K103651 · Cardio Medical Products, Inc. · LDD · Oct 4, 2011 · Cardiovascular
Device Facts
Record ID
K103651
Device Name
CARDIOMED
Applicant
Cardio Medical Products, Inc.
Product Code
LDD · Cardiovascular
Decision Date
Oct 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5300
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Cardio Med Pediatric Radiotranslucent Multifunction Electrodes, (Various Models) are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac paccmaker and the patient's skin. The electrode is intended for use on pediatric patients whose weight is less than 10 kg (22 lbs). When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (30 joule maximum).
Device Story
Single-use, non-sterile, radiotranslucent defibrillator electrodes; conductive interface between patient skin and external defibrillator/pacemaker. Applied by clinician to pediatric patients (<10 kg); connects to low-power (max 30J) instruments via specific connector styles. Facilitates delivery of electrical therapy (pacing/defibrillation) and signal acquisition for monitoring. Benefits include safe, effective electrical contact for pediatric cardiac intervention.
Clinical Evidence
Bench testing only. Evaluated DC offset, offset instability, AC small signal/large signal impedance, and defibrillator overload per ANSI/AAMI DF80:2003. Accelerated shelf-life testing (42 months) performed. Biocompatibility testing conducted on patient-contact hydrogel. All tests passed.
Technological Characteristics
Single-use, radiotranslucent electrodes; conductive adhesive hydrogel patient interface. Dimensions: 46.43 cm² (7.1967 in²). Connectors meet FDA touch-proof standards. Compliant with ANSI/AAMI DF80:2003. Passive device; no software or active electronic components.
Indications for Use
Indicated for pediatric patients weighing <10 kg (22 lbs) requiring external pacing, defibrillation, cardioversion, or ECG monitoring. Compatible with low-power defibrillators (max 30 joules).
Regulatory Classification
Identification
Low-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. This generic type of device includes low energy defibrillators with a maximum electrical output of less than 360 joules of energy that are used in pediatric defibrillation or in cardiac surgery. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body. High-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of greater than 360 joules of energy used for defibrillating the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body.
Predicate Devices
PadPro 2603 Pediatric Multifunction Electrodes
Related Devices
K992977 — HEWLETT PACKARD MULTIFUNCTION PEDIATRIC DEFIB FUNCTION, MODEL M3503A & M3504A · Hewlett-Packard Co. · Dec 1, 1999
K032421 — AMBU PEDIATRIC MULTI-FUNCTION DEFIBRILLATION ELECTRODE · Ambu, Inc. · Feb 27, 2004
K081371 — SKINTACT PEDIATRIC MULTIFUNCTION ELECTRODES WITH DH03 GEL · Leonhard Lang GmbH · Sep 24, 2008
K120536 — HEART SYNC PEDIATRIC PHYSIO AED PAD · Heart Sync, Inc. · Aug 6, 2012
{0}------------------------------------------------
## 510(k) Summary K103651
OCT - 4 2011
(1) Submitter: Cardio Medical Products, Inc. 385 Franklin Avenue - Suite L Rockaway, NJ. 07866 Phone: 973-586-9500 Toll Free: 800-227-3633 Fax: 973-586-9624 Contact: Nick Mendise, V.P. Sales Date prepared: 21 June 2011
(2) The name of the device: Cardio Med Pediatric Radiotranslucent Multifunction Electrodes, Various Models (Various Models) Common or usual name: Defibrillator Pad Classification name: 870.5300 DC-defibrillator (including paddles).
(3) Predicate devices: Identical in function and construction to the PadPro 2603 Pediatric Multifunction Electrodes.
(4) Description of the device: These are single use, non-sterile, self stick defibrillator electrodes packaged in pairs. Effective electrode area = 46.43 cm² (7.1967 in²) They are radiotranslucent. They come in various connector styles to match the specific defibrillator. (See intended use statement below.) The construction and materials employed are identical to the predicate. The patient contact material is a conductive adhesive hydrogel identical to the material used in the predicate devices. These pads meet the AAMI Standard ANSI/AAMI DF80:2003 and the connectors meet the FDA performance standard for touch proof ECG connectors.
(5) Statement of the intended use of the device: Cardio Med Pediatric Radiotranslucent Multifunction Electrodes, (Various Models) are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac paccmaker and the patient's skin. The electrode is intended for use on pediatric patients whose weight is less than 10 kg (22 lbs). When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (30 joule maximum).
Compatible with:
- a. PZ300 = Zoll
- b. PO301 = Medtronic-Physio Control
- C. PA302 = Cardiac Science (Anderson)
- d. PP303 = Phillips Heartstream
- PH304 = Philips Barrel (HP) e.
{1}------------------------------------------------
(6) This device has the same technological characteristics (i.c., design, material, chemical composition, energy source) as the predicate device. The defibrillator pads meet the AAMI and FDA performance standards for this type of device.
(7) A brief discussion of the nonclinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence:
Bench testing summary: Various different lots containing multiple samples each were subjected to the AAMI tests for DC offset, offset instability, AC small signal impedance, AC large signal impedance, defibrillator overload. The same testing was performed for post-pacing characteristics. All units passed these tests. The same testing routine was applied to a 42 month accelerated age shelf life test. All units passed the tests. Biocompatibility testing was performed on the patient contact material Hydrogel. The material passed biocompatibility testing.
Compliance with the FDA performance standard was verified by inspection of the connectors. (Connectors must be "touch-proof.")
(8) Conclusion: Based on the results of the nonclinical tests (that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device) we conclude that these defibrillator pads are as safe and effective as the predicates identified in paragraph (3). Furthermore, the materials and construction methods are identical to the predicate.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cardio Medical Products, Inc. c/o Mr. Nick Mendise Vice President Sales 385 Franklin Avenue - Suite L Rockaway, NJ 07866
ORT - A. COM
Re: K103651
Trade Name: Cardio Med Radiotranslucent Multifunction Electrodes Regulation Number: 21 CFR 870.5300 Regulation Name: DC-defibrillator (including paddles) Regulatory Class: Class II Product Code: LDD Dated: September 12, 2011 Received: September 20, 2011
Dear Mr. Mendise:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{3}------------------------------------------------
Page 2 - Mr. Nick Mendise
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDcvices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
[signature]
Bran D. Zuckerman M.D. 10-Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
## Indications for Use
510(k) Number(if known): K103651
Device Name: Cardio Med Pediatric Radiotranslucent Multifunction Electrodes, Various Models
Indications for Use:
The Cardio Med Pediatric Radiotranslucent Multifunction Electrodes, Various Models are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. The electrode is intended for use on pediatric patients whose weight is less than 10 kg (22 lbs). When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (30 joule maximum).
Compatible with:
- PZ300 = Zoll a.
- b. PO301 = Medtronic-Physio Control
- PA302 = Cardiac Science (Anderson) C.
- PP303 = Phillips Heartstream ರ.
- PH304 = Philips Barrel (HP) e.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
�
510(k) Number K103651
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.