The IQoolTM System

K232844 · Braincool AB · DWJ · Oct 4, 2023 · Cardiovascular

Device Facts

Record IDK232844
Device NameThe IQoolTM System
ApplicantBraincool AB
Product CodeDWJ · Cardiovascular
Decision DateOct 4, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5900
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The IQool System is a thermal regulating system indicated for monitoring and controlling patient temperature.

Device Story

Non-invasive thermal regulating system; monitors/controls patient core temperature (33°C–38.5°C). System comprises ECU 100 refrigeration/control unit, BC COOL (monopropylene glycol) coolant, single-use biocompatible cooling pads, and neoprene stabilization insulation. ECU 100 circulates coolant (4°C–40°C) through pads applied to patient skin; heat exchange regulates core temperature via PT1 probe feedback. Operated via touchscreen monitor; displays temperature graphs/alarms. BC Stick (USB) used for configuration, software updates, and log retrieval. Used in clinical settings to maintain target temperature; healthcare providers adjust settings via touchscreen and monitor patient status via visual feedback. Benefits include precise temperature management.

Clinical Evidence

Bench testing only. Verification testing confirmed that software changes maintain performance and safety equivalent to the predicate device.

Technological Characteristics

Thermal regulating system; ECU 100 refrigeration unit; monopropylene glycol coolant; biocompatible, latex-free cooling pads; neoprene stabilization insulation. Operates via closed-loop circulation of temperature-controlled fluid. Connectivity via USB (BC Stick) for configuration/logs. Software-controlled temperature regulation.

Indications for Use

Indicated for monitoring and controlling patient temperature in patients requiring thermal regulation.

Regulatory Classification

Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 4, 2023 BrainCool AB % Adam Harris Senior Director, Regulatory and Strategic Development Target Health LLC 450 Commerce Boulevard Carlstadt, New Jersey 07072 Re: K232844 Trade/Device Name: The IQool System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: September 12, 2023 Received: September 14, 2023 Dear Adam Harris: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely. Kalkidan Molla -S for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K232844 Device Name The IQool™ System Indications for Use (Describe) The IQool System is a thermal regulating system indicated for monitoring and controlling patient temperature. Type of Use (Select one or both, as applicable) < Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ The IQool™ System Special 510(k) # 5. 510(k) Summary #### 5.1 Sponsor Information Brain Cool AB Scheelevägen 2 Medicon Village Lund, Sweden SE-223 81 Telephone: +46 733 93 70 76 #### 5.2 Contact Person Adam Harris, MM, RAC Senior Director, Regulatory and Strategic Development Target Health LLC 450 Commerce Boulevard Carlstadt, NJ 07072 Telephone: 646-479-2437 Fax: (212) 682-2105 aharris@targethealth.com #### 5.3 Date of Summary October 2, 2023 #### 5.4 Device Name and Classification | Common or Usual Name | Targeted Temperature Management System (TTM) | |-----------------------|-----------------------------------------------| | Proprietary Name | The IQool™ System | | Classification Name | Thermal regulating system (21 CFR § 870.5900) | | Device Classification | Class II | | Product Code | DWJ | #### ર્સ્ટ Predicate Device/Reference Device - Predicate Device K180375 IQool™ Warm System ● - Reference Device K200225 Arctic Sun (for the addition of alarm to protect skin from ● extended cooling exposure) {5}------------------------------------------------ #### Description of Device 5.6 The IQool™ System is a non-invasive, thermal regulating system that monitors and controls patient temperature within a range of 33°C to 38.5°C (91.4°F to101.3°F). The IOooL™ System consists of: - ECU 100 Refrigeration and Control Unit an integrated control system operated via a . touch screen monitor. - BC COOL a cooling liquid consisting of diluted monopropylene glycol (MPG5). The ● dilution is made by BrainCool AB to optimally serve the IQool™ System. Five liters of BC COOL are delivered with the system. - Cooling Pads The Cooling Pads are made of non-sterile, biocompatible material and do ● not contain latex. The Cooling Pads are intended for single use only. - . Stabilization Insulation - The patented stabilization is made of insulating and moisture-absorbent neoprene. It insulates against the ambient environment while counteracting condensation. The elasticity of the stabilization keeps the Cooling Pad in place during treatment and ensures maximum contact between the skin and the surface of the Cooling Pad. The stabilizing insulation is intended for single use only. ## Accessories: - BC Stick a USB flash drive used to save system configurations, specifically prepared to ● communicate with the program of the IQool™ System. The BC stick does not store or capture any user identifying information. It can also be used to save a log file for system troubleshooting or to update the software. - Filling Pitcher - The filling pitcher should be used to fill or refill the coolant tank. Fill the tank with coolant before or directly after starting to enable start of the system. The IQool™ System is a thermoregulatory device that monitors and controls patient temperature within a range of 33°C to 38.5°C. The IQool System refrigeration and control unit (ECU 100) pushes a temperature-controlled cooling agent (BC COOL), ranging between 4°C and 40°C, through the Cooling Pads applied to the patient. This results in heat exchange between the BC COOL and the patient. Patient core temperature is monitored and controlled based on the temperature probe (PT1) attached to the system. The IQool™ System maintains a controlled patient temperature during the entire treatment period. The default treatment settings for temperature and time can be changed via the touchscreen monitor. Alarms and notifications are activated if any errors are detected. Temperature graphs for each treatment are shown on the touchscreen display for visual monitoring. After a treatment, a detailed log file can be collected with the memory stick (BC STICK). {6}------------------------------------------------ #### 5.7 Indications for Use The IQool™ System is a thermal regulating system indicated for monitoring and controlling patient temperature. ### Comparative Data for Determining Substantial Equivalence of the New Device to the 5.8 Predicate Device Information provided in this 510(k) submission for software changes documents that the IQooT™ System has the same technologicalcharacteristics (i.e., same design, materials and energy source) as the predicate device. In addition, the testing summaries provided document that the changes were verified and that the subject device provides the same temperature management performance and safety as the predicate device. #### 5.9 Conclusion It is the opinion of the Sponsor that the IQool™ System does not raise new questions of safety and effectiveness in comparison to the Predicate. This is supported by the equivalence and comparison information, and the data summaries presented. The differences in technological characteristics between the IQool™ System and the Predicate meet the requirements of 21 CFR 807.87(f). Therefore, the sponsor has determined that the updated IQool™ System is substantially equivalent to its Predicate, IOoolTM Warm System K180375.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...