CURRIE MEDICAL REPROCESSED ALP CALF/THIGH/FOOT COMPRESSION SLEEVE (ALL SIZES)

K140092 · Currie Medical Specialties, Inc. · JOW · Apr 16, 2014 · Cardiovascular

Device Facts

Record IDK140092
Device NameCURRIE MEDICAL REPROCESSED ALP CALF/THIGH/FOOT COMPRESSION SLEEVE (ALL SIZES)
ApplicantCurrie Medical Specialties, Inc.
Product CodeJOW · Cardiovascular
Decision DateApr 16, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of this device, as well as the predicate device, is to provide external limb compression in order to artificially imitate the pumping action of the leg muscles. This provides muscle contraction required by the venous return system, thereby helping to prevent venous stasis and subsequent thrombosis and embolism. The cyclic and alternating inflation of the garments closely simulates the normal healthy pumping action of the limb muscles to stimulate deep venous blood flow and the reactivation or increase in the body's fibrinolytic system.

Device Story

Reprocessed compression sleeves (calf, thigh, foot) designed for use with ALP 501 Pump System. Sleeves inflate/deflate to provide intermittent, sequentially gradient pressure to limbs; mimics leg muscle pumping action to stimulate venous blood flow, prevent venous stasis, and reduce DVT/pulmonary embolism risk. Operated by healthcare providers in clinical settings. Pump controller dictates inflation/deflation cycles. Benefits include improved venous return and fibrinolytic system activation. Device is a reprocessed version of existing compression garments.

Clinical Evidence

Bench testing only. Tests included bladder leak testing at 3 psi, bladder seal pull testing (>5 lbs force), velcro adhesion testing at 2 psi, and inflation-deflation time testing at 40 mm Hg. Results demonstrated that reprocessed devices maintain performance, safety, and efficacy equivalent to predicate devices.

Technological Characteristics

Pneumatic compression sleeves for calf, thigh, or foot. Operates via external ALP 501 Pump System controller. Materials and design identical to predicate. Bench testing verified integrity of bladders, seals, and velcro fasteners post-reprocessing. No electronic components or software within the sleeve itself; relies on external pump for energy and control.

Indications for Use

Indicated for patients requiring external compression therapy using the ALTERNATING LEG PRESSURE (ALP) SYSTEM to reduce the incidence of deep vein thrombosis and pulmonary embolism due to the presence of risk factors for thrombosis formation.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a symbol and some text. The symbol is an eight-pointed star with a circle in the center. The text to the right of the symbol is cut off, but it appears to be the beginning of some words. The text begins with the letters 'CU' and 'M'. | Section 11: 510(k) Summary | | K140092 pg. 1 of 2 | | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--| | Submitters Name and Address: | Currie Medical Specialties, Inc.<br>8758 Hellman Avenue<br>Rancho Cucamonga, CA 91730<br>Phone: (909) 912-0900<br>Fax: (909) 944-3030 | APR 16 201 | | | FDA Registration Number: | 2023637 | | | | Contact Person: | Owen J. Bry, Director of Quality Assurance | | | | Date Summary Prepared: | 12/17/2013 | | | | Trade or Proprietary Name(s): | Currie Medical Reprocessed ALP° Calf, Thigh and Foot<br>Compression Sleeves (all sizes) . | | | | Common Name: | Compression sleeve limb | | | | Product Code: | JOW | | | | Panel: | Cardiovascular | | | | Regulation Number: | 870.5800 | | | | Classification: | Class II | | | | Predicate Device(s): | K955853 - Healthcare Service & Supply Pump ALP 501 System<br>K964188 - Healthcare Service & Supply PVA (Pneumatic Venous<br>Augmentation) Foot Garment | | | ## Device Description: Currie Medical Specialties (CMS) ALP reprocessed compression sleeves are compression devices, which when attached to the ALP 501 Pump System, provide intermittent, sequentially gradient pressure to a patients leg/foot for the prevention of deep vein thrombosis (DVT). When the compression sleeve is inflated, the veins collapse which forces blood to move upward the heart. After compression is complete, the sleeves deflate which allows the veins to reopen and bring oxygenated blood to the calf, thigh or foot. The inflation and deflation sequence is predicated by the ALP 501 Pump System controller. #### Intended Use: The intended use of this device, as well as the predicate device, is to provide external limb compression in order to artificially imitate the pumping action of the leg muscles. This provides muscle contraction required by the venous return system, thereby helping to prevent venous stasis and subsequent thrombosis and embolism. The cyclic and alternating inflation of the garments closely simulates the normal healthy pumping action of the limb muscles to stimulate deep venous blood flow and the reactivation or increase in the body's fibrinolytic system. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a black and white drawing of a geometric shape. The shape is a square with a circle in the center. The square is surrounded by an eight-pointed star. The star is made up of eight triangles that are arranged around the square. CURRIE MEDICAL SPECIALTIES, INC. This submission is intended for use of the pump at pressure levels of 40 mm Hg. ## Indications for Use: The Currie Medical Reprocessed ALP Calf, Thigh and Foot Compression Sleeves (all sizes) are recommended for the use in patients for whom external compression therapy using the ALTERNATING LEG PRESSURE® (ALP®) SYSTEM is indicated to reduce the incidence of deep vein thrombosis and resulting pulmonary embolism due to the presence of risk factors for thrombosis formation. Technological characteristics of the CMS Reprocessed ALP Calf, Thigh and Foot Compression Sleeves: The CMS ALP Reprocessed Calf, Thigh and Foot Compression Sleeves indications for use, fundamental scientific technology, overall design, materials, energy source, mode of operation, performance characteristics are identical and no different than the predicate device. # Summary of Comparison Tests (Non-Clinical Tests): Bench testing was conducted to ensure that reprocessing did not compromise the performance or safety and efficacy of the device in a manner that is substantially equivalent to that of the predicate devices (K955853 & K964188). Bench testing was conducted to demonstrate a bladder leak test where the bladders were pressurized to 3 psi to ensure the bladders did not leak at this pressure and were not affected by reprocessing. A bladder seal pull test was performed, where the bladder seal was pulled apart at > 5 lbs of force to demonstrate the seals were intact and were not affected by reprocessing. A velcro functionality test was performed where bladders were pressurized to 2 psi to determine velcro adhesion remained intact at this pressure and were not affected by reprocessing. Finally, an inflation-deflation time test was performed to demonstrate product integrity remained the same in comparison to the predicate device. This required the devices to be inflated to 40 mm Hg and monitor the pressure over time. The series of data demonstrated that the reprocessing did not affect the performance, or the safety and efficacy of the device. ## Biocompatibility Reprocessing did not affect the biocompatibility of the device. ## Conclusion The Currie Medical Specialties ALP® Reprocessed Calf, Thigh, and Foot Compression Sleeves have demonstrated that they perform as safe and effective as the ALP® non-sterile predicate devices (K955853 & K964188) and are substantially equivalent to the predicate devices. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ## April 16, 2014 Currie Medical Specialties, Inc. Mr. Owen J. Bry Director of Quality Assurance 8758 Hellman Ave Rancho Cucamonga, CA 91730 Re: K140092 > Trade/Device Name: Currie Medical Reprocessed ALP® Calf, Thigh and Foot Compression Sleeves (all sizes) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: February 18, 2014 Received: February 19, 2014 Dear Mr. Bry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may on ontrols provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulterstion. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I heast be advised that i Dr i issualite of our device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Mr. Owen J. Bry or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to . http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuc Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo with a star-shaped design on the left and the text "CURRENT MEDICAL SPECIAL" on the right. The star-shaped design has a circle in the center with some text inside. The text on the right is vertically stacked, with each word on a separate line. #### INDICATIONS FOR USE | Applicant: | Currie Medical Specialties, Inc. | |---------------------------|--------------------------------------------------------------------------------------| | 510(k) Number (if known): | K140092 | | Device Name: | Currie Medical Reprocessed ALP. Calf, Thigh and Foot Compression Sleeves (all sizes) | | Indications for Use: | | The Currie Medical Reprocessed ALP Calf, Thigh and Foot Compression Sleeves (all sizes) are recommended for the use in patients for whom external compression therapy using the ALTERNATING LEG ·PRESSURE® (ALP®) SYSTEM is indicated to reduce the incidence of deep vein thrombosis and resulting pulmonary embolism due to the presence of risk factors for thrombosis formation. (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) | Prescription Use | | OR | Over-the-Counter | | |--------------------|--|----|------------------|--| | Per 21 CFR 801.109 | | | | | Bram D. Zuckerman -S 2014.04.16 10:20:46 -04'00'
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...