FLOWTRON EXCEL PUMP
Device Facts
| Record ID | K961166 |
|---|---|
| Device Name | FLOWTRON EXCEL PUMP |
| Applicant | Hne Healthcare, Inc. |
| Product Code | JOW · Cardiovascular |
| Decision Date | Oct 11, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.5800 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The AC550 Flowtron®Excel pump, coupled with the Flowtron DVT compression garments, is an external pneumatic compression system used to prevent deep vein thrombosis (DVT) in patients deemed to be at risk by their physicians.
Device Story
External pneumatic compression system; consists of AC550 pump and Flowtron DVT compression garments (calf/thigh). Used in clinical settings to prevent DVT. Pump utilizes compressor, timing motor, and valves to inflate garments. Microprocessor-controlled pressure transducer monitors output; provides audible/visible fault alarms and real-time pressure display. Physician-prescribed; operator-managed. Output informs clinician of system performance and pressure delivery; aids in DVT prophylaxis.
Clinical Evidence
No clinical data provided; bench testing only. Device relies on established history of external pneumatic compression therapy.
Technological Characteristics
Pneumatic compression pump; microprocessor-controlled pressure transducer; audible/visible LED fault monitoring; real-time pressure display. Compliance: UL listed, European Medical Directive (EMC).
Indications for Use
Indicated for patients at risk of deep vein thrombosis (DVT) requiring external pneumatic compression therapy.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
- Flowtron®DVT (K88xxxx)
- Flowtron DVT AC500
Related Devices
- K133119 — FLOWTRON ACS800 PUMP AND TRI PULSE GARMENTS · Arjohuntleigh Polska SP Z.O.O. · Feb 28, 2014
- K143438 — Flowtron ACS900 · Getinge (Suzhou) Co., Ltd. · Jun 23, 2015
- K173012 — ANGIO-PRESS LITE DVT Compression Device · Pegasus Medical Supply, Inc. · May 10, 2018
- K061814 — VASOPRESS DVT SYSTEM, MODEL VP500D · Compression Therapy Concepts, Inc. · Nov 30, 2006
- K110358 — VENODYNE V810 · Microtek Medical, Inc. · May 5, 2011