WARMTOUCH PATIENT WARMING SYSTEM, MODEL 5300

K020604 · Nellcor Puritan Bennett, Inc. · DWJ · Apr 23, 2002 · Cardiovascular

Device Facts

Record IDK020604
Device NameWARMTOUCH PATIENT WARMING SYSTEM, MODEL 5300
ApplicantNellcor Puritan Bennett, Inc.
Product CodeDWJ · Cardiovascular
Decision DateApr 23, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5900
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The WarmTouch Model 5300 Patient Warming System (warming unit and blanket) is intended for prevention and treatment of hypothermia, for example, with the surgical patient, the patient in the pre-operative holding area, the pregnant woman who shivers during epidural anesthesia due to hypothermia, or any patient who is uncomfortable anywhere in the cold critical care environment.

Device Story

WarmTouch Model 5300 is a convective air warming system; consists of a warming unit and specialized blankets (CareQuilt or CareDrape). Device warms air and delivers it into the blanket placed over or under the patient to transfer heat across the skin. Used in clinical settings (OR, pre-operative holding, critical care) to prevent or treat hypothermia. Operated by healthcare professionals. Modification from predicate involves updated temperature settings. Benefits include patient comfort and maintenance of normothermia.

Clinical Evidence

Bench testing only. Performance testing demonstrated safe maximum temperature compliance and conformity to ASTM standards for forced air patient temperature management devices. Electrical safety and EMC testing demonstrated compliance with UL544, IEC 60601-1-2, IEC 60601-2-35, and CAN/CSA C22.2 No. 125.

Technological Characteristics

Convective air warming system. Components: warming unit and blankets (CareQuilt/CareDrape). Standards: UL544, IEC 60601-1-2:1993, IEC 60601-2-35, CAN/CSA C22.2 No. 125, and draft ASTM Standard Specification for Circulating Liquid and Forced Air Patient Temperature Management Devices. Energy source: electrical. Form factor: warming unit with air-delivery blankets.

Indications for Use

Indicated for prevention and treatment of hypothermia in surgical patients, pre-operative patients, pregnant women shivering during epidural anesthesia, and patients in cold critical care environments.

Regulatory Classification

Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K62004 280 Hacienda Dri Pleasanton, CA 9458 Tele: 925 463-4000 Fax: 925 463-4020 # tuco Healthcare Nellcor # 510(k) Summary APR 2 3 2002 | Submitted by: | Nellcor Puritan Bennett Incorporated<br>(A business unit of Mallinckrodt Inc.,<br>a division of Tyco Healthcare Group LP)<br>4280 Hacienda Drive<br>Pleasanton, CA 94588 | |----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company Contact: | Gina To<br>Senior Regulatory Affairs Project Manager<br>(925) 463-4427<br>(925) 463-4020 - FAX | | Date Summary Prepared: | February 22, 2002 | | Trade Name: | WarmTouch Model 5300 Patient Warming System | | Common/Usual Name: | Convective Air Warming System | | Classification Name: | Thermal Regulating System per 21 CFR §870.5900 | | Product Code: | DWJ | | Legally Marketed Predicate<br>Device (Unmodified): | WarmTouch, Advanced Warming Systems, Inc.,<br>510(k) #K913016 | ## Device Description The WarmTouch Model 5300 Patient Warming System surrounds the patient with warm air and convectively transfers heat across the skin. With the WarmTouch Model 5300 Patient Warming System, air is warmed and delivered into a lightweight blanket (CareQuilt or CareDrape Blanket) that rests over or under the patient. The modification to the WarmTouch Patient Warming Unit consists of design changes relating to temperature settings. #### Intended Use The WarmTouch Model 5300 Patient Warming System (warming unit and blanket) is intended for prevention and treatment of hypothermia, for example, with the surgical patient, the patient in the pre-operative holding area, the pregnant woman who shivers during epidural anesthesia due to hypothermia, or any patient who is uncomfortable anywhere in the cold critical care environment. NPB CONFIDENTIAL {1}------------------------------------------------ # Summary of Technological Characteristics of the Device Compared to the Legally Marketed (Unmodified) Device The WarmTouch Model 5300 Patient Warming System has the same technological characteristics as the above referenced predicate device. The only changes relate to temperature settings. #### Testing Performance testing of the WarmTouch Model 5300 with the CareQuilt and CareDrape Blankets demonstrated safe maximum temperature compliance. Testing of the WarmTouch Model 5300 Patient Warming System for electrical safety and EMC testing demonstrated compliance to applicable sections in the November 1993 FDA Review Guidance document, UL544 Standard for Safety for Medical and Dental Equipment, IEC 60601-1-2:1993 1st Edition requirements, IEC 60601-2-35 Particular Standard for Patient Warming Systems, and CAN/CSA C22.2 No. 125 for Electromedical Equipment. Performance testing demonstrated conformity to the draft ASTM Standard Specification for Circulating Liquid and Forced Air Patient Temperature Management Devices. #### Conclusions The WarmTouch Model 5300 Patient Warming System and the results of testing do not raise new questions of safety or effectiveness when compared to the legally marketed predicate device. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 3 2002 Ms. Gina To Senior Regulatory Affairs Project Manager Nellcor Puritan Bennett Incorporated 4280 Hacienda Drive Pleasanton, CA 94588 Re: K020604 Trade Name: WarmTouch Model 5300 Patient Warming System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II (two) Product Code: DNJ Dated: March 27, 2002 Received: March 28, 2002 Dear Ms. To: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2 - Ms. Gina To Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of acrease a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or unf I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It Fart 6077, accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) rms letter will and in your e FDA finding of substantial equivalence of your device to a legally promance motification " results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the O invision of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Qarla Tull Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): WarmTouch Model 5300 Patient Warming System Device Name: ## Indications For Use: The WarmTouch Model 5300 Patient Warming System (warming unit and blanket) is intended for prevention and treatment of hypothermia, for example, with the surgical patient, the patient in the pre-operative holding area, the pregnant woman who shivers during epidural anesthesia due to hypothermia, or any patient who is uncomfortable anywhere in the cold critical care environment. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **Division of Cardiovascular & Respiratory Devices** 510(k) Number (Optional Format 3-10-98) **Prescription Use** (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...