POWERHEART AED G3, MODEL 9390E, POWERHEART AED G3 AUTOMATIC, MODEL 9390A
Applicant
Cardiac Science Corporation
Product Code
MKJ · Cardiovascular
Decision Date
Sep 10, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The Powerheart AED G3 and the Powerheart AED G3 Automatic devices are intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response. The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will change automatically and advise the operator to deliver therapy (G3) or automatically deliver the charge (G3 Automatic). When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the device should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.
Device Story
Portable, battery-operated AEDs (G3/G3 Automatic) for emergency treatment of sudden cardiac arrest. Input: ECG signals via chest electrodes. Operation: Device analyzes ECG rhythm; advises operator to shock (G3) or automatically delivers shock (G3 Automatic) if shockable ventricular tachyarrhythmia detected. Used by trained BLS/emergency personnel. Provides voice prompts, audible alerts, and visual indicators to guide rescue. Includes accessory adapter cable for AAMI DF-80 compliant electrodes. Benefits: Rapid defibrillation therapy for cardiac arrest victims.
Clinical Evidence
Bench testing only. Representative samples of the device and adapter cables were tested in accordance with system, safety, functional, and performance specifications; all samples passed.
Technological Characteristics
Portable, battery-operated AED. Sensing: ECG via chest electrodes. Energy: Defibrillation shock. Connectivity: Supports AAMI DF-80 compliant electrodes via adapter cable. Software: Automated ECG analysis and rhythm detection.
Indications for Use
Indicated for emergency treatment of unresponsive, non-breathing victims of sudden cardiac arrest. Pediatric use (up to 8 years or 55 lbs) requires Model 9730 Pediatric Attenuated Defibrillation Electrodes.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
K113787 — DDU-120 FULLY AUTOMATIC EXTERNAL DEFIBRILLATOR AND ACCESSORIES · Defibtech, LLC · Jan 4, 2013
P160034 — Powerheart® AED G3 Pro · ZOLL Medical Corporation · Dec 6, 2018
K052057 — LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR · Medtronic Emergency Response Systems, Inc. · Feb 17, 2006
P160015 — AED Plus and Fully Automatic AED Plus · ZOLL Medical Corporation · May 26, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 1: 0 2009
:
K691943 Cardiac Science Corporation.
Bothell, WA Special 510(k)
June 4, 2009
1/2
:
#### ���. 510(k) Summary
| Submitter: | Cardiac Science Corporation<br>3303 Monte Villa Parkway<br>Bothell, WA 98021-8969 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Beverly Magrane<br>Phone: (425) 402-2365<br>Fax: (425) 402-2017 |
| Date Prepared: | June 4, 2009 |
| Trade Name: | Powerheart® AED G3 (Model 9390E)<br>Powerheart® AED G3 Automatic (Model 9390A) |
| Classification<br>Name and Number: | Automated External Defibrillator<br>Class III, 21CFR 870.5310 |
| Product Code: | MKJ |
| Predicate<br>Device(s): | The Powerheart® AED G3 manufactured by Cardiac<br>Science, Inc is substantially equivalent to the Cardiac<br>Science Powerheart AED G3, K052161 (10/21/2005) and<br>to the FirstSaveTM STAR BiphasicTM AED, K010214,<br>(2/22/2001), manufactured by Survivalink Corporation. |
| Device<br>Description: | The Powerheart® AED G3 and G3 Automatic are<br>portable, battery operated, automated external<br>defibrillators (AED). After applying the AED's electrodes<br>(pads) to the patient's bare chest, the AED automatically<br>analyzes the patient's electrocardiogram (ECG) and<br>advises the operator to press the button and deliver a<br>shock if needed. The AED uses one button and guides<br>the operator through the rescue using a combination of<br>voice prompts, audible alerts, and visible indicators. For<br>the Powerheart AED G3 Automatic, the AED<br>automatically delivers a shock if needed.<br>The Cardiac Science G3 AEDs may be used in<br>conjunction with an adaptor cable that allows the use of<br>AAMI DF-80 compliant electrodes other than the Cardiac |
:
4
{1}------------------------------------------------
Cardiac Science Corporation. Bothell, WA Special 510(k) June 4, 2009
691943
Indications For Use:
Functional and Safety Testing:
Conclusion:
Science brand. This adaptor cable is provided as an accessory to the Cardiac Science AEDs.
The Powerheart AED G3 and the Powerheart AED G3 Automatic devices are intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response.
The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will change automatically and advise the operator to deliver therapy (G3) or automatically deliver the charge (G3 Automatic).
When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the device should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.
Representative samples of the device and adapter cables were tested in accordance with the system, safety, functional and performance specifications. All samples successfully passed.
Based on the results of the testing described above, it is concluded that the modifications to the Powerheart AED do not raise any new questions regarding the safety or effectiveness as compared with the predicate device. The Cardiac Science, Inc. Powerheart AED is substantially equivalent to the Powerheart AED cleared in K052161 and FirstSave STAR Biphasic AED cleared in K010214 in terms of indications for use, features and functions.
Confidential
5
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
SEP 1 0 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cardiac Science Corporation c/o Ms. Beverly Magrane Sr. Manager RA/RC 3303 Monte Villa Pkwy. Bothell, WA 98021
K091943 Re:
> Trade/Device Name: 9055 Electrode Adapter Cable Regulation Number: 21 CFR 870.5310 Regulation Name: Automated external defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: August 25, 2009 Received: August 27, 2009
Dear Ms. Magrane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Beverly Magrane
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
W.M.P.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K0919443
# Indications for Use Statement
### 510(k) Number:
Device Name: Powerheart® AED G3 (Model 9390E), Powerheart® AED G3 Automatic (Model 9390A)
### Indications for Use:
The Powerheart AED G3 and the Powerheart AED G3 Automatic devices are intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physicianauthorized emergency medical response.
The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Postresuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will charge automatically and advise the operator to deliver therapy (G3) or automatically deliver the charge (G3 Automatic).
When the patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the device should be used with the Model 9730 Pediatric Attenuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.
Prescription Use (Per 21 CFR 801.109) OR
Over-The Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off) |
|------------------------------------|
| Division of Cardiovascular Devices |
| 510(k) Number | K091943 |
|---------------|---------|
|---------------|---------|
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.