K251070 · Akura Medical · QEW · Nov 25, 2025 · Cardiovascular
Device Facts
Record ID
K251070
Device Name
Akura Thrombectomy System
Applicant
Akura Medical
Product Code
QEW · Cardiovascular
Decision Date
Nov 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Akura Thrombectomy Catheter System As part of the Akura Thrombectomy System, the Akura Thrombectomy Catheter System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. The Akura Thrombectomy Accessory Kit As part of the Akura Thrombectomy System, the Akura Accessory Kit is indicated to connect the Akura Thrombectomy Catheter to the Akura Thrombectomy Console. The Akura Thrombectomy Console The Akura Thrombectomy Console is indicated as a vacuum source for the Akura Thrombectomy System.
Device Story
System removes thrombus from peripheral vasculature via percutaneous endovascular procedure; operated by trained interventionalists. Components include catheter, sheath, dilator, console, accessory kit, saline pump, and aspiration canister. Catheter delivered to target site under fluoroscopic guidance using 16F introducer sheath and 0.035 in guidewire. System utilizes internal saline jets for clot fragmentation combined with vacuum aspiration provided by console. Console remains outside sterile field. Output is physical removal of thrombus; healthcare provider monitors procedure via fluoroscopy. Benefits include minimally invasive thrombus removal, reducing need for surgical intervention.
Clinical Evidence
No clinical data generated. Substantial equivalence supported by comprehensive bench testing, including visual/dimensional, funnel integrity, radial force, deployment force, flexibility, torsion, fluidics, pressure sensing, simulated use, particulate, corrosion resistance, radiopacity, EMC, and electrical safety testing.
Technological Characteristics
Large-bore catheter system; materials meet ISO 10993-1 biocompatibility standards. Employs internal saline jetting for clot fragmentation and vacuum aspiration. System includes catheter, sheath, dilator, and console. Sterilized via ethylene oxide. Single-use sterile components. Connectivity involves console-to-catheter interface via accessory kit.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from peripheral arterial and venous systems in patients requiring percutaneous endovascular thrombectomy.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Indigo Aspiration System - Aspiration Catheter 12 and Separator 12 (K192981)
K220887 — Mini-ClotTriever Thrombectomy System · Inari Medical · Aug 4, 2022
K250775 — Symphony Thrombectomy System; Symphony 16F 82cm Thrombectomy System · Imperative Care, Inc. · May 14, 2025
K234073 — ICE Aspiration System · Expanse Medical, Inc. · Apr 19, 2024
K254208 — FLOWRUNNER Aspiration System · Expanse Medical, Inc. · Mar 20, 2026
K160533 — Penumbra Embolectomy Aspiration System (INDIGOTM Aspiration System) · Penumbra, Inc. · May 24, 2016
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 25, 2025
Akura Medical
Theresa Brandner
VP, Regulatory Affairs
170 Knowles Drive
Los Gatos, California 95032
Re: K251070
Trade/Device Name: Akura Thrombectomy System
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy Catheter
Regulatory Class: Class II
Product Code: QEW
Dated: October 30, 2025
Received: October 31, 2025
Dear Theresa Brandner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251070 - Theresa Brandner
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251070 - Theresa Brandner
Page 3
Sincerely,
GREGORY W. O'CONNELL -S
Digitally signed by
GREGORY W. O'CONNELL -S
Date: 2025.11.25 16:16:03
-05'00"
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251070 | |
| Device Name Akura Thrombectomy System | |
| Indications for Use (Describe) The Akura Thrombectomy Catheter System As part of the Akura Thrombectomy System, the Akura Thrombectomy Catheter System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. The Akura Thrombectomy Accessory Kit As part of the Akura Thrombectomy System, the Akura Accessory Kit is indicated to connect the Akura Thrombectomy Catheter to the Akura Thrombectomy Console. The Akura Thrombectomy Console The Akura Thrombectomy Console is indicated as a vacuum source for the Akura Thrombectomy System. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251070
# 510(K) SUMMARY FOR THE AKURA THROMBECTOMY SYSTEM
510(k) Number: K251070
## A. SPONSOR
Akura Medical, Inc.
170 Knowles Drive
Los Gatos, CA 95032
USA
## B. CONTACT
Theresa Brandner
VP, Regulatory Affairs
Tel: 650-868-7268
Email: theresa@akuramed.com
## C. DEVICE NAME
Trade Name: Akura Thrombectomy System
Common/Usual Name: Embolectomy/Thrombectomy Catheter
Classification Name: Embolectomy Catheter
(21 CFR § 878.5150, Class II, Pro-Code QEW)
Classification Panel: Cardiovascular
## D. PREDICATE DEVICE
510(k): K192981
Trade Name: Indigo Aspiration System - Aspiration Catheter 12 and Separator 12
Common/Usual Name: Embolectomy Catheter
Classification Name: Embolectomy Catheter
(21 CFR § 878.5150, Class II, Pro-Code QEW)
Classification Panel: Cardiovascular
## E. REFERENCE DEVICE
510(k): K183403
Trade Name: JETi 88 Peripheral Thrombectomy System
Common/Usual Name: Embolectomy/Thrombectomy Catheter
Classification Name: Embolectomy Catheter
(21 CFR § 878.5150, Class II, Pro-Code QEZ)
Classification Panel: Cardiovascular
Page 1 of 4
{5}
K251070
## F. DEVICE DESCRIPTION
The subject Akura Thrombectomy System is designed to remove thrombus from the vasculature. The System is used by interventionalists trained in percutaneous endovascular thrombectomy procedures. The subject Akura Thrombectomy System is comprised of several devices.
- Akura Thrombectomy Catheter System
- Akura Thrombectomy Catheter
- Akura Thrombectomy Sheath
- Akura Thrombectomy Dilator
- Akura Thrombectomy Console
- Akura Accessory Kit
- Saline Pump
- Irrigation Tubing
- Aspiration Canister Assembly
During use, the Akura Thrombectomy Catheter System and Akura Thrombectomy Accessory Kit are connected to the Akura Thrombectomy Console that is placed outside the sterile field. The Akura Thrombectomy System is introduced through percutaneous access using a commercially available 16F vascular introducer sheath and a 0.035 in exchange length guidewire. The Akura Thrombectomy Catheter System is delivered to the target treatment site under fluoroscopic guidance. The Akura Thrombectomy Dilator is removed from the Akura Thrombectomy Sheath. The Akura Thrombectomy Catheter is inserted through the Akura Thrombectomy Sheath, and the Akura Thrombectomy Catheter funnel is deployed. Thrombus is removed by fragmentation using internal saline jets and aspiration. After the procedure, the Akura Thrombectomy Catheter System is removed from the vasculature and disposed of using standard hospital practices. The Akura Thrombectomy Console is cleaned and stored in accordance with the Akura Thrombectomy System User Manual.
## G. INDICATIONS FOR USE
**The Akura Thrombectomy Catheter System**
As part of the Akura Thrombectomy System, the Akura Thrombectomy Catheter System is indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
**The Akura Thrombectomy Accessory Kit**
As part of the Akura Thrombectomy System, the Akura Accessory Kit is indicated to connect the Akura Thrombectomy Catheter to the Akura Thrombectomy Console.
**The Akura Thrombectomy Console**
The Akura Thrombectomy Console is indicated as a vacuum source for the Akura Thrombectomy System.
Page 2 of 4
{6}
K251070
## H. STERILIZATION AND SHELF LIFE
The Akura Thrombectomy System is sterilized using ethylene oxide. The suitability of the packaging to protect the subject Akura Thrombectomy System and to ensure sterility within the stated shelf life was verified through a series of tests performed by independent test laboratories. These tests confirm the packaging integrity, sterility, and distribution cycle. This testing demonstrated that the packaging is sufficiently robust to withstand extreme distribution and environmental conditions while maintaining packaging integrity and sterility.
## I. BIOCOMPATIBILITY
The Akura Thrombectomy System includes components that are provided sterile for single use and disposal. The sterile components, the Akura Thrombectomy Catheter and Akura Thrombectomy Accessory Kit, have met the biocompatibility testing requirements identified in ISO 10993-1, including the tests for cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility.
## J. TECHNOLOGY CHARACTERISTICS
The predicate device, the Indigo Aspiration System - Aspiration Catheter 12 and Separator 12 (K192981), and the reference device, the JETi 88 Peripheral Thrombectomy System (K183403), were used to support substantial equivalence of the subject Akura Thrombectomy System. The subject device, the predicate device, and the reference device include the following similar technical characteristics.
- Both the proposed device and predicate device are designed for the non-surgical removal of thrombi or emboli from vasculature.
- Both the subject device and predicate device are intended to be used with commonly available vascular access tools (e.g., guidewire, vascular introducer) to facilitate the removal of thrombus, embolus, and clot during minimally invasive percutaneous procedures.
- Both the subject device and predicate device are large bore catheters with a pump aspiration source.
- Both the subject device and predicate device have similar operating principles by being advanced through a sheath and over a guidewire to the target treatment location along with a mechanical aspiration source that is used to aspirate thrombus/emboli from vasculature.
- Both the subject device and the reference device use internal saline jetting for clot fragmentation during aspiration.
The technological characteristics of the proposed Akura Thrombectomy System are substantially equivalent with respect to the basic system design, function, materials, dimensions, and manufacturing to that of the predicate device.
Page 3 of 4
{7}
K251070
# K. PERFORMANCE DATA
Comprehensive bench testing (integrity and functional performance) was performed to support substantial equivalence to the predicate devices. The Akura Thrombectomy System met the specified design and performance requirements based on the following tests.
- Visual And Dimensional
- Funnel Integrity
- Funnel Radial Force
- Deployment Force
- Flexibility
- Torsion
- Fluidics
- Pressure Sensor
- Pressure/Leak
- Water Resilience
- Contrast Injection
- Tensile
- Simulated Use
- Particulate
- Corrosion Resistance
- Radiopacity
- Electromagnetic Compatibility
- Electrical Safety
- Electrical Emitter/Immunity
- Human Factors Evaluation/Usability Evaluation
# L. CLINICAL TESTING
No clinical data were generated to establish substantial equivalence. Bench data are considered adequate to support a determination of substantial equivalence.
# M. CONCLUSIONS
The results of the non-clinical testing and a comparison of similarities and differences demonstrate that the Akura Thrombectomy System and predicate device are substantially equivalent.
Page 4 of 4
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.