
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111390
Prescription Use x
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.