JOBST ATHROMBIC PUMP/SLEEVE

K970922 · Jobst A Beiersdorf Co. · JOW · Jan 13, 1998 · Cardiovascular

Device Facts

Record IDK970922
Device NameJOBST ATHROMBIC PUMP/SLEEVE
ApplicantJobst A Beiersdorf Co.
Product CodeJOW · Cardiovascular
Decision DateJan 13, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Increase blood flow from lower extremities in recumbent patients to help prevent deep vein thrombosis (DVT) and pulmonary embolism (PE)

Device Story

Sterile sleeve for Jobst Athrombic Pump System 2500; used to prevent deep vein thrombosis and pulmonary embolism in recumbent patients. Device functions by providing intermittent pneumatic compression to lower limb muscles, increasing deep venous blood flow. Operated in clinical settings; intended for prescription use. Sterile version of previously marketed non-sterile sleeve; provides same mechanical compression therapy as predicate system.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use compared to the predicate device.

Technological Characteristics

Compressible limb sleeve; material: PVC; sterilization: ethylene oxide; energy source: pneumatic (via Jobst Athrombic Pump System 2500).

Indications for Use

Indicated for recumbent patients to increase lower extremity blood flow for the prevention of deep vein thrombosis (DVT) and pulmonary embolism (PE).

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 174 10422 JAN 13 1998 # 510(k) Summary of Safety and Effectiveness This submission is being made for a sterile version of the sleeve used for the Jobst Athrombic Pump System 2500. This product is a Class II device and is classified as a compressible limb sleeve under 21 CFR 870.5800. The sterile sleeve/Jobst Athrombic Pump System 2500 combination is substantially equivalent to the Kendall Healthcare Product's sterile sleeve/SCD compression system in its mode of operation and indications for use. Both products are used to help prevent deep vein thrombosis and pulmonary embolism in recumbent patients by increasing deep venous blood flow from the lower extremities. This is accomplished by intermittent compression of the muscles in the lower limbs. Both sleeves are made of PVC and are sterilized with ethylene oxide. The Jobst Athrombic Pump System 2500 with the non-sterile sleeves has been on the market since 1994 and has proven to be safe and effective in that time. The sterile version is being offered for situations where sterile product is preferred. Date: March 11, 1997 Prepared by: Angelo R. Pereira Beiersdorf-Jobst, Inc. 5825 Carnegie Boulevard Charlotte, NC 28209 Phone: (704) 551-7178 {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service JAN 13 1998 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Mr. Angelo R. Pereira Manager, Regulatory Affairs Beiersdorf-Jobst, Inc. 5825 Carnegie Boulevard Charlotte, NC 28209-4633 Re: K970922 Jobst Athrombic Pump/Sleeve Regulatory Class: II (Two) Product Code: 74 JOW Dated: March 28, 1997 Received: March 31, 1997 Dear Mr. Pereira: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2} Page 2 - Mr. Angelo R. Pereira This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} JUN 02 97 11:11AM JOB:1 CHK Page 2 Or 2 510 (k) Number (if known): K970922 Device Name: Sterile Sleeve for Jobst Athrombic Pump Indications for Use: Increase blood flow from lower extremities in recumbent patients to help prevent deep vein thrombosis (DVT) and pulmonary embolism (PE) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use Optional Format 1-2-96) DA Sygh (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K970922
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...