The CAPERE® Thrombectomy System is indicated for: -Non-surgical removal of soft emboli and thrombi from blood vessels. -Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
Device Story
CAPERE® Thrombectomy System is a percutaneous mechanical thrombectomy device for peripheral vasculature. System comprises 9Fr delivery catheter and 12Fr guide catheter. Physician advances catheters to therapy site via transfemoral or jugular venous access. Nitinol wire basket, deployed by retracting outer sheath, captures thrombus via mechanical retraction; basket includes suture support to maintain perpendicularity. Device does not use active aspiration for clot removal, though guide catheter side port allows fluid injection/aspiration. Clot-loaded basket is retracted into guide catheter funnel for removal. Device benefits patients by enabling non-surgical clot extraction.
Clinical Evidence
Bench testing only. Functional testing evaluated compression resistance, bond joints, basket retraction force, and clot capture/removal efficacy using simulated-use methods established for the predicate. Results confirmed the device meets performance requirements and maintains predicate characteristics. Biocompatibility, sterilization, and packaging remain unchanged from the predicate.
Technological Characteristics
System includes 9Fr delivery catheter, 12Fr guide catheter, and 11.4cm nitinol wire basket. Features suture support for basket assembly. Compatible with 0.014" guidewire. Mechanical capture principle. Complies with ISO 10555-1 and ISO 594-2 standards. Non-powered, manual operation.
Indications for Use
Indicated for non-surgical removal of soft emboli and thrombi from peripheral blood vessels and for injection, infusion, or aspiration of fluids/contrast media. Contraindicated for use in pulmonary arteries.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
K183179 — CAPERE Thrombectomy System · Vascular Medcure, Inc. · Feb 25, 2019
K180722 — CAPERE Thrombectomy System · Kp Medcure, Inc. · Jun 11, 2018
K200314 — Capere Thrombectomy System · Vascular Medcure, Inc. · Mar 4, 2020
K203476 — CAPERE Thrombectomy System · Vascular Medcure, Inc. · Dec 22, 2020
K192935 — CAPERE Thrombectomy System · Vascular Medcure, Inc. · Nov 15, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 8, 2020
Vascular Medcure, Inc. % Ms. Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, PA 19103
Re: K201216
Trade/Device Name: CAPERE® Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: June 12, 2020 Received: June 12, 2020
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201216
Device Name CAPERE® Thrombectomy System
| Indications for Use ( <i>Describe</i> ) |
|-----------------------------------------|
|-----------------------------------------|
The CAPERE® Thrombectomy System is indicated for:
-Non-surgical removal of soft emboli and thrombi from blood vessels.
-Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel.
The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
Type of Use (Select one or both, as applicable)
| <span><span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
# Vascular Medcure, Inc.'s CAPERE® Thrombectomy System
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Jeffrey P. DuMontelle Vascular Medcure, Inc. 1500 S. Sunkist Street, Suite H Anaheim, CA 92806 Office: 657-549-5175 714-915-0886 Cell: Date Prepared: May 28, 2020
#### Name of Device
CAPERE® Thrombectomy System
#### Common or Usual Name
Embolectomy Catheter
#### Classification
21 CFR 870.5150, Class II, product code QEW, KRA
#### Predicate Devices
CAPERE® Thrombectomy System (K192935)
## Intended Use / Indications for Use
The CAPERE® Thrombectomy System is indicated for:
- Non-surgical removal of soft emboli and thrombi from blood vessels. o
- Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. O
The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
#### Device Description
The CAPERE® Thrombectomy System primarily consists of an 9Fr Delivery Catheter and 12Fr Guide Catheter. The CAPERE® Thrombectomy System is delivered percutaneously via transfemoral or jugular venous access. Once delivered, the System's fine mesh nitinol wire basket is used to capture and mechanically remove emboli and thrombi. The CAPERE® System does not use aspiration to pull out the thrombus but does have a side port in the funnel catheter that allows aspiration or injection of saline or fluids if needed.
{4}------------------------------------------------
# Technological Characteristics
The CAPERE® Thrombectomy System has the same technological characteristics as its predicate device CAPERE Thrombectomy Catheter (K192935). Both CAPERE® Thrombectomy Systems have either the same components or substantially equivalent components (addition of suture support for the basket assembly). Each device system includes a guide catheter, delivery catheter, and a mechanism for capturing/removing the soft emboli or thrombi. The CAPERE® Thrombectomy System uses a 9Fr delivery catheter while the predicate uses a 6Fr delivery catheter. The CAPERE® Thrombectomy System utilizes and the predicate utilizes an 8Fr quide catheter. In addition, the CAPERE® Thrombectomy System and the predicate both use a 0.014" guidewire during the procedure.
The CAPERE® System includes a Guide Catheter and a Delivery Catheter. In the same manner as the previously cleared device, the Guide Catheter and Delivery Catheter are advanced to the therapy site. The funnel of the Guide Catheter is unsheathed, and the Delivery Catheter is then advanced passed the obstruction. A nitinol basket connected to the Delivery Catheter is deployed by retracting the outer sheath of the catheter.
To capture the soft emboli or thrombi ("clot"), the deployed basket is retracted while simultaneously being extended to surround and capture the length of the clot. The Delivery Catheter and clot-load is then retracted into the funnel of the Guide Catheter where the clot is captured and can be removed.
The basket design in the CAPERE® Thrombectomy Systems is the same (nitinol,11.4cm length) as the basket design in the predicate. The only difference between the baskets of the two systems is the addition of the suture support to the basket to aid in maintaining basket perpendicularity during basket retraction.
Therefore, the subject CAPERE® Thrombectomy System has the same technological characteristics as its predicate.
# Performance Data
The following nonclinical performance testing has been conducted, using the established test methods used in the cleared predicate device, to support the substantial equivalence of the CAPERE® Thrombectomy System to its predicate device. In all instances, the CAPERE® Thrombectorny System functioned as intended under the previously established test methods.
- . Functional bench testing was conducted (including demonstrated compliance with relevant standards such as ISO 10555-1 and ISO 594-2) on the compression resistance and bond joints affected by the change in diameter of the CAPERE® Thrombectomy System with respect to the predicate and demonstrated that the CAPERE® Thrombectomy System maintained the predicate characteristics. Specifically,
- The force to retract the basket into the funnel catheter was evaluated in о accordance with the same simulated-use test method used to clear the predicate and the CAPERE® was determined to meet the same retraction force requirements as the predicate.
- The ability of the CAPERE® to capture and remove clot was evaluated in the o accordance with the same simulated-use test method used to clear the predicate and the CAPERE® was determined to completely capture and
{5}------------------------------------------------
remove the simulated clot and therefore maintains the characteristics of the predicate.
- . The materials and packaging are unchanged from the predicate therefore the biocompatibility, sterilization, and packaging characteristics of the predicate were maintained.
# Substantial Equivalence
The CAPERE® Thrombectomy System has the same intended use and indications for use, technological characteristics, and principles of operation as its predicate devices. Nonclinical testing, including functional testing and simulated use testing demonstrated that the minor differences between the device and predicate (i.e. addition of suture support) do not raise new types of safety or effectiveness questions. Thus, the CAPERE® Thrombectomy System is substantially equivalent to the predicate devices.
# Conclusion
Vascular Medcure's CAPERE® is an Embolectomy Catheter, Class II device that has been evaluated in nonclinical testing in accordance with FDA's recognized standards and preestablished acceptance criteria. Testing demonstrated that the device performs as intended. The CAPERE® is substantially equivalent to its predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.