BIO-DETEK MODIFIED ADULT MULTI-FUNCTION ELECTRODE WITH CONNECTOR

K033771 · Bio-Detek, Inc. · MKJ · Feb 27, 2004 · Cardiovascular

Device Facts

Record IDK033771
Device NameBIO-DETEK MODIFIED ADULT MULTI-FUNCTION ELECTRODE WITH CONNECTOR
ApplicantBio-Detek, Inc.
Product CodeMKJ · Cardiovascular
Decision DateFeb 27, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The BIO-DETEK Modified Adult Multi-Function Electrode with Connector is indicated for use with Medtronic Physio-Control Defibrillator Models: LIFEPAK 9, LIFEPAK 10C, LIFEPAK 11, LIFEPAK 500, LIFEPAK 12, and LIFEPAK 20, all fitted with QUIK-COMBO Therapy Cables, for defibrillation, cardioversion, non-invasive pacing, and electrocardingraph monitoring.

Device Story

Multi-function electrodes designed for use with Medtronic Physio-Control LIFEPAK defibrillators; interface via QUIK-COMBO therapy cables. Electrodes facilitate delivery of electrical therapy (defibrillation, cardioversion, pacing) and acquisition of ECG signals. Used in clinical settings by healthcare professionals. Device serves as conductive interface between patient skin and defibrillator unit; enables therapeutic energy delivery and diagnostic monitoring. Benefits include standardized connectivity for emergency cardiac care.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-2-4, ANSI/AAMI/ISO DF39-1993, and ANSI/AAMI EC53:1995 standards demonstrated.

Technological Characteristics

Multi-function electrode pads with integrated connector. Complies with IEC 60601-2-4, ANSI/AAMI/ISO DF39-1993, and ANSI/AAMI EC53:1995. Passive conductive interface; no internal software or active electronic components.

Indications for Use

Indicated for adult patients requiring defibrillation, cardioversion, non-invasive pacing, or ECG monitoring using compatible Medtronic Physio-Control LIFEPAK defibrillators.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## BIO-DETEK Modified Adult Multi-Function Electrodes with Connector 510(k) Premarket Notification Submission #### 10.0 BIO-DETEK Modified Adult Multi-Function Electrodes with Connector 510(k) Summary #### Company: BIO-DETEK 525 Narragansett Park Drive Pawtucket, RI 02861-4323 #### Contact: Robert Morse QA/QC Manager ### Date Prepared: December 1, 2003 ### Name of Device: BIO-DETEK Adult Multi-Function Electrodes (#K931801) ### Predicate Device: BIO-DETEK Adult Multi-Function Electrodes # Device Description and Intended Use: The BIO-DETEK Modified Adult Multi-Function Electrodes with Connector is indicated for use with Medtronic Physio-Control Defibrillator Models: LIFEPAK 9, LIFEPAK 10C, LIFEPAK 11, LIFEPAK 12, and LIFEPAK 20, all fitted with QUIK-COMBO Therapy Cable, for defibrillation, cardioversion, non-invasive pacing, and electrocardiograph monitoring. # Technological Characteristics The BIO-DETEK Modified Adult Multi-Function Electrodes with Connector is designed to comply with the applicable portions of the following standards: - IEC 60601-2-4 Medical Electrical Equipment . - ANSI/AAMI/ISO DF39-1993 . - ANSI/AAMI EC53:1995 . {1}------------------------------------------------ ## BIO-DETEK Modified Adult Multi-Function Electrodes with Connector 510(k) Premarket Notification Submission - 21 CFR Part 898 Performance Standards for Electrode Lead Wires and Patient . Cables ## Basis for Substantial Equivalence: Operation and technological characteristics form the basis for the determination of substantial equivalence of The BIO-DETEK Modified Adult Multi-Function Electrodes with Connector with legally marketed predicate devices. Information supplied in this premarket notification includes descriptive information about the intended use, operation, and technological characteristics. This information is provided pursuant to the requirements of the Safe Medical Devices Act of 1990 (SMDA). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle is blank. In the center of the seal is an abstract image of an eagle. FEB 27 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Bio-Detek, Inc. c/o Mr. Robert Morse QA/QC Manager 525 Narragansett Park Drive Pawtucket, RI 02861-4323 Re: K033771 > Trade Name: Bio-Detek Modified Adult Multi-Function Electrodes with Connector Regulation Number: 21 CFR 870,1025 Regulation Name: Arrythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: 74 MKJ, MLN Dated: December, 08 2003 Received: December 09, 2003 Dear Mr. Morse: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Robert Morse Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Dma R. Lochner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number: K033771 Device Name: BIQ-DETEK Modified Adult Multi-Function Electrode with Connector Intended Use: The BIO-DETEK Modified Adult Multi-Function Electrode with Connector is indicated for use with Medtronic Physio-Control Defibrillator Models: LIFEPAK 9, LIFEPAK 10C, LIFEPAK 11, LIFEPAK 500, LIFEPAK 12, and LIFEPAK 20, all fitted with QUIK-COMBO Therapy Cables, for defibrillation, cardioversion, non-invasive pacing, and electrocardingraph monitoring. Prescription Use V (Pari 21 CFR 801 Subpan D) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) - Dwing R. bodanes. (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number k033771
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...