EMBOLECTOMY, IRRIGATION - OCCLUSION BALLOON CATHETER BY J-LLOYD MEDICAL, INC.
K974335 · J-Lloyd Medical, Inc. · DXE · Feb 13, 1998 · Cardiovascular
Device Facts
Record ID
K974335
Device Name
EMBOLECTOMY, IRRIGATION - OCCLUSION BALLOON CATHETER BY J-LLOYD MEDICAL, INC.
Applicant
J-Lloyd Medical, Inc.
Product Code
DXE · Cardiovascular
Decision Date
Feb 13, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1. Temporary use in vessel occlusion. 2. Injection of contrast media and other fluids. 3. The removal of thrombi and emboli.
Device Story
Balloon catheter for non-surgical removal of emboli and thrombi; also used for vessel occlusion and fluid/contrast media injection. Device features distal balloon, shaft with 10cm markings, and proximal manifold with two extensions (hub and stopcock). Available in 3-9 French sizes. Operated by trained physicians in clinical settings. Physician inflates balloon to occlude vessel or remove thrombi/emboli; uses lumen for irrigation or contrast delivery. Provides minimally invasive alternative to surgery for vascular clearance.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on engineering testing, mechanical strength, and balloon performance testing compared to the predicate device.
Technological Characteristics
Balloon catheter; 3-9 French sizes; distal balloon; shaft with 10cm markings; proximal manifold with two extensions (hub and stopcock). Materials, assembly methods, and biological/chemical/sterilization specifications are identical to predicate. Multi-lumen design (additional lumen compared to predicate).
Indications for Use
Indicated for temporary blood vessel occlusion, irrigation with heparin or contrast media, and non-surgical removal of emboli and thrombi in patients requiring such interventions.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
J-LLoyd Medical, Inc. Embolectomy Catheter (K935405)
J-LLoyd Medical, Inc. Embolectomy Catheter (K920871/B)
Submission Summary (Full Text)
{0}------------------------------------------------
K974335
FEB 1 3 1998
510 K SUMMARY (as required by 807.92 c)
Date: November 15, 1997
Submitter: James L. Skaggs, Sr. 415 Commerce Lane, Suite 6 Berlin, New Jersey 08009 Phone: 609-753-8339 FAX: 609-753-8340
Contact Person: Same as above.
Product: Embolectomy Irrigation - Ocelusion Balloon Catheter (EIOBC)
I. Predicate Device:
The J-LLoyd Medical, Inc. Embolectorny Catheter has been approved by FDA 510K #K935405 and K920871/B. The function of this device is to provide for non-surgical removal of emboli and thrombi.
II. Description of New Device:
This balloon catheter is furnished in three, four, five, six, seven, eight and nine French sizes.
The device has an balloon mounted at the distal tip. The shaft has bands every ten centimeters. The proximal end has a manifold and one extension with a hub and a second extension with a stopcock.
III. Intended Use of New Device:
This catheter is designed for use as follows:
1. Temporary use in vessel occlusion.
2. Injection of contrast media and other fluids.
3. The removal of thrombi and emboli.
IV. Technological Characteristics of New Device and Predicate Device (510K #K935405 and K920871/B).
1. The Embolectorny Irrigation - Occlusion Balloon Catheter and the Predicate Catheter are basically the same, using the same materials, the same basic design and the same methods of assembly.
{1}------------------------------------------------
2. The difference is the addition of a lumen to the Predicate Device which has one lumen. The addition of the lumens increases the functions of the catheters.
3. The performance data such as mechanical strengths and balloon tests remain the same as with the Predicate Device.
4. The Biological, Chemical. and Sterilization specifications remain the same as the Predicate Device, in that the only physical difference is the addition of a lumen presently used on the Predicate Device.
V. Safety Statement:
1. As with any invasive procedure, there are certain inherent hazards, however, these hazards have been identified in the instruction sheet under the titles of Precautions and Warnings. It is important that the physicians be aware of the basic principles involved prior to using these devices in their varied application.
2. Based on engineering testing, the J-LLoyd Medical, Inc. devices and the fact that the basic design has been in use since before May 28, 1976, by thousands of physicians, in hundreds of hospitals world wide, it is our judgement that this device presents an acceptable level of safety when properly used by a trained physician.
James L. Skaggs, Jr.
General Manager
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 3 1998
James L. Skaggs, Sr. General Manager J-Lloyd Medical, Inc. 415 Commerce Lane, Suite 6 08009 Berlin, New Jersey
K974335 Re: Embolectomy, Irrigation - Occlusion Balloon Catheter by J-LLoyd Medical, Inc. Regulatory Class: II Product Code: DXE November 15, 1997 Dated: Received: November 18, 1997
Dear Mr. Skaggs:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts_800_to_895 ... A ........................................................................................................................ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2 - James L. Skaggs, Sr.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
74335
Page ೧೯
SIO(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Narne: __Embolectomy_Irrigation_occlusion_Balloon_Catheter
Indications For Use:
This catherer is intended for:
- Temporarily occlude blood vessel. 】.
- 2. Irrigating facility for heparin or contrast media.
- 3. USed for the non-surgical removal of emboli and thrombi,
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEBDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K974335
Prescription Use
(Per 21 CFR 801.109)
OR
Over-Tuc-Counter Use