EUREDUC PNEUMATIC DRAINAGE SYSTEM MODEL TP03

K960902 · Eureduc · JOW · Sep 26, 1996 · Cardiovascular

Device Facts

Record IDK960902
Device NameEUREDUC PNEUMATIC DRAINAGE SYSTEM MODEL TP03
ApplicantEureduc
Product CodeJOW · Cardiovascular
Decision DateSep 26, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Lymphedema of lower extremities; post surgical or congenital ; Postmastectomy lymphedema ; Dysfunction of the muscle pump ; Pre, post phases of surgery.

Device Story

Pneumatic drainage system; microprocessor-controlled pump in plastic casing; connects to leg boots or arm sleeves; provides intermittent pneumatic compression; used in clinical or home settings; operated by patient or clinician; output is controlled limb compression; aids lymphatic drainage and muscle pump function; benefits include edema reduction and improved circulation.

Clinical Evidence

No clinical data provided; substantial equivalence based on technical characteristics and performance comparisons.

Technological Characteristics

Microprocessor-controlled pneumatic pump; plastic casing; 15-99 mmHg pressure range; 3-compartment sleeve compatibility; 24 Watts power; 30-minute treatment session capability.

Indications for Use

Indicated for patients with lymphedema of lower extremities (post-surgical or congenital), postmastectomy lymphedema, muscle pump dysfunction, and pre/post-surgical phases.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K960902 SEP 26 1996 EUREDUC TP03 510(k) 1996-01-15 Page 11 5- 510(k) SUMMARY as required by 807.92 (c) Submitter's name EUREDUC Submitter's address 35, rue des Broderies 78310 COIGNIERES FRANCE Phone # (33) 1 34 61 01 61 Fax # (33) 1 34 61 40 61 Name of contact person: Mr Didier CAUZOT Device Trade name: EUREDUC TP03 Device common name: Pneumatic Drainage System Classification name: Sleeve Limb Compressible 74JOW CFR Section 870 58 00 Predicate device name: JOBST EXTREMITY PUMP(R) SYSTEM 7500 Predicate device 510(k): K882683 EUREDUC TP03 is comprised of: a pneumatic unit in a plastic casing with : - a microprocessor controlled pump; - a control panel a range of garments connectable to the unit: - leg boots; - arm sleeves; The EUREDUC TP03 is intended for : - Lymphedema of lower extremities; post surgical or congenital ; - Postmastectomy lymphedema ; - Dysfunction of the muscle pump ; - Pre, post phases of surgery. .../... {1} EUREDUC TP03 510(k) 1996-01-15 Page 12 Compared technical characteristics and performances (from attached labeling) | DEVICE NAME | JOBST EXTREMITY PUMP (R) | EUREDUC TP03 | | --- | --- | --- | | DEVICE 510(k) | K8826 83 | NA | | PRESSURE RANGE | 30 to 100 mm Hg | 15 to 99 mm Hg | | TREATMENT CYCLE | 200 s | Up to 37 s | | TREATMENT SESSION | NA | 30 mn | | POWER | 15 Watts | 24 Watts | | SLEEVES: No OF COMPARTMENTS | Up to 3 | Up to 3 | The EUREDUC TP03 has similar technical characteristics and performances, identical indications and contraindications to the JOBST EXTREMITY PUMP(R). .../...
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...