DIVER C.E. CATHETER

K051917 · Invatec Innovative Technologies, S.R.L. · QEZ · Aug 8, 2005 · Cardiovascular

Device Facts

Record IDK051917
Device NameDIVER C.E. CATHETER
ApplicantInvatec Innovative Technologies, S.R.L.
Product CodeQEZ · Cardiovascular
Decision DateAug 8, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DIVER C.E. Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system.

Device Story

Aspiration catheter for removal of fresh, soft emboli and thrombi from arterial vessels; features central aspiration lumen and atraumatic tip with radiopaque marker; rapid exchange design compatible with 0.014" guide wires; includes extension tubing, stopcock, syringe, and 40u filter basket for post-procedure thrombi examination; used by physicians in clinical settings; device provides mechanical aspiration to clear arterial obstructions; benefits patient by restoring blood flow.

Clinical Evidence

Bench testing only. Sterilization validation and performance verification testing demonstrated that the device and accessories meet acceptance criteria and perform equivalently to predicate devices.

Technological Characteristics

Aspiration catheter with central lumen and atraumatic tip; radiopaque marker band at 1mm from distal end; rapid exchange design; compatible with 0.014" guide wires; available with/without sideholes; includes accessories (tubing, stopcock, syringe, 40u filter basket).

Indications for Use

Indicated for removal of fresh, soft emboli and thrombi from arterial vessels.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 8, 2021 Invatec Innovative Technologies Stephanie Isgrigg Robinson Regulatory Affairs Specialist 4600 Nathan Lane North Plymouth, Minnesota 55442 Re: K051917 Trade/Device Name: Diver C.E. Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEZ Dear Stephanie Isgrigg Robinson: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 8, 2005. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory. Oconnell(@FDA.HHS.gov. Sincerely. Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Date: 2021.10.08 Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ Public Health Service Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 8 - 2005 Invatec Innovative Technologies, S.R. c/o Ms. Stephanie Isgrigg Robinson Regulatory Affairs Specialist 4600 Nathan Lane North Plymouth, MN 55442 Re: K051917 DIVER C.E. Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II (two) Product Code: DXE Dated: July 14, 2005 Received: July 15, 2005 Dear Ms. Robinson: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bhimma for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications For Use 510(k) Number (if known): K051917 DIVER™ C.E. Catheter Device Name: Device Name Indications for Use: The DIVER C.E. Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system. Prescription Use Over-The-Counter Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) x Page 1 of 1 almanian and the many of the may be the may be the may be the may be the may be the may be the may sion Sign-Off) Division of Cardiovascular Devices 510(k) Number_ DIVERTM C.E. Catheter Traditional 510(k) Confidential Document {4}------------------------------------------------ # AUG 8 - 2005 510(k) Summary DIVERTM C.E. Catheter 510(k) Number: 4051917 This 510(k) summary is being submitted in accordance with the requirements of 21 CFR \$807.92. | Submitter/Contact Person:<br>Applicant<br>(Manufacturer): | Invatec Innovative Technologies<br>Via Martiri della Libertà, 7<br>25030 Roncadelle (BS) Italy<br>Tel: +39 030 258 93 11<br>Fax: +39 030 258 93 12 | |-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Name: | ev3 Inc.<br>4600 Nathan Lane North<br>Plymouth, MN 55442<br>Tel: (763) 398-7000<br>Fax: (763) 398-7200 | | Official Contact: | Stephanie K. Isgrigg Robinson<br>Regulatory Affairs Specialist<br>ev3 Inc.<br>4600 Nathan Lane North<br>Plymouth, MN 55442<br>Tel: (763) 398-7487<br>Fax: (763) 398-7200<br>srobinson@ev3.net | | Summary Preparation Date: | July 15, 2005 | | Device Name and Classification:<br>Trade Name:<br>Common Name/Usual Name:<br>Classification Name:<br>Class: | DIVERTM C.E. Catheter<br>Catheter, Embolectomy<br>Catheter, Embolectomy<br>Class II, 21 CFR 870.5150 | | Predicate Devices: | DIVERTM C.E. Catheter – K050276<br>Vascular Solutions Inc. PRONTOTM Extraction Catheter – K032763<br>VacLokTM Syringe - K994253 | Hi-Flex™ High Pressure Tubing - K883718 Stopcock (One-port Manifold) - K934123 {5}------------------------------------------------ #### Device Description: The DIVER C.E. Catheter is an aspiration catheter indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system. The catheter features an aspiration lumen running through the full length of the catheter. A luer lock type hub in the proximal end allows the connection of the catheter to the extension line (High Pressure Tubing), stopcock, and syringe that are included in the catheter package. The DIVER C.E. Catheter has a central aspiration lumen and a soft, atraumatic tip. The smooth, soft, atraumatic tip has a radiopaque marker band located at 1mm from its distal end. The DIVER C.E. Catheter is available in two configurations - with and without sideholes. The catheter is a rapid exchange design, with the guide wire lumen running from the distal tip to 22.7 cm proximally in the catheter body, where the exit port is located. The catheter is compatible with guide wires with a maximum diameter of 0.014". Additionally, a 40u filter basket is also included to allow the physician to perform a post-procedure examination of the aspirated thrombi/emboli material removed during the procedure. #### Intended Use: The DIVER C.E. Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system. ## Summary of Testing: Sterilization: Results of sterilization validation demonstrate that the DIVER C.E. Catheter and its accessories are adequately sterilized. Performance Data: The results of verification testing demonstrated that the DIVER C.E. Catheter and its accessories met the established acceptance criteria and performs in a manner equivalent to the predicate devices. No new safety or effectiveness issues were raised during the testing program. ## Statement of Equivalence: The DIVER C.E. Catheter and its accessories are substantially equivalent to the currently marketed DIVER C.E. Catheter (K050276), VacLokTM Syringe (K994253), Hi-Flex™ High Pressure Tubing (K883718) Stopcock (One-port Manifold) (K934123) and Vascular Solutions Pronto™ Extraction Catheter (K032763) in intended use, performance, technological characteristics and materials.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...